K202224 · Boston Scientific Corporation · OVB · Aug 28, 2020 · Radiology
Device Facts
Record ID
K202224
Device Name
SpaceOAR System
Applicant
Boston Scientific Corporation
Product Code
OVB · Radiology
Decision Date
Aug 28, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SpaceOAR System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this the intent of SpaceOAR System to reduce the radiation dose delivered to the anterior rectum. The SpaceOAR System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.
Device Story
SpaceOAR System is a sterile, single-use kit for creating a temporary perirectal spacer. It consists of a synthetic, absorbable polyethylene glycol (PEG)-based hydrogel and a delivery mechanism. A clinician injects the hydrogel between the anterior rectal wall and the prostate. The hydrogel creates a physical space that persists for the duration of prostate radiotherapy (approx. 3 months) and is subsequently absorbed by the body via hydrolysis and renal excretion (approx. 6 months). By increasing the distance between the prostate and the rectum, the device reduces the radiation dose delivered to the anterior rectal wall, potentially minimizing radiation-induced side effects for the patient.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including sterilization validation, shelf life testing, needle-to-hub tensile strength, fluid connection testing, preparation/assembly time, vial puncture force, and packaging performance testing.
Technological Characteristics
Synthetic, absorbable PEG-based hydrogel; 10cc volume. Components include needle, vial adapter, and hydrogel precursors. Sterile, single-use. PEG hydrogel degrades via hydrolysis and is excreted renally. Packaging uses Tyvek and Biaxially Oriented Nylon/PE/COEX layers.
Indications for Use
Indicated for patients undergoing radiotherapy for prostate cancer to temporarily position the anterior rectal wall away from the prostate to reduce radiation dose to the rectum.
Regulatory Classification
Identification
An absorbable perirectal spacer is composed of biodegradable material that temporarily positions the anterior rectal wall away from the prostate during radiotherapy for prostate cancer with the intent to reduce the radiation dose delivered to the anterior rectum. The absorbable spacer maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.
Special Controls
An absorbable perirectal spacer must comply with the following special controls: (1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided. Performance bench testing must demonstrate appropriate perirectal space (i) creation and maintenance for the duration of prostate radiotherapy; Performance bench testing must demonstrate that therapeutic radiation levels (ii) do not alter the performance of the device; (iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment; (iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity: Sterility testing must demonstrate the sterility of the device and the effects of (v) the sterilization process on the physical characteristics of the spacer; (vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling: and. (vii) The device must be demonstrated to be biocompatible. (2) The risk management activities performed as part of the manufacturer's 21 CFR 820.30 design controls must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique. (3) The device labeling must include the following: A detailed summary of reported or observed complications related to the use (i) of the device; (ii) Appropriate warnings; Detailed instructions for system preparations and detailed implant procedure (iii) instructions; and, (iv) An expiration date that is supported by performance data as specified in subparagraph (b)(i)(vi).
*Classification.* Class II (special controls). The special controls for this device are:(1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided.
(i) Performance bench testing must demonstrate appropriate perirectal space creation and maintenance for the duration of prostate radiotherapy.
(ii) Performance bench testing must demonstrate that therapeutic radiation levels do not alter the performance of the device.
(iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment.
(iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity.
(v) Sterility testing must demonstrate the sterility of the device and the effects of the sterilization process on the physical characteristics of the spacer.
(vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling.
(vii) The device must be demonstrated to be biocompatible.
(2) The risk management activities performed as part of the manufacturer's § 820.10(c) of this chapter design and development activities must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique.
(3) The device labeling must include the following:
(i) A detailed summary of reported or observed complications related to the use of the device;
(ii) Appropriate warnings;
(iii) Detailed instructions for system preparations and detailed implant procedure instructions; and
(iv) An expiration date that is supported by performance data as specified in paragraph (b)(1)(vi) of this section.
{0}------------------------------------------------
August 28, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Boston Scientific Corporation % Ms. Jeanne O'Toole Principal Regulatory Affairs Specialist 100 Boston Scientific Way MARLBOROUGH MA 01752
Re: K202224
Trade/Device Name: SpaceOAR System Regulation Number: 21 CFR 892.5725 Regulation Name: Absorbable Perirectal Spacer Regulatory Class: Class II Product Code: OVB Dated: August 6, 2020 Received: August 7, 2020
Dear Ms. O'Toole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
{1}------------------------------------------------
2
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202224
Device Name SpaceOAR System
Indications for Use (Describe)
SpaceOAR System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this the intent of SpaceOAR System to reduce the radiation dose delivered to the anterior rectum.
