The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8/VS 8A Vital Signs Monitor are not intended for helicopter transport, hospital ambulance, or home use.
Device Story
VS series (VS 9, VS 8, VS 8A) are multiparameter patient monitors for clinical settings. Inputs include physiological signals: temperature (thermal resistance), SpO2 (Mindray/Masimo/Nellcor modules), PR, NIBP (cuff-based), and CO2 (sidestream). Device processes signals to display, store, and alarm on patient status. Operated by trained clinicians in healthcare facilities. Output provides real-time vital signs and early warning scores (MEWS, NEWS, NEWS2) to assist in recognizing patient deterioration. Benefits include continuous monitoring and clinical decision support for patient management. Features include touchscreen interface, wireless connectivity (Wi-Fi/Bluetooth), and battery/AC power. VS 9/VS 8 support NIBP venipuncture; VS 8A does not. Device integrates with hospital workflows for data transfer.
Clinical Evidence
Clinical investigation conducted for NIBP measurement accuracy per ISO 81060-2:2018. Study compared device NIBP measurements against reference invasive blood pressure monitoring and auscultation sphygmomanometer in neonate, infant, child, and adult populations. Results demonstrated compliance with standard requirements for automated measurement accuracy.
Technological Characteristics
Multiparameter monitor; touchscreen display (8"-10.1"); Wi-Fi/Bluetooth connectivity; rechargeable Li-ion battery/AC power. Parameters: Temp (thermal resistance), SpO2 (Mindray/Masimo/Nellcor), PR, NIBP (inflation/deflation), CO2 (sidestream). Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 62133-2, ISO 80601-2-55/56/61, IEC 80601-2-30/49. Software-based alarm system and early warning scoring (MEWS/NEWS/NEWS2).
Indications for Use
Indicated for monitoring, displaying, storing, and alarming of Temp, SpO2, PR, NIBP, and CO2 in adult, pediatric, and neonatal patients in healthcare facilities by trained clinical professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 5, 2021
Shenzhen Mindray Bio-Medical Electronics Co., LTD Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China
Re: K211475
Trade/Device Name: Vital Signs Monitors Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, CCK, DQA, DXN, FLL Dated: September 2, 2021 Received: September 3, 2021
Dear Yanhong Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211475
Device Name Vital Signs Monitors
### Indications for Use (Describe)
The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8A Vital Signs Monitors are not intended for helicopter transport, hospital ambulance, or home use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for the VS series Patient Monitors is provided below.
#### 1. SUBMITTER
| Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Contact Person: Yanhong Bai<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
- Date Prepared: August 1, 2021
#### 2. DEVICE
| Device Trade Name: | Vital Signs Monitors |
|---------------------------------------|-----------------------------------------------------------------------------------|
| Device Common Name: | VS series Vital Signs Monitors (Including VS 9, VS 8,<br>VS 8A) |
| Classification Name and<br>Regulation | 21 CFR 870.2300, Cardiac monitor (including<br>cardiotachometer and rate alarm) |
| Primary Product Code: | MWI - Monitor, physiological, patient (without<br>arrhythmia detection or alarms) |
| Regulatory Class | Class II |
| Panel | Cardiovascular |
{4}------------------------------------------------
| Product<br>Code | Regulation<br>Number | Panel | Regulation description | Device Common Name |
|-----------------|----------------------|----------------|-----------------------------------------------------|---------------------------------------------------|
| CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas<br>analyzer | analyzer, gas, carbon-dioxide, gaseous-phase |
| DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | oximeter |
| DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood<br>pressure measurement<br>system | system, measurement, blood-pressure, non-invasive |
| FLL | 21 CFR<br>880.2910 | Cardiovascular | Clinical electronic<br>thermometer | thermometer, electronic, clinical |
#### Table 1: Secondary Product Codes
### 3. PREDICATE DEVICE
K191769 - ePM Series Patient Monitors (including ePM 10/ePM 12/ePM Predicate Device: 15/ePM 10M/ePM 12M/ePM 15M) (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
Reference Device: K182821 - Accutorr 7/VS-900 Vital Signs Monitor (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
> K193391 - TM80 telemetry module component of the BeneVision Central Monitoring System (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
#### DEVICE DESCRIPTION 4.
The subject VS series Vital Signs Monitors includes three monitors:
- VS 9 Vital Signs Monitor ●
- VS 8 Vital Signs Monitor ●
- VS 8A Vital Signs Monitor ●
The VS series Vital Signs Monitors are for use for adult, pediatric, and neonatal patients. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. The monitors should only be used by persons who have received adequate training in their use.
