K221195 · Active Life Scientific, Inc. · QGQ · Jun 28, 2022 · Orthopedic
Device Facts
Record ID
K221195
Device Name
OsteoProbe
Applicant
Active Life Scientific, Inc.
Product Code
QGQ · Orthopedic
Decision Date
Jun 28, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.1600
Device Class
Class 2
Indications for Use
The OsteoProbe is indicated for use as a measure bone tissue resistance to microindentation on the tibia in adults. The clinical significance of resistance to microindentation is unknown. The device is not intended to diagnose or treat any clinical condition.
Device Story
OsteoProbe is a bone microindentation measurement tool; used to measure bone tissue resistance on the tibia in adults. Device consists of reusable components and a single-use, sterile disposable component with a sterile cover. Reusable components require cleaning and intermediate-level disinfection between uses. Device is prescription-only. Output provides a measurement of bone tissue resistance; clinical significance of this measurement is unknown. Device is not intended for diagnosis or treatment of any clinical condition.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including sterilization validation, distribution testing, shelf-life testing, and biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, and rabbit pyrogen testing).
Technological Characteristics
Bone indentation device. Includes reusable components and single-use sterile disposable component. Sterilization per AAMI TIR28, ISO 10993-7, ISO 11135. Biocompatibility per ISO 10993-5, -10, -11. Distribution/Shelf-life per ASTM D4169, D4332-1, F1980, F2096, F88.
Indications for Use
Indicated for measuring bone tissue resistance to microindentation on the tibia in adults. No specific disease state or contraindications listed.
Regulatory Classification
Identification
A bone indentation device is a device that measures resistance to indentation in bone.
Special Controls
In combination with the general controls of the FD&C Act, the bone indentation device is subject to the following special controls:
- (1) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate the risk of bone fracture, soft tissue damage, pain, discomfort, bruising, or bleeding.
- (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of the accuracy and precision of the device with respect to resistance to bone indentation.
- (3) Human factors testing must demonstrate that the intended user(s) can correctly use the device, based on the instructions for use.
- The patient-contacting components of the device must be demonstrated to be (4) biocompatible.
- (ર) Performance testing must demonstrate:
- The sterility of the patient-contacting components of the device; and (i)
- (ii) Validation of reprocessing instructions for any reusable components of the device.
- Performance data must support the shelf life of the device by demonstrating continued (6) sterility and device functionality over the identified shelf life.
- Software verification, validation, and hazard analysis must be performed. (7)
- (8) Performance data must be provided to demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
- (9) Labeling must include:
- (i) Instructions for use:
- (ii) Validated methods and instructions for reprocessing of any reusable components;
- (iii) A shelf life for any sterile components;
- (iv) Information regarding limitations of the clinical significance of the device output; and
- A detailed summary of the accuracy and precision of the device. (v)
*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate the risk of bone fracture, soft tissue damage, pain, discomfort, bruising, or bleeding.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of the accuracy and precision of the device with respect to resistance to bone indentation.
(3) Human factors testing must demonstrate that the intended user(s) can correctly use the device, based on the instructions for use.
(4) The patient-contacting components of the device must be demonstrated to be biocompatible.
(5) Performance testing must demonstrate:
(i) The sterility of the patient-contacting components of the device; and
(ii) Validation of reprocessing instructions for any reusable components of the device.
(6) Performance data must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life.
(7) Software verification, validation, and hazard analysis must be performed.
(8) Performance data must be provided to demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
(9) Labeling must include:
(i) Instructions for use;
(ii) Validated methods and instructions for reprocessing of any reusable components;
(iii) A shelf life for any sterile components;
(iv) Information regarding limitations of the clinical significance of the device output; and
(v) A detailed summary of the accuracy and precision of the device.
{0}------------------------------------------------
June 28, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, there is a blue square with the letters "FDA" in white. Next to the square, the words "U.S. FOOD & DRUG" are written in blue, followed by the word "ADMINISTRATION" in a smaller font size.
Active Life Scientific, Inc. Alexander Proctor Chief Technology Officer 1027 Garden Street Santa Barbara, California 93101
Re: K221195
Trade/Device Name: OsteoProbe Regulation Number: 21 CFR 888.1600 Regulation Name: Bone Indentation Device Regulatory Class: Class II Product Code: QGQ Dated: June 6, 2022 Received: June 7, 2022
Dear Alexander Proctor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, PhD Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221195
Device Name OsteoProbe
Indications for Use (Describe)
The OsteoProbe is indicated for use as a measure bone tissue resistance to microindentation on the tibia in adults. The clinical significance of resistance to microindentation is unknown. The device is not intended to diagnose or treat any clinical condition.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | OsteoProbe |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Active Life Scientific, Inc. |
| | 1027 Garden Street, Santa Barbara, CA 93101 |
| Contact: | Alexander Proctor |
| | Chief Technology Officer |
| | Phone: 805-770-2600 x109 |
| | Email: alex@activelifescientific.com |
| Date Prepared: | 6/23/2022 |
| Classification: | 21 CFR §888.1600, Bone indentation device |
| Class: | II |
| Product Code: | QGQ |
| Predicate Devices: | OsteoProbe - DEN210013 |
| Indications for Use: | The OsteoProbe is indicated for use as a measurement tool to measure bone tissue<br>resistance to microindentation on the tibia in adults. The clinical significance of<br>resistance to microindentation is unknown. The device is not intended to diagnose or<br>treat any clinical condition. |
| Device Description: | OsteoProbe® is a bone microindentation measurement tool. It is a prescription device<br>per 21 CFR Part 801.109. The device includes a single-use disposable component and<br>reusable components. The single-use disposable component has a Spaulding<br>classification of critical and is provided sterile. The reusable components have a<br>Spaulding classification of non-critical and must be reprocessed (cleaning and<br>intermediate-level disinfection) between each use. The device has one accessory: a<br>single-use, disposable sterile cover. |
## Substantial Equivalence:
The subject OsteoProbe device is substantially equivalent to the predicate device with respect to indications, function, and performance. The sterilization, distribution, shelf-life, and biocompatibility of the single-use disposable of the subject device are substantially equivalent to the predicate device and do not raise different questions of safety and effectiveness.
## Preclinical Testing:
The following validations and testing were performed on the OsteoProbe device:
- . Sterilization validation testing (AAMI TIR28, ISO 10993-7, & ISO 11135)
- . Distribution Testing (ASTM D4169, ASTM D4332-1)
- Shelf Life (ASTM F1980, ASTM F2096, ASTM F88)
- Biocompatibility
- o Cytotoxicity (ISO 10993-5)
{4}------------------------------------------------
- O Sensitization (ISO 10993-10)
- Irritation/Intracutaneous Reactivity (ISO 10993-10) O
- o Acute Systemic Toxicity (ISO 10993-11)
- Rabbit Pyrogen-Material Mediated (ISO 10993-11) O
## Clinical Testing:
Clinical testing was not necessary to support equivalence.
#### Conclusion:
The subject OsteoProbe device is substantially equivalent for its intended use to the previously-cleared, predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.