Smart Correction System Rings and Compatible HA-Coated Half Pins
K221366 · Wishbone Medical, Inc. · KTT · Jun 6, 2022 · Orthopedic
Device Facts
Record ID
K221366
Device Name
Smart Correction System Rings and Compatible HA-Coated Half Pins
Applicant
Wishbone Medical, Inc.
Product Code
KTT · Orthopedic
Decision Date
Jun 6, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The WishBone Medical Smart Correction System is intended for use in pediatric subgroups (except newborns) and adult patients for the treatment of open and closed fractures, arthrodesis and pseudoarthrosis of long bones, limb lengthening, deformity and angular correction, bony or soft tissue defects, and joint contractures. The Smart Correction System is indicated for pediatric subpopulations (excluding newborns) and adults for the following: Joint contracture resulting in loss of range of motion. Fractures and disease which generally may result in joint contractures or loss of range of motion. Fractures requiring distraction. Open and closed fracture fixation, including fractures of long bones (intracapsular, intertrochanteric, supracondylar, condylar). Correction of bony or soft tissue defects. Correction of bony or soft tissue deformities. Joint arthrodesis. Infected fractures or nonunion. Limb Lengthening by epiphyseal or metaphyseal distraction. Pseudoarthrosis of long bones.
Device Story
Multilateral hexapod circular external fixator; stabilizes/aligns fractured bones or corrects congenital deformities. System comprises rings connected by six independent telescopic struts; allows six-axis adjustment. Anchored to bone via crossed tensioned wires and half-pins. Includes web-based software for preoperative planning and postoperative adjustment scheduling. Surgeon inputs radiographic images and frame parameters into software; software calculates adjustment schedule to guide osteogenesis. Used in clinical settings by orthopedic surgeons. Benefits include controlled, steady bone correction and stabilization.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including compressive stiffness, construct stiffness, and fatigue/load-to-failure testing per ASTM 1541-17, alongside biocompatibility and sterilization validation.
Indicated for pediatric (excluding newborns) and adult patients requiring treatment for joint contractures, fractures (open/closed, long bone, requiring distraction), bony/soft tissue defects or deformities, joint arthrodesis, infected fractures/nonunion, limb lengthening, and pseudoarthrosis.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
ST.A.R. 90 F4 External Fixation Screws with Hydroxyapatite (K150661)
Submission Summary (Full Text)
{0}------------------------------------------------
June 6, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
WishBone Medical, Inc. Kellie Myers Regulatory Affairs Manager 100 Capital Drive Warsaw, Indiana 46582
Re: K221366
Trade/Device Name: Smart Correction System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT Dated: May 11, 2022 Received: May 12, 2022
Dear Kellie Myers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221366
Device Name Smart Correction System
#### Indications for Use (Describe)
- The Smart Correction System is indicated for pediatric subpopulations (excluding newborns) and adults for the following: · Joint contracture resulting in loss of range of motion.
- · Fractures and disease which generally may result in joint contractures or loss of range of motion.
- · Fractures requiring distraction.
- · Open and closed fracture fixation, including fractures of long bones (intracapsular, intertrochanteric, supracondylar, condylar).
- · Correction of bony or soft tissue defects.
- · Correction of bony or soft tissue deformities.
- · Joint arthrodesis.
- · Infected fractures or nonunion.
- · Limb Lengthening by epiphyseal or metaphyseal distraction.
- · Pseudoarthrosis of long bones.
| Type of Use (Select one or both, as applicable) | <div> <input type="checkbox"/> <input type="checkbox"/> </div> |
|-------------------------------------------------|----------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------|
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for WishBone Medical. The logo consists of a blue circle with a white wishbone inside, followed by the text "WishBone" in a sans-serif font. Below "WishBone" is the word "MEDICAL" in a smaller, sans-serif font.
In accordance with 21 CFR \$807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the WishBone Medical Smart Correction System Small Rings and compatible HA-Coated Half-Pins Special 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document The Special 510(k) Program, issued September 13, 2019.
| SUBMITTER INFORMATION | |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | WishBone Medical, Inc. |
| Address | 100 Capital Drive<br>Warsaw, IN 46582 |
| Phone Number | (574)306-4006 |
| Establishment Registration Number | 3013680140 |
| Name of Contact Person | Kellie Myers |
| Date Prepared | June 6, 2022 |
| NAME OF DEVICE | |
| Trade or Proprietary Name | Smart Correction System |
| Common or usual name | External Skeletal Fixation Device, |
| Classification Name | Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, |
| Regulatory Classification | II |
| 510(k) Review Panel | Orthopedic Devices (OHT6) |
| Regulation | 888.3030 Single/multiple component metallic bone fixation appliances and<br>accessories. |
| Product Code(s) | KTT |
| Legally marketed device to<br>which equivalence is claimed | Smart Correction System, (K193368) |
| Reference Device | ST.A.R. 90 F4 External Fixation Screws with Hydroxyapatite (K150661) |
| Device Description | The WishBone Medical Smart Correction System: a multilateral hexapod<br>circular external fixator device used to stabilize and maintain alignment of<br>complicated fractured bones, soft tissues and/or congenital deformity<br>repairs of an extremity. The basic system consists of a minimum of two rings<br>connected by six (6) telescopic struts that are lengthened and shortened<br>independently. The struts' independent motion allows the surgeon to adjust<br>the position of the proximal and distal ring. The system allows for<br>movement in six different axes to correct difficult trauma extremity<br>situations and/or congenital limb deformity correction. The Smart<br>Correction System capitalizes on the body's natural ability of osteogenesis,<br>guiding the orientation and position of this new bone to the desired corrected<br>location in a steady controlled fashion. In addition to the hardware, the<br>Smart Correction System has a web-based software treatment planning tool<br>with Radiographic Navigation. The surgeon enters data from direct<br>examination, radiographic images and the fixator parameters into the |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image contains the logo for WishBone Medical. The logo consists of a blue circle with a white wishbone inside of it on the left. To the right of the circle is the company name, "WishBone" in a larger font and "MEDICAL" in a smaller font below it.
