The cobas® pulse blood glucose monitoring system consists of the cobas® pulse instrument and the cobas® GLU test strips. This system is intended for in vitro diagnostic, point of care, multiple-patient use within professional healthcare settings, including patients receiving intensive medical intervention/therapy. The cobas® GLU test strips provide an in vitro diagnostic test that quantitatively measures glucose in arterial whole blood, neonate arterial whole blood, or neonate heel stick whole blood, on the cobas® pulse system. For neonate heel stick, the system should only be used with single-use, auto-disabling lancing devices. This system is not intended for use with capillary finger stick, venous whole blood, or neonate cord blood specimens. This system is not intended for screening or diagnosis of diabetes, but is indicated for use in determining dysglycemia.
Device Story
Hand-held, battery-powered glucose monitoring system; uses single-use cobas® GLU test strips with FAD-dependent glucose dehydrogenase (FAD-GDH) chemistry. Input: arterial whole blood, neonate arterial whole blood, or neonate heel stick whole blood (lithium heparin anticoagulant). Operation: blood applied to test strip; chemical reaction generates current proportional to glucose concentration; instrument calculates and displays plasma-equivalent glucose concentration. Used in professional healthcare settings (OR, ICU, NICU, general ward) by clinicians/nurses. Includes laser barcode scanner for patient ID. Output: quantitative glucose result on meter display. Benefits: rapid point-of-care dysglycemia assessment in critical care environments. System includes QC lockout feature requiring successful control solution testing to ensure accuracy.
Clinical Evidence
Clinical study conducted at 14 sites with 873 patients (adult, pediatric, neonate). Evaluated arterial whole blood (n=622), neonate arterial (n=104), and neonate heel stick (n=147). Compared against cobas 6000 system. For arterial samples ≥75 mg/dL, 99.0% were within ±20% of comparator. For neonate heel stick ≥75 mg/dL, 100% were within ±20%. Performance validated across 709 medications and 1617 patient conditions. Bench testing confirmed linearity (10-600 mg/dL), precision (CV <2.4%), and interference resistance to numerous endogenous/exogenous substances.
Indicated for professional healthcare settings to quantitatively measure glucose in arterial whole blood, neonate arterial whole blood, or neonate heel stick whole blood for determining dysglycemia. Not for capillary finger stick, venous whole blood, or neonate cord blood. Not for diabetes screening or diagnosis.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K220272
B Applicant
Roche Diagnostics
C Proprietary and Established Names
cobas® pulse blood glucose monitoring system
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PZI | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
This submission is a Dual 510(k) and CLIA Waiver by Application (Dual Submission) for a new device, tracked as K220272 and CW220002
B Measurand:
Glucose
C Type of Test:
Quantitative aperometric assay (FAD-dependent glucose dehydrogenase)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K220272 - Page 2 of 20
# III Intended Use/Indications for Use:
## A Intended Use(s):
See Indications for Use below.
## B Indication(s) for Use:
The cobas® pulse blood glucose monitoring system consists of the cobas® pulse instrument and the cobas® GLU test strips.
This system is intended for in vitro diagnostic, point of care, multiple-patient use within professional healthcare settings, including patients receiving intensive medical intervention/therapy.
The cobas® GLU test strips provide an in vitro diagnostic test that quantitatively measures glucose in arterial whole blood, neonate arterial whole blood, or neonate heel stick whole blood, on the cobas® pulse system.
For neonate heel stick, the system should only be used with single-use, auto-disabling lancing devices.
This system is not intended for use with capillary finger stick, venous whole blood, or neonate cord blood specimens.
This system is not intended for screening or diagnosis of diabetes, but is indicated for use in determining dysglycemia.
## C Special Conditions for Use Statement(s):
- Rx - For Prescription Use Only
- For in vitro diagnostic use only
- This system is not intended for use with capillary finger stick, venous whole blood, or neonate cord blood specimens.
*See section VII.C.3 for additional information on sample limitations.
- Not for alternative site testing (AST)
- This system is not intended for screening or diagnosis of diabetes.
- Use only fresh whole blood or whole blood collected in lithium heparin collection devices for arterial, neonate arterial, or neonate heel stick whole blood.
- Blood collection devices containing iodoacetate or fluoride should not be used.
- Do not use serum or plasma.
- Use only an auto-disabling single-use lancing devices for heel stick samples.
- Altitudes above 14,107 feet (4300 meters) above sea level have not been evaluated.
## D Special Instrument Requirements:
cobas® pulse instrument
{2}
K220272 - Page 3 of 20
## IV Device/System Characteristics:
### A Device Description:
The cobas® pulse blood glucose monitoring system consists of a hand-held, battery powered cobas® pulse instrument, single use cobas® GLU test strips, instrument charging station, quick start guide and user guide for CLIA waived users as well as cobas® GLU test strip insert and user assistance guide for non-CLIA waived users. The cobas GLU Linearity Kit and GLU QC kit are sold separately.
