The CenterVue DRSplus is a confocal scanning ophthalmoscope indicated for color and infrared imaging of a human retina without the use of a mydriatic agent.
Device Story
DRSplus is a confocal scanning ophthalmoscope for non-mydriatic retinal imaging. Device uses infrared (825-870 nm) and white (420-675 nm) LEDs to capture retinal images. Software version 2.0 enables acquisition, storage, and review of infrared images, which were previously used only for internal alignment/focusing. Operation is automated, featuring auto-alignment, auto-focus, auto-exposure, and auto-capture. Used in clinical settings by eye care professionals to visualize retinal structures. Output is displayed on a 10.1-inch multi-touch screen for clinician review. Device aids in retinal assessment by providing color, red-free, and infrared imaging modalities without pharmacological dilation, improving workflow efficiency and patient comfort.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing and image comparisons. Compliance with ISO 10940 (fundus cameras), ISO 15004-1 (ophthalmic instruments), and ANSI Z80-36 (light hazard protection) confirmed.
Technological Characteristics
Confocal, line-scanning ophthalmoscope. Light sources: Infrared LED (825-870 nm), White LED (420-675 nm). Field of view: 45° (H) x 40° (V). Sensor: 10 MP. Connectivity: SSD storage. Power: 12V DC. Class I, Type B applied part. IPX0. Software version 2.0.
Indications for Use
Indicated for color and infrared imaging of the human retina in patients without the use of a mydriatic agent.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2021
CenterVue SpA Luca Scienza Quality and Regulatory Affairs Manager Via S. Marco 9H Padova, 35129 Italy
Re: K213705
Trade/Device Name: DRSplus Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: November 19, 2021 Received: November 24, 2021
Dear Luca Scienza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Elvin Ng
Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213705
Device Name DRSplus
Indications for Use (Describe)
The CenterVue DRSplus is a confocal scanning ophthalmoscope indicated for color and infrared imaging of a human retina without the use of a mydriatic agent.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| Proprietary Name(s): | DRSplus |
|------------------------|------------------------------------------------------------------------|
| Type of submission: | Special |
| 510(k) submission: | K213705 |
| Date of submission: | November 19th, 2021 |
| Manufacturer: | CENTERVUE S.p.A. |
| | Via San Marco 9H |
| | 35129 Padova - ITALY |
| Applicant and contact: | Mr. Luca Scienza |
| | Centervue S.p.A. |
| | Manager of Quality and Regulatory Affairs |
| | Via San Marco 9H, 35129 Padova - ITALY |
| | Phone: +39 049 501 8399 |
| | Fax: +39 049 501 8398 |
| | Email: luca.scienza@icare-world.com |
| Product Code: | MYC |
| Regulation Number: | 886.1570 |
| Common Name: | Ophthalmoscope |
| Panel: | Ophthalmic |
| Class: | Class II |
| Indications for Use: | The CenterVue DRSplus is a confocal scanning ophthalmoscope |
| | indicated for color and infrared imaging of a human retina without the |
| | use of a mydriatic agent. |
{5}------------------------------------------------
### Predicate Device Information
| Proprietary name: | DRSplus |
|---------------------|--------------------------------------------------------------------|
| 510(k): | K192113 |
| Product Code: | MYC |
| Regulation number: | 886.1570 |
| Common name: | Ophthalmoscope |
| Panel: | Ophthalmic |
| Class: | II |
| Indication for use: | The CenterVue DRSplus is a confocal scanning ophthalmoscope |
| | indicated for color imaging of a human retina without the use of a |
| | mydriatic agent. |
#### Reference Device Information
| Proprietary name: | EIDON |
|---------------------|-------------------------------------------------------------------------------------------------------------------|
| 510(k): | K142047 |
| Product Code: | MYC |
| Regulation number: | 886.1570 |
| Common name: | Ophthalmoscope |
| Panel: | Ophthalmic |
| Class: | II |
| Indication for use: | The CenterVue EIDON is intended for taking digital images of a human retina without the use of a mydriatic agent. |
The reference device has been included only in the context of image comparison, as no comparison of Infrared Imaging Modality is possible between the subject device and the predicate device.
{6}------------------------------------------------
### Device description
The CenterVue DRSplus with Software version 2.0 is a modification of the CenterVue DRSplus (K192113).
The DRSplus (K192113) is a scanning ophthalmoscope which uses infrared and white light to obtain confocal images of the retina, without pharmacological dilation.
With respect to the previous cleared device, DRSplus with software version 2.0 (subject of this submission) provides one additional software feature that enable to acquire, store and review infrared retinal images, in addition to color and red-free photos which were the only imaging modalities of the previously cleared device.
This feature does not require any hardware modification, because the infrared illumination is already present in the previously cleared device: in fact, the infrared retinal pictures were automatically taken by the device for alignment and focusing purposes, but they were not presented in the patient's image collection. In the proposed device modification, the user can decide to acquire also infrared pictures that are stored and displayed amongst the other, color and red-free images.
The modified device uses the same base technology and maintains the same Intended Use of the previously cleared device; the Indications for Use of the modified device are amended to add the infrared imaging modality.
