Elvie Stride

K210936 · Chiaro Technology Limited · HGX · Aug 25, 2021 · Obstetrics/Gynecology

Device Facts

Record IDK210936
Device NameElvie Stride
ApplicantChiaro Technology Limited
Product CodeHGX · Obstetrics/Gynecology
Decision DateAug 25, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5160
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Elvie Stride is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Stride is intended for a single user.

Device Story

Elvie Stride is a wearable, electric, powered breast pump for lactating women. Device consists of a hub, breast shield, milk collection cup, and tubing. Operates by applying negative pressure to the nipple to express milk into a collection bottle. Designed for in-bra use; hands-free; battery-powered. Controlled via physical interface on the hub or a mobile companion app. Used in home environments by the patient. Provides adjustable suction levels and cycle speeds to facilitate milk expression. Benefits include portability and convenience for the user. Healthcare providers do not interact with the device output; it is a patient-operated device for personal use.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), software verification/validation, and performance testing (vacuum pressure, cycle rate, backflow, and battery life).

Technological Characteristics

Powered breast pump; rotary electric pump mechanism. Components: hub, breast shield, cup, diaphragm, valve, tubing. Materials: reusable, dishwasher-safe. Power: Li-Ion battery, USB charging. Connectivity: wireless (mobile app). Standards: ISO 10993 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-11 (home healthcare), IEC 60601-1-2 (EMC), IEC 62133 (battery safety).

Indications for Use

Indicated for lactating women to express and collect milk from their breasts. Intended for single-user, home-use.

