K220641 · Palette Life Sciences · OVB · May 26, 2022 · Radiology
Device Facts
Record ID
K220641
Device Name
Barrigel Injectable Gel
Applicant
Palette Life Sciences
Product Code
OVB · Radiology
Decision Date
May 26, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5725
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K220641 · May 26, 2022
Barrigel Injectable Gel
Palette Life Sciences
Retrospective clinical study (Goni); Retrospective clinical study (RPAH1)
Retrospective clinical data from studies of a predicate-equivalent device (Restylane SubQ) were used to characterize the resorption profile and long-term safety of the Barrigel injectable gel.
27 male subjects who underwent perirectal injections of Restylane SubQ with radiotherapy and/or brachytherapy for prostate cancer; Sample Size: 27
Not applicable for this study
Resorption of gel and adverse tissue reaction
RPAH1 clinical study; Retrospective clinical study; Follow-up/Duration: Up to 2 years post-treatment
36 male subjects injected with Restylane SubQ prior to IMRT and IGRT for prostate cancer; Sample Size: 36
Not applicable for this study
Material stability and gel resorption
Indications for Use
Barrigel is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space, it is the intent of Barrigel to reduce the radiation dose delivered to the anterior rectum. Barrigel is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is intended to be absorbed by the patient's body over time.
Device Story
Barrigel is a sterile, biodegradable, cross-linked hyaluronic acid (HA) gel injected into the perirectal space. It acts as a mechanical spacer to separate the anterior rectal wall from the prostate during radiotherapy for prostate cancer. The device is supplied in a single-use glass syringe and administered by a healthcare professional using transrectal ultrasound (TRUS) guidance. By creating this physical distance, the device reduces the radiation dose delivered to the rectum, potentially lowering the risk of radiation-induced gastrointestinal toxicity. The gel remains in place throughout the radiotherapy course and is gradually absorbed by the body over time. It requires no pre-mixing or hydro-dissection prior to injection.
Clinical Evidence
Clinical evidence includes a randomized, controlled, single-blinded multicenter IDE study (G190206) with 14 sites. Primary effectiveness endpoint: 98.6% of Barrigel subjects achieved ≥25% reduction in rectal volume receiving 90% of prescription dose (LCL 0.923, p<0.0001). Key secondary safety objective: Barrigel was non-inferior to control for acute Grade 2+ GI toxicity at 3 months (Barrigel 2.8% vs. Control 14.7%, p=0.0348). Additional retrospective/prospective data from Restylane SubQ studies (Goni, RPAH1) provided long-term resorption evidence (complete resorption observed by 60 months).
Technological Characteristics
Sterile, transparent, biodegradable gel of stabilized hyaluronic acid (HA) at 20 mg/mL in phosphate buffered saline. Cross-linked with 1,4-butanediol diglycidyl ether (BDDE). Supplied in 3 mL single-use glass syringe. Delivered via 18G stainless-steel needle (20 cm). Sterilization: moist heat (gel), radiation (needle). Biocompatibility per ISO 10993. Shelf-life: 24 months. Storage: up to 25°C.
Indications for Use
Indicated for adult human males undergoing radiotherapy for prostate cancer to temporarily position the anterior rectal wall away from the prostate to reduce radiation dose to the rectum.
Regulatory Classification
Identification
An absorbable perirectal spacer is composed of biodegradable material that temporarily positions the anterior rectal wall away from the prostate during radiotherapy for prostate cancer with the intent to reduce the radiation dose delivered to the anterior rectum. The absorbable spacer maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.
Special Controls
An absorbable perirectal spacer must comply with the following special controls: (1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided. Performance bench testing must demonstrate appropriate perirectal space (i) creation and maintenance for the duration of prostate radiotherapy; Performance bench testing must demonstrate that therapeutic radiation levels (ii) do not alter the performance of the device; (iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment; (iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity: Sterility testing must demonstrate the sterility of the device and the effects of (v) the sterilization process on the physical characteristics of the spacer; (vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling: and. (vii) The device must be demonstrated to be biocompatible. (2) The risk management activities performed as part of the manufacturer's 21 CFR 820.30 design controls must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique. (3) The device labeling must include the following: A detailed summary of reported or observed complications related to the use (i) of the device; (ii) Appropriate warnings; Detailed instructions for system preparations and detailed implant procedure (iii) instructions; and, (iv) An expiration date that is supported by performance data as specified in subparagraph (b)(i)(vi).
