Retrospective clinical database of paired critical care capillary and plasma glucose specimens
Retrospective analysis of over 16,000 paired clinical specimens was used to demonstrate device performance in critical care settings, supporting the expansion of indications for use to include patients receiving intensive medical intervention.
Retrospective analysis; Critical care; Clinical performance; Real-world clinical data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of paired clinical specimens
Patients in critical care departments; Sample Size: >16,000 paired specimens
Roche Cobas Modular P800 Hexokinase System (central laboratory)
Accuracy of capillary whole blood glucose results compared to plasma glucose results
Indications for Use
The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy. The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick. It is not intended for use with neonate cord blood specimens. It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.
Device Story
Handheld electrochemical biosensor system; measures glucose in capillary, venous, arterial, and neonatal whole blood. Input: 1.2 μL blood sample applied to test strip via capillary action; glucose oxidase enzyme chemistry generates current proportional to glucose concentration. Output: quantitative glucose readout on meter display. Used in hospital settings (ICU, OR) by nursing staff/CLIA-waived operators. Data stored on-device; bi-directional transfer via docking station to data management systems. Supports clinical decision-making by providing rapid bedside glucose monitoring. Benefits: enables point-of-care testing for patients receiving intensive therapy, reducing turnaround time compared to central lab testing.
Clinical Evidence
Two clinical studies evaluated capillary fingerstick performance in critical care settings. Study 1 (n=568) compared meter results to central lab hexokinase method; 96.8% of results ≥75 mg/dL were within ±20% of comparator. Study 2 (retrospective, n>16,000) compared bedside capillary results to central lab plasma results; 96.4% of results ≥75 mg/dL were within ±20% of comparator. Bench testing (precision, linearity, interference, hematocrit, altitude, temperature/humidity) previously established in k060345/k063821/k132121.
Technological Characteristics
Electrochemical biosensor; glucose oxidase enzyme; ferricyanide mediator. Handheld form factor; 3.7V Lithium battery; 1.2 µL sample volume; 6-second test time. Connectivity via IEEE 802.11 a/b/g Wi-Fi. Non-volatile data storage for 1,500 results. Automatic calibration (no code).
Indications for Use
Indicated for quantitative glucose determination in capillary finger stick, venous, arterial, neonate arterial, and neonate heel stick whole blood specimens in all hospital/professional healthcare settings, including patients receiving intensive medical intervention/therapy. Not for neonate cord blood, diabetes screening, or diagnosis.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k181043
B. Purpose for Submission:
This submission is a Dual 510(k) and CLIA Waiver by Application (Dual Submission) tracked as k181043 and CW180005. This 510(k) is to expand the indications for use for the StatStrip Glucose Hospital Meter System to include capillary whole blood samples for use in all hospitalized patients.
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative amperometric assay, glucose oxidase
E. Applicant:
Nova Biomedical Corporation
F. Proprietary and Established Names:
StatStrip Glucose Hospital Meter System
G. Regulatory Information:
| Product Code | Regulation Name | Classification | Regulation Section | Panel |
| --- | --- | --- | --- | --- |
| PZI | Glucose Test System | II | 21 CFR 862.1345 | Clinical Chemistry |
H. Intended Use:
1. Intended Use:
See Indications for Use below.
{1}
2. Indications for Use:
The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens throughout all hospital and all professional healthcare settings including patients receiving intensive medical intervention/therapy.
The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick. It is not intended for use with neonate cord blood specimens. It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia.
3. Special conditions for use statement(s):
- For prescription use only
- For in vitro diagnostic use only
- The system has not been evaluated for use with neonate venous blood.
- Temperature and humidity extremes - Test results may be inaccurate when test strips are stored outside of the storage and handling conditions.
- Altitudes above 15,000 feet (4500 meters) above sea level have not been evaluated.
- Specimens - Only fresh whole blood or whole blood collected in lithium heparin collection devices should be used for arterial and venous specimens.
- Fluoride, EDTA, Sodium, and Ammonium blood collection devices should not be used.
- Use only whole blood. Do not use serum or plasma.
- Should only be used with single-use, auto-disabling lancing devices
- Caution should be exercised when testing capillary whole blood due to potential pre-analytical variability in capillary specimen collection.
- A capillary whole blood specimen relies upon an adequate, non-compromised capillary blood flow. The healthcare provider must be aware that a capillary whole blood specimen glucose result may not always be the same as an arterial or a venous whole blood glucose result, especially when the patient’s condition is rapidly changing.
- If a capillary whole blood glucose result is not consistent with a patient’s clinical signs and symptoms, glucose testing should be repeated with either an arterial or venous specimen on the StatStrip Glucose Hospital Meter System.
4. Special instrument requirements:
StatStrip Blood Glucose Hospital Meter
I. Device Description:
The StatStrip Glucose Hospital Meter System (previously cleared under k060345, k063821 and k132121 and k150281) consists of a hand held StatStrip Glucose Hospital meter,
{2}
StatStrip Test Strips (sold separately), Nova StatStrip Control Solutions (Levels 1, 2 and 3; sold separately), Nova StatStrip Linearity Test Kit solutions (5 levels; sold separately), docking station, Quick Reference Guide, and User Manual.
