Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702
K220572 · Ceragem Co., Ltd. · JFB · Jun 17, 2022 · Physical Medicine
Device Facts
Record ID
K220572
Device Name
Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702
Applicant
Ceragem Co., Ltd.
Product Code
JFB · Physical Medicine
Decision Date
Jun 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for: -Temporary relief of minor muscle and joint pain stiffness -Temporary relief of minor joint pain associated with arthritis - -Temporary increase in local circulation where applied - Relaxation of muscles The Air Cell Massager (only CGM MB-1701) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas. The Air Cell Massager stimulates kneading of tissues by using an inflatable garment.
Device Story
Ceragem Automatic Thermal Massager is a multi-function physical therapy table for home use. It delivers heat and massage via internal ceramic rollers and external projectors. The CGM MB-1701 model includes an integrated Air Cell Massager for leg therapy. The Air Cell Massager uses an air pump, pressure sensors, and inflatable sleeves with 4 compression chambers to provide sequential or peristaltic kneading and stroking of tissues from foot to thigh. A microprocessor controls inflation/deflation cycles and pressure levels. Users operate the device via a remote control to select modes and adjust settings. The device provides radiant infrared heat and mechanical massage to promote muscle relaxation and local circulation. Benefits include temporary pain relief and improved circulation. The system includes safety features like overheating protection and emergency stop buttons.
Clinical Evidence
Bench testing only. Includes skin temperature study, usability study for lay users, electrical safety and EMC testing (ANSI/AAMI ES60601-1, IEC 60601-1-2/6/11), and biocompatibility testing (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Multi-function physical therapy table. Materials: Steel, ABS frame; Polyester/Cotton/Rayon/PU fabric; Nylon/PU sleeves. Energy: 100-127Vac, 50/60Hz. Sensing: Thermistor (temperature), air pressure sensor. Actuation: Geared DC motor (rollers), air pump (compression). Connectivity: Standalone. Software: Microcontroller-based. Sterilization: N/A. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2/6/11, ISO 10993-5/10.
Indications for Use
Indicated for individuals seeking temporary relief of minor muscle and joint pain, stiffness, and arthritis-associated joint pain, and for temporary increase in local circulation. The Air Cell Massager (CGM MB-1701) is indicated for temporary relief of minor muscle aches and pains and increased circulation in treated areas.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
Predicate Devices
Ceragem Automatic Thermal Massager Model CGM-MB-1701 & CGM MB-1702 (K202937)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2022
CERAGEM Co. Ltd. % Joyce Kwon CEO Provision Consulting Group, Inc. 100 Barranca St. Suite 700 West Covina, California 91791
Re: K220572
Trade/Device Name: Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702 Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB, IRP Dated: May 16, 2022 Received: May 19, 2022
Dear Joyce Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220572
Device Name
Ceragem Automatic Thermal Massager, Model CGM-MB-1701 & CGM-MB-1702
Indications for Use (Describe)
The intended use of the Ceragem Automatic Thermal Massager,
Model CGM MB-1701 & CGM MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
-Temporary relief of minor muscle and joint pain stiffness
-Temporary relief of minor joint pain associated with arthritis
- -Temporary increase in local circulation where applied
- Relaxation of muscles
The Air Cell Massager (only CGM MB-1701) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas. The Air Cell Massager stimulates kneading of tissues by using an inflatable garment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Provision Consulting Group. The logo features a stylized green bird with its wings spread, facing left. To the right of the bird is the word "PROVISION" in bold, followed by "CONSULTING GROUP" in a smaller font size underneath. The logo has a clean and professional design.
# 510(k) Summary
# 510(k) Submitter
Hyeonggeun Hwang CERAGEM Co, Ltd. 10, Jeongja 1-gil, Seonggeo-eup, Seobuk-gu, Cheonan-si, Chungcheongnam-do 31041 Republic of Korea
# Official Correspondent / Contact Person
Joyce Kwon, CEO/Principal Consultant Provision Consulting Group, Inc. 100 N. Barranca St. Suite 700 West Covina, CA 91791 +1-909-493-3276 joyce@provisionfda.com
# Date Prepared
June 17, 2022
## Device Information
- Trade Name: Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702 .
- . Common Name: Physical Therapy Table
- Classification Name: Multi-function physical therapy table .
- . Regulation Number: 21 CFR 890.5880
- Device Class: Class II .
- . Product Code: JFB, IRP
- . 510(k) Identification Number: K220572
### Primary Predicate Devices
Ceragem Automatic Thermal Massager Model CGM-MB-1701 & CGM-MB-1702, K202937 The predicate has not been subject to a design-related recall.
### Reference Device:
- Air Compression Therapy Device, K193354 .
### Prior Submission Information
None
### Indication for Use
The intended use of the Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
- Temporary relief of minor muscle and joint pain stiffness
- Temporary relief of minor joint pain associated with arthritis
- Temporary increase in local circulation where applied
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Provision Consulting Group. The logo features a stylized bird with outstretched wings on the left side. To the right of the bird is the word "PROVISION" in a bold, sans-serif font, and below that is "CONSULTING GROUP" in a smaller font.
- Relaxation of muscles
The Air Cell Massager (only CGM MB-1701) is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas. The Air Cell Massager stimulates kneading and stroking of tissues by using an inflatable garment.
# Device Description
The Ceragem Automatic Thermal Massager with Air Cell Massager is an updated version of the predicate device with an addition of a leg massaging feature.
This feature utilizes an "air compression transfer," and its principle is as follows:
Air Cell Massager consists of an air pump, air pressure sensor, and sleeves working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; each sleeve has 4 compression chambers. The compression massage direction is from foot to thigh. By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. Software controls the timing and pressure cognized to the sensor, cycling airflow into and out of sleeves to compress body.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Provision Consulting Group. The logo features a stylized bird with outstretched wings in a muted green color. To the right of the bird is the company name, "PROVISION," in a larger, bolder font, with the words "CONSULTING GROUP" underneath in a smaller font.