#### 5. INTENDED USE/INDICATIONS FOR USE
The VS 9/VS 8/VS 8A Vital Signs Monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure
{5}------------------------------------------------
(NIBP), Carbon Dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The VS 9/VS 8/VS 8A Vital Signs Monitor are not intended for helicopter transport, hospital ambulance, or home use.
### SUBSTANTIAL EQUIVALENCE 6.
# Comparison of Indications
Both the predicate devices and the subject devices are multi parameter monitors intended to be used under the direction of clinical professionals. The monitoring parameters supported by the VS series are a subset of those supported by the predicate ePM series monitors (K191769).
In conclusion, the minor differences of the indications for use do not change the fundamental intended use of the VS Series as a multiparameter monitor.
# Technological Comparisons
The table below compares the key technological feature of the subject device to the primary predicate device (ePM series Patient Monitors, K191769). The features in gray are features which are different between the predicate device and the subject devices.
Note: In order to maintain readability, Table 2 only compares the subject device to the predicate ePM 10 model from K191769.
| | Predicate Devices (K191769) | Subject VS Devices | | |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Feature | ePM 10 | VS 9 | VS 8 | VS 8A |
| Display and<br>touchscreen | 10.1" 1280*800 pixels | 10.1" 1280*800 pixels | | 8" 1024*768 pixels |
| Wireless | Wifi: The ePM series patient<br>monitors support the laird<br>MSD45N WiFi module. | Wifi: VS series monitors support the silex SX-<br>SDMAC-2832S+ WiFi module.<br>Bluetooth: VS series monitors support Bluetooth. | | |
| Power supply | Battery or AC power | same | | |
| Battery | Rechargeable Lithium-Ion,<br>10.8VDC, 5600 mAh<br>Rechargeable Lithium-Ion,<br>10.95 VDC, 4500 mAh<br>Rechargeable Lithium-Ion,<br>10.95 VDC, 2600 mA | Only support<br>Rechargeable Lithium-<br>Ion, 10.8VDC, 5600<br>mAhwhich is as same<br>as ePM 10. | Support soft package<br>Rechargeable Lithium-<br>Ion, 10.95VDC, 5000<br>mAh<br>Rechargeable Lithium-<br>Ion, 10.95VDC, 2500<br>mAh. | |
| Data storage | Embedded Multi Media<br>Card(eMMC) | same | | |
| Data Recorder | Supports internal thermal<br>recorder | same | | |
| Speaker | Give alarm tones (45 to 85 dB),<br>key tones, QRS tones; support | same | | |
Table 2: Technological Comparison
{6}------------------------------------------------
| Feature | Predicate Devices (K191769) | | Subject VS Devices | | | | | | | | | | | | | | | | | |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | ePM 10 | VS 9 | VS 8 | VS 8A | | | | | | | | | | | | | | | | |
| | PITCH TONE and multi-level tone modulation. | | | | | | | | | | | | | | | | | | | |
| Alarm system | The alarm lamp is cyan, yellow, or red depending on alarm type.<br>Supports Alarm Volume Escalation | | same | | | | | | | | | | | | | | | | | |
| Temperature (Temp) | Use internal temperature module of MPM 3.0<br>Measurement range: 0 to 50°C (32 to 122°F)<br>Accuracy: ±0.1°C or ±0.2°F (without probe). | | Different. Supports 2 types of temperature module, but only one module can be integrated at one time.<br>Mindray Temp Module:<br>Technique: Thermal resistance<br>Measurement range:<br>Monitor mode:25 to 44 °C (77 to 111.2 °F)<br>Predictive mode: 34 to 43°C (93.2 to 109.4 °F)<br>Accuracy (Monitor mode):<br>25 to 32°C (not include 32°C): ± 0.2 °C<br>32 to 44°C (include 32°C): ±0.1 °C(± 0.2 °F)<br>or<br>77 to 89.6 °F (not include 89.6°F): ± 0.4 °F<br>89.6 to 111.2 °F (include 89.6°F): ± 0.2 °F<br>Statistical Results of Clinical Investigation Data (Predictive mode)<br>Clinical BIAS (Δcb) Limits of Agreement (LA) Clinical Repeatability (σr) Oral 0.02°C 0.33°C 0.1°C Axilla 0.06°C 0.38°C 0.13°C Rectum -0.05°C 0.48°C 0.14°C Integrated an optional OEM temp module (the Exergen Temporal Scanner Thermometer, cleared in K011291)<br>Measurement range:<br>16.0°C to 43.0°C (61°F~110°F) | | | | | | | | | | | | | | | | | |
{7}------------------------------------------------
| Feature | Predicate Devices (K191769)<br>ePM 10 | VS 9 | VS 8 | VS 8A |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|-------|
