Disposable surgical face mask

K220302 · Shandong Aida Medical Products Co., Ltd. · FXX · Nov 16, 2022 · General, Plastic Surgery

Device Facts

Record IDK220302
Device NameDisposable surgical face mask
ApplicantShandong Aida Medical Products Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateNov 16, 2022
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Disposable surgical face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

Disposable, flat-pleated surgical face mask; three-layer construction; outer/inner layers spun-bond polypropylene; middle layer melt-blown polypropylene filter. Features nose piece (polypropylene/iron) and ear loops (polyester/spandex). Used in clinical settings by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Functions via particle adsorption, mechanical sieve retention, inertial deposition, and electrostatic adsorption. Provided non-sterile; single-use only.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-2020, ASTM F2101 (BFE 99.9%), ASTM F2299 (Particulate Filtration Efficiency), EN 14683 (Differential Pressure 4.2-4.8 mmH2O/cm2), ASTM F1862 (Synthetic blood penetration 120 mmHg), and 16 CFR 1610 (Flammability Class I). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Three-layer nonwoven polypropylene mask; melt-blown filter layer. Dimensions: 17.5 cm x 9.5 cm. Nose piece: polypropylene/iron. Ear loops: polyester/spandex. Non-sterile. Meets ASTM F2100-2020 Level 2 standards. No software or electronic components.