### B Principle of Operation:
The detection chemistry used by the cobas® pulse blood glucose monitoring system for the blood glucose result is FAD-dependent glucose dehydrogenase (FAD-GDH). When the blood is applied to the test strip, the blood induces a chemical reaction which produces a current that is transmitted through the electrodes to the instrument. The magnitude of the resultant current is proportional to the glucose concentration in the sample. The detected current signal is then calculated by the meter and the glucose concentration is then displayed on the meter. The meter reports plasma equivalent glucose concentrations.
### C Instrument Description Information:
1. Instrument Name:
cobas pulse® instrument
2. Specimen Identification:
Arterial whole blood, neonatal arterial whole blood, and neonatal heel stick whole blood anticoagulated with lithium heparin can be used with this test system. The meter contains a laser barcode scanner that allows for scanning patient identification information that may also be entered manually. The meter prompts the user to choose from a dropdown menu of sample types prior to testing.
3. Specimen Sampling and Handling:
All samples must be tested immediately upon collection.
4. Calibration:
The system does not require calibration by the end-user. When the test strip is inserted into the instrument, the instrument reads the lot information from the strip that includes the calibration and expiration information for that test strip lot.
5. Quality Control:
The cobas® GLU QC kit control solutions are aqueous solutions available at two different glucose levels (Level 1 and Level 2). Instructions on when to perform quality control testing
{3}
using both levels are included in the labeling. If the quality control testing using both QC levels is not performed when required as described in the labeling, the user is cautioned that a QC lockout is implemented and blood glucose cannot be tested until the QC tests are run successfully. The meter displays a pass or fail result once both QC tests are performed. The acceptable ranges are communicated to the meter by the user scanning the barcode on the QC bottle. The user is cautioned not to use the meter and to contact the system administrator if the quality control test fails after repeating the QC test.
K220272 - Page 4 of 20
V Substantial Equivalence Information:
A Predicate Device Name(s):
StatStrip Glucose Hospital Meter System
B Predicate 510(k) Number(s):
K181043
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220272 | K181043 |
| --- | --- | --- |
| Device Trade Name | cobas pulse blood glucose monitoring system | StatStrip Glucose Hospital Meter System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Quantitative determination of glucose for use in determining dysglycemia in professional healthcare settings | Same |
| Measuring Range | 10-600 mg/dL | Same |
| General Device Characteristic Differences | | |
| Enzyme | FAD-dependent glucose dehydrogenase | Glucose Oxidase |
| Sample Types | Arterial whole blood, neonate arterial whole blood, or neonate heel stick whole blood | Capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens |
{4}
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline
CLSI EP07 3rd Edition Interference Testing in Clinical Chemistry.
CLSI EP37 1st Edition Supplemental Tables for Interference Testing in Clinical Chemistry
CLSI EP09c 3rd Edition Measurement Procedure Comparison and Bias Estimation Using Patient Samples
CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
CLSI EP21 2nd Edition Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision evaluations were conducted to evaluate repeatability (within-run precision) and intermediate precision of the cobas pulse blood glucose monitoring system according to CLSI guideline EP05-A3 as well as the FDA Guidance document: Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use (issued September 29, 2020).
Within-run Precision
Within-run precision was evaluated using whole blood samples adjusted (spiked or allowed to glycolyze) to achieve five glucose concentration ranges (30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, and 251-400 mg/dL). Each glucose level was tested with three lots of cobas GLU test strips in replicates of 10 on 10 cobas pulse instruments for a total of 300 results per glucose level. The summary of within-run precision results are presented in the table below:
| Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | %CV |
| --- | --- | --- | --- | --- | --- |
| 30-50 | 1 | 100 | 40.0 | 0.91 | 2.28 |
| | 2 | 100 | 39.6 | 0.871 | 2.20 |
| | 3 | 100 | 40.7 | 0.755 | 1.86 |
