K101250 · Given Imaging , Ltd. · NEZ · Mar 7, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K101250
Device Name
GIVEN PILLCAM PLATFORM WITH PILLCAM SB CAPSULES
Applicant
Given Imaging , Ltd.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
Mar 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PillCam Platform with a PillCam SB capsules is intended for visualization of the small bowel mucosa. The PillCam Platform with PillCam SB capsules may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. The PillCam Platform with PillCam SB capsules may be used in the visualization of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. The PillCam Platform with PillCam SB capsules may be used in the visualization of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam Platform with PillCam SB capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.
Device Story
System consists of ingestible PillCam capsule, external DataRecorder, and RAPID software on a workstation. Capsule acquires GI tract video images during natural propulsion; transmits via RF to external SensorArray and DataRecorder. RAPID software processes, analyzes, stores, and displays images; generates clinical reports and patient forms. RAPID Real Time handheld device allows live viewing. Used in clinical settings by healthcare providers to visualize small bowel mucosa; aids in detecting lesions, Crohn's disease, obscure bleeding, and iron deficiency anemia. Output assists clinicians in diagnostic decision-making; provides non-invasive alternative to traditional endoscopy.
Clinical Evidence
Clinical data summarized to demonstrate safety and effectiveness for the proposed indications. Device meets requirements of the 'Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System'.
Technological Characteristics
System comprises ingestible capsule, RF-based data transmission, external SensorArray, DataRecorder, and workstation-based software. Operates as a telemetric GI imaging system. Complies with Class II Special Controls Guidance for Ingestible Telemetric Gastrointestinal Capsule Imaging Systems.
Indications for Use
Indicated for visualization of small bowel mucosa in adults and children (≥2 years) to detect abnormalities, monitor Crohn's disease, identify sources of obscure GI bleeding, or investigate iron deficiency anemia when upper/lower endoscopy is non-diagnostic.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
Predicate Devices
Given® Diagnostic System with PillCam® SB2 Capsule (K090557)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for ULTEN IMAGING. The logo consists of the word "ULTEN" in large, bold, sans-serif font, with a stylized image of a camera lens or optical device between the "L" and "T". Below the word "ULTEN" is the word "IMAGING" in a smaller, sans-serif font. The logo is black and white.
K101250 Pgz/
MAR - 7 2011
New Industrial Park PO Box 258, Yoqneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
| Submitter Name and<br>Address: | Given Imaging Ltd.<br>Hermon Building<br>New Industrial Park<br>PO Box 258<br>Yokneam 20692<br>Israel<br>Tel.: 011-972-4-9097936<br>Fax: 011-972-4-9938060 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tim Thomas<br>Vice President,<br>Regulatory Affairs and Quality Assurance<br>Email: tim.thomas@givenimaging.com |
| Phone Number: | 770-662-0870 ext. 1006 |
| Fax Number: | 770-662-0510 |
| Establishment<br>Registration Number: | 9710107 |
| Date Prepared: | May 3, 2010 |
| Device Trade Name(s): | Given PillCam® Platform with PillCam® SB Capsules |
| Device Common Name: | Ingestible telemetric gastrointestinal capsule imaging system |
| Classification: | Regulation No: 876.1300<br>Class: II<br>Panel: Gastroenterology<br>NEZ - System, Imaging, Gastrointestinal, Wireless, Capsule |
| Predicate Device(s): | Given® Diagnostic System with PillCam® SB2 Capsule<br>(K090557) |
| General Device<br>Description: | The Given PillCam® Platform is comprised of three main<br>subsystems; (1) the ingestible PillCam capsule, (2) the<br>RAPID® software, and (3) the Given® Workstation and<br>Hardware. |
Given Imaging Ltd. New Industrial Park, PO Box 258, Yokneam 20692, Israel
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Given Imaging. The logo consists of the word "GIVEN" in large, bold letters, with a stylized image of a goblet or chalice in between the "I" and the "E". Below the word "GIVEN" is the word "IMAGING" in smaller, bold letters. The logo is simple and modern, and the use of the goblet or chalice suggests a connection to medicine or health.