The SpaceOAR System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5
#### 510(k) Summary for SpaceOAR Hydrogel System
#### A. Sponsor
Boston Scientific Corporation Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01752
#### B. Contact
Jeanne O'Toole Principal Regulatory Affairs Specialist 508-683-4271 jeanne.otoole@bsci.com
#### C. Proposed Device
Trade Name: SpaceOAR Hydrogel System, Model Number: SO-2101 Common / Usual Name: Hvdrogel Spacer Regulation Number: 892.5725 Classification Name: Absorbable Perirectal Spacer Classification: Class II Product Code: OVB Review Panel: Radiology
## D. Predicate Device
Trade Name: SpaceOAR Hydrogel System, Model Number: SO-2101 Common / Usual Name: Hydrogel Spacer Regulation Number: 892.5725 Classification Name: Absorbable Perirectal Spacer Classification: Class II Product Code: OVB Identification of Predicate Device: SpaceOAR Hydrogel System, K181465
## E. Device Description
The SpaceOAR Hydrogel System consists of components for the preparation of a synthetic, absorbable hydrogel spacer and a delivery mechanism provided in a sterile, single use package. The SpaceOAR hydrogel is a synthetic, absorbable polyethylene glycol (PEG)based hydrogel that upon injection creates a space that temporarily positions the anterior rectal wall away from the prostate during radiotherapy for prostate cancer, and in creating this space it is the intent of the perirectal spacer to reduce the radiation dose delivered to the anterior rectum. SpaceOAR hydrogel is completely synthetic with no animal or human derived components. It is composed of biodegradable material and maintains space for the
{4}------------------------------------------------
## SECTION 5
entire course of prostate radiotherapy treatment (approximately 3 months) and is completely absorbed by the patient's body over time (approximately 6 months).
## F. Indications for Use
SpaceOAR System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR System to reduce the radiation dose delivered to the anterior rectum. The SpaceOAR System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.
The Indications for Use Statement for the proposed SpaceOAR System is identical to the predicate device.
## G. Technological Characteristics Compared to Predicate
The principles of operation and underlying technological characteristics of the proposed SpaceOAR System and the predicate device are identical. Both devices are systems that facilitate implantation of 10cc PEG biodegradable hydrogel between the anterior rectal wall and prostate prior to radiotherapy. This temporarily creates space between the anterior rectal wall and prostate during radiotherapy. The PEG hydrogel is broken down through hydrolysis and excreted from the body through renal filtration.
The differences between the subject device and the predicate are minor. Differences include:
- The Bioset component is being replaced by an aluminum cap vial stopper, and vented ● vial adapter (VVA), that includes a hollow spike and a flange that enables the spike to be centered over the cap, and stopper.
- The 18g X 15cm needle hub is being modified to add a white colorant and a blue ● stamped marking to indicate needle bevel direction. There are no changes to the design specifications of the hub or needle.
- The system packaging tray is being modified to account for the change in components from the Bioset to the VVA. The tray material is unchanged.
- The outer pouch which acts as the sterile barrier is being modified. The bottom web material of the outer pouch has been changed from 48ga (0.48mil) PET layer and 200 Ga (2 ml) LDPE layer to 100 Ga (1ml) Biaxially Oriented Nylon/PE/COEX layer, 10# LDPE and 200 Ga (2 ml) Peelable Sealant. The top web Tyvek material is identical to that of the predicate device.
- The device labeling, including the instructions for use (IFU), is being modified to ● reflect system component changes.
The technological characteristics remain equivalent to the predicate device because the modifications that are the subject of this submission are limited to replacing the Bioset component, which has been discontinued by the manufacturer, as well as improvements to the existing design.
## H. Substantial Equivalence
{5}------------------------------------------------
## SECTION 5
The modified SpaceOAR System is substantially equivalent to the predicate SpaceOAR System (K181465). It has the same intended use of creating a temporary space between the prostate and the rectum and the same indication for use. The system design and principles of operation remain the same.
# I. Performance Data
The modifications to the subject SpaceOAR System have been tested to ensure compliance to the initial specifications. Based on the change assessment including risk analysis, the design verification tests outlined below were repeated. The test methods used were the same as those submitted for the predicate device.
- Sterilization Validation
- Shelf Life Testing ●
- System Needle to Hub Tensile Strength
- Fluid Connection
- SpaceOAR Hydrogel System Prep and Assembly Time
- PEG Vial Puncture Force ●
- . Packaging Performance Testing
The predicate and proposed device are predominantly identical in device design with limited changes to device components, packaging and labeling. A risk-based biocompatibility assessment of these limited changes indicates that no new biocompatibility testing is required to assess the proposed device change. The proposed SpaceOAR System is biologically safe for its intended use.
The principles of operation and underlying device technology are identical for both the predicate and proposed device. The clinical workflow for device implantation is identical for both the predicate and proposed device. Therefore, clinical data were not required to support a determination of substantial equivalence.
Verification and validation of the proposed device were completed and confirmed that the proposed SpaceOAR System meets the same functional and performance specifications as the predicate SpaceOAR System.
The conclusion of the assessments demonstrates that the modified device continues to function as intended in a manner equivalent to the predicate device. The modified device raises no new issues of safety or effectiveness compared to the predicate.
## J. Conclusion
Based on the test data, the same intended use, and same indications for use, the modified SpaceOAR System is substantially equivalent to its predicate device, K181465.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.