K221088 · Double Medical Technology, Inc. · HRS · Jan 18, 2023 · Orthopedic
Device Facts
Record ID
K221088
Device Name
Anatomic Bone Plate
Applicant
Double Medical Technology, Inc.
Product Code
HRS · Orthopedic
Decision Date
Jan 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Anatomic Bone Plates are intended for the internal fixation of fractures (VAL Distal Clavicle, VAL Proximal Humeral, VAL Distal Humeral, Distal Ulnar Locking, Distal Radial Locking, VAL Proximal Tibia, VAL Distal Tibia, VAL Distal Lateral Fibular, VAL Distal Femoral, Pelvic Bone Plate), fracture dislocations (VAL Proximal Humeral), osteotomies, and mal/non-unions (VAL Distal Clavicle, VAL Proximal Humeral, VAL Distal Humeral) of the clavicle, pelvis, and the upper (i.e., humerus, ulna and radius) and lower (i.e., femur, tibia, and fibula) extremities. Additional plates included in the system with specific indications as follows: VAL (Variable Angle Locking) Distal Clavicle Plate is indicated for fractures of the shaft and lateral clavicle, malunions and non-unions ; VAL Distal Humeral Plate is indicated for fractures (including supracondylar), and nonunions ; VAL Distal Femoral Plate is indicated for distal femoral fractures (diaphyseal, supracondylar, and intra-articular); Pelvic Bone Plate is indicated for pelvic fractures (e.g., acetabular , iliac wing, and symphysis pubis).
Device Story
Anatomic Bone Plate system; metallic bone fixation appliance; includes locking plates and screws. Used for internal fixation of fractures, dislocations, and osteotomies; provides temporary internal stent to stabilize fracture ends; promotes healing. Used in clinical settings by orthopedic surgeons. Implants are single-use; instruments are reusable. Provided non-sterile.
Clinical Evidence
No clinical trials performed. Substantial equivalence supported by non-clinical mechanical testing (bending stiffness, strength, and structural stiffness per ASTM F382-17) and biocompatibility evaluation per ISO 10993-1. Clinical literature provided to support equivalence.
Technological Characteristics
Materials: Titanium alloy (ASTM F1472) or unalloyed titanium (ASTM F67). Mechanical testing per ASTM F382-17. Form factor: Anatomically shaped locking plates and screws. Non-sterile, single-use implants; reusable instruments.
Indications for Use
Indicated for internal fixation of fractures, fracture dislocations, osteotomies, and mal/non-unions of the clavicle, pelvis, and upper/lower extremities in patients requiring orthopedic stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Double Engine Bone Plate and Bone Screw System (K130108)
Submission Summary (Full Text)
{0}------------------------------------------------
January 18, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Double Medical Technology Inc. Yan Zuo Deputy International RA Supervisor No.18, Shanbianhong East Road, Haicang District Xiamen, Fujian 361026 China
Re: K221088
Trade/Device Name: Anatomic Bone Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: September 16, 2022 Received: September 23, 2022
Dear Yan Zuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Shumaya Ali-S" in a large, clear font. The text is black and stands out against a background that has a faded, light blue watermark. The watermark is only partially visible and does not obscure the name.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221088
Device Name Anatomic Bone Plate
### Indications for Use (Describe)
The Anatomic Bone Plates are intended for the internal fixation of fractures (VAL Distal Clavicle, VAL Proximal Humeral, VAL Distal Humeral, Distal Ulnar Locking, Distal Radial Locking, VAL Proximal Tibia, VAL Distal Tibia, VAL Distal Lateral Fibular, VAL Distal Femoral, Pelvic Bone Plate), fracture dislocations (VAL Proximal Humeral), osteotomies, and mal/non-unions (VAL Distal Clavicle, VAL Proximal Humeral) of the clavice, pelvis, and the upper (i.e., humerus, ulna and radius) and lower (i.e., femur, tibia, and fibula) extremities. Additional plates included in the system with specific indications as follows:
VAL (Variable Angle Locking) Distal Clavicle Plate is indicated for fractures of the shaft and lateral clavicle, malunions and non-unions ;
VAL Distal Humeral Plate is indicated for fractures (including supracondylar), and nonunions ;
VAL Distal Femoral Plate is indicated for distal femoral fractures (diaphyseal, supracondy)ar, , and intra-articular);
Pelvic Bone Plate is indicated for pelvic fractures (e.g., acetabular , iliac wing, and symphysis pubis).