| | Combined | 300 | 40.1 | 0.945 | 2.36 |
| | 1 | 100 | 80.5 | 1.04 | 1.29 |
K220272 - Page 5 of 20
{5}
| Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | %CV |
| --- | --- | --- | --- | --- | --- |
| 51-110 | 2 | 100 | 79.6 | 1.13 | 1.42 |
| | 3 | 100 | 80.8 | 1.22 | 1.51 |
| | Combined | 300 | 80.3 | 1.23 | 1.54 |
| 111-150 | 1 | 100 | 129 | 1.62 | 1.25 |
| | 2 | 100 | 129 | 1.39 | 1.07 |
| | 3 | 100 | 130 | 1.74 | 1.34 |
| | Combined | 300 | 130 | 1.63 | 1.26 |
| 151-250 | 1 | 100 | 194 | 1.89 | 0.97 |
| | 2 | 100 | 195 | 1.80 | 0.92 |
| | 3 | 100 | 194 | 2.42 | 1.25 |
| | Combined | 300 | 194 | 2.09 | 1.08 |
| 251-400 | 1 | 100 | 328 | 5.24 | 1.6 |
| | 2 | 100 | 329 | 3.07 | 0.93 |
| | 3 | 100 | 327 | 3.34 | 1.02 |
| | Combined | 300 | 328 | 4.05 | 1.23 |
## Intermediate Precision
Intermediate precision was evaluated using five levels of control solutions with the following glucose concentration ranges: 30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL. Multiple operators tested the control solutions over 10 days using three lots of cobas GLU test strips and 10 cobas pulse instruments with two replicates per glucose concentration, instrument and day. The combined intermediate precision results are summarized below:
| Glucose Level (mg/dL) | N | Mean (mg/dL) | SD (mg/dL) | %CV |
| --- | --- | --- | --- | --- |
| 30-50 | 600 | 41.6 | 0.712 | 1.71 |
| 51-110 | 600 | 80.3 | 1.17 | 1.45 |
| 111-150 | 600 | 130 | 1.66 | 1.28 |
| 151-250 | 600 | 201 | 2.03 | 1.01 |
| 251-400 | 600 | 316 | 2.42 | 0.76 |
K220272 - Page 6 of 20
{6}
K220272 - Page 7 of 20
2. Linearity:
The linearity of the cobas pulse blood glucose monitoring system was evaluated using whole blood samples with 11 glucose concentrations between 8-720 mg/dL. Each sample was tested using three lots of cobas GLU test strips and results were compared to those obtained using the comparator method (the cobas 6000 system). The results from the regression analysis are summarized below:
| Lot | Linear Regression Equation | R2 |
| --- | --- | --- |
| 1 | y =1.01x +3.99 | 1.00 |
| 2 | y =1.01x -4.10 | 1.00 |
| 3 | y =1.03x -4.55 | 1.00 |
The results of the study support that the test system is linear from 10 – 600 mg/dL glucose. If the concentration of a sample is less than 10 mg/dL glucose, the result is flagged by the meter as "Lo". If a sample exceeds 600 mg/dL glucose, the result is flagged by the meter as "Hi". The "Lo" and "Hi" functions were validated and demonstrated to function as intended.
3. Analytical Specificity/Interference:
Interference testing was conducted to evaluate the effect of common endogenous and exogenous substances expected in the intended use population on the cobas pulse blood glucose monitoring system. Additional potential interferents were included in the testing following a detailed analysis of medications and medical conditions of the subjects evaluated in the method comparison study (please see the tables in Section VII.C.3 below for a detailed list of medications and conditions).
Whole blood samples were adjusted to achieve three glucose concentration ranges: 50-70 mg/dL, 110-130 mg/dL, 225-270 mg/dL. Each of these samples was divided into a test pool with the potential interferent added and a control sample with no added interferent. Results from the test samples on the candidate meter were compared to results obtained from the control sample on the candidate meter using three test strip lots. The % difference between the test sample and the control sample was calculated using the mean of 10 replicates. The highest tested concentrations at which no significant interference was observed (defined by the sponsor as less than ±10% bias between the test and control samples) are presented in the following table:
| Endogenous Substance | Highest concentration with no significant interference (mg/dL) |
| --- | --- |
| β-Hydroxybutyrate sodium | 265 |
| Acetoacetic acid (Acetoacetate) | 20 |
| Acetone | 100 |
| Albumin human | 6000 |
| Biotin | 3 |
| Bilirubin, conjugated | 50 |
| Bilirubin, unconjugated | 40 |
{7}
| Endogenous Substance | Highest concentration with no significant interference (mg/dL) |
| --- | --- |
| Calcium (chloride) | 20 |
| Cholesterol | 500 |
| Citric acid | 384.3 |
| Creatinine | 15 |
| Fructose | 18 |
| Galactose | 60 |
| Galactose-1-phosphate | 500 |
| Glutathione, reduced | 5 |
| Glycerol | 17 |
| Hemoglobin | 1000 |
| Insulin | 127 mU/L |
| Intralipid | 2000 |
| Lactate (Lactic acid) | 200 |
| Magnesium (chloride) | 10 |
| Maltose | 480 |
| Mannose | 500 |
| Phosphate | 12 |
| Potassium (chloride) | 59.64 |
| Sodium chloride | 414 |
| Urea | 120 |