*Classification.* Class II (special controls). The special controls for this device are:(1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided.
(i) Performance bench testing must demonstrate appropriate perirectal space creation and maintenance for the duration of prostate radiotherapy.
(ii) Performance bench testing must demonstrate that therapeutic radiation levels do not alter the performance of the device.
(iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment.
(iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity.
(v) Sterility testing must demonstrate the sterility of the device and the effects of the sterilization process on the physical characteristics of the spacer.
(vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling.
(vii) The device must be demonstrated to be biocompatible.
(2) The risk management activities performed as part of the manufacturer's § 820.10(c) of this chapter design and development activities must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique.
(3) The device labeling must include the following:
(i) A detailed summary of reported or observed complications related to the use of the device;
(ii) Appropriate warnings;
(iii) Detailed instructions for system preparations and detailed implant procedure instructions; and
(iv) An expiration date that is supported by performance data as specified in paragraph (b)(1)(vi) of this section.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
Palette Life Sciences % David Goodnough Regulatory Affairs Manager 27 East Cota Street, Suite 402 Santa Barbara, California 93101
# Re: K220641
Trade/Device Name: Barrigel Injectable Gel Regulation Number: 21 CFR 892.5725 Regulation Name: Absorbable Perirectal Spacer Regulatory Class: Class II Product Code: OVB Dated: March 4, 2022 Received: March 4, 2022
# Dear David Goodnough:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Sullivan, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220641
Device Name Barrigel Injectable Gel
### Indications for Use (Describe)
Barrigel is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space, it is the intent of Barrigel to reduce the radiation dose delivered to the anterior rectum. Barrigel is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is intended to be absorbed by the patient's body over time.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span> ☑ </span> |
|----------------------------------------------|------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> ☐ </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Palette Life Sciences. The logo consists of a stylized "P" symbol on the left, with the word "Palette" in a dark blue sans-serif font to the right of the symbol. Below the word "Palette" are the words "LIFE SCIENCES" in a smaller, lighter font.
# Section 5. 510(k) Summary for Barrigel® Injectable Gel
This 510(k) Summary for Barrigel Injectable Gel has been generated in accordance with 21 CFR §807.92 Content and format of a 510(k) summary. In response to the Safe Medical Devices Act of 1990, the following summarizes the information upon which substantial equivalence is based.
#### 5.1 Sponsor
Palette Life Sciences, Inc. 27 East Cota Street, Suite 402 Santa Barbara, CA 93101
#### 5.2 Contact
David Goodnough Regulatory Affairs Manager Palette Life Sciences Phone: 805-869-7087 Email: dgoodnough(@palettelifesciences.com
#### Date of 510(k) Summary Preparation 5.3
5 May, 2022
#### Proposed Device 5.4
Trade Name: Barrigel Injectable Gel Common/Usual Name: Barrigel Regulation Number: 892.5725 Classification Name: Absorbable Perirectal Spacer Classification: Class II (Special Controls) Product Code: OVB Review Panel: Radiology
#### ર્સ્ટ Predicate Device
Trade Name: SpaceOAR Hydrogel System, Model Number: SO-2101 Common/Usual Name: Hydrogel Spacer Regulation Number: 892.5725 Classification Name: Absorbable Perirectal Spacer Classification: Class II Product Code: OVB Identification of Predicate Device: SpaceOAR Hydrogel System, K202224
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" made of overlapping circles in shades of blue and orange, followed by the text "Palette" in a dark blue sans-serif font. Below the word "Palette" is the text "LIFE SCIENCES" in a smaller, also dark blue, sans-serif font.
#### 5.6 Device Description
Barrigel Injectable Gel is a sterile, transparent, biodegradable gel of stabilized hyaluronic acid (HA) at a concentration of 20 mg/mL in phosphate buffered saline.
The HA, formulated utilizing Q-Med's patented NASHA™ technology, is produced from non-animal hyaluronic acid by fermentation of Streptococcus species of bacteria. The HA gel is cross-linked with 1,4-butanediol diglycidyl ether (BDDE) under alkaline conditions, thereby creating ether bonds between the HA chains, resulting in a three-dimensional network. The gel is insoluble in water and organic solvents.