Three levels of control solutions (Level 1, Level 2, Level 3) and five levels of linearity solutions (Level 1, Level 2, Level 3, Level 4, Level 5) are available for use with the StatStrip Glucose Hospital Meter System and were previously cleared in k060345.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Nova StatStrip Glucose Hospital Meter System
2. Predicate 510(k) number(s):
k150281
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device (k150281) | Candidate Device (k181043) |
| Indications for Use/Intended Use | Intended for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood and neonate heel stick specimens for use in determining dysglycemia | Same |
| Population limitation | Capillary whole blood specimens (e.g. obtained by finger stick) should not be used in patients receiving intensive medical intervention/therapy because of the potential for pre-analytical collection error and specifically in patients with decreased peripheral blood flow, as it may not truly reflect the patient's true physiological state. Examples include, but are not limited to, severe | Cleared for use with capillary fingerstick samples in all hospital patients including those receiving intensive medical intervention/therapy. |
{3}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate Device (k150281) | Candidate Device (k181043) |
| | hypotension, shock, hyper-osmolar-hyperglycemia (with or without ketosis) and severe dehydration. | |
| Enzyme | Glucose Oxidase | Same |
| Test Principle | Electro-chemical biosensor | Same |
| Sample type | Capillary finger stick, venous and arterial whole blood, neonatal arterial whole blood and neonatal heelstick. | Same |
| Measuring range | 10-600 mg/dL | Same |
| Measuring time | 6 sec | Same |
| Sample volume | 1.2 μL | Same |
| Data storage | 1000 Patient Test
200 QC Tests
4000 Operators | Same |
| Wi-Fi network connectivity | Yes | Same |
K. Standard/Guidance Document Referenced (if applicable):
IEC 61010-1:2010, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
EN 60601-1-2:2007, Medical electrical equipment. General requirements for basic safety and essential performance. Collateral standard. Electromagnetic compatibility. Requirements and tests
L. Test Principle:
The Nova StatStrip Hospital Meter System is based on electrochemical biosensor technology and the principle of capillary action. The system quantitatively measures blood glucose levels using glucose oxidase enzyme chemistry. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Previously established in k060345.
b. Linearity/assay reportable range:
As established in k063821, the reportable range for the Nova StatStrip Glucose Hospital Meter System is 10 to 600 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability to NIST Standard SRM917B, as established in k060345.
Test Strip:
Test strip stability protocols and acceptance criteria were evaluated in k060345 and were found to be acceptable to support the claimed shelf life of 24 months at 33-86°F and 10-90% relative humidity (RH) and the claimed open-vial stability of 180 days when stored at the recommended storage temperatures 33-86°F and 10-90% RH or until the expiration date printed on the label, whichever comes first. The labeling instructs the users not to freeze the test strips.
d. Detection limit:
This range was verified by the linearity established in k063821; section M.1.b.
e. Analytical specificity:
Potential interference from common endogenous and exogenous substances, as well as in vivo interference, was evaluated in k060345 and k132121.
f. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison study
Performance for capillary finger stick, venous whole blood and arterial whole blood samples from non-hospitalized patients was established in k060345.
Performance for neonatal heelstick and neonatal arterial samples was established in k063821.
{5}
Performance for venous and arterial, neonatal heelsticks and neonatal arterial samples from patients throughout the hospital was established in k132121.
In this submission, additional comparison studies were performed using fingerstick capillary samples at three hospital sites as follows:
## Study 1
For Study 1, 568 capillary whole blood fingerstick specimens were obtained from patients within three different critical care units including the cardiovascular intensive care unit (CVICU), medical intensive care unit (MICU), and the operating room (OR). This study included 80 unique patient conditions receiving a total of 3,785 medications representing 17 parent drug classes. All testing with the StatStrip Blood Glucose Hospital Meter (StatStrip Meter) was performed by CLIA waived operators (non-laboratory personnel, typically nursing staff) within each of these three critical care settings. Capillary whole blood glucose results on the StatStrip Meter were compared to matched arterial or venous plasma results obtained on a comparator method, the Roche Cobas Modular P800 Hexokinase System, located in a central laboratory. The glucose ranges of the samples, according to the comparator method, ranged from 74 mg/dL to 379 mg/dL glucose. The results from the capillary fingerstick samples obtained from the StatStrip Hospital Meter compared to the results from the comparator method are summarized below:
Fingertip capillary samples with glucose concentrations <75 mg/dL
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 12 mg/dL | Within ± 15 mg/dL | Exceeds ± 15 mg/dL |
| --- | --- | --- | --- | --- |
| 1/1 (100%) | 1/1 (100%) | 1/1 (100%) | 1/1 (100%) | 0/1 (0%) |
Fingertip capillary samples with glucose concentrations ≥75 mg/dL
| Within ± 5 % | Within ± 10 % | Within ± 12 % | Within ± 15 % | Within ± 20 % | Exceeds ± 20 % |
| --- | --- | --- | --- | --- | --- |
| 277/567 (48.9%) | 450/567 (79.4%) | 484/567 (85.4%) | 516/567 (91.0%) | 549/567 (96.8%) | 18/567 (3.2%) |
## Study 2
Over 16,000 paired critical care capillary glucose specimens were retrospectively identified and met the following criteria:
Within critical care departments, a capillary fingerstick specimen, and a venous/arterial glucose result were measured at the bedside by a CLIA Waived operator using the BGMS.