| Uric acid | 23.5 |
| Exogenous Substance | Highest concentration with no significant interference (mg/dL) | Exogenous Substance | Highest concentration with no significant interference (mg/dL) |
| --- | --- | --- | --- |
| Acetaminophen (Paracetamol) | 20 | Calcium gluconate | 120 |
| Acetate | 30.1 | Canagliflozin | 1.4 |
| Acetylsalicylic Acid | 3 | Captopril | 0.265 |
| Acyclovir | 6.6 | Carbamazepine | 4.5 |
| Amikacin | 14.4 | Cefaclor | 8.7 |
| Amiodarone | 4.2 | Cefazolin | 45 |
| Amoxicillin | 5.4 | Cefepime | 48 |
| Ampicillin | 7.5 | Cefotaxime | 52.8 |
| Anakinra | 30.86 | Cefoxitin | 600 |
| Anidulafungin | 3.3 | Ceftazidime | 60.6 |
| Ascorbic acid (Ascorbate) | 8 | Ceftriaxone | 84 |
K220272 - Page 8 of 20
{8}
K220272 - Page 9 of 20
| Exogenous Substance | Highest concentration with no significant interference (mg/dL) | Exogenous Substance | Highest concentration |
| --- | --- | --- | --- |
| Azithromycin | 1.11 | Cefuroxime | 54.3 |
| Aztreonam | 72.6 | Celecoxib | 0.879 |
| Benzylpenicillin | 72 | Chlorothiazide | 2.7 |
| Bicarbonate | 62 | Cimetidine | 3 |
| Bictegravir | 1.43 | Clavulanate | 1.2 |
| Bromide | 300 | Clindamycin | 5.1 |
| Bupivacaine | 1.2 | Clopidogrel carboxylic acid | 3.67 |
| Cafedrine | 1.8 | Cyclosporin A | 0.18 |
| Caffeine | 10.8 | Dexamethasone | 1.2 |
| Diazepam | 3 | Ipragliflozin | 0.52 |
| Diclofenac | 2.4 | Irbesartan | 2.31 |
| Dipyridamole | 3.54 | Iron saccharate | 3 |
| Docusate | 2.37 | Isomalt | 100 |
| Dopamine | 0.09 | Isosorbide dinitrate | 0.593 |
| Doxycycline | 1.8 | Isosorbide mononitrate | 0.221 |
| D-Sorbitol | 70 | Ketamine | 1.89 |
| EDTA | 0.1 | Ketoprofen | 6 |
| Emtricitabine | 0.9 | Ketorolac | 2.055 |
| Enoxaparin | 435 IU/dL | Lacosamide | 2.61 |
| Ephedrine | 1 | Lactitol | 100 |
| Erythromycin | 13.8 | Lactose | 500 |
| Esmolol | 0.48 | Lansoprazole | 20 |
| Ethanol | 600 | Levetiracetam | 27 |
| Fe(II) (sulfate) | 0.879 | Levodopa (L-Dopa) | 0.75 |
| Fenofibrate | 4.5 | Levofloxacin | 3.6 |
| Flecainide | 0.513 | Lidocaine | 1.5 |
| 5-Fluorouracil | 9 | L-Leucine | 2.82 |
| Fluconazole | 2.55 | L-Methionine | 0.9 |
| Fosinopril | 1.68 | Lopinavir/Ritonavir | 4.05/1.0125 |
| Fosphenytoin | 38.7 | L-Phenylalanine | 2.23 |
| Furosemide | 1.59 | Maltitol | 20.2 |
| Fusidate | 60 | Mannitol | 1800 |
| Gabapentin | 2.67 | Meropenem | 33.9 |
| Gentamicin | 3 | Metamizole sodium | 3.3 |
| Gentisic acid | 2 | Metformin | 1.2 |
| Gliclazid | 1.11 | Methadone | 0.318 |
| Guaifenesin | 0.45 | Methocarbamol | 12.3 |
| Heparin | 330 IU/dL | Methotrexate | 136 |
{9}
K220272 - Page 10 of 20
| Exogenous Substance | Highest concentration with no significant interference (mg/dL) | Exogenous Substance | Highest concentration with no significant interference (mg/dL) |
| --- | --- | --- | --- |
| HES (Hydroxyethyl starch) | 3600 | Methyldopa | 2.25 |
| Hydralazine | 1.44 | Methylprednisolone | 0.783 |
| Hydroxychloroquine | 0.731 | Metronidazole | 12.3 |
| Ibuprofen | 50 | Micafungin | 6.6 |
| Iohexol | 35.7 | Midazolam | 0.376 |
| Morphine | 0.78 | Ranitidine | 1.05 |
| Mycophenolate | 4.2 | Remdesivir | 1.63 |
| N-Acetyl-L-cysteine | 15 | Rifampicin | 4.8 |
| Nafcillin | 11.1 | Rocuronium bromide | 1.8 |
| Naproxen | 36 | Salicylate | 60 |
| Nateglinide | 4.86 | Sodium fluoride | 0.12 |
| Neomycin | 1.2 | Succinylmonocholin chloride (suxamethonium metabolite) | 5 |
| Nicotinic acid (sorbinicate metabolite) | 4.23 | Sucrose | 500 |
| Omeprazole | 0.84 | Sulbactam | 24 |
| Pancuronium bromide | 0.4 | Sulfamethoxazole | 40 |
| Pantoprazole | 3 | Tazobactam | 3.05 |
| Phenobarbital | 69 | Tetracycline | 2.4 |
| Phenol | 4.6 | Thalidomide | 1.03 |
| Phenylbutazone | 32.1 | Theophylline | 6 |
| Phytomenadione | 0.534 | Thiocyanate (nitroprusside metabolite) | 5.22 |
| Pioglitazone | 0.476 | Tolazamide | 9 |
| Piperacillin | 110 | Tolbutamide | 72 |
| Pirfenidone | 3.78 | Tramadol | 0.314 |
| Pralidoxime iodide (PAM) | 25 | Trimethoprim | 4.2 |
| Pregabalin | 2.25 | Valproate | 31.8 |
| Primidone | 5.7 | Vancomycin | 12 |
| Probenecid | 44.7 | Warfarin | 7.5 |
| Procainamide | 4.8 | Xylitol | 200 |
| Propofol | 4.8 | Xylose | 600 |
| 2-pyridyl acetic acid (betahistine metabolite) | 0.579 | Zonisamide | 11 |
| Quinidine | 1.5 | γ-Globuline, human | 1600 |
{10}
4. **Assay Reportable Range:**
The reportable range is 10 - 600 mg/dL.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
**Traceability**
Each measurement result is traceable to the NIST SRM 917 glucose reference material. A method comparison was performed using the candidate device and the cobas 6000 system as the comparator method (see section VII.C.3 below).