Barrigel is supplied in a single-use glass syringe, containing 3 mL of product. Each syringe is terminally sterilized by moist heat in a heat-sealed pouch of PET/Tyvek® and packaged in a cardboard carton. Barrigel is intended for use by health-care professionals only and should be stored up to 25° C (77° F). Barrigel is manufactured for Palette Life Sciences by O-Med AB, a subsidiary of Galderma AB.
The Barrigel needle is a sterile 18G stainless-steel needle, 20 cm in length, provided with an optional stylet and protected by a polyester sheath which is removed prior to use. The needle is sterilized by radiation and two (2) needles are provided in each heat-sealed pouch of PET/Tyvek®, packaged in a cardboard carton. The Barrigel needle is identical to the needle used during the Barrigel IDE study and is manufactured for Palette Life Sciences by R.K. Manufacturing.
#### 5.7 Indications for Use
Barrigel is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and, in creating this space, it is the intent of Barrigel to reduce the radiation dose delivered to the anterior rectum. Barrigel is composed of biodegradable material and maintains the space
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" made up of overlapping circles in shades of orange and blue, followed by the word "Palette" in a dark blue, sans-serif font. Below "Palette" are the words "LIFE SCIENCES" in a smaller, sans-serif font, also in dark blue.
for the entire course of prostate radiotherapy treatment and is intended to be absorbed by the body over time.
#### Technological Characteristics Compared to the Predicate 5.8
The intended use and principles of operation of the proposed Barrigel device and the predicate device are identical and meet the regulatory definition for Absorbable Rectal Spacers outlined in 21 CFR 892.75. Both devices are systems that facilitate implantation of biodegradable materials between the anterior rectal wall and the prostate prior to radiotherapy for prostate cancer, mechanically creating space between the anterior rectal wall and the prostate. Both devices maintain the separation between the prostate and the anterior rectal wall during the entire course of treatment and are absorbed by the body over time. The target population of both devices is adult human males being treated for prostate cancer. Both devices are single-use and are provided sterile to licensed medical professionals (only) and utilize Transrectal Ultrasound (TRUS) to provide safe and accurate placement of the device.
The shelf-life of 24 months for Barrigel and 24 months for SpaceOAR; intended volume of treatment of 9-12 ml for Barrigel and 10 ml for SpaceOAR; and 18G needles (20 cm length for Barrigel and 15 cm length for SpaceOAR) are considered to represent technological characteristics of the devices that are similar and support the claim of substantial equivalence between the subject and predicate devices.
The differences between the subject and predicate device materials, syringes, treatment process, and absorption rates of the devices are as follows:
- . Material:
- Barrigel Injectable Gel is comprised of non-animal hyaluronic o acid (NASHA) sourced from the fermentation of Streptococcus species of bacteria and produced by cross-linking HA with BDDE,
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" shape in gradient colors of orange and blue, followed by the word "Palette" in a dark blue sans-serif font. Below "Palette" is the text "LIFE SCIENCES" in a smaller, lighter font.
resulting in a three-dimensional network. Barrigel does not require pre-mixing prior to use.
- SpaceOAR System is comprised of synthetic polyethylene glycol o (PEG); a hydrophilic polymer, that when cross-linked produces a three-dimensional hydrogel. SpaceOAR is prepared by mixing PEG powder with a Diluent solution which is mixed with a saltbuffer accelerator solution during injection.
- Syringe: .
- Barrigel utilizes a single sterile glass syringe containing 3 ml of o Barrigel cross-linked HA material in buffered saline solution.
- SpaceOAR System utilizes a dual syringe system containing 10 о ml of hydrogel material: One syringe is a mixture of PEG powder (provided in a separate vial) with a Diluent solution which is then combined during injection with second syringe containing an accelerant solution of buffered saline. The two syringes are injected together via a Y-connector attached to the two syringes.
- Process: ●
- Barrigel does not require hydro-dissection prior to injection. o
- SpaceOAR System requires hydro-dissection prior to injection. о
- Absorption: .
