VS 8 Vital Signs Monitor, VS 8A Vital Signs Monitor, VS 9 Vital Signs Monitor

K221267 · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · MWI · Dec 15, 2022 · Cardiovascular

Device Facts

Record IDK221267
Device NameVS 8 Vital Signs Monitor, VS 8A Vital Signs Monitor, VS 9 Vital Signs Monitor
ApplicantShenzhen Mindray Bio-Meidcal Electronics Co., Ltd.
Product CodeMWI · Cardiovascular
Decision DateDec 15, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Indications for Use

The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8/VS 8A Vital Signs Monitor are not intended for helicopter transport, hospital ambulance, or home use.

Device Story

VS Series Vital Signs Monitors (VS 9, VS 8, VS 8A) are bedside physiological monitors for healthcare facilities. Inputs include temperature (via TrueTemp/SmarTemp modules), SpO2 (via Mindray, Masimo, or Nellcor modules), NIBP (oscillometric), and CO2. Device processes signals to display, store, and alarm on patient vitals. Operated by trained clinicians; output viewed on integrated touchscreen. NIBP cuff also supports venipuncture. Provides real-time monitoring to assist clinical decision-making and patient assessment. Connectivity includes WiFi and Bluetooth for data transfer.

Clinical Evidence

Clinical accuracy study for the TrueTemp module conducted per ISO 80601-2-56:2017/AMD1:2018. Evaluated 106 oral, 110 axillary, and 109 rectal temperature cases across infants, children, and adults (ages 4 days to 67 years). Compared against WelchAllyn SureTemp PLUS 690. Results: Oral bias 0.03°C (LA 0.37°C), Axilla bias 0.03°C (LA 0.32°C), Rectal bias -0.06°C (LA 0.38°C). Study confirmed equivalence to comparator; no adverse reactions reported.

Technological Characteristics

Materials: TrueTemp probe covers (biocompatible). Sensing: Thermal resistance (Temp), Oscillometric (NIBP), SpO2/CO2 modules. Power: Rechargeable Li-ion battery or AC. Connectivity: WiFi (IEEE 802.11 a/b/g/n/ac), Bluetooth 5. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, ISO 80601-2-56, ASTM E1112, ASTM E1104. Software: Verification/validation per FDA guidance.