**Test Strip Stability**
Test strip stability was assessed using real time stability studies. Protocols and acceptance criteria were reviewed and found to be acceptable to support the labeling claims of a 24 month closed vial shelf life and open vial use life when stored at 4 - 30°C (39 - 86°F) and relative humidity of 10-90%. The labeling instructs the users not to freeze the test strips.
6. **Detection Limit:**
Not applicable
7. **Assay Cut-Off:**
Not applicable
8. **Accuracy (Instrument):**
See Section VII.C.3. Other Clinical Supportive Data.
9. **Carry-Over:**
Not Applicable.
B. **Comparison Studies:**
1. **Method Comparison with Predicate Device:**
See Section VII.C.3. Other Clinical Supportive Data.
2. **Matrix Comparison:**
Not Applicable.
K220272 - Page 11 of 20
{11}
K220272 - Page 12 of 20
# C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable
2. Clinical Specificity:
Not Applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
The performance of the cobas pulse glucose monitoring system was assessed in professional healthcare settings using lithium heparin arterial samples from 622 patients (adult, pediatric and neonate that included neonate arterial samples from 104 patients) and neonate heel stick samples from 147 neonates for a total of 873 patients. The study was conducted at 14 clinical sites that included patients from the general ward, operating rooms, medical intensive care units, surgical intensive care units, pediatric intensive care units, neonatal intensive care units and nurseries. Results from the cobas pulse glucose monitoring system were obtained by a total of 75 untrained operators who were also untrained in the use of the cobas pulse glucose monitoring system.
Samples from patients, less than 24 hours old to over 81 years of age, were analyzed using a total of 3 different test strip lots throughout the study. The glucose levels of the patient samples were as follows (according to the comparator method; the cobas 6000 system): 49.0 to 461.9 mg/L for all arterial (24 samples <80 mg/dL and 10 samples >300 mg/dL); 49.0 to 461.9 mg/L (16 samples <80 mg/dL and 4 samples >300 mg/dL) for neonate arterial; 18.5 mg/dL to 176.8 mg/dL (85 samples <80 mg/dL) for neonatal heelstick. In addition to glucose levels, patient conditions, medication information, pO2, sodium, and hematocrit levels were collected during the study. The study included 1617 unique patient conditions receiving a total of 709 medications representing 77 parent drug classes (see patient conditions and medication tables below).
The results of the cobas pulse glucose monitoring system were compared with those obtained on the comparator method (cobas 6000 system) and are as follows for each specimen type (all arterial combined, neonate arterial, and neonate heel stick):
# Arterial Whole Blood (Combined)
Glucose concentrations < 75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL | Exceeds ±15 mg/dL |
| --- | --- | --- | --- | --- |
| 6/16 (37.5%) | 15/16 (93.8%) | 16/16 (100.0%) | 16/16 (100.0%) | 0/16 (0.0%) |
{12}
Glucose concentrations ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±12% | Within ±15% | Within ±20% | Exceeds ±20% |
| --- | --- | --- | --- | --- | --- |
| 423/606
(69.8%) | 571/606
(94.2%) | 585/606
(96.5%) | 596/606
(98.3%) | 600/606
(99.0%) | 6/606
(1.0%) |
## Neonatal Arterial Study
Glucose concentrations < 75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL | Exceeds ±15 mg/dL |
| --- | --- | --- | --- | --- |
| 4/10
(40.0%) | 9/10
(90.0%) | 10/10
(100.0%) | 10/10
(100.0%) | 0/10
(0.0%) |
Glucose concentrations ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±12% | Within ±15% | Within ±20% | Exceeds ±20% |
| --- | --- | --- | --- | --- | --- |
| 49/94
(52.1%) | 87/94
(92.6%) | 91/94
(96.8%) | 93/94
(98.9%) | 93/94
(98.9%) | 1/94
(1.1%) |
## Neonatal Heel Stick Study
Glucose concentrations < 75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL | Exceeds ±15 mg/dL |
| --- | --- | --- | --- | --- |
| 27/69
(39.1%) | 61/69
(88.4%) | 66/69
(95.7%) | 68/69
(98.6%) | 1/69
(1.4%) |
Glucose concentrations ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±12% | Within ±15% | Within ±20% | Exceeds ±20% |
| --- | --- | --- | --- | --- | --- |
| 43/78
(55.1%) | 70/78
(89.7%) | 74/78
(94.9%) | 74/78
(94.9%) | 78/78
(100.0%) | 0/78
(0.0%) |
K220272 - Page 13 of 20
{13}
Usability Assessment/Operator Questionnaire
Upon completion of the clinical studies, operators completed a questionnaire to evaluate the ease of understanding of the test procedure. The participants found the cobas pulse glucose monitoring system easy to use and the instructions in the user manual and QRG clear and easy to follow.