- o Barrigel As only a limited number (6% = 6/98) of patients were followed to complete resorption at 18 months, it is unknown if there are potential late complications or side effects of incompletely resorbed gel. Table 1 below, represents resorption data available for 96 patients at the 3-month visit, 55 patients at the 12-month visit, and 20 patients at the 18-month visit. There is currently no evidence that reflects potential late complications or side effects of incompletely resorbed gel.
- SpaceOAR Completely absorbed approximately 6 months after o implantation.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for Palette Life Sciences. The logo consists of a stylized "P" symbol on the left, with a gradient from orange to blue. To the right of the symbol is the word "Palette" in a bold, blue sans-serif font. Below "Palette" is the text "LIFE SCIENCES" in a smaller, lighter sans-serif font, also in blue.
| Visit | Statistic | Percent<br>Resorption |
|--------------------------|--------------------|-----------------------|
| Immediate Post-Injection | N (patients) | 98 |
| | MIN | 0% |
| | MAX | 0% |
| | MEAN | 0% |
| | STANDARD DEVIATION | 0 |
| 3-Month Visit | N (patients) | 96 |
| | MIN | -11.4%* |
| | MAX | 66.8% |
| | MEAN | 18.4% |
| | STANDARD DEVIATION | 15.4 |
| 12-Month Visit | N (patients) | 55 |
| | MIN | 20.4% |
| | MAX | 90.3% |
| | MEAN | 57.2% |
| | STANDARD DEVIATION | 16.8 |
| 18-Month Visit | N (patients) | 20 |
| | MIN | 35.0% |
| | MAX | 93.9% |
| | MEAN | 73.6% |
| | STANDARD DEVIATION | 15.5 |
# Table 1. Resorption of Barrigel over Time (as reflected by patient visit)
* Negative resorption in some subjects due to the hydrophilic nature of Barrigel and the temporary increase in water present at the point of injection.
The technological characteristics of the subject device are substantially equivalent to the predicate device based on the intended use, delivery system and mode of operation of both devices. Differences between the predicate and subject devices are the base device material, syringe configuration, additional preinjection process step, and rates of absorption.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" made up of overlapping shapes in shades of blue and orange, followed by the text "Palette" in a dark blue sans-serif font. Below "Palette" is the text "LIFE SCIENCES" in a smaller, lighter font.
Both the subject and predicate devices have long histories of safe use and differences that may exist in the effectiveness of Barrigel are well addressed by the biocompatibility, performance, and clinical data. The safety profile of Barrigel has been previously reviewed as part of IDE G190206.
Injectable NASHA products such as Barrigel possess a safety profile based on more than 40 million patient treatments over the past 20 years of aesthetic, therapeutic and spacing indications. Barrigel is approved and has been used in over 1,250 treatments for the recommended indication in the EU and Australia with no reports of serious, device related adverse events.
#### Biocompatibility 5.9
Biocompatibility testing of Barrigel was performed per applicable standards on both irradiated and non-irradiated Barrigel samples, addressing the following evaluation endpoints:
- Cytotoxicity .
- Sensitization .
- Irritation/intracutaneous Reactivity .
- . Acute Systemic Toxicity
- Material Mediated Pyrogenicity ●
- Subacute/Sub-chronic Toxicity ●
- Genotoxicity ●
- Implantation ●
- Chronic Toxicity .
- Carcinogenicity .
Tests for reproductive and developmental toxicity, toxicokinetics, immunotoxicity and carcinogenicity were determined to not be applicable to Barrigel since Barrigel does not contain any levels of materials considered hazardous; does not contain materials of toxicological significance; and is nonimmunogenic. It was further determined that use of Barrigel does not represent a risk of tumor development and therefore does not present a cancer risk.
6
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" shape on the left, with the word "Palette" in a dark blue sans-serif font to the right of the shape. Below the word "Palette" are the words "LIFE SCIENCES" in a smaller, lighter sans-serif font. The "P" shape is made up of three overlapping semi-circles, with the left-most semi-circle being orange, the middle semi-circle being a gradient of orange to blue, and the right-most semi-circle being blue.
Examination of the risk-based biocompatibility assessment indicates Barrigel met the acceptance criteria of the biocompatibility endpoint tests listed above.