The fundamental scientific technology of the device is unchanged from the previously cleared device and remains confocal, line scanning, LED-based, imaging.
The functional differences between DRSplus with software version 2.0 and the previously cleared device are as follows:
- . DRSplus with software 2.0 allows the user to take also IR pictures of the retina;
- . DRSplus with software 2.0 allows the user to save and review IR pictures of the retina.
No technological differences between DRSplus with software version 2.0 and the previously cleared device exist. In particular:
- . DRSplus with software version 2.0 uses the same technological principle of retinal acquisitions of the cleared device;
- . DRSplus with software version 2.0 uses the same illuminator, illumination optics and viewing optics of the cleared device.
Other than the above, no other differences exist between the DRSplus with software version 2.0 and the previously cleared device, in particular the principle of operation, mechanism of action and interaction with the patient are unchanged.
{7}------------------------------------------------
### TECHNICAL SPECIFICATIONS of the DRSplus with Software version 2.0
#### Class and type of applied part
- Class I, Type B (according to IEC 60601-1).
#### IP classification:
- IPX0 (according to the degree of protection provided by the enclosure with respect to harmful penetration of particulate matter or water).
#### Image acquisition:
- . Minimum pupil size: 3.2 mm (non-mydriatic use)
- 45° (H) x 40° (V) captured in a single exposure ● Field of view:
- 3600 x 2910 (10 MP) ● Sensor size:
- infrared LED (825-870 nm), white LED (420-675 nm) ● Light source:
- color, red-free, infrared . Imaging modalities:
- 25 mm ● Working distance:
- 3.7 μm ● Pixel pitch:
#### Other features:
- Automatic operation: auto-alignment, auto-focus, auto-exposure, auto-capture
- . Focus adjustment range: -15 D to +15 D
- . Internal fixation target: 10 positions
- Display: 10.1″ multi-touch, color .
- Hard disk: SSD, 512 GB .
#### Dimensions:
- 11 Kg (24 lbs) · Weight: • Size (W x H x D): 300 X 450 X 650 mm (11.8" x 17.7" x 25.6")
#### Power supply:
- · Voltage: 12 V DC . Power consumption: 60 W
{8}------------------------------------------------
### Comparison with previously cleared device
The previously cleared device is identified as follows:
| Proprietary Name: | DRSplus |
|----------------------|---------------------------------|
| Manufacturer: | CenterVue S.p.A. |
| 510(k) Number: | K192113 |
| Clearance Date: | 25 November 2019 |
| FDA Product Code: | MYC |
| Classification Name: | Ophthalmoscope, Laser, Scanning |
| Regulation Number: | 886.1570 |
The Subject Device and the Predicate Device share many identical or similar properties and features. In particular, the fundamental scientific technology of the subject device is unchanged from the predicate and remains confocal, line scanning, LED-based, ophthalmoscopy. Also, the principle of operation, mechanism of action and interaction with the patient are unchanged.
With respect to the previous cleared device, DRSplus with software version 2.0 (subject of this submission) provides one additional software feature that enable to acquire, store and review infrared retinal images, in addition to color and red-free photos which were the only imaging modalities of the previously cleared device.
The functional differences between DRSplus with software version 2.0 and the previously cleared device are as follows:
- . DRSplus with software 2.0 allows the user to take also IR pictures of the retina;
- . DRSplus with software 2.0 allows the user to save and review IR pictures of the retina.
No technological differences between DRSplus with software version 2.0 and the previously cleared device exist. In particular:
- DRSplus with software version 2.0 uses the same technological principle of retinal acquisitions of the . cleared device;
- . DRSplus with software version 2.0 uses the same illumination optics and viewing optics of the cleared device.
Other than the above, no other differences exist between the DRSplus with software version 2.0 and the previously cleared device, in particular the principle of operation, mechanism of action and interaction with the patient are unchanged.
None of these differences have any significant effect on safety or effectiveness of the Subject Device. Notably, with respect to the impact of these design modifications on a key part of the product's risk analysis (i.e. optical radiation safety), Centervue has confirmed conformance with the applicable recognized standard, ANSI Z80-36:2016, as no modification has been introduced in the illumination and viewing systems.
<sup>4</sup> Neither the CenterVue DRSplus with software 2.0 contain lasers but rather use LEDs for confocal imaging
{9}------------------------------------------------
### Performance data
Continued conformance with the following standards has been confirmed in support of the substantial equivalence determination:
#### ANSI Z80-26 (light Hazard Protection)
DRSplus with software version 2.0 fulfills the requirements for a Group 1 determination according to ANSI Z80-36.
#### ISO 15004-1
The subject device complies with the ISO 15004-1:2006 standard for ophthalmic instruments.
#### IEC 62304
The subject device software complies with IEC 62304.
#### ISO 10940
DRSplus with software version 2.0 fully complicable requirements of ISO 10940 - Ophthalmic instruments — Fundus Cameras.
#### Clinical data
No clinical data required to evaluate the proposed change, only image comparisons are provided.
#### Conclusion
Based on the information contained within this submission, it is concluded that the CenterVue DRSplus with software version 2.0 is substantially equivalent to the previously cleared device already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.