Ceragem Automatic Thermal Massager Model CGM-MB-1701 & CGM-MB-1702
K202937 · Ceragem Co., Ltd. · JFB · Mar 18, 2021 · Physical Medicine
Device Facts
Record ID
K202937
Device Name
Ceragem Automatic Thermal Massager Model CGM-MB-1701 & CGM-MB-1702
Applicant
Ceragem Co., Ltd.
Product Code
JFB · Physical Medicine
Decision Date
Mar 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Ceragem Automatic Thermal Massager, Model CGM-MB-1701 & CGM-MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for: - Temporary relief of minor muscle and joint pain stiffness - Temporary relief of minor joint pain associated with arthritis - Temporary increase in local circulation where applied - Relaxation of muscles
Device Story
Electrically powered, motorized physical therapy table for home use. Device inputs: user-selected settings via remote control. Principle of operation: internal geared DC motor moves ceramic rollers vertically along spinal line to provide massage pressure; heating pads and ceramic projectors deliver radiant infrared heat (5-20μm). Output: thermal massage and infrared heat. Operated by patient (self-use). Healthcare provider reviews output indirectly via patient report of symptom relief. Benefits: temporary pain relief, increased local circulation, muscle relaxation. Subject device adds air massage module for calves (CGM-MB-1701) and vibration feature to abdominal projector compared to predicate.
Clinical Evidence
Bench testing only. Includes skin temperature studies verifying maintenance of 40-45°C, usability testing for lay-user operation, EMC/electrical safety testing (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11), and biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 skin irritation/sensitization).
Indicated for individuals seeking temporary relief of minor muscle and joint pain, stiffness, and minor joint pain associated with arthritis, as well as for muscle relaxation and temporary increase in local circulation.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
Predicate Devices
Automatic Thermal Massager, Model CGM MB-1101 (K140592)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
March 18, 2021
Ceragem Co, Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 1150 Roosevelt STE 200 Irvine, California 92620
Re: K202937
Trade/Device Name: Ceragem Automatic Thermal Massager Model CGM-MB-1701 & CGM-MB-1702 Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB Dated: December 14, 2020 Received: December 22, 2020
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jitendra Virani Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K202937
#### Device Name
Ceragem Automatic Thermal Massager, Model CGM-MB-1701 & CGM-MB-1702
Indications for Use (Describe)
The intended use of the Ceragem Automatic Thermal Massager, Model CGM-MB-1701 & CGM-MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
- Temporary relief of minor muscle and joint pain stiffness
- Temporary relief of minor joint pain associated with arthritis
- Temporary increase in local circulation where applied
- Relaxation of muscles
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K202937)
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Feb 23, 2021
### 1. 510K Applicant / Submitter:
CERAGEM Co. Ltd. 10, Jeongja 1-gil, Seonggeo-eup, Seobuk-gu, Cheonan-si, Chungcheongnam-do 31041 Republic of Korea
#### 2. Submission Contact Person
LK Consulting Group USA, Inc. 1150 Roosevelt, STE 200, Irvine CA 92620 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
### 3. Device
- · Proprietary Name: Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702
- Regulation Number: 21 CFR 890.5880 .
- Regulation Name: Multi-Function Physical Therapy Table .
- Regulatory Class: Class II .
- · Product Code: JFB
### 4. Predicate Device
Automatic Thermal Massager, Model CGM MB-1101 (K140592) by Ceragem International, Inc.
### 5. Description:
The Ceragem Automatic Thermal Massager which offers two models, CGM MB-1701 & CGM MB-1702 is an electrically powered, motorized, multi-functional physical therapy table intended to use for temporary relief of muscle/joint pain and stiffness by applying massage
{4}------------------------------------------------
pressure from up and down movement of the internal motor, and heat from the internal and external projectors.
The main components of the subject device are a main table mat, a supporting mat, a 3-Sphere projector, an abdominal vibration projector, and a remote control. The main table has an internal projector which includes an up/down movement motor for pressure effect with ceramic balls as a medium for thermal massage effect. The internal projector is located in the center of the mat which is mainly for spine and up/down and vertically along the spinal line. The main table mat has a heating pad for thermal effect.
The 3-sphere projector is an optional accessory which the user can put anywhere needed additionally such as neck and armpit, and has 3 balls for thermal effect. The abdominal vibration projector is also an optional accessory mainly for abdominal area and offers vibration and thermal effect.
The only difference between the model CGM MB-1701 and the CGM MB-1702 is that the CGM MB-1701 has an air massage module for calf on the supporting mat. The supporting mat of the both models has a heating pad.
