A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
Device Story
Disposable nitrile (acrylonitrile butadiene) examination glove; worn on hands/fingers by clinicians or patients to provide a physical barrier; prevents cross-contamination between patient and examiner. Device is ambidextrous; single-use; non-sterile. Performance relies on material barrier properties, physical dimensions, and tensile strength. Used in clinical or home settings. Output is the physical barrier provided by the glove material. Benefits include reduced risk of pathogen transmission during examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D6319-19 (dimensions, physical properties, freedom from holes, thickness), ASTM D6124-06 (residual powder), and ISO 10993-10/11 (biocompatibility: skin irritation, sensitization, acute systemic toxicity). All results met acceptance criteria.
Technological Characteristics
Material: Nitrile (acrylonitrile butadiene). Construction: Ambidextrous. Standards: ASTM D6319-19 (physical properties/dimensions), ASTM D6124-06 (residual powder), ISO 10993-10/11 (biocompatibility). Form factor: Disposable glove. Energy source: None. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hands or fingers to prevent contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Blue Nitrile Examination Glove (K210375)
Related Devices
K211028 — Nitrile Patient Examination Gloves · Huayuan Medical Technology(Shangqiu) Co., Ltd. · Jun 4, 2021
K142283 — Non-Sterile, Powder Free Nitrile Examination Gloves (Blue, Black and White Color) · Yty Industry (Manjung) Sdn Bhd · Jan 16, 2015
K222713 — White Nitrile Powder Free Patient Examination Glove, Non Sterile · Safeskin Retailing (Hk) Limited · Dec 3, 2022
K210369 — Powder Free Nitrile Examination Glove · Pastel Glove Sdn Bhd · May 13, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 8, 2022
KGM Gloves Sdn. Bhd. Ooi Seng RA Manager Lot 18893 & 18894, Jalan Perusahaan 9, Kawasan Perusahaan Kamunting. Perak 34600 Malaysia
Re: K220284
Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 10, 2022 Received: May 16, 2022
Dear Ooi Seng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220284
Device Name Powder Free Nitrile Examination Gloves (Blue)
### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1.0 510 (K) SUMMARY 510(k) Number: K220284
#### 2.0 Submitter:
KGM Gloves Sdn. Bhd. Lot 18893 & 18894, Jalan Perusahaan 9, Kawasan Perusahaan Kamunting, 34600 Kamunting, Perak, Malaysia. Tel: +605-891 3612 Fax: +605-891 6612
Establishment Registration Number: 3020763216
## Contact Person:
Ooi Loon Seng (Mdm) i) (ii) Contact No: +6012-4612918 Email: loonsenq54@gmail.com
Melissa Oon Contact No: +6012-4730656 Email: melissaoon@gmail.com
Date of Preparation : January 15, 2022
#### 3.0 Name of the Device
Trade Name: Powder Free Nitrile Examination Glove (Blue) Classification Name: Patient Examination Glove Device Classification: Class I Regulation Number: 21 CFR 880.6250 Product Code: LZA
#### 4.0 Identification of The Legally Marketed Predicate Device
Predicate Device Name: Powder Free Blue Nitrile Examination Glove Predicate 510(K) Number: K210375 Manufacturer's Name: Duramitt Sdn. Bhd. Device Classification: Class I Regulation Number: 21 CFR 880.6250 Product Code: LZA
#### 5.0 Description of Device
Powder Free Blue Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.
#### 6.0 The Intended Use of Device
The Powder Free Nitrile Examination Glove (Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
{4}------------------------------------------------
#### 7.0 Summary of the Technological Characteristic Comparison of the Device
Provided below is a comparison of the subject device and the predicate device
| Characteristics and<br>Parameters | Proposed Device -<br>Powder Free Nitrile<br>Examination Gloves | Predicate Device -<br>Powder Free Blue<br>Nitrile Examination<br>Gloves (K210375) | Comparison |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | LZA | LZA | Identical |
| Intended Use | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands or fingers to<br>prevent contamination<br>between patient and<br>examiner. | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that is<br>worn on the hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Identical |
| Device Classification | Class I | Class I | Identical |
| Device material | Nitrile<br>(Acrylonitrile butadiene) | Nitrile<br>(Acrylonitrile butadiene) | Similar |
| Colour | Blue | Blue | Similar |
| Instruction for<br>Use on Labeling | Single Use only. | Single Use only. | Identical |
| Construction | Ambidextrous | Ambidextrous | Identical |
| Freedom from holes | Meet AQL 2.5 at G1 | Meet AQL 1.5 at G1 | Same |
| Length:<br>XS-Minimum 220mm<br>S-Minimum 220mm<br>M-Minimum 230mm<br>L-Minimum 230mm<br>XL-Minimum 230mm | Minimal Value for Length:<br>XS: 240mm<br>S: 246mm<br>M: 240mm<br>L : 250mm<br>XL: 251mm | Average<br>XS:-<br>S: 242mm<br>M: 242mm<br>L : 243mm<br>XL:- | Similar |