Indications for Use

Indicated for monitoring, displaying, storing, and alarming of Temp, SpO2, PR, NIBP, and CO2 in adult, pediatric, and neonatal patients. For use in healthcare facilities by trained clinical professionals. Not for home, ambulance, or helicopter use.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 15, 2022 Shenzhen Mindray Bio-Medcal Electronics Co., LTD Li Lei Manager Regulatory Affairs, Technical Regulation Department Mindrav Building, Keii 12th Road South, Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China Re: K221267 Trade/Device Name: VS 8 Vital Signs Monitor, VS 8A Vital Signs Monitor, VS 9 Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, CCK, DQA, DXN, FLL Dated: April 19, 2022 Received: May 2, 2022 Dear Li Lei: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Robert T. Kazmierski -S for LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K221267 Device Name VS Series Vital Signs Monitors ### Indications for Use (Describe) The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, storing, alarming, and transfering of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8/VS 8A Vital Signs Monitor are not intended for helicopter transport, hospital ambulance, or home use. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;"> <span style="color: black;">☑</span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | | | <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;"> <span style="color: black;">☐</span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for the VS series Patient Monitors is provided below. #### 1. SUBMITTER | Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 | |------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Contact Person: Li Lei<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885953<br>Fax: +86 755 26582680<br>E-mail: lilei.js@mindray.com | - Date Prepared: April 19, 2022 #### 2. DEVICE | Device Trade Name: | VS Series Vital Signs Monitors (Including VS 9/VS<br>8/VS 8A) | |---------------------------------------|-----------------------------------------------------------------------------------| | Device Common Name: | Vital Signs Monitor | | Classification Name and<br>Regulation | 21 CFR 870.2300, Cardiac monitor (including<br>cardiotachometer and rate alarm) | | Primary Product Code: | MWI - Monitor, physiological, patient (without<br>arrhythmia detection or alarms) | | Regulatory Class | Class II | | Panel | Cardiovascular | {4}------------------------------------------------ | Product<br>Code | Regulation<br>Number | Panel | Regulation description | Device Common Name | |-----------------|----------------------|----------------|-----------------------------------------------------|----------------------------------------------------------| | CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas<br>analyzer | analyzer, gas, carbon-<br>dioxide, gaseous-phase | | DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | oximeter | | DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood<br>pressure measurement<br>system | system, measurement,<br>blood-pressure, non-<br>invasive | | FLL | 21 CFR<br>880.2910 | Cardiovascular | Clinical electronic<br>thermometer | thermometer, electronic,<br>clinical | #### Table 1: Secondary Product Codes #### 3. PREDICATE DEVICE K211475 - VS Series Vital Signs Monitor (SHENZHEN MINDRAY Predicate Device: BIO-MEDICAL ELECTRONICS CO., LTD) #### 4. DEVICE DESCRIPTION The subject VS series Vital Signs Monitors includes three monitors: - VS 9 Vital Signs Monitor - VS 8 Vital Signs Monitor ● - VS 8A Vital Signs Monitor . The VS series Vital Signs Monitors are for use for adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. The monitors should only be used by persons who have received adequate training in their use. #### INTENDED USE/INDICATIONS FOR USE 5. The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8/VS 8A Vital Signs Monitor are not intended for helicopter transport, hospital ambulance, or home use. {5}------------------------------------------------ # Page 3 of 9 #### SUBSTANTIAL EQUIVALENCE 6. ### Comparison of Indications Indications of the predicate device and the subject device are the same. ### Technological Comparisons Table 2 compares the key technological feature of the subject device to the predicate device (VS series Vital Signs Monitor, K211475). The features in gray are features that have been modified since their previous clearances and that are the subject of this 510(k). | Feature | Predicate Devices (K211475) | | | Subject Devices | | | |---------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------|-----------------|------|-------| | | VS 9 | VS 8 | VS 8A | VS 9 | VS 8 | VS 8A | | Integrated<br>display and<br>touchscreen | 10.1" 1280*800 pixels | 8"<br>pixels | 1024*768 | Same | Same | | | Touch screen | Capacitive touch screen | | | Same | | | | Function<br>Button And<br>Rotary<br>Encoder | Not supported | Support | | Same | Same | | | Power supply | One or two rechargeable Lithium-ion battery or<br>AC power supply | | | Same | | | | Battery | Chargeable Smart Lithium-<br>Ion, 10.8 VDC, 5600 mAh | Chargeable non-<br>Smart Lithium-<br>Ion, 10.95 VDC,<br>2500 mAh<br>or<br>Chargeable non-<br>Smart Lithium-<br>Ion, 10.95 VDC,<br>5000 mAh | | Same | Same | | | Data<br>Recorder | The thermal recorder can be used to print patient<br>information, measurement numerics, and<br>waveforms. | | | Same | | | | Speaker | Provide alarm tones (45 to 85 dB), key tones;<br>support PITCH TONE and multi-level tone<br>modulation | | | Same | | | | WiFi | 2.4G/5G dual band WiFi, compatible with IEEE<br>802.11 a/b/g/n/ac | | | Same | | | | Bluetooth | Bluetooth 5 | | | Same | | | Table 2: Technological Comparison {6}------------------------------------------------ | Feature | Predicate Devices (K211475) | | | Subject Devices | | | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|--------------------------------| | | VS 9 | VS 8 | VS 8A | VS 9 | VS 8 | VS 8A | | Mindray<br>Temperature<br>Module | Mindray Temp Module (SmarTemp):<br>Technique: Thermal resistance<br>Measurement range:<br>Monitor mode:25 to 44 °C (77 to 111.2 °F)<br>Predictive mode: 34 to 43 °C (93.2 to 109.4 °F)<br>Accuracy (Monitor mode):<br>25 to 32°C(not include 32°C): ± 0.2 °C<br>32 to 44°C(include 32°C): ±0.1 °C(± 0.2 °F)<br>or<br>77 to 89.6 °F (not include 89.6°F): ± 0.4 °F<br>89.6 to 111.2 °F (include 89.6°F): ± 0.2 °F | | | Similar.<br>The performance of the module is validated by the<br>clinical trial.<br>New Mindray Temp Module (TrueTemp):<br>Technique: Thermal resistance<br>Monitor mode:25 to 44 °C (77 to 111.2 °F)<br>Predictive mode: 34 to 42 °C (93.2 to 107.6 °F)<br>Accuracy (Monitor mode):<br>25 to 44°C: ±0.1 °C(± 0.2 °F)<br>Or 77 to 111.2 °F: ± 0.2 °F | | | | | Statistical Results of Clinical Investigation<br>Data(Predictive mode) | | | Statistical Results of Clinical Investigation<br>Data(Predictive mode) | | | | | Clinical BIAS<br>(△cb) | Limits of Agreement<br>(LA) | Clinical Repeatability<br>(στ) | Clinical BIAS<br>(Acb) | Limits of Agreement<br>(LA) | Clinical Repeatability<br>(στ) | | | Oral<br>0.02°C | 0.33°C | 0.1°C | Oral<br>0.03°C | 0.37°C | 0.14°C | | | Axilla<br>0.06°C | 0.38°C | 0.13°C | Axilla<br>0.03°C | 0.32°C | 0.12°C | | | Rectum<br>-0.05°C | 0.48°C | 0.14°C | Rectum<br>-0.06°C | 0.38°C | 0.14°C | | Exergen<br>Temperature<br>(Temp)<br>Module | Integrated an optional OEM temp module (the<br>Exergen Temporal Scanner Thermometer, cleared<br>in K011291)<br>Measurement range: 15.5°C to 42°C<br>(60°F~107.6°F)<br>Measurement accuracy: ± 0.2°C or ± 0.4°F | | | Same | | | | Feature | Predicate Devices (K211475) | | | Subject Devices | | | | | VS 9 | VS 8 | VS 8A | VS 9 | VS 8 | VS 8A | | | Is compatible with the following 3 modules to<br>measure oxygen saturation:<br>• Mindray SpO2 Module<br>• Masimo SpO2 Module<br>• Nellcor SpO2 Module | | | Same | | | | Pulse oxygen<br>saturation<br>(SpO2) | Mindray SpO2 Module:<br>Measurement range: 0 to 100%<br>Accuracy:<br>70 to 100%: ±2% ABS (adult/pediatric mode)<br>70 to 100%: ±3% ABS (neonate mode)<br>0% to 69%: Not specified. | | | | | | | | Masimo SpO2 Module:<br>Measurement range: 1 to 100%<br>Accuracy:<br>70 to 100%: ±2% ABS (without motion in<br>adult/pediatric mode)<br>70 to 100%: ±3% ABS (without motion in neonate<br>mode)<br>70 to 100%: ±3% ABS (with motion)<br>1% to 69%: Not specified. | | | | | | | | Nellcor SpO2 Module:<br>Measurement range: 0 to 100%<br>Accuracy:<br>70 to 100%: ±2%ABS (adult/pediatric mode)<br>70 to 100%: ±3% ABS(neonate mode)<br>0% to 69%: Not specified. | | | | | | | | Uses the oscillometric method for measuring non-<br>invasive blood pressure (NIBP), support inflation<br>measuring mode. This measurement can be used<br>for adults, pediatrics and neonates. | | | Same | | | | Non-invasive<br>blood<br>pressure<br>(NIBP) | Measurement range:<br>Adult<br>Pediatric<br>Systolic: 25 to 290<br>25 to 240<br>Diastolic: 10 to 250<br>10 to 200<br>Mean: 15 to 260<br>15 to 215 | | Neonate<br>25 to 140<br>10 to 115<br>15 to 125 | | | | | | Accuracy:<br>Maximum average error: ±5 mmHg<br>Maximum standard deviation: 8mmHg | | | | | | | NIBP<br>Venipuncture | Uses the NIBP cuff to cause sub-diastolic pressure<br>to block the venous blood vessel and therefore help<br>venous puncture | | | Same | | | | Feature | Predicate Devices (K211475) | | | Subject Devices | | | | | VS 9 | VS 8 | VS 8A | VS 9 | VS 8 | VS 8A | | | Pulse rate may be obtained from the SpO2 module<br>or the NIBP module. | | | Same | | | | Pulse<br>rate<br>(PR) | PR from Mindray SpO2 Module<br>Measurement range: 20 to 254 bpm<br>Resolution: 1bpm<br>Accuracy: ±3 bpm | | | | | | | | PR from Masimo SpO2 Module<br>Measurement range: 25 to 240 bpm<br>Resolution: 1bpm<br>Accuracy:<br>±3 bpm (without motion)<br>±5 bpm (with motion) | | | | | | | | PR from Nellcor SpO2 Module<br>Measurement range: 20 to 300 bpm<br>Resolution: 1bpm<br>Accuracy:<br>20 to 250 bpm: ±3 bpm<br>251 to 300 bpm: not specified | | | | | | | | PR from NIBP Module<br>Measurement range: 30 to 300 bpm<br>Resolution: 1bpm<br>Accuracy: ±3bpm or ±3%, whichever is greater | | | | | | | Carbon<br>dioxide<br>(CO2) | CO2 parameter measuring.