# Substantial Equivalent Comparison Chart with Primary Predicate Device
| | Subject Device | Primary Predicate Device | Comparison | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-----------------------|
| Device Name | Ceragem Automatic Thermal Massager,<br>Model CGM MB-1701 & CGM MB-1702 | Ceragem Automatic Thermal Massager,<br>Model CGM MB-1701&<br>CGM MB-1702 | | |
| Manufacturer | Ceragem International, Inc. | Ceragem International, Inc. | | |
| 510(k) Number | K220572 | K202937 | | |
| Regulation Number | 21 CFR 890.5880,<br>21 CFR 890.5650 | 21 CFR 890.5880 | N/A | |
| Product Code | JFB, IRP | JFB | | |
| Indications for Use<br>(IFU) | The intended use of the Ceragem Automatic<br>Thermal Massager,<br>Model CGM MB-1701 & CGM MB-1702 is to<br>provide muscle relaxation therapy by delivering<br>heat and soothing massage. Additionally, the<br>product provides topical radiant infrared heat<br>for:<br>-Temporary relief of minor muscle and joint<br>pain stiffness<br>-Temporary relief of minor joint pain associated<br>with arthritis<br>-Temporary increase in local circulation where<br>applied<br>- Relaxation of muscles<br>The Air Cell Massager (only CGM MB-1701) is<br>indicated for the temporary relief of minor<br>muscle aches and pains and for temporary<br>increase in blood circulation to the treated areas.<br>The Air Cell Massager stimulates kneading and<br>stroking of tissues by using an inflatable<br>garment. | The intended use of the Ceragem Automatic<br>Thermal Massager,<br>Model CGM MB-1701 & CGM MB-1702<br>is to provide muscle relaxation therapy by<br>delivering heat and soothing massage.<br>Additionally, the product provides topical<br>radiant infrared heat for:<br>- Temporary relief of minor muscle and<br>joint pain stiffness<br>- Temporary relief of minor joint pain<br>associated with arthritis<br>- Temporary increase in local circulation<br>where applied<br>- Relaxation of muscles | Similar | |
| Rated Voltage | 100-127Vac 50/60Hz | 100-127Vac 50/60Hz | Same | |
| | Home Use | Yes | Yes | |
| | Style | Flood Model (Massage Bed) | Flood Model (Massage Bed) | |
| | Components | Main Table | Main Table | Same |
| | | Supporting Mat | Supporting Mat | |
| | | Remote Control | Remote Control | |
| | | 3-Sphere projector | 3-Sphere projector | |
| | | Abdominal Vibration Projector | Abdominal Vibration Projector | |
| | Components | Power cord | Power cord | |
| | | Outer fabric | Outer fabric | |
| | | Head cushion | Head cushion | |
| | | Projector Cover | Projector Cover | |
| | | Air Cell Massager | Calf Massager | Different<br>(Note 1) |
| | Remote Control | Yes | Yes | |
| | Operation Method | Auto / Manual | Auto / Manual | |
| | Infrared Emission<br>Spectrum | Ceramic: 5~20<br>Epoxy Carbon Panel: 5~20 | Ceramic: 5~20<br>Epoxy Carbon Panel: 5~20 | |
| Heating<br>Device | Voltage | 24V | 24V | |
| | Power | 28.8W | 28.8W | |
| | Temperature Range | Internal: 30°C - 65°C (86°F - 149°F)<br>External, Main, Auxiliary: 30°C - 60°C (86°F -<br>140°F) | Internal: 30°C - 65°C (86°F - 149°F)<br>External, Main, Auxiliary: 30°C - 60°C (86°F -<br>140°F) | |
| | Distance of the Internal<br>Projector | 710mm | 710mm | |
| | Intensity Level | 1~9<br>(12.9~69.2mm) | 1~9<br>(12.9~69.2mm) | |
| | Extra Overheating<br>Protection | Yes | Yes | |
| | Moving Device | Geared DC Motor | Geared DC Motor | |
| | Limit Detector | Limit Switch | Limit Switch | |
| | Temperature Sensor | Thermistor | Thermistor | Same |
| | Control Method | Microcontroller | Microcontroller | |
| | Tugging Method | Wire-Chain | Wire-Chain | |
| Material | Main/Supporting Frame | Steel, ABS | Steel, ABS | |
| | Outer Fabric | Polyester, Cotton, Rayon, Polyurethane | Polyester, Cotton, Rayon, Polyurethane | |
| | Main Mat/Supporting Mat | Nylon<br>Polyurethane | Nylon<br>Polyurethane | |
| | | | | |
| Projectors | Internal Projector | Ceramic Rollers | Ceramic Rollers | |
| | External Projector | Ceramic Heads Projector | Ceramic Heads Projector | |
| Mat Dimensions | | • When spread out:<br>: 728.2mm x 2012.4mm x 431.5mm (±5mm)<br>• When folded:<br>: 728.2mm x 2044mm x 431.5mm (±5mm) | • When spread out:<br>: 728.2mm x 2012.4mm x 431.5mm (±5mm)<br>• When folded:<br>: 728.2mm x 2044mm x 431.5mm (±5mm) | |
| | Weight | Weight: 62kg for CGM MB-1701,<br>57kg for CGM MB-1702 | Weight: 62kg for CGM MB-1701,<br>57kg for CGM MB-1702 | |
| | Mode | Mode A | Mode A | |
| | | Mode 1 | Mode 1 | |
| | | Mode 2 | Mode 2 | |
| | | Mode 3 | Mode 3 | |
| | | Mode 4…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.