K101250
Given Imaging Limited New Industrial Park PO Box 258, Yogneam 20692 Israel Voice 972 4 909 7777 Fax 972 4 959 2466
#### Ingestible PillCam Capsule 1.
The disposable, ingestible PillCam Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via a RF communication channel to the DataRecorder located outside the body.
#### 2 RAPID Software
The RAPID Software is a software application that is utilized to process, analyze, store, and view the acquired images collected from the DataRecorder to create a RAPID video of the images. The software also includes a reporting function to create detailed clinical reports, in-service training videos, and patient instruction forms.
#### 3 Given Workstation and Hardware
The Workstation is a modified standard personal computer that is the operational platform for the RAPID software. The DataRecorder is an external receiving/recording unit that receives acquired images from the capsule. The SensorArray receives data from the PillCam capsule and transfers the data to the DataRecorder. The RAPID Real Time is a handheld device that allows for real-time viewing of acquired images through the GI tract. Other accessories include a flat panel LCD monitor, a high-capacity mass storage device, and a high-capacity USB portable storage device.
Given Imaging Ltd. New Industrial Park, PO Box 258, Yokneam 20692, Israel
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "ULTEN IMAGING". The logo features the word "ULTEN" in bold, sans-serif font, with a stylized image of a lightbulb or lamp between the letters "L" and "T". Below "ULTEN", the word "IMAGING" is printed in a similar font, but slightly smaller in size.
K101250 Page 39
uren maging limited
Intended Use: SB Indications for Use: The PillCam Platform with a PillCam SB capsules is intended for visualization of the small bowel mucosa. The PillCam Platform with PillCam SB capsules may . be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. The PillCam Platform with PillCam SB capsules may . be used in the visualization of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. The PillCam Platform with PillCam SB capsules may . be used in the visualization of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas. The PillCam Platform with PillCam SB capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age. Technological The technology characteristics are exactly the same as the Characteristics: predicate devices. Performance Data: The devices meet the guidance entitled "Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA dated November 28, 2001. Clinical data has been summarized to show safety and effectiveness for the proposed indication for use. Conclusion: Based on the technological characteristics and clinical performance of the devices, Given Imaging Ltd. believes that the Given PillCam® Platform with PillCam® SB Capsules and the predicate devices selected are substantially equivalent and do not raise new issues of safety or effectiveness.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an abstract caduceus or a bird-like figure with three curved lines, representing health and human services. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and is evenly spaced around the circle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Tim Thomas Vice President, Regulatory Affairs & Quality Assurance Given Imaging Ltd. New Industrial Park, Hermon Building P.O. Box 258 Yogneam 20692 ISRAEL
MAR - 7 2011
Re: K101250
Trade/Device Name: Given PillCam® Platform with PillCam® SB Capsules Regulation Number: 21 CFR 8876.1300 Regulation Name: Ingestible telemetric gastro-intestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: February 26, 2011 Received: March 1, 2011
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability . warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Stubert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### INDICATIONS FOR USE
| 510(k) Number (if known): | K101250 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: Given PillCam® Platform with PillCam® SB Capsules
Indications for Use:
The PillCam Platform with a PillCam SB capsules is intended for visualization of the small bowel mucosa.
- The PillCam Platform with PillCam SB capsules may be used in the visualization and . monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
- The PillCam Platform with PillCam SB capsules may be used in the visualization of . lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
- The PillCam Platform with PillCam SB capsules may be used in the visualization of ● lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
The Suspected Blood Indicator (SBI) feature is intended to mark frames of the video suspected of containing blood or red areas.
The PillCam Platform with PillCam SB capsules may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults and children from two years of age.
Prescription Use AND/OR × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Concurrence of CDRH
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K101250
Given Imaging Ltd. 510(k) Submission Given PillCam® Platform with PillCam® SB Capsules February 26, 2011
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.