| Pulse oxygen<br>saturation<br>(SpO2) | Supports Mindray SpO2<br>function, Masimo SpO2<br>function and Nellcor SpO2<br>function from multi parameter<br>module.<br><br>The specifications for various<br>SpO2 functions are the same.<br><br>Mindray SpO2 function<br>Measurement range: 0 to<br>100%;<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric mode);<br>70 to 100%: ±3% (neonate<br>mode);<br>0% to 69%: Not specified.<br><br>Masimo SpO2 function<br>Measurement range: 1 to<br>100%;<br>Accuracy:70 to 100%: ±2%<br>(measured without motion in<br>adult/pediatric mode);<br>70 to 100%: ±3% (measured<br>without motion in neonate<br>mode);<br>70 to 100%: ±3% (measured<br>with motion);<br>1% to 69%: Not specified.<br><br>Nellcor SpO2 function<br>Measurement range: 0 to<br>100%;<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric);<br>70 to 100%: ±3% (neonate);<br>0% to 69%: Not specified.<br><br>Note: The specifictions of the<br>various SpO2 functions<br>provided by each manufacturer<br>are the same across platforms. | Supports Mindray SpO2 function, Masimo SpO2<br>function and Nellcor SpO2 function from multi<br>parameter module.<br><br>The specifications for various SpO2 functions are<br>the same.<br><br>The Masimo SpO2 module and the Nellcor SpO2<br>module are identical to the Masimo SpO2 module<br>and the Nellcor SpO2 module in K191769.<br><br>The internal Mindray SpO2 minor circuit<br>board layout is different compared to the<br>Mindray SpO2 module in K191769. | | |
| Pulse rate (PR) | PR from built-in Mindray<br>SpO2 Module Measurement<br>range: 20 to 254 bpm;<br>Accuracy: ±3 bpm. | Same | | |
| | PR from built-in Masimo SpO2 | | | |
{8}------------------------------------------------
| Feature | Predicate Devices (K191769) | Subject VS Devices | | | | | | | | | | | | | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|---------------|--|--|--|--|--|--|--|--|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | ePM 10 | VS 9 | VS 8 | VS 8A | | | | | | | | | | | | |
| | Module Measurement range:<br>25 to 240 bpm;<br>Accuracy: $\pm$ 3 bpm (measured<br>without motion);<br>$\pm$ 5 bpm (measured with<br>motion).<br><br>PR from built-in Nellcor SpO2<br>Module Measurement range:<br>20 to 300 bpm;<br>Accuracy:20 to 250 bpm:<br>$\pm$ 3 bpm;<br>251 to 300 bpm, not specified. | | | | | | | | | | | | | | | |
| Non- invasive<br>blood<br>pressure(NIBP) | Measurement range:<br>Adult Pediatric Neonate<br>Systolic 25~290<br>25~140 25~240 Diastolic 10~250<br>10~115 10~200 Mean 15~260<br>15~125 15~215 Accuracy:<br>Max mean error: $\pm$ 5 mmHg;<br>Max standard deviation: 8<br>mmHg.<br>Measure in deflation phase<br>Supports NIBP Venipuncture | | | | | | | | | | | | | The specifications are same as ePM10.<br>Neonate is measured in the deflation phase, such<br>as in the predicate device,<br>Adult and Pediatric is measured in the inflation<br>phase (inflation measuring mode) If the<br>measurement fails in the inflation phase, the<br>blood pressure will be measured in deflation<br>phase as the predicate device.<br>The ink used on the cuffs accessories has been<br>modified.<br>VS9 and VS8 support NIBP Venipuncture, but VS<br>8A doesn't support the feature. | | |
| Carbon<br>dioxide(CO2) | Compatible with 3 internal<br>CO2 modules:<br>Sidestream CO2 2.0 module<br>Mainstream CO2 module<br>MicroStream CO2 module<br>The internal CO2 modules are<br>identical to the external CO2<br>module in K182075<br>Type: Sidestream CO2 Module<br>(CO2 2.0):<br>Measurement range:<br>0~150mmHg Accuracy:<br>0~40 mmHg: $\pm$ 2mmHg,<br>41~76 mmHg: $\pm$ 5% of reading,<br>77~99 mmHg: $\pm$ 10% of<br>reading, | Compatible with 1<br>internal CO2 modules:<br>Sidestream CO2 2.0<br>module<br>Sidestream CO2<br>Module (CO2 2.0):<br>Measurement range:<br>0~150mmHg<br>Accuracy:<br>0~40 mmHg:<br>$\pm$ 2mmHg,<br>41~76 mmHg: $\pm$ 5% of<br>reading, 77~99 mmHg:<br>$\pm$ 10% of reading,<br>100~150mmHg:<br>$\pm$ (3mmHg + 8% of | | Not supported | | | | | | | | | | | | |
{9}------------------------------------------------
| Feature | Predicate Devices (K191769) | Subject VS Devices | | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------|
| | ePM 10 | VS 9 | VS 8 | VS 8A |
| | 100~150mmHg: ±(3mmHg +<br>8% of reading),<br>ISO accuracy mode:<br>Add ±2mmHg to the full<br>accuracy mode<br>AwRR measurement:<br>awRR measurement range:<br>0 to 150rpm;<br>awRR: <60rpm, ±1rpm,<br>60~150rpm, ±2rpm. | reading),<br>ISO accuracy mode:<br>Add ±2mmHg to the<br>full accuracy mode<br>AwRR measurement:<br>awRR measurement<br>range: 0 to 150rpm;<br>awRR: <60rpm,<br>±1rpm, 60~150rpm,<br>±2rpm.<br>Some electronic<br>components have<br>been replaced because<br>they are nearing end<br>of life or Mindray has<br>changed suppliers. | | |
