K220611 · Uro-1, Inc. · KNW · Dec 2, 2022 · Gastroenterology, Urology
Device Facts
Record ID
K220611
Device Name
SUREcore Plus Biopsy Instrument
Applicant
Uro-1, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Dec 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The SUREcore Plus biopsy instrument is intended for use in obtaining biopsies from soft tissue such as liver, kidney, and prostate. It is not intended for use in bone.
Device Story
SUREcore Plus is a single-use, spring-loaded biopsy instrument used in hospitals, clinics, and physician offices. It consists of a handle and a needle set. The physician manually cocks the handle to energize internal springs, then triggers the device to rapidly advance the needle into target tissue under imaging guidance (e.g., ultrasound, X-ray, CT). The needle's core collector captures a tissue sample. The device allows for multiple samples to be taken from different locations by re-cocking the handle. The collected tissue is removed for pathological analysis. The device is disposed of after the procedure. It benefits patients by facilitating minimally invasive tissue collection for disease diagnosis.
Clinical Evidence
No clinical or animal studies were conducted. Performance was established through bench testing, including ex vivo tissue collection capacity in liver, kidney, and prostate models, and mechanical force testing for needle deformation. Biocompatibility was established via reference to K201650.
Technological Characteristics
Manual spring-loaded biopsy instrument. Components: handle and 18-gauge needle set. Materials: biocompatible (ISO 10993 compliant). Dimensions: 5.5 in x 1.5 in x 1 in handle; 10-25 cm needle assembly; 19 mm sample notch. Operation: 2-stroke cocking lever, button-activated needle release. Sterilization: Gamma radiation. Single-use. No electronic, software, or electromagnetic components.
Indications for Use
Indicated for obtaining soft tissue biopsies (liver, kidney, prostate) in male and female patients. Not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2022
Uro-1, Inc. Thomas Lawson Regulatory Consultant 3701-A Alliance Drive Greensboro, North Carolina 27407
Re: K220611
Trade/Device Name: SUREcore Plus Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: February 28, 2022 Received: March 3, 2022
Dear Thomas Lawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Jessica Carr -S" in a large, bold font. The name is written in black and is centered in the image. The background of the image is a light blue color. The text is clear and easy to read.
for Long Chen Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220611
#### Device Name
SUREcore Plus Biopsy Instrument
#### Indications for Use (Describe)
The SUREcore Plus Biopsy Instrument is intended for use in obt tissue such as liver, kidney, and prostate. It is not intended for use in bone.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| [X] Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------|---------------------------------------------|
|--------------------------------------------------|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
# SECTION 5.
# 510(k) SUMMARY
# General Information
| Submitter | Uro-1, Inc. |
|-----------------------|---------------------------------------------------------|
| Address | 3703-A Alliance Drive<br>Greensboro, NC 27404 |
| Correspondence Person | Thomas Lawson, PhD<br>Regulatory Affairs<br>Uro-1, Inc. |
| Contact Information | Email: drthomlawson@gmail.com<br>Phone: 510-206-1794 |
| Date Prepared | 1 December 2022 |
## Proposed Device
| Trade Name | SUREcore Plus Biopsy Instrument |
|----------------------------------------------|---------------------------------------------------------------|
| Common Name | SUREcore Plus |
| Regulation Number and<br>Classification Name | 21 CFR§876.1075 Gastroenterology-urology biopsy<br>instrument |
| Product Code | KNW |
| Regulatory Class | II |
## Predicate Device
| Trade Name | Max-Core Biopsy Instrument |
|--------------------------------------------------------------------------|-------------------------------------------------|
| Common Name | Max-Core |
| Premarket Notification | K133948 |
| Regulation Number and | 21 CFR§876.1075 Gastroenterology-urology biopsy |
| Classification Name | instrument |
| Product Code | KNW |
| Regulatory Class | II |
| Note: This predicate device has been subject to a design-related recall. | |
{5}------------------------------------------------
### Device Description
The SUREcore Plus biopsy instrument facilitates collection of tissue for analysis by pathology in order to assist with a diagnosis of a disease condition in a patient. The SUREcore Plus device consists of two elements: (1) a handle and (2) a needle set. The handle contains springs that energize the biopsy needles and when used will cause rapid advance of the needles into target tissue. This action causes tissue to be held within a component of the needle (the core collector) intact from the body. The tissue samples are removed from the needle set and prepared for transfer to a pathologist or lab. The user can re-energize the springs by pulling back on the cocking slide so that multiple tissue samples can be collected from different locations within the target tissue. When the user has determined that sufficient amount of tissue has been obtained, the SUREcore Plus device is disposed in accordance with local and facility policies and procedures. The device is single-use only and is not to be resterilized by the user.
#### Indications for Use
The indication for use for the SUREcore Plus biopsy instrument is:
The SUREcore Plus biopsy instrument is intended for use in obtaining biopsies from soft tissue such as liver, kidney, and prostate. It is not intended for use in bone.