A readability assessment demonstrated that the quick reference guide (QRG) and the instructions for the CLIA waived user in the user manual were written at a 7th grade reading level.
Patient Conditions and Medications
For all arterial, neonate arterial and neonate heel stick patients combined, there were a total of 709 different medications administered to study subjects within 24 hours of blood draw and a total of 1617 different diagnoses were observed among the study subjects. A summary of the medications and diagnosis for each sample type is shown below:
| Sample Type | Number of Patients | Different Medications | Medication Classes | Different Diagnoses | System Organ Classes |
| --- | --- | --- | --- | --- | --- |
| Arterial | 622 | 699 | 77 | 1562 | 27 |
| Neonate Arterial | 104 | 128 | 34 | 228 | 25 |
| Neonate Heel Stick | 147 | 100 | 27 | 158 | 21 |
The number of patients in the study by condition (organ classes) and medication classes are detailed below:
Patient Conditions
| System Organ Class | Number of Patients |
| --- | --- |
| Blood and lymphatic system disorders | 252 |
| Cardiac disorders | 560 |
| Congenital, familial and genetic disorders | 330 |
| Ear and labyrinth disorders | 29 |
| Endocrine disorders | 70 |
| Eye disorders | 113 |
| Gastrointestinal disorders | 352 |
| General disorders and administration site conditions | 125 |
| Hepatobiliary disorders | 197 |
| Immune system disorders | 69 |
| Infections and infestations | 231 |
| Injury, poisoning and procedural complications | 306 |
K220272 - Page 14 of 20
{14}
| System Organ Class | Number of Patients |
| --- | --- |
| Metabolism and nutrition disorders | 669 |
| Musculoskeletal and connective tissue disorders | 308 |
| Neoplasms benign, malignant and unspecified (incl cysts and polyps) | 148 |
| Nervous system disorders | 362 |
| Pregnancy, puerperium and perinatal conditions | 377 |
| Psychiatric disorders | 227 |
| Renal and urinary disorders | 221 |
| Reproductive system and breast disorders | 85 |
| Respiratory, thoracic and mediastinal disorders | 549 |
| Skin and subcutaneous tissue disorders | 61 |
| Social circumstances | 26 |
| Surgical and medical procedures | 399 |
| Vascular disorders | 458 |
## Medications
| Medication Class | Number of Patients |
| --- | --- |
| All other non-therapeutic products | 59 |
| All other therapeutic products | 45 |
| Anabolic agents for systemic use | 1 |
| Analgesics | 1187 |
| Anesthetics | 332 |
| Anti-parkinson drugs | 5 |
| Antianemic preparations | 35 |
| Antibacterials for systemic use | 688 |
| Antibiotics and chemotherapeutics dermatological use | 73 |
| Antidiarrheals, intestinal antiinflammatory/antiinfective | 13 |
| Antiemetics and antinauseants | 70 |
| Antiepileptics | 54 |
| Antifungals for dermatological use | 9 |
| Antigout preparations | 13 |
| Antihemorrhagics | 166 |
| Antihistamines for systemic use | 15 |
| Antihypertensives | 52 |
| Antiinflammatory and antirheumatic products | 33 |
| Antimycotics for systemic use | 44 |
| Antineoplastic agents | 1 |
| Antiobesity preparations, excl. Diet products | 1 |
| Antiprotozoals | 1 |
| Antipruritics, incl. Antihistamines, anesthetics, etc. | 47 |
| Antiseptics and disinfectants | 12 |
| Antithrombotic agents | 582 |
K220272 - Page 15 of 20
{15}
K220272 - Page 16 of 20
| Medication Class | Number of Patients |
| --- | --- |
| Antivirals for systemic use | 37 |
| Beta blocking agents | 201 |
| Bile and liver therapy | 7 |
| Blood substitutes and perfusion solutions | 1337 |
| Calcium channel blockers | 120 |
| Calcium homeostasis | 1 |
| Cardiac therapy | 602 |
| Contrast media | 49 |
| Corticosteroids for systemic use | 237 |
| Corticosteroids, dermatological preparations | 7 |
| Cough and cold preparations | 11 |
| Digestives, incl. Enzymes | 2 |
| Diuretics | 155 |
| Drugs for acid related disorders | 324 |
| Drugs for constipation | 249 |
| Drugs for functional gastrointestinal disorders | 19 |
| Drugs for obstructive airway diseases | 157 |
| Drugs used in diabetes | 187 |
| Emollients and protectives | 9 |
| Endocrine therapy | 3 |
| General nutrients | 19 |
| Immune sera and immunoglobulins | 3 |
| Immunostimulants | 1 |
| Immunosuppressants | 100 |
| Lipid modifying agents | 139 |
| Mineral supplements | 68 |
| Muscle relaxants | 409 |
| Nasal preparations | 139 |
| Ophthalmological and otological preparations | 2 |
| Ophthalmologicals | 115 |