#### Performance Testing – Bench 5.10
Nonclinical testing was performed on Barrigel components and materials. These tests included physio/chemical characteristics of Barrigel injectable gel and were performed utilizing three (3) production lots of Barrigel to ensure consistency and reproducibility throughout the manufacturing process. Physio/chemical characterization was performed in accordance with EN ISO 10993-18 and included the following:
- HA Content
- Gel Content ●
- Extractable Carbohydrates ●
- Swelling Factor ●
- pH
- Particle Size Distribution ●
- Bacterial endotoxins ●
- Sterility ●
- Viscoelastic Properties .
- Osmolality .
- Elemental Impurities .
- . Residual Proteins
- Enzymatic Degradation ●
- Extrusion Force .
Results of the physio/chemical characterization indicated all acceptance criteria were met and illustrated that there is minimal variation between production lots.
7
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" shape on the left, with the word "Palette" in a dark blue sans-serif font to the right of the shape. Below the word "Palette" are the words "LIFE SCIENCES" in a smaller, lighter font. The "P" shape is made up of three overlapping semi-circles, with a gradient of colors ranging from orange to blue.
#### 5.11 Performance Testing - Animal
A rabbit implantation study involving subcutaneous injections of both irradiated and non-irradiated Barrigel was conducted to identify early tissue response, local effects secondary to degradation as well as characterization and degree of resorption over the 52-week test period.
Test results demonstrated minimal to no reaction to irradiated Barrigel when compared to the non-irradiated control. In addition, and using the standard set forth in ISO 10993-6, Barrigel Injectable Gel can be considered completely absorbed, when material is significantly resorbed (≥ 90%) and tissue reactivity has subsided.
#### Performance Testing - Clinical 5.12
IDE G190206 is a clinical study conducted to determine the extent to which Barrigel Injectable Gel is able to mechanically increase the distance between the prostate and the rectum and thereby decrease the exposure of the rectum to the radiation dose in patients receiving definitive radiation therapy for localized prostate cancer. The study is a randomized, controlled, single-blinded multicenter study, conducted over 14 study sites. The Barrigel IDE was conducted to support the 510(k) submission and represented the first stage of a two-stage study.
In the second stage of the study and upon approval, all study subjects will continue with a post-market surveillance (PMS) study for longer-term follow-up.
All study subjects were treated with transrectal, ultra-sound guided trans-perineal prostate fiducial marker placement. Randomly assigned subjects followed fiducial placement with Barrigel injection between the rectum and the prostate. Study subjects not receiving Barrigel are the control group. All subjects received radiation therapy using IMRT.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" shape on the left, with the word "Palette" in a sans-serif font to the right of it. Below the word "Palette" is the phrase "LIFE SCIENCES" in a smaller, sans-serif font. The "P" shape is colored with a gradient that transitions from orange to blue.
The primary effectiveness endpoint for the Barrigel subjects in the study is the achievement of a 25% reduction in the volume of the rectum receiving 90% of the prescription radiation dose.
The key secondary safety objective for the study is determination that the Barrigel spacer does not increase the evidence of acute Grade 2+ GI toxicity within the first 3 months following treatment, assessed using the proportion of subjects experiencing one or more incidents of acute Grade 2+ toxicity within the first 3 months following treatment.
The primary effectiveness endpoint of the study was met and demonstrated that the proportion of subjects achieving a 25% reduction in the volume of the rectum receiving 90% of the prescribed radiation dose (based on pre-injection and immediate post-treatment CT and MRI scans), was greater than the minimally acceptable success rate of 70%, established by the De Novo application of the predicate device. The success rate for stage 1 of the Barrigel IDE indicates that 98.6% of the complete cases in the Barrigel group achieved at least a 25% reduction in the volume of the rectum receiving 90% of the prescription dose, the lower boundary of the 95% confidence interval (LCL) is 0.923 and the one-sided p-value is < 0.0001.
Therefore, study success has been demonstrated and has been shown to be effective in achieving a clinically relevant reduction in rectal radiation
The key secondary safety objective, demonstrating that the proportion of subjects with one or more acute Trade 2+ GI toxicities at 3 months for the Barrigel group is no worse than (non-inferior to) the proportion of subjects in the control group, using a non-inferiority margin of 10%, has been met.