Regulatory Classification

Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 25, 2021 Chiaro Technology Limited % Marc C. Sanchez, Esq. Regulatory Attorney Contract In-House Counsel and Consultants, LLC d/b/a FDA Atty 53516 Bickett Chapel Hill, NC 27517 Re: K210936 Trade/Device Name: Elvie Stride Regulation Number: 21 CFR\$ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 21, 2021 Received: July 28, 2021 Dear Marc C. Sanchez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210936 Device Name Elvie Stride Indications for Use (Describe) The Elvie Stride is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Stride is intended for a single user. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K210936 Page 1 of 4 # 510(K) Summary # K210936 ### Submitter Information: | Sponsor: | Chiaro Technology Limited<br>63-66 Hatton Garden<br>London EC1N 8LE United Kingdom<br>Establishment Registration: 3012098706 | |---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Sanmina-SCI Systems De Mexico SA De CV<br>Km 15.5 No. 29, Plant 06<br>Carr. Chapala-Guadalajara Jalisco, MX 45640<br>Establishment Registration: 3006544299 | | Contact: | Marc C. Sanchez, Esq.<br>Contract In-House Counsel and Consultants, LLC<br>53516 Bickett Chapel Hill NC 27517<br>Ph: 202.765.4491<br>E-mail: msanchez@fdaatty.com | ## Date 510(k) Summary Prepared: August 23, 2021 #### Device Information: | Device/Trade Name: | Elvie Stride | |--------------------|-----------------------------| | Common Name: | Powered breast pump | | Regulation Name: | Powered breast pump | | Regulation Number: | 21 CFR 884.5160 | | Product Code: | HGX (Pump, Breast, Powered) | | Regulatory Class: | Class II | | Review Panel: | Obstetrics/Gynecology | #### Predicate Device: | 510(k) Number: | K181863 | |----------------|---------------------------| | Manufacturer: | Chiaro Technology Limited | | Device Name: | Elvie Pump | The predicate device has not been subject to a design-related recall. ## Device Description: Elvie Stride is an electric breast pump that is intended for lactating women to express and collect breast milk. It is intended for a single user. It may be operated as a single or double pumping system. {4}------------------------------------------------ Elvie Stride has adopted the in-bra aesthetic. It is designed to work in the user's nursing bra and has a rechargeable battery so it can be used hands-free without any external power cords. The pump is located off the body and the milk collection bottle is located on the breast. It expresses milk by applying negative pressure to the nipple so that expressed milk may collect in the milk collection bottle. It can be controlled through the physical interface on the pump or through a mobile companion app. Elvie Stride consists of 15 components: Hub, Breast Shield, Cup Front, Diaphragm Seal, Valve, Stopper, Tubing, Tube Splitter, Hub Connector, Cover, Cup Seal, Cap, Cap Seal, Clip and USB Charging Cable. All components are reusable and may be manually cleaned or placed on the top shelf of a dish washer. #### Indications for Use: The Elvie Stride is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Elvie Stride is intended for a single user. #### Comparison of Intended Use and Technological Characteristics: The following table compares the subject device to the predicate device with respect to the indications for use and technological characteristics: | Attribute | Subject Device<br>Elvie Stride<br>K210936 | Predicate Device<br>Elvie Pump<br>K181863 | Comparison | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Indications for Use | The Elvie Stride is a powered<br>breast pump to be used by<br>lactating women to express<br>and collect milk from their<br>breasts. The Elvie Stride is<br>intended for a single user. | The Elvie Pump is a powered<br>breast pump to be used by<br>lactating women to express and<br>collect milk from their breasts.<br>The Elvie Pump is intended for a<br>single user. | Identical | | Intended Use | Express milk from breast | Express milk from breast | Identical | | Single User Device | Yes | Yes | Identical | | Single / Double Pumping | Both | Both | Identical | | Patient Population | Lactating women | Lactating women | Identical | | Environment of Use | Home | Home | Identical | | Over the Counter | Yes | Yes | Identical | | Power Source | USB cable (battery charger)<br>and Li-Ion Battery | USB cable (battery charger) and<br>Li-Polymer Battery | Different | | Adjustable Suction<br>Levels | Yes | Yes | Identical | | Backflow Protection | Yes | Yes | Identical | | Suction Strength | 35-260 mmHg (double)<br>55-300 mmHg (single) | 40-220 mmHg | Different | | Maximum Suction<br>Strength | 300 mmHg | 220 mmHg | Different | | Cycle Speed | 31 to 83 cycles/min<br>(Stimulation)<br>18 to 51 cycles/min<br>(Expression) | 1.1 to 2.0 cycles per/sec<br>(Stimulation)<br>0.6 to 1.0 cycles/sec<br>(Expression) | Different | | Expression levels | 10 | 7 | Different | | Pump Type | Rotary Electric Pump | Piezoelectric Pump | Different | | Two-phase Expression | Yes | Yes | Identical | {5}------------------------------------------------ | User Controls | On pump body and/or through<br>app | On pump body and/or through<br>app | Identical | |------------------------------|------------------------------------|------------------------------------|-----------| | Control Mechanism | Microcontroller | Microcontroller | Identical | | Mobile App | Yes | Yes | Identical | | Milk Quantity<br>Measurement | No | Yes | Different | The subject and predicate device have the same intended use – to express and collect milk from lactating women. The subject and predicate device have different technological features, including the power source, suction strength, pump type, and milk quantity measurement. These technological differences do not raise different questions of safety or effectiveness. ### Summary of Non-Clinical Testing: Non-clinical tests were conducted to verify that the proposed device met all design specifications and that the subject devices are substantially equivalent to the predicate device. - Biocompatibility a. Biocompatibility testing in accordance with the FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020. Testing included the following assessments: - Cytotoxicity per ISO 10993-5:2009 Biological evaluation of medical devices --● Part 5: Tests for in vitro cytotoxicity - . Sensitization per ISO 10993-10:2010 Biological evaluation of medical devices --Part 10: Tests for irritation and skin sensitization - Irritation per ISO 10993-10:2010 Biological evaluation of medical devices -- Part . 10: Tests for irritation and skin sensitization The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing. - b. Electrical Safety, Electromagnetic Compatibility, and Wireless Technology - IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical electrical ● equipment - Part 1: General requirements for basic safety, and essential performance with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012. - IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - . IEC 62133:2012 - Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Documentation provided in accordance with the FDA Guidance document, "Radio Frequency Wireless Technology in Medical Devices" dated August 14, 2013. {6}------------------------------------------------ - Software Verification c. Software verification and validation in accordance with the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005. - d. Performance Testing - Vacuum pressure and cycle rate testing was conducted at all settings and . demonstrated that the device met its specifications. - . Backflow testing was conducted to demonstrate that liquid does not backflow into the tubing/pump. - . Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life. - Battery performance testing was conducted to demonstrate that the battery ● remains functional during its stated battery use-life. - Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life. #### Conclusion: The comparison and analysis above demonstrate that the Elvie Stride is as safe and effective as the predicate device and supports a determination of substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...