| Other alimentary tract and metabolism products | 4 |
| Other dermatological preparations | 1 |
| Other gynecologicals | 1 |
| Other hematological agents | 1 |
| Other nervous system drugs | 12 |
| Other respiratory system products | 117 |
| Otologicals | 2 |
| Peripheral vasodilators | 23 |
| Pituitary and hypothalamic hormones and analogues | 52 |
| Preparations for treatment of wounds and ulcers | 8 |
| Psychoanaleptics | 148 |
| Psycholeptics | 405 |
| Sex hormones and modulators of the genital system | 1 |
| Stomatological preparations | 58 |
| Thyroid therapy | 44 |
{16}
| Medication Class | Number of Patients |
| --- | --- |
| Topical products for joint and muscular pain | 1 |
| Urologicals | 24 |
| Vaccines | 56 |
| Vasoprotectives | 1 |
| Vitamins | 85 |
## Accuracy at extreme glucose concentrations
An additional study was conducted to further assess the performance of the cobas pulse blood glucose monitoring system at the extreme upper and lower ends of the claimed measuring range. The study was performed using whole blood samples altered to achieve glucose concentrations of less than 80 mg/dL (52 samples) and greater than 300 mg/dL (52 samples). Samples were tested on the cobas pulse instrument by intended use operators using test strips from 3 lots and results compared to the results obtained on the comparator method (cobas 6000 system). The results are summarized below:
Results for Glucose Concentrations < 80 mg/dL:
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL |
| --- | --- | --- | --- |
| 52/52 (100.0%) | 52/52 (100.0%) | 52/52 (100.0%) | 52/52 (100.0%) |
Results for Glucose Concentrations > 300 mg/dL:
| | Within ±10% | Within ±12% | Within ±15% | Within ±20% |
| --- | --- | --- | --- | --- |
| 50/52 (96.2%) | 52/52 (100.0%) | 52/52 (100.0%) | 52/52 (100.0%) | 52/52 (100.0%) |
## Accuracy in samples representative of neonate samples
An additional study was performed using adult whole blood samples altered to achieve glucose concentrations between 10 and 50 mg/dL and adjusted to two levels of hematocrit at 40% and 65%, in order to simulate the hematocrit levels of neonatal blood. A total of 94 distinct samples were used for the study (48 samples at 40% hematocrit and 46 samples at 65% hematocrit). Samples were tested on the cobas pulse meter by intended use operators using test strips from 3 lots and results were compared to the results obtained on the comparator method (cobas 6000 system). The results are summarized below:
Results for 40% Hematocrit Concentrations
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL |
| --- | --- | --- | --- |
| 48/48 (100.0%) | 48/48 (100.0%) | 48/48 (100.0%) | 48/48 (100.0%) |
K220272 - Page 17 of 20
{17}
K220272 - Page 18 of 20
Results for 65% Hematocrit Concentrations
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±12 mg/dL | Within ±15 mg/dL |
| --- | --- | --- | --- |
| 46/46
(100.0%) | 46/46
(100.0%) | 46/46
(100.0%) | 46/46
(100.0%) |
## Capillary Fingerstick and Venous Whole Blood Limitations
The sponsor includes limitations against the use of the cobas pulse with capillary finger stick and venous whole blood. The labeling states the following:
- In comparing capillary whole blood glucose measurements using the cobas® pulse to venous plasma measurements using the comparator method, a large percentage of samples had significant bias to the comparator method.
- In comparing venous whole blood on the cobas® pulse to venous plasma on the comparator method for samples with glucose less than 75 mg/dL, 2 outliers showed significant bias to the comparator method, and protocol deviations led to a large number of venous sample exclusions resulting in uncertainty in the study.
These limitations were put in place due to risks posed by the performance of this device in these two sample types. The results of the cobas pulse glucose monitoring system were compared with those obtained on the comparator method and are as follows for each specimen type:
For capillary fingerstick samples, 54.5% (12/22) of samples with glucose concentrations <75 mg/dL were within ±12 mg/dL and 45.5% (10/22) exceeded a difference of ±15 mg/dL relative to the comparator method result. Additionally, for capillary fingerstick samples with glucose concentrations ≥75 mg/dL, 82.2% (562/684) of the samples were within ±12 mg/dL and 7.0% (48/684) exceeded a difference of ±20% relative to the comparator method results.