| EWS | The EWS is a set of early<br>warning scores that are<br>intended to assist clinicians in<br>recognizing the early signs of<br>deterioration in patients based<br>on vital signs and clinical<br>observations. The four types of<br>EWS provided are Modified<br>Early Warning Score (MEWS),<br>National Early Warning Score<br>(NEWS), National Early<br>Warning Score2 (NEWS2) and<br>user configurable Custom<br>Score. | The EWS is a set of early warning scores that are<br>intended to assist clinicians in recognizing the<br>early signs of deterioration in patients based on<br>vital signs and clinical observations. The four<br>types of EWS provided are Modified Early<br>Warning Score (MEWS), National Early Warning<br>Score (NEWS), National Early Warning Score2<br>(NEWS2) and user configurable Custom Score.<br>The VS series monitors remove the History<br>total scores and Operator ID function from the<br>EWS Scores. | | |
| Glasgow Coma<br>Scale (GCS) | The GCS a well-established<br>scoring system used to assess<br>the state of consciousness<br>based three sub- components:<br>eye-opening response, verbal<br>response, and limb movement. | The GCS a well-established scoring system used to<br>assess the state of consciousness based three sub-<br>components: eye-opening response, verbal<br>response, and limb movement<br>The VS series monitors have a modify interface<br>and does not include the Review and Interval<br>Function. | | |
| Pain Score | Not support | Support | Support | Support |
| Vitalslink | Not support | Support | Support | Support |
| Rolling Stand | without Worktop and Power<br>Supply | with Worktop and Power Supply | | |
{10}------------------------------------------------
#### 6.1. Substantial Equivalence Conclusion
In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness.
To establish the substantial equivalence of the VS Series Vital Signs Monitors, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its specifications and is substantially equivalent to the predicate device.
In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
#### PERFORMANCE DATA 7.
# Biocompatibility Testing
The VS Series Vital Signs Monitors are not patient contacting. The only change to patient contacting devices is an ink change to the NIBP cuffs. Therefore, in accordance with FDA guidance, "Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" dated September 4, 2020, cytotoxicity, sensitization and irritation testing were performed.
# Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation was provided, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the VS Series Vital Signs Monitors was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
# Electrical safety and electromagnetic compatibility (EMC)
The VS Series Vital Signs Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply:
- ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and ● A2:2010/(R) 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements ● for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.
- . IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them for use in portable applications - Part 2: Lithium systems.
{11}------------------------------------------------
# Bench Testing
To establish the substantial equivalence of the VS Series Vital Signs Monitors, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification, and is substantially equivalent to the predicate device.
In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
- IEC 60601-1-6:2010. AMD1:2013 Medical electrical equipment Part 1-6 General requirements for safety - Collateral Standard: Usability (Edition 3.1)
- IEC 60601-1-8:2006+Am1:2012 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 80601-2-49:2018 Medical electrical equipment -Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
- ISO 80601-2-55: 2018 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
- ISO 80601-2-56: 2017, AMD1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- . ISO 80601-2-61: 2017, COR1:2018 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
- IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering to medical ● devices
# Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
# Clinical Data
To meet the requirements for the validity and accuracy of the NIBP measurement of automated measurement type, Mindray conducted clinical investigation with Reference Invasive Blood Pressure Monitoring Equipment and Auscultation reference sphygmomanometer according to the
{12}------------------------------------------------
requirements of the standard ISO 81060-2:2018 Non-invasive sphygmomanometers - part 2: clinical validation of automated measurement type. Study results demonstrated that the VS Series Vital Signs Monitors comply with the requirements of the standard in the neonate, infant, child and adult populations.
#### 8. CONCLUSION
Based on the detailed comparison between the predicate devices and the subject devices, the performance testing and conformance with applicable standards, the VS Series Vital Signs Monitors can be found substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.