The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.
Device Story
System consists of ingestible capsule, external data recorder, locator, and ESView software. Capsule acquires video images of small bowel mucosa during natural GI transit; transmits images via RF to external data recorder. Locator activates capsule and confirms presence in body. Physician uses ESView software to process, analyze, store, and view images; software generates clinical reports and atlas. Device provides non-invasive visualization for detecting small bowel abnormalities. Benefits include diagnostic imaging without invasive endoscopy.
Clinical Evidence
Prospective clinical study (NCT05086471) compared NaviCam to PillCam SB3. Results showed diagnostic Overall Percent Agreement of 89.66% (95% CI: 81.50%, 94.46%) and Kappa of 0.6652 (95% CI: 0.4653, 0.8652).
Technological Characteristics
System includes ingestible capsule, RF data recorder, and ESView software. Capsule features optical imaging, battery (min 8 hours, >57,500 images), and RF transmitter. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62471, and UL 1642. Bench testing verified biting resistance, MTF/angular resolution, temperature safety, pH integrity, image uniformity, frame rate, geometric distortion, FOV, and data integrity.
Indications for Use
Indicated for visualization of small bowel mucosa in adults to detect abnormalities.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
NaviCam Capsule Endoscope System with NaviCam Stomach Capsule (DEN190037)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2022
ANKON Technologies Co., Ltd. % Shoshana Friedman Senior Regulatory Affairs Consultant ProMedoss Inc 3521 Hatwynn Road Charlotte, NC 28269
Re: K221590
> Trade/Device Name: NaviCam Small Bowel Capsule Endoscopy System Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: Class II Product Code: NEZ Dated: November 2, 2022 Received: November 2, 2022
Dear Shoshana Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K221590
Device Name
NaviCam Small Bowel Capsule Endoscopy System
Indications for Use (Describe)
The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized, light blue, rounded shape resembling a 'U' with a small vertical line inside, suggesting a power button symbol. Below the symbol is the word "ankon" in a bold, sans-serif font, with the tagline "INNOVATION FOR BETTER HEALTH" in smaller letters underneath.
### 510(K) SUMMARY NaviCam Small Bowel Capsule Endoscopy System 510(k) NumberK221590
#### SUBMITTER 1.
### Applicant's Name:
ANKON Technologies Co., Ltd. B3-2, B3-3, D3-4 Biolake, No.666, Hi-Tech Road, East Lake New Technology Development Zone, Wuhan, 430075 Hubei, China Phone: +86- 27-87056201 Fax: +86- 27-87705217
### Primary Contact:
Shoshana (Shosh) Friedman Senior Regulatory Affairs Consultant Phone: (704) 430-8695 s.friedman@promedoss.com
#### DATE PREPARED 2.
11/20/2022
#### DEVICE 3.
Trade Name: NaviCam Small Bowel Capsule Endoscopy System
### Classification: Name: System, Imaging, Gastrointestinal, Wireless, Capsule Product Code: NEZ Regulation No: 876.1300 Class: II Classification Panel: Gastroenterology/Urology
#### PREDICATE DEVICES 4.
- · Primary Predicate Device: Given PillCam (SB3) Endoscopy System cleared under K123864 (hereafter referred to as "PillCam SB3").
- · Reference Device: NaviCam Capsule Endoscope System with NaviCam Stomach Capsule (also referred to as "NaviCam Stomach System") granted De-Novo under DEN190037
#### 5. DEVICE DESCRIPTION
The NaviCam Small Bowel Capsule Endoscopy System is an endoscopic capsule imaging system intended to obtain images of the small bowel. It is comprised of the following components:
- [1.] Capsule (AKES-11SW, AKES-11SI): The disposable, ingestible NaviCam Small Bowel Capsule is designed to acquire video images during the natural propulsion
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized, rounded shape in a light blue color, resembling a water droplet or a stylized letter 'U'. Below the symbol, the word "ankon" is written in a modern, sans-serif font. Underneath the company name, the tagline "INNOVATION FOR BETTER HEALTH" is displayed in smaller, uppercase letters.
through the GI tract. The capsule transmits the acquired images via an RF communication channel to the NaviCam Data Recorder located outside the body.
- [2.] Data recorder (AKR-1, AKRI-1): The Data Recorder is an external receiving and recording unit that receives and stores the acquired images from the capsule.
- [3.] ESView Software: The ESView is a software application for processing, analyzing, storing, and viewing the acquired images collected from the NaviCam Data Recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports and a capsule endoscopy atlas.
- [4.] Locator: The Locator is a handheld device that is used to turn the NaviCam Capsule on. It is also used for determining if the capsule is still in the body when the patient is not sure whether he/she expelled it.
#### INDICATIONS FOR USE 6.
The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.
#### SUBSTANTIAL EQUIVALENCE 7.
The NaviCam Small Bowel Capsule Endoscopy System is substantially equivalent to the predicate devices based on the following:
### Indications and Contraindications
The indications for use of the NaviCam Small Bowel Capsule Endoscopy System are included within the indications for use of the PillCam SB3 Capsule Endoscopy System primary predicate device. The contraindications for use of the NaviCam Small Bowel Capsule Endoscopy System are the same as these of the PillCam SB3 Capsule Endoscopy System
### Technology
The capsule, data recorder, and software of the NaviCam Small Bowel Capsule Endoscopy System are substantially equivalent to these of the PillCam SB3 Capsule Endoscopy System primary predicate system.