### 8. Indications for Use
The intended use of the Ceragem Automatic Thermal Massager,
Model CGM MB-1701 & CGM MB-1702 is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:
- Temporary relief of minor muscle and joint pain stiffness
- Temporary relief of minor joint pain associated with arthritis
- Temporary increase in local circulation where applied
- Relaxation of muscles
#### 9. Substantial Equivalence Discussion:
#### 9.1. Comparison Chart
| | Subject Device | Predicate Device | |
|----------------|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Device Name | Ceragem Automatic Thermal Massager,<br>Model CGM MB-1701&CGM MB-1702 | CGM MB-1101<br>Automatic Thermal Massager | |
| Manufacturer | Ceragem International, Inc. | Ceragem International, Inc. | |
| 510(k) Number | K202937 | K140592 | |
| Rated Voltage | 100-127Vac 50/60Hz | 100-127Vac 50/60Hz | |
| Home Use | Yes | Yes | |
| Style | Flood Model (Massage Bed) | Flood Model (Massage Bed) | |
| Components | Main Table<br>Supporting Mat<br>Remote Control<br>3-Sphere projector | Main Table<br>Supporting Mat<br>Remote Control<br>3-Sphere projector | |
| | | Abdominal Vibration Projector | 9-Sphere projector |
| | | Power cord | Power cord |
| | | Outer fabric | Outer fabric |
| | | Head cushion | Head cushion |
| | | Projector Cover | Projector Cover |
| | | Calf Massager | |
| | Remote Control | Yes | Yes |
| | Operation Method | Auto / Manual | Auto / Manual |
| | Infrared Emission Spectrum | Ceramic: 5~20μm | Jade: 5~20µm |
| | | Epoxy Carbon Panel: 5~20 µm | Epoxy Carbon Panel: 5~20 μm |
| Heating | Voltage | 24V | 24V |
| Device | Power | 28.8W | 15W |
| | Temperature Range | • Internal: 30°C - 65°C (86°F - 149°F)<br>• External, Main, Auxiliary: 30°C - 60°C(86°F - 140°F) | • Internal: 30°C - 65°C (86°F - 49°F)<br>• External, Main, Auxiliary: 30°C - 60°C(86°F - 140°F) |
| | Distance of the Internal Projector | 710mm | 710mm |
| | Intensity Level | 1~9<br>(12.9~69.2mm) | 1~6<br>(21~75mm) |
| | Extra Overheating Protection | Yes | Yes |
| | Moving Device | Geared DC Motor | Geared DC Motor |
| | Limit Detector | Limit Switch | Limit Switch |
| | Temperature Sensor | Thermistor | Thermistor |
| | Control Method | Microcontroller | Microcontroller |
| | Tugging Method | Wire-Chain | Wire-Chain |
| Material | Main/Supporting<br>Frame | Steel, ABS | Steel, ABS |
| | Outer Fabric | Polyester, Cotton, Rayon, Polyurethane | Polyester, Cotton, Rayon,<br>Polyurethane |
| | Main<br>Mat/Supporting Mat | Nylon<br>Polyurethane | Polyurethane |
| Projectors | Internal<br>Projector | Ceramic Rollers | Jade Rollers |
| | External<br>Projector | Ceramic Heads Projector | Jade Heads Projector |
| Mat Dimensions | | • When spread out<br>728.2mmx2012.4mmx431.5mm(±5mm)<br>• When folded<br>: 728.2mmx2044mmx431.5mm(±5mm) | • When spread out<br>700mmx2016mmx450mm(±5mm)<br>• When folded<br>700mmx1258mmx450mm(±5mm) |
| | Weight | Weight: 62kg for CGM MB-1701,<br>57kg for CGM MB-1702 | 53 kg |
| | | Mode A | Mode 1 |
| | | Mode 1 | Mode 2 |
| | | Mode 2 | Mode 3 |
| | | Mode 3 | Mode 4 |
| | | Mode 4 | Mode 5 |
| Mode | | Mode 5 | Mode 6 |
| | | Mode 6 | Mode 7 |
| | | Mode 7 | Mode 8 |
| | | Mode 8 | Mode 9 |
| | | Mode 9 | Semi-Automatic Mode |
| | | Mode 10 | Manual Mode |
| | | Mode 11 | |
{5}------------------------------------------------
{6}------------------------------------------------
| Mode 12 | |
|------------------------------------|--|
| Mode 13 | |
| Mode 14 | |
| Mode 15 | |
| Intensive Mode | |
| Semi-Automatic Mode | |
| Semi-Automatic Master Mode | |
| Manual Mode | |
| Manual Master Mode | |
| Abdominal Vibration Projector Mode | |
| Calf Mode | |
### 9.2. Substantial Equivalence Discussion
The subject device is identical to the predicate device in the indications for use, principle of operation, and technological characteristics. The performance specifications are nearly the same, yet the subject device has more choosing options. For example, the intensity level specification is the same between the subject device and the predicate device, but the subject device divided the range into 9 options instead of 6 to better meet user's intensity preferences. The same thing applies to the massage mode that the overall massage specification is the same, but the subject device offers more variety of the massage patterns and options.
The major difference is the material of the internal heating element has changed. We have performed the skin temperature to validate the performance of the heating massage and the test results support that the subject device is substantially equivalent to the predicate device. Other differences are that the subject device has upgraded the 9-Sphere projector to have vibration feature for massage effect, and the subject model, CGM MB-1701, has an air massage module for calf on the supporting mat. These vibration and air massage functions are just additional features and still the main function is heating. The skin temperature test result supported that the Abdominal Vibration Projector and the supporting mat perform as well as the predicate device. Similar to the predicate device, the subject device was able to heat skin to 40-45 degree C within 15-20 minutes. During the intended treatment the subject device maintained skin temperature between 40-45 degree C.
### 10. Performance Tests (Non-clinical)
- The skin temperature study was performed to verify that the subject device maintained skin temperature between 40-45 degree C during the intended treatment.
- The usability study was performed to validate that the lay users can use the subject ● device only with the User Manual provided.
- . The EMC and electrical safety testing were conducted on the subject device in accordance with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, andIEC 60601-1-11.
- The cytotoxicity test per ISO 10993-5 and skin irritation & sensitization tests per ISO ● 10993-10 were performed to evaluate the safety of the new materials which contact the
{7}------------------------------------------------
patients.
- The level of concern for the subject device SW is moderate and the validation tests were . performed to verify that the firmware works as intended. The test results of all the tests supported that it is substantially equivalent to the predicate device.
### 11. Conclusions:
Based on the information provided in this premarket notification. CERAGEM Co, Ltd. concludes that the Ceragem Automatic Thermal Massager, Model CGM MB-1701 & CGM MB-1702 are substantially equivalent to the predicate device as described herein in.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.