The proportion of subjects with one or more acute Grade 2 or higher toxicities was 0.028 in the Barrigel group and 0.147 in the control group. The upper limit of the one-sided 95% CI (UCL) is for the difference in proportions (Barrigel minus control), -0.1189 and the two-sided Fisher's Exact p-value for the differences in proportion was 0.0348. As the UCL was less than the non-
{12}------------------------------------------------
inferiority margin of 0.10, Barrigel is shown to be non-inferior to the control and study success is demonstrated. In addition, as the two-sided Fisher's Exact test is < 0.05, Barrigel is shown to be superior to the control.
As indicated, per study protocol a sensitivity analysis of the key secondary safety objective was performed, and the safety objective was met. This supports the robustness and study success finding of the key secondary safety results. To assess any difference in treatment effect across subgroups, the key secondary safety objective was analyzed separately for three factors: ADT usage, erectile quality, and study site. The results show that Barrigel is consistently safe in all subgroups analyzed. These sensitivity and subgroup analysis support the robustness and study success finding of the key secondary safety results.
There were no reports of unexpected adverse device effects (UADEs) or Barrigelrelated adverse events.
Resorption of Barrigel following implantation was determined based on clinical data from independent retrospective and prospective clinical studies (Goni and RPAH1) of Restylane SubQ, which is identical in all respects to Barrigel except the amount of product contained in syringes of each product: 3 ml per syringe for Barrigel and 2 ml per syringe for Restylane SubQ. The typical amount of both Barrigel and Restylane SubQ per treatment is 9 ml, with variation according to individual patient physiology and at the discretion of the administering physician.
The Goñi data represented 27 male subjects who underwent perirectal injections of Restylane SubQ (2-6 ml) with subsequent radiotherapy and/or brachytherapy for prostate cancer. Follow-up MRIs were performed at a minimum of 60 months post injection. The MRIs indicated complete resorption, as evidenced by the absence of detectable gel by MRI, at 60 months post-injection. There were no indications of any adverse tissue reaction during degradation and absorption of the gel.
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" shape on the left, with the word "Palette" in a dark blue sans-serif font to the right of the "P". Below the word "Palette" is the phrase "LIFE SCIENCES" in a smaller, lighter font. The "P" shape is made up of three overlapping circles, with the colors transitioning from orange at the bottom to blue at the top.
The RPAH1 clinical study represented 36 male subjects injected with Restylane SubQ (10 ml) prior to IMRT (intensity-modulated radiation therapy) and IGRT (image-guided radiation therapy) for prostate cancer. Follow-up digital rectal examinations indicated material stability for the first 6 months following treatment with complete gel resorption for 18.5% of patients at 1-year posttreatment and complete gel resorption for 42.3% of patients after 2 years. No adverse tissue reactions were reported.
#### Substantial Equivalence 5.13
Biocompatibility and performance testing of Barrigel Injectable Gel, combined with results of the clinical studies, indicate that Barrigel is effective compared to control without increased risk and is substantially equivalent to the findings that supported FDA's approval and subsequent clearance of the predicate device.
Considering the complications and symptomology of rectal toxicity associated with radiation treatment for prostate cancer, which may last 6-18 months posttherapy, the benefits of Barrigel considerably outweigh the risks.
Both the subject and predicate device are low-to-moderate risk devices that are neither life-supporting nor life-sustaining, with risks that can be mitigated with appropriate use by medical professionals.
It is therefore reasonable to conclude that while Barrigel may have some different technological characteristics when compared to SpaceOAR, the benefits outweigh the potential risk of injury when used in accordance with the published instructions for use. The use of Barrigel does not represent new questions of safety or effectiveness and all attendant risks have been addressed by the provided test data and results.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Palette Life Sciences. The logo consists of a stylized "P" symbol on the left, with the word "Palette" in a sans-serif font to the right of the symbol. Below the word "Palette" are the words "LIFE SCIENCES" in a smaller font. The "P" symbol is made up of three overlapping shapes, with a gradient of colors ranging from orange to blue.
Based on this information, Barrigel is substantially equivalent to SpaceOAR in terms of intended use and technological characteristics. Performance data summarized above supports that any differences in technology do not affect the safety and effectiveness of Barrigel and the risk/benefit profile for Barrigel supports substantial equivalence with the proposed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.