K220834 · Daxwell, LLC · LZA · Jun 29, 2022 · General Hospital
Device Facts
Record ID
K220834
Device Name
Powder Free Nitrile Examination Glove
Applicant
Daxwell, LLC
Product Code
LZA · General Hospital
Decision Date
Jun 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The Powder Free Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Disposable nitrile examination glove; worn on examiner's hands; prevents cross-contamination between patient and examiner. Available in blue and indigo colors; various sizes (XS-XL). Non-sterile; single-use. Used in clinical or general medical settings. Provides physical barrier protection. Benefits patient and examiner by reducing risk of contamination during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ASTM D6319 (physical properties/dimensions), ASTM D5151 (freedom from holes), and ASTM D6124 (residual powder) demonstrated. Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (systemic toxicity).
Technological Characteristics
Material: Nitrile. Non-sterile. Single-use. Dimensions and physical properties comply with ASTM D6319-19. Biocompatibility tested per ISO 10993-10:2010 and ISO 10993-11:2017. No software or electronic components.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hands to prevent contamination between patient and examiner. Suitable for use by healthcare professionals and in over-the-counter settings.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
K232576 — Medical Nitrile Examination Gloves · Jiangsu Complete Sincerity Medical New Materials Co., Ltd. · Nov 21, 2023
K212924 — Nitrile Examination Gloves · Shanxi Nacosa Medical Technology Co.,Ltd · Dec 13, 2021
K212363 — Disposable Medical Nitrile Examination Gloves · Lanhuo Medical Technology (Jiangsu) Co.,Ltd · Oct 26, 2021
K223280 — Disposible Nitrile Gloves · Holik Asia Group Co.,Ltd · May 8, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 29, 2022
Daxwell, LLC % Ray Wang Gerneral Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China
Re: K220834
Trade/Device Name: Powder Free Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 19, 2022 Received: May 26, 2022
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220834
Device Name Powder Free Nitrile Examination Glove
Indications for Use (Describe)
The Powder Free Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K220834
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K220834
- 1. Date of Preparation: 06/29/2022
- 2. Sponsor
## DAXWELL, LLC
2825 Wilcrest Dr #500 Houston, TX 77042 US Contact Person: Frank Zhang Position: Head of Firm Tel: 01-281-699-0622 Fax: 01-281-699-0617 Email: scm.asst@Daxwell.com
- Submission Correspondent 3.
### Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email:information(@believe-med.com
- Proposed Device Identification 4.
Trade Name: Powder Free Nitrile Examination Glove Common Name: Powder Free Nitrile Examination Glove Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital
{4}------------------------------------------------
Indication for Use Statement:
The Powder Free Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
- Predicate Device Identification 5.
510(k) Number: K212497 Product Name: Nitrile Examination Glove (Powder free, Blue) Manufacturer: Jiangsu Jinlian Medical Technology Co., Ltd
- Device Description 6.
The proposed device, Powder Free Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed devices are Powder Free Nitrile Examination Gloves and include variations of different sizes from X-Small to X-large. The color of the proposed device is Blue and Indigo
The proposed device is not provided as sterilized The proposed device is made of Nitrile.
| Size<br>Model | Cuff<br>Thickness<br>(mm) | Palm<br>Thickness<br>(mm) | Finger<br>Thickness<br>(mm) | Width<br>(mm) | Length<br>(mm) | Color |
|---------------|---------------------------|---------------------------|-----------------------------|---------------|----------------|--------|
| XS | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 70±10 | ≥ 220 | |
| S | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 80±10 | ≥ 220 | Blue |
| M | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 95±10 | | |
| L | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 110±10 | ≥ 230 | |
| XL | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 120±10 | | |
| XS | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 70±10 | ≥ 220 | |
| S | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 80±10 | ≥ 220 | |
| M | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 95±10 | | Indigo |
| L | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 110±10 | ≥ 230 | |
| XL | ≥ 0.05 | ≥ 0.05 | ≥ 0.05 | 120±10 | | |