<br><br>CO2 measurement range: 0-<br>150mmHg<br><br>Resolution: 1mmHg<br><br>CO2 Accuracy :<br><br>Full accuracy mode:<br><br>0~40 mmHg: ±2mmHg<br><br>41~76 mmHg: ±5%<br>of<br>reading<br><br>77~99 mmHg: ±10%<br>of<br>reading<br><br>100~150mmHg: ±(3mmHg<br>+ 8% of reading)<br><br>> 150mmHg CO2:<br>unspecified<br><br>ISO accuracy mode: Add<br>±2mmHg to the full accuracy<br>mode | Not supported | | Same | Same | | | Feature | Predicate Devices (K211475) | | | Subject Devices | | | | | VS 9 | VS 8 | VS 8A | VS 9 | VS 8 | VS 8A | | NIBP<br>VeniPuncture | Support | | Not<br>supported | Same | | Same | {7}------------------------------------------------ {8}------------------------------------------------ K221267 {9}------------------------------------------------ #### Substantial Equivalence Conclusion 6.1. In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness. To establish the substantial equivalence of the subject VS Series Vital Signs Monitors, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the software and bench testing show that the subject device meets its specifications and is substantially equivalent to the predicate device. In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards. #### 7. PERFORMANCE DATA ### Biocompatibility Testing The subject VS Series Vital Signs Monitors are not patient contacting. The subject VS Series Vital Signs Monitors have a new Mindray Temperature Module (TrueTemp) and TrueTemp temperature probe. The Mindray Temperature Module is non-patient contacting. The only component TrueTemp temperature probe that is patient contacting is the probe cover. Cytotoxicity, Sensitization, and Intracutaneous Reactivity testing was completed and passed. ### Software Verification and Validation Testing Software verification and validation testing was conducted, and documentation was provided, as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the Vital Signs Monitors was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product. # Electrical safety and electromagnetic compatibility (EMC) The VS Series Vital Signs Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply: - ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and . A2:2010/(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. {10}------------------------------------------------ - . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests. - AIM Standard 7351731 Rev. 2.00 2017-02-23 Medical Electrical Equipment and ● System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers - An AIM Standard # Bench Testing Mindray has conducted testing in accordance with the following standards to establish substantial equivalence, ensure the subject devices meet relevant consensus standards, and the device performs as intended. - . ISO 80601-2-56 Second edition 2017-03 + AMD1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement - . ASTM E1112-00 Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature (Reapproved 2018) - ASTM E1104-98 Standard Specification for Clinical Thermometer Probe Covers . and Sheaths (Reapproved 2016) ### Animal Testing Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device. # Clinical Data To meet the requirements for the validity and accuracy of the TEMP measurement of the new Mindray Temp Module (TrueTemp), Mindray conducted clinical investigation according to the requirements of the standard ISO 80601-2-56 Second edition 2017-03 + AMD1:2018 : Medical electrical equipment- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. A clinical accuracy study evaluated 106, 110, 109 valid cases of oral, axillary and rectal temperature, which were performed on the following three age groups: infants (newborn to one year), children (greater than one to five years), and adults (greater than five years old) in accordance with ISO 80601-2-56 :2017/Amd.1:2018(E) to compare the direct mode of WelchAllyn SureTemp PLUS 690. The age of subjects is from 4 days to 67 years old. The total number of febrile subjects are not less than 30 % and not greater than 50 % of all subjects in the selected age group and body site. Statistical results show that, the temperature Measurement function of the TrueTemp module of the VS 9 Vital Signs Monitor in Predictive mode meets the requirements of ISO 80601-2-56:2017/ Amd.1:2018(E) for temperature measurement and meets the acceptance criteria in clinical protocol; the performances of the test device and the RCT are equivalent. During the entire clinical trial, all subjects are in good condition, and none of the subjects have discomfort {11}------------------------------------------------ symptoms or adverse reactions due to the use of Mindray VS 9 Vital Signs Monitor for temperature measurement. #### CONCLUSION 8. Based on the detailed comparison between the predicate devices and the subject devices, the performance testing and conformance with applicable standards, the VS Series Vital Signs Monitors can be found substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...