software. The software is used preoperatively to plan the reconstruction/correction and identify the frame construction. Post operatively, the surgeon enters the X-ray images and the current frame parameters to establish an adjustment schedule for the patient during the healing process.
The Smart Correction System is modular and facilitates a multitude of frame configurations to serve a wide variety of patient needs. Listed below are the high-level components and accessories:
- . The fixator bridge is constructed of two (2) or more ring components, and each ring component is connected to another via six (6) telescopic struts. Full, 2/3, and 1/3 ring components are available, along with standard and rapid adjust struts in multiple lengths. Femoral arches and threaded rods are used as needed to provide added frame stability. Rings and femoral arches are manufactured from aluminum material; struts from titanium, stainless steel, and aluminum; and threaded rods are made from titanium material.
- . The fixator bridge is anchored to the patient's bone by crossed tensioned wires and half pins that are secured to the rings by connector elements (wire clamps, pin clamps, cubes, bolts, nuts, washers, and twisted plates). Standard, olive wires, and threaded wires are available, as well as multiple diameters and styles of half pins. Pins and wires may also be used to secure fragments of bone and are made from stainless steel and titanium. Connector elements are manufactured out of titanium material. This submission also includes a line extension to add HA-coated titanium Half Pin components.
- . A foot ring is available and connected to the distal ring when a procedure such as ankle arthrodesis is performed. The foot rings are manufactured out of aluminum material.
- . Patient comfort accessories are also included: strut ID bands (polycarbonate), foot walking attachment (POM-C), and pin/wire caps (silicone, PVC) are also included.
- . The Smart Correction System includes reusable surgical instruments to facilitate surgical assembly of the fixator construct. The non-sterile implants and other fixator elements are contained within sterilization cases, along with the reusable instruments Intended Use of the Device The WishBone Medical Smart Correction System is intended for use in pediatric subgroups (except newborns) and adult patients for the treatment of open and closed fractures, arthrodesis and pseudoarthrosis of long bones, limb lengthening, deformity and angular correction, bony or soft
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for WishBone Medical. The logo consists of a blue circle with a white wishbone inside on the left. To the right of the circle is the text "WishBone" in a sans-serif font, with the word "MEDICAL" in smaller letters below.
| Indications for Use | radiographic measurements, the software provides a schedule of<br>adjustments for the fixator frame.<br>The Smart Correction System is indicated for pediatric subpopulations<br>(excluding newborns) and adults for the following:<br>Joint contracture resulting in loss of range of motion. Fractures and disease which generally may result in joint contractures<br>or loss of range of motion. Fractures requiring distraction. Open and closed fracture fixation, including fractures of long bones<br>(intracapsular, intertrochanteric, supracondylar, condylar). Correction of bony or soft tissue defects. Correction of bony or soft tissue deformities. Joint arthrodesis. Infected fractures or nonunion. Limb Lengthening by epiphyseal or metaphyseal distraction. Pseudoarthrosis of long bones. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE
The rationale for substantial equivalence is based on consideration of the following characteristics:
- a. Intended Use: The subject device and predicate systems have the same intended use. No new or increased risks are identified.
- b. Indications for Use: The subject device and predicate systems have the same intended use. No new or increased risks are identified
- c. Materials: The subject device is manufactured from similar materials. No new or increased risks have been identified.
- d. Design Features: The subject device design is similar to the predicates. No new or increased risks are identified.
- e. Sterilization: The Smart Correction System is supplied nonsterile for the end user to sterilize in the provided sterilization cases. While the predicate system offers some pins sterile, there are pins that are provided nonsterile. The subject system has the same sterilization method for similar elements as the predicate. Therefore, no new or increased risks have been identified.
# PERFORMANCE DATA NON-CLINICAL TESTING
Engineering analyses were conducted in compliance with ASTM 1541-17: Standard Specification and Test Methods for External Skeletal Fixation Devices. Evaluations conducted include:
- Ring Compressive Stiffness Justification
- Construct Stiffness Test Justification
- Construct Fatigue & Load to Failure Test Justification
Engineering analyses were also conducted for the following:
- . Clinical Cleaning & Sterilization Validation Justification
- Cleaning for Biocompatibility Test Justification
- Biocompatibility Assessment Justification
# CLINICAL TESTING
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for WishBone Medical. The logo consists of a blue circle with a white wishbone inside, followed by the text "WishBone" in a sans-serif font. Below the word "WishBone" is the word "MEDICAL" in a smaller, sans-serif font.
Clinical testing was not deemed necessary to demonstrate substantial equivalence.
CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject device has the same intended use and indications as the predicate Smart Correction System (K193368). It also has similar technological characteristics as the predicate device, and the performance data and analyses demonstrate that any differences do not raise new questions of safety and effectiveness. Therefore, we conclude that the proposed device is at least as safe and effective and performs as well or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.