Type of Use (Select one or both, as applicable)
| <div><span style="font-family: Arial, sans-serif;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: Arial, sans-serif;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 1. Submitter | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Name | Double Medical Technology Inc. |
| Address | No. 18, Shanbianhong East Road, Haicang District, Xiamen,<br>361026, P. R. China |
| Phone | +86 592 6885079 |
| Contact person | Yan Zuo |
| Date prepared | Jan 16th, 2023 |
| 2. Proposed Device | |
| Trade/proprietary name | Anatomic Bone Plate |
| Common or usual name | Anatomic Bone Plate |
| Classification name | Single/multiple component metallic bone fixation appliance<br>and accessories |
| Regulation number | 21 CFR 888.3030 |
| Product code | HRS |
| Regulatory class | II |
| Classification panel | Orthopedic |
| 3. Predicate Device | |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | Predicate device:<br>K130108<br>Double Engine Bone Plate and Bone Screw System |
| Reason for 510(k)<br>submission | New device |
# K221088 - 510(k) Summary
#### Device Description 4.
Anatomic Bone Plate is mainly used for internal fixation of fracture. According to the anatomical characteristics of human bones, locking plates of different shapes and locking screws of different diameters are designed. In clinical practice, locking bone plate and locking bone screw are used in combination to play the role of temporary internal stent, providing a stable local environment for fracture ends and creating conditions for the healing of fracture ends. Anatomic Bone Plate is provided as non-sterile. The implants are intended for single-use only, while the instruments are reusable.
Anatomic Bone Plate are made of titanium alloy following ASTM F1472 or unalloyed titanium following ASTM F67.
{4}------------------------------------------------
#### న. Indications for Use
The Anatomic Bone Plates are intended for the internal fixation of fractures (VAL Distal Clavicle, VAL Proximal Humeral, VAL Distal Humeral, Distal Ulnar Locking, Distal Radial Locking, VAL Proximal Tibia, VAL Distal Tibia, VAL Distal Lateral Fibular, VAL Distal Femoral, Pelvic Bone Plate), fracture dislocations (VAL Proximal Humeral), osteotomies, and mal/non-unions (VAL Distal Clavicle, VAL Proximal Humeral, VAL Distal Humeral) of the clavicle, pelvis, and the upper (i.e., humerus, ulna and radius) and lower (i.e., femur, tibia, and fibula) extremities. Additional plates included in the system with specific indications as follows:
- A VAL (Variable Angle Locking) Distal Clavicle Plate is indicated for fractures of the shaft and lateral clavicle, malunions and non-unions ;
- > VAL Distal Humeral Plate is indicated for fractures (including supracondylar), and nonunions :
- A VAL Distal Femoral Plate is indicated for distal femoral fractures (diaphyseal, supracondylar, and intra-articular);
- > Pelvic Bone Plate is indicated for pelvic fractures (e.g., acetabular , iliac wing, and symphysis pubis).
#### 6. Comparison of Technological Characteristics with the Predicate Device
The rationale for substantial equivalence is based on consideration of the following characteristics:
Regulatory Classification: Same as the predicate devices Indications for Use: Substantially equivalent (SE) to the predicate devices Materials: Substantially equivalent (SE) to the predicate devices Design Features: Substantially equivalent (SE) to the predicate devices
#### 7. Non-Clinical Performance Data
## 7.1 Biocompatibility testing
The biocompatibility evaluation for the Anatomic Bone Plate was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
{5}------------------------------------------------
## 7.2 Mechanical testing
The mechanical tests were performed per ASTM F382-17 Standard Specification and Test Methods for Metallic bone plate and three-point bending test method on Anatomic Bone Plate to demonstrate substantially equivalent of safety and efficacy with the predicate device. According to ASTM F382-17 and three-point bending test method, test methods are used to determine the bending stiffness, bending strength and bending structural stiffness.
#### 8. Clinical Data
Clinical literature was provided to demonstrate substantially equivalence.
#### 9. Conclusion
The Anatomic Bone Plate is compared to the predicate devices in respect of safety and effectiveness. The information provided within this premarket notification demonstrates that Anatomic Bone Plate is determined to be substantially equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.