Table 1 Device Size Specifications
{5}------------------------------------------------
| Before Aging | | After Aging | | Pinhole AQL |
|---------------------|------------------------|---------------------|------------------------|-------------|
| Tensile<br>Strength | Ultimate<br>Elongation | Tensile<br>Strength | Ultimate<br>Elongation | 1.5 |
| 15 MPa, min | 500 % min | 14 MPa, min | 500 % min | |
Table 2 Performance and Physical Specifications
# 7. Technological Characteristic Comparison Table
| | Table 3 General Comparison |
|--|----------------------------|
| | |
| ITEM | Proposed Device(K220834)<br>Powder Free Nitrile Examination Glove | Predicate Device (K212497)<br>Nitrile Examination Glove (Powder free,<br>Blue) | Remark |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | LZA | LZA | SAME |
| Regulation No. | 21 CFR 880.6250 | 21 CFR 880.6250 | SAME |
| Class | I | I | SAME |
| Intended Use /<br>Indications for<br>Use | The Powder Free Nitrile Examination<br>Glove is a disposable device intended for<br>medical purposes that is worn on the<br>examiner's hands to prevent contamination<br>between patient and examiner. | The Nitrile Examination Glove (Powder<br>free, Blue) is a disposable device intended<br>for medical purposes that is worn on the<br>examiner's hands to prevent contamination<br>between patient and examiner. | SAME |
| Sterility | Non-sterile | Non-sterile | SAME |
| Single use or<br>reuses | Single use | Single use | SAME |
| Powdered or<br>Powered free | Powdered free | Powdered free | SAME |
{6}------------------------------------------------
| Proposed Device (K220834) | Designation | Size | | | | | Tolerance |
|------------------------------------------------------------------|----------------|------|-----|-----|-----|-----------|-----------|
| Powder Free Nitrile<br>Examination Glove<br>(Blue, Indigo) | Length, mm | XS | S | M | L | XL | |
| | | 230 | 230 | 230 | 230 | 230 | ±10 |
| | Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 |
| | Thickness, mm: | | | | | | |
| | Finger | 0.12 | | | | | ±0.03 |
| | Palm | 0.10 | | | | | ±0.03 |
| | Cuff | 0.09 | | | | | ±0.03 |
| Predicate Device (K212497) | Designation | Size | | | | Tolerance | |
| Nitrile Examination Glove<br>(Powder free, Purple-Blue,<br>Blue) | Length, mm | S | M | L | XL | min | |
| | | 230 | 230 | 230 | 230 | | |
| | Width, mm | 80 | 95 | 110 | 120 | ±10 | |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | | | | | min |
| | Palm | 0.05 | | | | | min |
| | Cuff | 0.05 | | | | | min |
| Remark | Analysis 1 | | | | | | |
Table 4 Device Dimensions Comparison
## Analysis 1:
The proposed device has different size specification to the predicate device, but all proposed devices meet the specifications of ASTM D 6319.
### Table 5 Performance Comparison
| ITEM | | | Proposed Device(K220834)<br>Powder Free Nitrile Examination<br>Glove | Predicate Device (K212497)<br>Nitrile Examination Glove<br>(Powder free, Purple-Blue,<br>Blue) | Remark |
|------------------------|-----------------|------------------------|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------|
| Colorant | | | Blue, Indigo | Blue | Analysis 2 |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 15 MPa, min | 14 MPa, min | Analysis 3 |
| | | Ultimate<br>Elongation | 500 % min | 500 % min | SAME |
| | After<br>Aging | Tensile<br>Strength | 14 MPa, min | 14 MPa, min | SAME |
| | | Ultimate | 500 % min | 400 % min | Analysis 4 |
{7}------------------------------------------------
| | | Elongation | | | |
|--|------------------------|------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|------|
| | Comply with ASTM D6319 | | | Comply with ASTM D6319 | SAME |
| | Freedom from Holes | | Be free from holes when tested in<br>accordance with ASTM D5151<br>(AQL:1.5) | Be free from holes when tested in<br>accordance with ASTM D5151<br>(AQL:1.5) | SAME |
| | Powder Content | | Less than 2 mg per glove when tested<br>in accordance with ASTM D6124 | Less than 2 mg per glove when<br>tested in accordance with ASTM<br>D6124. | SAME |
#### Analysis 2:
The proposed device has different color to the predicate device, this difference may cause potential biocompatibility risk, for this risk we conducted the biocompatibility testing according to ISO 10993-11 and ISO 10993-10, the test results showed that the proposed devices with blue colorant and indigo colorant did not induce skin irritation, and showed no significant evidence of causing skin sensitization and acute system toxicity reactions.
#### Analysis 3:
The proposed device has different Tensile Strength before aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.
