GIVEN PILLCAM ENDOSCOPY SYSTEM WITH RAPID 8.0, GIVEN PILLCAM SB 3 CAPSULE ENDOSCOPY SYSTEM

K123864 · Given Imaging , Ltd. · NEZ · Aug 12, 2013 · Gastroenterology, Urology

Device Facts

Record IDK123864
Device NameGIVEN PILLCAM ENDOSCOPY SYSTEM WITH RAPID 8.0, GIVEN PILLCAM SB 3 CAPSULE ENDOSCOPY SYSTEM
ApplicantGiven Imaging , Ltd.
Product CodeNEZ · Gastroenterology, Urology
Decision DateAug 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1300
Device ClassClass 2
AttributesPediatric

Indications for Use

With PillCam SB Capsule The PillCam SB capsule is intended for visualization of the small bowel mucosa. - It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam SB capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age. With PillCam ESO Capsule The PillCam capsule endoscopy system with PillCam ESO capsules is intended for the visualization of esophageal mucosa.

Device Story

System comprises ingestible PillCam capsule, DR 3 recorder, RAPID 8.0 software, and workstation. Capsule acquires video images during GI tract transit; transmits via RF to external DR 3 recorder. RAPID 8.0 software processes, analyzes, and stores images; creates RAPID video and clinical reports. Used in clinical settings by healthcare professionals. Features include Suspected Blood Indicator (SBI) to mark potential blood/red areas, adaptive frame rate (AFR) for improved coverage in fast-moving segments (e.g., duodenum), and bi-directional communication. Real-time viewing enabled via handheld device. Output assists clinicians in detecting small bowel abnormalities, aiding diagnosis of Crohn's, obscure bleeding, and iron deficiency anemia.

Clinical Evidence

System validated with clinical data. Bench testing included field of view, battery life, wireless communication, color reproduction, image quality, frame rate verification, and optical resolution. Complies with FDA guidance for ingestible telemetric GI capsule imaging systems.

Technological Characteristics

Ingestible telemetric capsule imaging system. Components: disposable capsule, external DR 3 recorder, workstation, sensor array/belt. Features: RF communication, adaptive frame rate (AFR), bi-directional communication, real-time SB detection. Software: RAPID 8.0. Validated via bench testing (FOV, battery, wireless, optical resolution).

Indications for Use

Indicated for visualization of small bowel mucosa (SB capsule) or esophageal mucosa (ESO capsule) in adults and children ≥2 years. Used for monitoring/detection of Crohn's disease, obscure GI bleeding, or iron deficiency anemia not identified by standard endoscopy.