For venous whole blood samples, 94.6% (35/37) of samples with glucose concentrations <75 mg/dL were within ±12 mg/dL and 5.4% (2/37) exceeded a difference of ±15 mg/dL relative to the comparator method result. Additionally, protocol deviations led to the exclusion of more than 10% of venous whole blood samples from the study resulting in uncertainty in the results from this study.
## D Clinical Cut-Off:
Not applicable.
{18}
K220272 - Page 19 of 20
# E Expected Values/Reference Range:
The sponsor includes that following in the labeling: Expected blood glucose values for adults without diabetes is between 74-100 mg/dL. The normal glucose level for an adult without diabetes 2 hours post-meal is less than 140 mg/dL.
## Reference:
1. Tietz (2018) Textbook of Clinical Chemistry and Molecular Diagnostics, 6th edition, ISBN:978-0-323-35921-4.
2. ElSayed NA, Aleppo G, Aroda VR, et al., American Diabetes Association. 2. Classification and diagnosis of diabetes: Standards of Care in Diabetes—2023. Diabetes Care 2023;46(Suppl. 1):S19–S40.
# F Other Supportive Instrument Performance Characteristics Data:
1. Oxygen Study:
The impact of oxygen partial pressure (pO2) in blood on measurements with the cobas pulse blood glucose monitoring system was evaluated using whole blood samples adjusted to three concentration levels of glucose (target ranges 50 - 70 mg/dL, 110 - 130 mg/dL and 225 - 270 mg/dL). Each glucose range was tested at four pO2 levels of about 30, 60, 90, and 200 mm Hg. The values were compared with the glucose measurements obtained from the comparator method (cobas 6000 system). The results of the study demonstrate that pO2 does not significantly impact the results.
2. Hematocrit Study:
The effect of different hematocrit levels on the cobas pulse blood glucose monitoring system was evaluated using whole blood samples adjusted to hematocrit levels in 5% intervals in the range from 5 to 70% and at 5 glucose concentrations (30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL). The results were compared to those obtained from the comparator method (cobas 6000 system). The results of the study support the labeled hematocrit claimed range of 5-70%.
3. Cleaning and Disinfection:
The device is intended for multiple-patient use. Disinfection efficacy studies were performed on the exterior meter materials by an outside commercial testing laboratory, demonstrating complete inactivation of Hepatitis B Virus (HBV) with the chosen disinfectants, Clorox Germicidal Wipes (EPA registration # 67619-12) and Super Sani-Cloth Germicidal Disposable Wipe (EPA registration # 9480-4). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or in the external materials of the meter after 12,045 cleaning and disinfection cycles using each of the chosen disinfectants. The robustness studies were designed to simulate cleaning and disinfection over the 3-year multiple-patient use life of the meter. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
4. System Operating Conditions Testing:
To assess the performance of the cobas pulse glucose monitoring system when used under various operating temperature and humidity conditions, the system was tested at different temperature and relative humidity (RH) conditions including 10°C/45%RH, 42°C/45% RH, 12°C/10%RH, 12°C/90%RH, 30°C/90%RH and 40°C/10%RH. Each condition was tested
{19}
using whole blood at five blood glucose concentrations (30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL). The results from the cobas pulse blood glucose monitoring system under each test condition were compared to nominal control conditions. The study results support the labeled operating conditions claim of 12°C to 40°C (54 to 104 °F) and relative humidity range of 10% to 90%.
5. Altitude Study:
To assess the effect of altitude on the performance of the cobas pulse glucose monitoring system, the system was tested at an altitude of about 15,420 ft (4,700 meters) above sea level. Whole blood samples adjusted to 3 glucose concentrations (50-65 mg/dL, 100-120 mg/dL, and 200-250 mg/dL) were tested. The results obtained from the candidate system at 15,420 ft were compared to those obtained at an altitude of 295 ft (90 meters) above sea level. The results of the study support the claim that the cobas pulse glucose monitoring system can be operated at altitudes of up to 15,420 ft (4,700 meters).
6. Sample Volume Study:
The sponsor conducted a sample volume study using whole blood samples at 3 glucose concentrations (50-65, 100-120 and 200-250 mg/dL) tested at 5 sample volumes (0.2, 0.4, 0.6, 1.2, and 2.4 μL). Values obtained using the candidate system were compared to those obtained using the comparator method (cobas 6000 system). Results support the claimed minimum sample volume of 0.6 μL for the cobas pulse glucose monitoring system. The meter has an error message displayed if enough blood is not added to the test strip. This feature was validated and was shown to function as intended.
7. Electromagnetic Compatibility and Electrical Safety:
The sponsor provided documentation certifying that acceptable electrical safety and electromagnetic compatibility (EMC) testing had been performed and the system was found to be compliant.
8. Flex Studies:
Intermittent sampling, sample perturbation, drop testing, testing with used test strips, quality control lockout, transport, and vibration testing was completed by the sponsor. The testing performed demonstrated that the cobas pulse glucose monitoring system is robust to these expected use scenarios.
9. Test Strip Lot Release Protocol:
The test strip lot release protocol and acceptance criteria were reviewed and found to be acceptable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220272 - Page 20 of 20
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.