Analysis 4:
The proposed device has different Ultimate Elongation after aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.
| ITEM | | Proposed Device(K220834) | Predicate Device (K212497) | Remark |
|--------------------|----------------------|-----------------------------------------------------------|-----------------------------------------------------------|--------|
| | | Powder Free Nitrile Examination<br>Glove | Nitrile Examination Glove<br>(Powder free, Blue) | |
| Material | | Nitrile | Nitrile | SAME |
| Biocompatibility | Irritation | Under the conditions of the study,<br>not an irritant | Under the conditions of the study,<br>not an irritant | SAME |
| | Sensitization | Under conditions of the study, not a<br>sensitizer. | Under conditions of the study, not a<br>sensitizer. | |
| | Systemic<br>toxicity | Under conditions of the study, not a<br>systemic toxicity | Under conditions of the study, not a<br>systemic toxicity | |
| Label and Labeling | | Meet FDA's Requirements | Meet FDA's Requirements | SAME |
#### Table 6 Safety Comparison
{8}------------------------------------------------
# 8.0 Summary of Non-Clinical Testing
Bench tests were conducted to demonstrate that the proposed device complies with the following standards:
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves.
ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
| Test<br>Method | Purpose | Acceptance Criteria | Color | Results |
|-----------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM<br>D5151 | Testing for Freedom from holes | Freedom from holes (AQL:1.5) | Blue<br>Indigo | No water leakage is inspected<br>form 200 samples |
| ASTM<br>D6124 | Determine the powder residue for powder free gloves | <2.0 mg per glove | Blue<br>Indigo | Residual Powder: Average 0. 38 Mg; |
| ASTM<br>D6319 | Testing for Physical property characteristics | Before Aging<br>Tensile Strength: 15 MPa, min<br>Ultimate Elongation: 500 % min<br><br>After Aging<br>Tensile Strength: 14 MPa, min<br>Ultimate Elongation: 500 % min | Blue<br>Indigo | Before Aging:<br>Tensile Strength: ≥ 17 MPa;<br>Ultimate Elongation: ≥ 512%.<br><br>After Aging:<br>Tensile Strength: ≥ 17 MPa;<br>Ultimate Elongation: ≥ 508%. |
| | Testing For physical dimensions specification | Length: 230 mm ±10 for all size (XS, S, M, L, XL);<br>Width: 70±10 mm for XS; 80±10 mm for S; 95±10 mm for M; 110±10 mm for L; 120±10 mm for XL.<br>Finger Thickness: 0.12±0.03 mm;<br>Palm Thickness: 0.10±0.03 mm;<br>All acceptance criteria above meet the requirements in Table 2 Dimensions and Tolerances of ASTM D6319 | Blue/<br>Indigo | Length of Size XS: ≥ 229 mm;<br>Width of Size XS: ≥70 (70-73) mm;<br>Palm Thickness of Size XS: ≥0.09 mm;<br>Finger Thickness of Size XS: ≥0.11 mm.<br><br>Length of Size S: ≥ 230 mm;<br>Width of Size S: ≥85 (85-86) mm;<br>Palm Thickness of Size S: ≥0.09 mm; |
| | | | | |
| | | | | Finger Thickness of Size S:<br>≥0.11 mm. |
| | | | | Length of Size M: ≥ 230 mm;<br>Width of Size M: ≥95 (95-96) mm;<br>Palm Thickness of Size M:<br>≥0.09 mm;<br>Finger Thickness of Size M:<br>≥0.11 mm. |
| | | | | Length of Size L: ≥ 230 mm;<br>Width of Size L: ≥108 (108-110) mm;<br>Palm Thickness of Size L: ≥0.09 mm;<br>Finger Thickness of Size L:<br>≥0.11 mm. |
| | | | | Length of Size XL: ≥ 230mm;<br>Width of Size XL: ≥120 (120-121) mm;<br>Palm Thickness of Size XL:<br>≥0.09 mm;<br>Finger Thickness of Size XL:<br>≥0.11 mm. |
| ISO<br>10993-11 | Evaluate the endpoint of<br>systemic toxicity for<br>biocompatibility | The test article showed “negative”<br>systemic toxicity | Blue<br>Indigo | Under the conditions of the<br>study, the test article showed<br>“negative” systemic toxicity. |
| ISO<br>10993-10 | Evaluate the endpoint of<br>irritant for<br>biocompatibility | The response of the test article has no<br>skin irritation | Blue<br>Indigo | Under the experimental<br>conditions, the test article has no<br>skin irritation on rabbits. |
| | Evaluate the endpoint of<br>sensitization for<br>biocompatibility | The test article showed no evidence of<br>causing delayed dermal contact<br>sensitization. | Blue<br>Indigo | The test article showed no<br>evidence of causing delayed<br>dermal contact sensitization in<br>the guinea pig. |
Table 7 Performance Test Results Summary
{9}------------------------------------------------
# 9.0 Summary of Clinical Testing
Clinical testing is not needed for this device.
{10}------------------------------------------------
# 10.0 Conclusion
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Nitrile Examination Glove (Powder free, Blue) cleared under K212497.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.