Regulatory Classification

Identification

An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Given Imaging. The logo consists of the word "GIVEN" in large, bold letters, with a stylized image of a capsule between the "I" and "E". Below the word "GIVEN" is the word "IMAGING" in smaller, bold letters. Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466 # 510(k) Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92. | Submitter Name<br>and Address: | Given Imaging Ltd.<br>Hermon Building<br>New Industrial Park<br>PO Box 258<br>Yokneam 20692<br>Israel<br>Tel.: 011-972-4-9097936<br>Fax: 011-972-4-9938060 | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Tim Thomas<br>SVP, Regulatory, Clinical, and Quality<br>Email: tim.thomas@givenimaging.com | | Phone Number: | 770-662-0870 ext. 1006 | | Fax Number: | 770-662-0510 | | Establishment<br>Registration<br>Number: | 9710107 | | Date Prepared: | August 9, 2013 | | Device Trade<br>Name(s): | Given® PillCam® SB 3 capsule endoscopy system<br>Given® PillCam® endoscopy system with RAPID® 8.0 | | Device Common<br>Name: | Ingestible telemetric gastrointestinal capsule imaging system | | Classification: | Regulation No: 876.1300, Class: II<br>Panel: Gastroenterology/Urology<br>NEZ -- System, Imaging, Gastrointestinal, Wireless, Capsule | | Predicate<br>Device(s): | • Given PillCam Platform with PillCam RAPID 6.5 (K103025)<br>• Given PillCam Platform System with PillCam SB Capsules<br>(K101250) | {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for Given Imaging. The logo consists of the text "GI-EN" in a stylized font, with a small registered trademark symbol next to the "N". Below the text is an image of a medical device, followed by the word "IMAGING" in block letters. The logo is simple and professional, and it is likely used to represent a medical imaging company. Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 0692 Israel ice 972 4 909 777 Fax 972 4 959 2466 General Device Description: The Given PillCam Endoscopy system is comprised of four main subsystems: (1) the ingestible PillCam capsule, (2) the DR 3 PillCam Recorder, (3) the RAPID software, and (4) the Given Workstation and Accessories. - Ingestible PillCam Capsule 1. The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the PillCam Recorder located outside the body. - 2. DR 3 PillCam Recorder The DR 3 PillCam Recorder is and external receiving/recording unit that receives and stores the acquired images from the capsule. - 3. RAPID Software The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the PillCam Recorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, inservice training videos, and patient instruction forms. RAPID 8.0 supports PillCam capsule endoscopy of the GI tract with all PillCam video capsules. In addition, RAPID 8.0 supports PillCam Recorder DR 2C and PillCam Recorder DR 3. - 4. Given Workstation and Accessories The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. The PillCam Recorder is an external receiving/recording unit that receives acquired images from the capsule. The Sensor Array or Sensor Belt receives data from the PillCam capsule and transfers the data to the PillCam Recorder. The RAPID Real Time is a handheld device that allows for real-time viewing of acquired images through the GI tract. Other accessories include a flat panel LCD monitor, a highcapacity mass storage device, and a high-capacity USB portable storage device. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for Given Imaging. The logo consists of the word "GIVEN" in large, bold letters, with a stylized image of a capsule between the "I" and "E". Below the word "GIVEN" is the word "IMAGING" in smaller, bolder letters. The logo is simple and modern, and it is likely used to represent a medical imaging company. Given Imaging Limited New Industrial Park PO Box 258, Yogneam 20602 Israel Voice 972 4 909 7777 Fax 972 4 959 2466 Indication for use: # 1. Given PillCam endoscopy system with RAPID 8.0 ## With PillCam SB Capsule The PillCam SB capsule is intended for visualization of the small bowel mucosa. - · It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of . lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of . lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam SB capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age. # With PillCam ESO Capsule The PillCam ESO capsule is intended for the visualization of esophageal mucosa. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Given Imaging. The logo consists of the word "GIVEN" in large, stylized letters, with a graphic of a capsule endoscopy device between the "I" and "E". Below the word "GIVEN" is the word "IMAGING" in smaller letters. Given Imaging Limited New Industrial Park PO Box 258, Yoqneam 0602 Israe Voice 972 4 909 777 Fax 972 4 959 2466 The technological characteristics are similar to the predicate Technological devices, except for the differences that are listed in Section 12 of Characteristics: this submission. However, it may be concluded from the Substantial Equivalence Summary that none of the presented differences raise any new safety or efficacy issues. > The primary improved features of PillCam SB 3 capsules are the capsule optical spatial resolution, to achieve better image quality, and higher effective and adaptive frame rate (AFR), to achieve better coverage of small bowel tissue in segments at which the capsule moves fast such as in the duodenum. The main new features implemented in PillCam recorder DR 3, are support of the new introduced SB 3 capsule and support of its functionalities such as: Real-time SB detection, Bi-directional communication and Adaptive Frame Rate (AFR) activation. The safety of PillCam SB 3 capsules was validated and no safety or efficacy issues have been raised. It may be concluded that the new features do not raise new safety issues. - Since PillCam SB 3 capsule and the predicate device are same Bench testing: capsules in terms of external components and technology. The biting test and pH resistance test for the predicate device are applicable for PillCam SB 3 capsules as well. Additional bench testing included field of view (FOV), battery life, wireless communication, color reproduction, image quality, frame rate verification, and optical resolution. - The devices meet the guidance entitled "Class II Special Performance Data: Guidance Document: Ingestible Telemetric Controls Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA dated November 28, 2001. The Given PillCam endoscopy system with RAPID 8.0 was validated with clinical data. The Advanced A-Mode software feature was also validated with clinical data. The proposed changes in this submission do not raise new performance or safety issues. - Conclusion: Based on the technological characteristics of the devices, Given Imaging Ltd. believes that the Given PillCam endoscopy system with RAPID 8.0 and Given PillCam SB 3 capsule endoscopy system and the predicate devices selected are substantially equivalent and do not raise new issues of safety or effectiveness. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized emblem with three curved lines, resembling a person with outstretched arms. The emblem is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WQ66-G609 Silver Spring, MD 20993-0002 August 12, 2013 Given Imaging, Ltd. % Tim Thomas Senior Vice President Hermon Building New Industrial Park, P.O. Box 258 Yokneam 20692 Israel Re: K123864 Trade/Device Name: Given PillCam endoscopy system with RAPID 8.0 Regulation Number: 21 CFR§ 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: July 11, 2013 Received: July 15, 2013 Dear Tim Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {6}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## INDICATIONS FOR USE #### 510(k) Number : K123864 Device Name: Given PillCam endoscopy system with RAPID 8.0 #### Indications for Use: #### With PillCam SB Capsule The PillCam SB capsule is intended for visualization of the small bowel mucosa. - It may be used in the visualization and monitoring of lesions that may indicate . Crohn's disease not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be a source of ● obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be potential . causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam SB capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age. ### With PillCam ESO Capsule The PillCam capsule endoscopy system with PillCam ESO capsules is intended for the visualization of esophageal mucosa. Prescription Use ______________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH, Office of Device Evaluation (ODE) Herbert P. Lerner -S (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123864 510(k) Number_ Given Imaging Ltd. K123864 - Response Letter PillCam® SB 3 capsule endoscopy system Supplement #1 April 28, 2013
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...