### Comparison to the Predicate Device
Uro-1, Inc. has identified the Max-Core Biopsy Instrument (Bard Medical) as the predicate device. The SUREcore Plus Biopsy Instrument is substantially equivalent to the predicate device based upon the following similarities:
- 1. The intended use of both the predicate device and the SUREcore Plus device are equivalent for the two devices, which is to obtain biopsies from soft tissues such as liver, kidney, and prostate. Neither are not intended for use in bone.
- Both devices introduce a biopsy needle into the body under imaging control (e.g., 2. ultrasound, X-Ray, CT, etc.)
- 3. Both devices are designed to collect multiple samples.
- Both devices are made from biocompatible materials. 4.
{6}------------------------------------------------
Table 1 shows the comparison of the SUREcore Plus Biopsy Instrument (the subject of this submission) to the predicate device, Max-Core Biopsy Instrument (K133948). The similarities between the two devices satisfy the criteria for a 510(k) notice.
| | Subject device | Predicate device |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| | SUREcore Plus<br>Biopsy Instrument<br>URO-1, Inc.<br><br>(this submission) | Max-Core Biopsy<br>Instrument<br>Bard Medical<br>(K133948) |
| Indication for use | For use in obtaining<br>biopsies from soft<br>tissues | Same |
| Intended use | The introduction of<br>the needle into the<br>body should be<br>carried out under<br>imaging control<br>(ultrasound, X-Ray,<br>CT, etc.). It is<br>provided sterile for<br>single use. | Same |
| Route of<br>advancement | Percutaneous | Same |
| Target populations | Male & Female | Same |
| Location of biopsy | Liver, kidney,<br>and prostate. | Liver, kidney, prostate,<br>spleen, lymph nodes and<br>various soft tissue<br>tumors. |
| Site of use | Hospitals, clinics,<br>and physician<br>offices | Same |
| Device Features | | |
| Components | (1) Handle<br>(2) Needle set consisting<br>of a core collector<br>and an outer cannula | Same |
| Dimension of the handle | 5.5 in X 1.5 in X 1 in | 6 in X 1.4 in (cylinder diameter) |
| Mechanics of energizing the needle | 2-stroke cocking action (using a lever on the handle) | Same |
| Mechanics of releasing the needle to puncture tissue | Pressing on an activator button on the handle | Same |
| Size of needle | 18 gauge | 14 to 20 ga |
| Length of needle tissue collection trough (sample notch) | 19 mm | Same |
| Length of needle assembly | 10-25 cm | Same |
| <b>Performance</b> | | |
| Depth of penetration | 22 mm | Same |
| Sterilization | Gamma Radiation | ETO |
| Frequency of use | Single patient use | Same |
| Tissue contact materials | Compliant with ISO 10993 | Same |
Table 1. Comparison of the SUREcore Plus Biopsy Instrument to the predicate device, the Bard Max-Core Biopsy Instrument.
{7}------------------------------------------------
## Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the subject device are substantially equivalent to those of the predicate device in terms of the following:
- . Equivalent intended use;
- Equivalen indications for use; ●
- Similar penetration depth;
- Similar sample notch; .
- Same mechanics of action; ●
- Same mode of action;
- Same energy used/delivered;
- . Similar patient-contacting materials;
- . Same fundamental scientific technology;
- . Same patient population;
- Single-use device; .
{8}------------------------------------------------
- Supplied sterile; and ●
- Similar packaging.
#### Performance Data
The performance testing conducted establishes that SUREcore Plus Biopsy Instrument does not raise new questions of the safety and effectiveness for a biopsy system.
#### Biocompatibility testing
The handle of the SUREcore Plus Biopsy Instrument does not come into contact with the patient, but it has been assessed for cytotoxicity and found to pass such testing. The VMCore needle set does contact patient tissue and passed all tests, as noted in K201650:
- Cytotoxicity, O
- O Sensitization,
- Irritation, and O
- Systemic toxicity. o
In this respect, the VMCore needle set is a reference device for the biocompatibility information for this submission related to the needle set.
### Electrical safety and electromagnetic compatibility (EMC)
The subject and predicate devices do not have electronic components, so such testing was not required.
### Software Verification and Validation Testing
Neither the subject nor predicate devices contain software.
### Mechanical Testing
The mechanical testing of the subject device included:
- . Capacity to collect tissue in ex vivo models of liver, kidney & prostate tissue, and
- Force necessary to deform the needle set.
{9}------------------------------------------------
#### Preclinical (Animal) Studies
Bench testing was sufficient to demonstrate performance of the device. No preclinical testing of the subject device was necessary.
#### Clinical Studies
Bench testing was sufficient to demonstrate performance of the device. No clinical testing of the subject device was necessary.
#### Conclusion
The information submitted in this premarket notification confirms that the SUREcore Plus Biopsy Instrument raises no new questions of safety and effectiveness and that the SUREcore Plus Biopsy Instrument is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.