K153496 · Xiantao Rayxin Medical Products Co., Ltd. · FXX · Jun 15, 2016 · General, Plastic Surgery
Device Facts
Record ID
K153496
Device Name
Disposable Surgical Face Mask
Applicant
Xiantao Rayxin Medical Products Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Jun 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Device Story
Disposable surgical face mask; flat pleated design; three-layer construction (spun-bond polypropylene outer/inner layers; melt-blown polypropylene filter middle layer). Features malleable aluminum nose piece for fit; available in tie-on or elastic ear loop configurations. Non-sterile; single-use. Used in clinical settings for infection control; provides barrier protection against blood, body fluids, and particulates. Healthcare personnel or patients wear device over mouth and nose to reduce exposure risk.
Clinical Evidence
No clinical data. Bench testing only: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ASTM F2100-11 (performance), ASTM F1862-13 (fluid resistance: 32/32 pass at 120 mmHg), ASTM F2101-14 (BFE: 98.7%), ASTM F2299-03 (PFE: 98.46%), MIL-M-36945C (Delta P: 4.2 mmH2O/cm²), and 16 CFR 1610 (flammability: Class 1).
Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material in infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in black. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Xiantao Rayxin Medical Products Co., Ltd. June 15, 2016 % Mr. Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 1-202, Build 3, Beijing New World, No.5 Chaoyang Rd., Chaoyang District Beijing, 100024 CHINA
Re: K153496
Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: May 11, 2016 Received: May 16, 2016
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) k153496
Device Name Disposable Surgical Face Mask Ear Loop and Tie-On
Indications for Use (Describe)
The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
# Tab #3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K153496
- 1. Date of Preparation:2016/06/11
- 2. Sponsor Identification
# Xiantao Rayxin Medical Products Co., Itd. No. 258 Pengchang Road, Middle Street, Xiantao city, Hubei 433018, China
Establishment Registration Number: Not yet registered
Contact Person: Liang Zhang Position: General Manager Tel: +86-728-2617592 Fax: +86-728-26177777 Email: zl@rayxin.com
- 3. Designated Submission Correspondent
Mr. Ray Wang
# Beijing Believe Tech. Service Co., Ltd
5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China
Tel: +86-21-50313932, Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Disposable Surgical Face Mask Common Name: Surgical Face Mask Model(s): Tie-On, Ear Loops
Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product Code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel
Indication for use Statement:
The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
### Device Description
The proposed device(s) are Blue color, and Flat Pleated type mask, utilizing Tie-On or Ear Loops way for wearing, and they all has Nose Piece design for fitting the facemask around the nose. The blue colorant is polypropylene (PP) master batch.
The proposed device(s) are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.
The model of proposed device, tie-on, is held in place over the uers's mouth and nose by four ties welded to the facemask. The tie is made of spun-bond polypropylene.
The model of proposed device, ear loops, is held in place over the uers's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex.
The nose piece contained in the proposed device(s) is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of malleable aluminum wire.
The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.
#### 5. Identification of Predicate Device(s)
Predicate Device K133070 Surgical Face Mask
{5}------------------------------------------------
BH Medical Products Co., Ltd.
- Non-Clinical Test Conclusion Q.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- > ASTM F2100-11, Standard Specification For Performance Of Materials Used In Medical Face Masks.
- > ASTM F1862-13, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity)
- > MIL-M-36945C, Method 1 Military Specifications: Surgical Mask disposable;
- ASTM F2101-14, Standard Test Method For Evaluating The Bacterial Filtration Efficiency > (Bfe) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus;
- > ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- A 16 CFR 1610, Standard for the Flammability of clothing textiles;
- > Bench Testing for the performance of Dimensions.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{6}------------------------------------------------
#### Substantially Equivalent (SE) Comparison 8.
| ITEM | Proposed Device | Predicate Device | Remark | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|----|
| Intended Use | The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | SE | |
| Ear loop model and tie-on model | Ear Loops, Tie-On, Flat Pleated, 3 layers | Ear Loops, Tie-On, Flat Pleated, 3 layers | SE | |
| Materials | Outer Facing Layer | Spun-bond polypropylene | Spun-bond polypropylene | SE |
| | Middle Layer | Melt blown polypropylene filter | Melt blown polypropylene filter | |
| | Inner Facing Layer | Spun-bond polypropylene | Spun-bond polypropylene | |
| | Tieon | Spun-bond polypropylene | Spun-bond polypropylene | |
| | Nose Piece | Malleable aluminum wire | Malleable aluminum wire | |
| | Ear Loops | Polyester | Polyester | |
| Color | Blue | Blue, Green | SE | |
| Dimension (Width) | 17.5 cm +/- 1cm | 6.8" +/-0.25" (17.27 cm +/- 0.63cm) | SE | |
| Dimension (Length) | 9.5 cm +/- 1cm | 4.2" +/-0.25" (10.67 cm +/- 0.63cm)<br>3.5" +/-0.25" (8.89 cm +/- 0.63cm) | | |
| OTC use | Yes | Yes | SE | |
| Single Use | Yes | Yes | SE | |
| Sterile | No | No | SE | |
| ASTM F2100 Level | Level 2 | Level 2 | SE | |
Table 1 General Comparison
Table 2 Performance Characteristic Comparision
| ITEM | Proposed Device | Predicate Device | ASTM F2100<br>Requirements for<br>Level 2<br>Classification | Remark |
|--------------------------------------------------|-------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------|--------|
| Fluid Resistance<br>Performance<br>ASTM F1862-13 | 32 out of 32 pass at 120 mmHg | Meet the ASTM F2100<br>Requirements for Level 2<br>Classification | 29 out of 32 pass at 120 mmHg | SE |
| Particulate<br>Filtration<br>Efficiency ASTM | 98.46% | Meet the ASTM F2100<br>Requirements for Level 2<br>Classification | $\ge$ 98% | |
{7}------------------------------------------------
| F2299 | | | | |
|-------------------------------------------------------|--------------------------|-------------------------------------------------------------------|-----------------|----|
| Bacterial<br>Filtration<br>Efficiency ASTM<br>F2101 | 98.7% | Meet the ASTM F2100<br>Requirements for Level 2<br>Classification | ≥ 98% | |
| Differential<br>Pressure (Delta<br>P)<br>MIL-M-36954C | 4.2 mmH2O/cm² | Meet the ASTM F2100<br>Requirements for Level 2<br>Classification | < 5.0 mmH2O/cm² | |
| Flammability<br>16 CFR 1610 | Class 1<br>Non Flammable | Class 1 | Class 1 | SE |
| | Table 3 Biocompatibility Comparison | |
|--|-------------------------------------|--|
| | | |
| ITEM | Proposed Device | Predicate Device | Remark |
|---------------|---------------------------------------------------------------|--------------------------|--------|
| Cytotoxicity | Under the conditions of the study, not<br>cytotoxicity effect | Comply with ISO 10993-5 | SE |
| Irritation | Under the conditions of the study, not an<br>irritant | Comply with ISO 10993-10 | SE |
| Sensitization | Under conditions of the study, not a<br>sensitizer. | | SE |
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
The Xianto Rayxin Medical Products Disposable Surgical Face Mask is substantially equivalent to the BH Medical Products Surgical Face Mask. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, BH Medical Products Surgical Face Mask cleared under K133070.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.