Subsequently a plasma glucose test was performed on the same subject on the central laboratory hexokinase method within 15 minutes.
6
{6}
Capillary whole blood glucose results on the StatStrip Meter were compared to matched arterial or venous plasma results obtained on a comparator method, Roche Cobas Modular P800 Hexokinase System. The glucose ranges of the samples, according to the comparator method, ranged from 27 mg/dL to 667 mg/dL glucose. The results from the capillary fingerstick samples obtained from the StatStrip Hospital Meter compared with the results from the comparator method are summarized below:
Fingertip capillary samples with glucose concentrations <75 mg/dL
| Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 12 mg/dL | Within ± 15 mg/dL | Exceeds ± 15 mg/dL |
| --- | --- | --- | --- | --- |
| 907/1894 (47.9%) | 1470/1894 (77.6%) | 1614/1894 (85.2%) | 1737/1894 (91.7%) | 157/1894 (8.3%) |
Fingertip capillary samples with glucose concentrations ≥75 mg/dL
| Within ± 5 % | Within ± 10 % | Within ± 12 % | Within ± 15 % | Within ± 20 % | Exceeds ± 20 % |
| --- | --- | --- | --- | --- | --- |
| 7473/14884 (50.2%) | 11087/14884 (74.5%) | 12799/14884 (86.0%) | 13712/14884 (92.1%) | 14350/14884 (96.4%) | 534/14884 (3.6%) |
b. Matrix comparison
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Normal (non-diabetic) adult fasting: Less than 100 mg/dL (5.55 mmol/L) and less than
{7}
140 mg/dL (7.77 mmol/L) 1-2 hours after meals.
American Diabetes Association. Classification and Diagnosis of Diabetes: Standards of Medical care in Diabetes. Diabetes Care (2018), Volume 41, Supplement 1.
## N. Instrument Name:
Nova StatStrip Glucose Hospital Meter, previously cleared in k150281.
## O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ X or No ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
The Nova StatStrip Glucose Hospital Meter memory will store 1000 patient tests, 200 QC tests, and 4000 operators. The meter contains a laser barcode scanner that allows for scanning patient identification information that may also be entered manually.
4. Specimen Sampling and Handling:
The glucose test is intended to be used with capillary fingerstick whole blood, arterial, venous, neonatal heel stick and neonatal arterial. The blood sample is applied directly to the test strip by capillary action.
The meter stores patient test data, quality control test data, and other information relating to the patient, patient sample, operator, reagents, and meter. Meter setup options relating to authorized operators, reagent lots, quality control preferences, and other operational settings are customizable. Data is transferred bi-directionally between the meter, data docking station, and separate data management system each time a meter is placed in to a data docking station.
8
{8}
5. Calibration:
The meter does not require the user to input a test strip code or perform any other calibration.
6. Quality Control:
Three levels of aqueous ready to use glucose control solutions are available with this system (Level 1, Level 2, and Level 3). Control solution testing can be performed by pushing the QC key, entering (or scanning) the test strip lot number. Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the vial label of the control being used.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1) Hematocrit Study:
As established in k060345 and k063821 to support the claimed hematocrit range of 20-65%.
2) Altitude Study:
As established in k060345 to support the use of the device up to 15,000 ft.
3) Temperature and humidity studies:
As established in k060345 to support the claimed operating condition range of 59°F - 104°F and 10-90% relative humidity.
4) Infection Control Studies:
The device is intended for multiple-patient use. Disinfection efficacy studies were performed (in k132121) on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, Clorox Germicidal Wipes, EPA registration # 67619-12 was validated for use with the meter. Robustness studies were also performed by the sponsor (in k132121) demonstrating that there was no change in performance or in the external materials of the modified StatStrip Glucose Hospital Meter after 10,950 cleaning and disinfection cycles (one cycle includes one cleaning wipe plus one disinfecting wipe) using the Clorox Germicidal Wipes to simulate 3 years of device use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
5) Electromagnetic Compatibility and Electrical Safety:
Established in k150281.
6) Wireless Data Transmission Test:
As established in k150281 to support functional use in a hospital environment.
9
{9}
7) Clinical Chemistry and Clinical Toxicology Devices Panel:
A FDA meeting of the Clinical Chemistry and Clinical Toxicology Panel was held on March 30th, 2018, to discuss the benefits and risks of measuring capillary blood using blood glucose meters in patients receiving intensive medical intervention/therapy. The data from the 2 studies above using the Nova StatStrip meter were part of the data presented to the panel. The panel concluded that with respect to the performance data generated in the studies presented during the panel, the benefits outweigh the risk in this patient population receiving intensive medical intervention/therapy.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirement of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.