The capsule, data recorder, and locator of the NaviCam Small Bowel Capsule Endoscopy System are also substantially equivalent to these of the reference device, the NaviCam Stomach System.
### Performance
The performance of the NaviCam Small Bowel Capsule Endoscopy System was compared to this of the PillCam SB3 Capsule Endoscopy System primary predicate device and found to be substantially equivalent.
### Conclusion
The NaviCam Small Bowel Capsule Endoscopy System has substantially similar indications and contraindications, technological, and performance characteristics as these of the predicate and reference devices. Any minor differences that may exist do not affect its safety and effectiveness when used as intended.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Ankon. The logo features a stylized, rounded shape in a gradient of blue, resembling a water droplet or a stylized letter 'U'. Inside the shape is a vertical line. Below the shape is the word "ankon" in a bold, sans-serif font. Underneath the word "ankon" is the phrase "INNOVATION FOR BETTER HEALTH" in a smaller, sans-serif font.
#### 8. PERFORMANCE DATA
The NaviCam Small Bowel Capsule Endoscopy System was tested in accordance with the FDA guidance titled "Ingestible Telemetric Gastrointestinal Capsule Imaging System - Final Class II Special Controls Guidance Document for Industry and FDA".
## Bench/In-Vitro Testing
The NaviCam Small Bowel Capsule Endoscopy System successfully passed the following test:
| Test | Purpose |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biting Test | To demonstrate that the NaviCam SB capsule can withstand applied forces as may occur in case of accidental biting. |
| Angular Resolution Test | To measure the MTF using ISO 12233 slanted edge methodology, and to provide a new angular resolution method using LEDs in the capsule. |
| Temperature Safety Test | To test the temperature change during NaviCam SB capsule operation. |
| pH Test | To evaluate the integrity of the NaviCam SB capsule during exposure to simulated condition of extreme pH levels. |
| Image Intensity Uniformity | To evaluate the intensity uniformity of the NaviCam SB capsule image. |
| Image Frame Rate Test | To demonstrate that the higher frame rate of NaviCam SB capsule image provides good transmission property. |
| Geometric Distortion Test | To determine geometric distortion of the NaviCam SB capsule and provide the local magnification of the image. |
| Field of View (FOV) Test | To determine the FOV value of the NaviCam SB capsule. |
| Battery Life Test | To demonstrate that the battery life of the NaviCam SB capsule lasts at least 8 hours and captures over 57500 images. |
| Image Resolution Test | To test the image resolution of the NaviCam SB capsule. |
| Magnetic Field Test | To measure the maximum value of magnetic flux density on the surface of the NaviCam SB capsule, the magnetic flux density on the non-optical bottom of the capsule, and to determine the magnetic field safety distance of the capsule. |
| DOV Test | To measure the MTF in air and underwater at different distances within the claimed DOV range using ISO 12233 slanted edge methodology and angular resolution method. |
| Color and Gray Scale Test | To evaluate the optical performance of the NaviCam SB capsule. |
| Data Integrity Test | To test the data transmission between the capsule through the data recorder and the ESView software. |
The NaviCam Small Bowel Capsule System was tested for compliance with the following standards and regulations:
- IEC 60601-1:2005+A1:2012 / EN 60601-1:2006 + A1:2013 + A12:2014 ●
- ANSI/AAMI ES60601-1 (2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012) - Amendment 1 - Revision Date 2012/08/21.
- CAN/CSA-C22.2 No. 60601-1:14 Edition 3 Revision Date 2014/03. ●
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized blue "O" shape with a smaller white "i" inside, resembling a power button. Below the symbol is the word "ANKON" in a modern, sans-serif font, and beneath that is the tagline "INNOVATION FOR BETTER HEALTH" in smaller letters.
- IEC 60601-1-11:2015. ANSI/AAMI HA60601-1-11: 2015 ●
- CAN/CSA-C22.2 No. 60601-1-11: 15
- IEC 60601-1-2:2014 ●
- IEC 61000-3-2:2014
- IEC 61000-3-3:2013 ●
- IEC 62471 (1st edition 2006-07) ●
- IEC 62133:2012; IEC 62133-2:2017
- ANSI C63.27-2017
- UL 1642:2020 ●
- . 47 CFR Part 15
The system was found to comply with all applicable requirements of these standards.
### Clinical Data
The NaviCam Small Bowel Capsule Endoscopy System was evaluated in a prospective study, comparing its performance to that of the FDA cleared PillCam SB3 Capsule Endoscopy System (NCT05086471).
In this prospective clinical study, the NaviCam Small Bowel Capsule Endoscopy System and PillCam SB3 Capsule Endoscopy System had a diagnostic Overall Percent Agreement rate of 89.66% (81.50%, 94.46%), and an overall percent agreement Kappa of 0.6652 (0.4653, 0.8652). The results of this comparative study demonstrate that the NaviCam Small Bowel Capsule Endoscopy System performs similarly to the PillCam SB3 Capsule Endoscopy System and provides an effective non-invasive method for the detection of small bowel diseases.
#### 9. CONCLUSION
ANKON Technologies Co., Ltd. believes that the NaviCam Small Bowel Capsule Endoscopy System is substantially equivalent to its primary predicate device, the Given PillCam (SB3) Endoscopy System and its reference device, the NaviCam Stomach System. It has substantially equivalent indications and contraindication, technological characteristics, and performance characteristics to these of the primary and reference systems, and therefore does not introduce any new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.