| Width:-<br>XS-70 ± 10mm<br>S-80 ± 10mm<br>M-95 ± 10mm<br>L-110 ± 10mm<br>XL-120 ± 10mm | Minimal Value for Width:<br>XS: 73mm<br>S: 84 mm<br>M: 93 mm<br>L: 104 mm<br>XL: 111mm | Average<br>XS:-<br>S: 84mm<br>M: 96mm<br>L : 107mm<br>XL:- | Similar |
| Palm Thickness<br>Minimum 0.05mm | Palm Minimal Value:<br>XS: 0.07mm<br>S: 0.07mm<br>M: 0.07mm | Average<br>XS:-<br>S: 0.06mm<br>M: 0.06mm | Similar |
| | L : 0.07mm | L : 0.06mm | |
| | XL: 0.07mm | XL:- | |
| Finger Thickness<br>Minimum 0.05mm | Finger Minimal Value:<br>XS: 0.11mm<br>S: 0.11mm<br>M: 0.11mm<br>L : 0.11mm<br>XL: 0.12mm | Average<br>XS:-<br>S: 0.08mm<br>M: 0.08mm<br>L : 0.08mm<br>XL:- | Similar |
| Tensile Strength<br>(Before age)<br>Minimum 14.0MPa | Minimum 19.1MPa | Average 19.04MPa | Similar |
| Tensile Strength<br>(After Age)<br>Minimum 14.0MPa | Minimum 19.0MPa | Minimum 14.0MPa | Different |
| Ultimate Elongation<br>(Before age)<br>Minimum 500% | Minimum 520% | Average 560% | Similar |
| Ultimate Elongation<br>(After age)<br>Minimum 400% | Minimum 480% | Average 510% | Similar |
| Residual powder test<br>(<2mg/glove) | Average 0.10mg/glove | Average 0.59mg/glove | Similar |
| Primary Skin Irritation | Under the condition of<br>study, not an irritant | Under the condition of<br>study, not an irritant | Same |
| Dermal Sensitization | Under the condition of<br>study, not a sensitizer | Under the condition of<br>study, not a sensitizer | Same |
| Acute Systemic<br>Toxicity | Not induce systemic<br>toxicity | Not induce systemic<br>toxicity | Same |
{5}------------------------------------------------
### Summary of Non-Clinical Performance Data 8.0
Provided in the table below is the summary of the non-clinical testing performed with the subject device. The results demonstrate that the subject device met the acceptance criteria.
| Test Methodology | Purpose | Acceptance Criteria | Results | | | Width:<br>XS-70 ± 10mm<br>S-80 ± 10mm<br>M-95 ± 10mm<br>L-110 ± 10mm<br>XL-120 ± 10mm | Minimal Value for Width:<br>XS: 73mm<br>S: 84 mm<br>M: 93 mm<br>L: 104 mm<br>XL: 111mm<br>PASS |
|------------------------------|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| ASTM D6319-19:<br>Dimensions | To measure the<br>physical length and<br>width of the gloves | Length:<br>XS-Minimum 220mm<br>S-Minimum 220mm<br>M-Minimum 230mm<br>L-Minimum 230mm<br>XL-Minimum 230mm | Minimal Value for<br>Length:<br>XS: 240mm<br>S: 246mm<br>M: 240mm<br>L : 250mm<br>XL: 251mm | ASTM D6319-19:<br>Physical<br>Properties | To measure the tensile strength and elongation before and after aging | Tensile Strength: Before Aging: 14.0 MPa After Aging: 14.0 MPa | Tensile Strength: Before Aging: Minimal Value: 19.10 MPa After Accelerated Aging: Minimal Value: 19.0 MPa |
| | | Ultimate Elongation:<br>Before Aging: min. 500%<br>After Aging: min. 400% | Ultimate Elongation: Before Aging Minimal Value: 520% After Aging Minimal Value: 480%<br>PASS | | | | |
| ASTM D6319-19:<br>Thickness | To measure the physical palm and finger thickness of the gloves | Palm-Min. 0.05mm | Palm Minimal Value:<br>XS: 0.07mm<br>S: 0.07mm<br>M: 0.07mm<br>L : 0.07mm<br>XL: 0.07mm | | | | |
| | | Finger-Min. 0.05mm | Finger Minimal Value:<br>XS: 0.11mm<br>S: 0.11mm<br>M: 0.11mm<br>L : 0.11mm<br>XL: 0.12mm<br>PASS | | | | |
{6}------------------------------------------------
{7}------------------------------------------------
| ASTM D6319-19:<br>Freedom from<br>Holes<br>Testing | To detect the<br>presence of hole<br>in glove | Meet AQL 2.5 at G1 | Meet AQL 2.5 at G1<br>PASS |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------|-----------------------------------------------------------|
| ISO 10993-10:<br>Primary Skin<br>Irritation | To<br>determine<br>the<br>irritation potential of<br>glove when expose to<br>skin surface. | Not an irritant | Under the condition of<br>study, not an irritant<br>PASS |
| ISO 10993-10:<br>Dermal<br>Sensitization | To<br>determine<br>the<br>sensitization<br>potential of glove<br>when expose to skin<br>surface | Not a Sensitizer | Under the condition of<br>study, not a sensitizer<br>PASS |
| ISO 10993-11:<br>Acute Systemic<br>Toxicity | To<br>evaluate<br>the<br>adverse<br>systemic<br>reaction<br>potential<br>from glove | Not induce systemic toxicity | Did not induce systemic<br>Toxicity<br>PASS |
| ASTM D6124-06:<br>Residual Powder<br>Analysis | To check the amount<br>of powder residue<br>from glove surface | Residual powder test<br>( $<2$ mg/glove) | Average 0.10mg/glove<br>PASS |
#### Summary of Clinical Performance Data 9.0
Not applicable - Clinical data is not needed for the subject device, gloves.
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K210375.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.