Indications for Use

Indicated for use by healthcare personnel and patients as a single-use, disposable barrier to reduce exposure to microorganisms, body fluids, and particulate materials during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 16, 2022 Shandong Aida Medical Products Co., Ltd Wang Bing Technical Manager Wangjiazhuang, Wangliu Street Office, Weicheng District Weifang, Shandong 261053 China Re: K220302 Trade/Device Name: Disposable surgical face mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 17, 2022 Received: October 17, 2022 Dear Wang Bing: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K220302 Device Name Disposable surgical face mask #### Indications for Use (Describe) The Disposable surgical face mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### K220302 page 1 of 8 ## B-2. 510(K) Summary Summary Date: July 23, 2021 | 1 Submitted by : | Shandong Aida Medical Products Co.,Ltd<br>Wangjiazhuang, Wangliu Street Office, Weicheng District,<br>Weifang City, Shandong province, P.R.China, 261053<br>0086-536-8131936<br>zoro@aidamasks.com | |------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| 2 Contact Person: Primary Contact: WangBing Wangjiazhuang, Wangliu Street Office,Weicheng District, Weifang City, Shandong Province, P.R. China 0086-15263613795 zoro@aaidamasks.com > Secondary Contact: Yu guangjiang Wangjiazhuang, Wangliu Street Office, Weicheng District, Weifang City, Shandong Province P.R. China 0086-13792660238 914424647@qq.com #### 3 Device Information Trade Name: Disposable surgical face mask Common/Usual Name: Disposable surgical mask Model: AD-PR02, AD-PR02-2,AD-PR02-S01,AD-PR02-S02 {4}------------------------------------------------ Regulatory Information Classification Name: Mask, Surgical Classification: Class II Product Code: FXX Regulation Number: 21 CFR 878.4040 Regulation Description: Surgical Apparel #### 4 Legally Marketed Predicate device: Trade Name: Disposable Surgical Face Mask 510(K) Number: K153496 Product Code: FXX Manufacturer: Xiantao Rayxin Medical Products Co.,Ltd. #### 5 Device Description: The proposed device is blue color, and Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the facemask around the nose. The proposed device is manufactured with three layers, the inner and outer layers are made of spun-bond Nonwoven fabric polypropylene, and the middle layer is made of melt blown polypropylene filter. The filter fiber is used for particle adsorption, mechanical sieve retention, inertial deposition, and electrostatic adsorption deposition, so that the air containing harmful substances is filtered through the filter material of the mask and then inhaled or exhaled. The style of the device, ear loops, is held in place over the users' mouth and nose by two ear loops welded to the facemask. The ear loops are not made with natural rubber latex but made of polyester and spandex. The nose piece contained in the proposed device is in the layers of facemask to allow {5}------------------------------------------------ the user to fit the facemask around their nose, which is made of polypropylene and iron. The proposed device should be stored in a dry, ventilated and pollution-free area; the temperature should be 0-40 ℃, and the humidity should not be higher than 80%; Keep away from fire sources, avoid dust exposure and direct sunlight. The proposed device is sold non-sterile and is intended to be a single use, disposable device. #### 6 Intended Use/Indications for Use: The Disposable surgical face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. #### 7 Performance Data ### Non-Clinical Test Conclusion: Non-clinical tests were conducted to verify that the device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the device complies with the following standards: #### Biocompatibility: The Disposable surgical face mask has been subjected to biocompatibility studies to demonstrate the safety of device. The biocompatibility studies are in accordance with ISO10993: - A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity - A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests {6}------------------------------------------------ for Irritation and Skin Sensitization. #### Performance: - A ASTM F2100-2020, Standard Specification for Performance of Materials Used in Medical Face Masks. - ASTM F2101-2019, Standard Test Method for Evaluating the Bacterial Filtration A Efficiency (BFE) Of Medical Face Mask Materials, Using a nebulizer and a culture suspension of staphylococcus aureus. - A EN 14683:2019+AC:2019 Annex C, Standard Test Method for differential pressure of Medical Face Masks - A ASTM F2299/F2299M-03(Reapproved 2017), Standard test method for determining the initial filtration efficiency of materials used in medical face masks to penetration by particulates using latex sphere aerosol. - ASTM F1862/F1862M-2017, Standard Test Method for Resistance of Medical △ Face Masks to penetration by synthetic blood. - A 16 CFR Part 1610-2008, Standard for the Flammability of clothing textiles. #### Clinical Test Conclusion : No clinical study is included in this submission. #### 8 Substantially Equivalent (SE) Comparison with The Predicate Device: Comparison with the predicate devices, the device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table | Table B-II-1 General Comparison | | |---------------------------------|--| | | | | | | | ITEM | Proposed Device | Predicate | Remark | |------|-----------------|------------------------------|--------| | | | Predicate<br>Device(K153496) | | {7}------------------------------------------------ ### K220302 page 5 of 8 | Manufacturer | Shandong Aida Medical<br>Products Co.,Ltd | Xiantao Rayxin Medical<br>Products Co.,Ltd. | | | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-----------| | Trade name | Disposable surgical face<br>mask | Disposable Surgical<br>Face Mask | | | | Classification | Class II Device, FXX (21<br>CFR § 878.4040) | Class II Device, FXX<br>(21<br>CFR § 878.4040) | SE | | | Indications for use | The Disposable surgical<br>face mask is intended to be<br>worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>materials in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This<br>is a single use, disposable<br>device(s), provided<br>non-sterile. | The Disposable Surgical<br>Face Masks are intended<br>to be worn to protect<br>both the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device(s),<br>provided non-sterile. | SE | | | Mask Styles and<br>Design features | Ear Loops, Flat Pleated, 3<br>layers | Ear Loops/Tie-on, Flat<br>Pleated, 3 layers | Different | | | Materials | | | | | | | Outer Facing<br>Layer | Spun-bond Nonwoven<br>fabric polypropylene | Spun-bond<br>polypropylene | SE | | | Middle layer | Melt blown polypropylene<br>filter | Melt blown<br>polypropylene filter | SE | | | Inner Facing<br>Layer | Spun-bond Nonwoven<br>fabric polypropylene | Spun-bond<br>polypropylene | SE | | | Nose clip | Polypropylene and iron | Malleable aluminum<br>wire | Different | | | Tie-one | | Spun-bond<br>polypropylene | Different | | | Ear loops | Polyester and spandex | Polyester | Different | | Color | | Blue | Blue | SE | | Dimension (Width) | | 17.5 cm +/- 5% | 17.5 cm +/- 1 cm | Different | | Dimension (Length) | | 9.5 cm +/- 5% | 9.5 cm +/- 1 cm | Different | | OTC use | | Yes | Yes | SE | | Sterility | | No | No | SE | | Single Use | | Yes | Yes | SE | | ASTM F2100-2020 | | Level 2 | Level 2 | SE | | Biocompatibility | | ISO10993 | ISO10993 | SE | {8}------------------------------------------------ ### K220302 page 6 of 8 Table B-II-2 Performance Characteristic Comparison | Test Methodology | Proposed Device | Predicate<br>Device(K153496) | ASTM F2100-2020<br>Requirements for Level<br>2 Classification | Remark | |----------------------------------------------------|----------------------------------|----------------------------------|---------------------------------------------------------------|--------| | Resistance to<br>Penetration by<br>Synthetic Blood | 32 out of 32 pass<br>at 120 mmHg | 32 out of 32 pass<br>at 120 mmHg | 29 out of 32 pass at<br>120 mmHg | Meet | {9}------------------------------------------------ K220302 page 7 of 8 | ASTM F1862 | | | | | |---------------------------------------------------|--------------------------------------------------|--------------------------|------------------|------| | Particulate<br>Filtration Efficiency<br>ASTMF2299 | Pass | 98.46% | $\u2265$ 98% | Meet | | Bacterial Filtration<br>Efficiency<br>ASTMF2101 | 99.9% | 98.7% | $\u2265$ 98% | Meet | | Differential<br>Pressure (Delta P)<br>EN 14683 | Max: 4.8 mmH₂O<br>/cm²<br>Min: 4.2 mmH₂O<br>/cm² | 4.2 mmH₂O /cm² | < 6.0 mmH₂O/ cm² | Meet | | Flame spread<br>16 CFR 1610 | Class I | Class I<br>Non-Flammable | Class 1 | Meet | Table B-II-3 Biocompatibility Comparison | Item | Subject device | Acceptance<br>Criteria | Remark | |---------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------| | Cytotoxicity | Under the conditions of this study, the test<br>article extract did not show potential toxicity to<br>L929 cells | No potential<br>cytotoxicity | SE | | Irritation | The test result showed that the response of the<br>test article extract was categorized as negligible<br>under the test condition. | Non-Irritating | SE | | Sensitization | Under the conditions of this study, no evidence<br>of skin sensitization in guinea pigs was found. | Non-Sensitizing | SE | {10}------------------------------------------------ All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion. #### 9 Conclusion: Based on the comparison and analysis above, the devices are determined to be Substantially Equivalent (SE) to the predicate devices. The Shandong Aida Medical Products Disposable surgical face mask is substantially equivalent to the Xiantao Rayxin Medical Products Disposable Surgical Face Mask. Based on the nonclinical tests performed, the subject device is as safe, as effective,and performs as well as the legally marketed predicate device, Xiantao Rayxin Medical Products Disposable Surgical Face Mask cleared under K153496.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...