The CenterVue EIDON is intended for taking digital images of a human retina without the use of a mydriatic agent.
Device Story
CenterVue EIDON is a standalone scanning ophthalmoscope for prescription use. It captures high-resolution color and infrared (IR) digital images of the human retina. The device uses an illumination system with IR (850nm, 940nm), white, blue, and green LEDs. An oscillating mirror scans a line of light across the retina (60° field of view). An imaging system with a focusing lens and 14-megapixel CMOS camera collects back-reflected light. An anterior segment alignment system uses two cameras and IR LEDs for stereoscopic pupil tracking. The device includes a motorized chin-rest and head-rest for patient positioning. Operated by clinicians in a clinic setting, the device provides digital retinal images to assist in ocular examination. The non-mydriatic operation benefits patients by eliminating the need for pupil-dilating drops, improving workflow and patient comfort.
Clinical Evidence
Bench testing only. The device conforms to IEC 60601-1:2005, IEC 60601-1-2:2007, ISO 10940:2009, ISO 15004-1:2006, and ISO 15004-2:2007. Performance metrics verified include 14-megapixel resolution, 60° x 55° field of view, 15-micron optical resolution, and a focus range of -12D to +15D.
Technological Characteristics
Scanning ophthalmoscope; IR (850nm, 940nm), white, blue, and green LED illumination; 14MP CMOS sensor; oscillating mirror scanning mechanism; motorized chin/head rest; standalone unit with tablet interface; conforms to IEC 60601-1, IEC 60601-1-2, ISO 10940, ISO 15004-1, and ISO 15004-2.
Indications for Use
Indicated for capturing digital retinal images in humans without mydriatic agents.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of three stylized human profiles facing to the right. The profiles are connected and appear to be layered on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
CenterVue S.p.A. Mr. Roger Gray V.P. Quality and Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 00153 Rome, Italy
Re: K142047
Trade/Device Name: CenterVue EIDON Ophthalmoscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: September 30, 2014 Received: October 2, 2014
Dear Mr. Gray:
This letter corrects our substantially equivalent letter of November 12, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142047
Device Name CenterVue EIDON
Indications for Use (Describe)
The CenterVue EIDON is intended for taking digital images of a human retina without the use of a mydriatic agent.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for centervue. The logo consists of the word "centervue" in lowercase letters, with a stylized globe above it. The globe is made up of gray squares and has a black circle in the center.
# 510(k) Summary in accordance with 21 CFR 807.92(c)
| Device Name: | CenterVue EIDON Ophthalmoscope |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional |
| Date of Submission: | 24 July 2014 (summary revised on 18 September 2014) |
| Manufacturer: | CenterVue S.p.A.<br>via San Marco 9h<br>35129 Padova<br>Italy |
| 510(k) Owner: | CenterVue S.p.A.<br>via San Marco 9h<br>35129 Padova<br>Italy |
| Phone:<br>Fax: | +39 049 7396 147<br>+39 049 7396 148 |
| 510(k) Submitter and Contact: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy<br>Phone: +39 06 578 2665<br>Fax: +30 06 574 3786<br>Email: rgray@donawa.com |
| FDA Product Code: | MYC |
| FDA Regulation Number: | 886.1570 |
| FDA Classification Name: | Ophthalmoscope |
| Classification Panel: | Ophthalmic |
| Common Name: | Ophthalmoscope |
| FDA Classification: | Class II |
| FDA Identification: | An ophthalmoscope is an AC-powered or battery-powered<br>device containing illumination and viewing optics intended to<br>examine the media (cornea, aqueous, lens, and vitreous) and<br>the retina of the eye. |
| Indications for Use/Intended Use: | The CenterVue EIDON is intended for taking digital images of a<br>human retina without the use of a mydriatic agent. |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Centervue. The logo consists of a stylized globe-like shape above the word "centervue" in lowercase letters. The globe is made up of several gray blocks arranged in a circular pattern, with a black circle in the center. The word "centervue" is written in a simple, sans-serif font.
#### Device Description:
The CenterVue EIDON scanning ophthalmoscope operates as a standalone unit, running a dedicated software application, intended for prescription use only, and includes:
- 1. Optical head, including a removable lens cap;
- 2. Patient head-rest, including removable front-rest;
- 3. Patient chin rest;
- Base, including touch-screen device (tablet with magnetic holder and USB cable), USB joystick 4. and an external power supply.
EIDON uses infrared and visible light to obtain colored confocal digital images of a human retina without a mydriatic agent.
The CenterVue EIDON device operates on the following principles:
- An illumination system consisting of infrared (IR) LEDs (850nm, 940nm), a white LED, a blue LED and a) a green LED illuminates the patient eye with the following functionality:
- The IR LED with a centroid wavelength of 850 nm allows the capture of IR photos. The patient retina is uniformly illuminated by a line in a horizontal direction. Along the optical path there is an oscillating mirror which scans the line in order to illuminate the retina with a field of view of 60°.
- • Two IR Leds with a centroid wavelength of 940 nm are seen from the eye in a free viewing system. The two LEDs are equally shifted with respect to the machine optical axis. The LEDS are switched on during all exams in order to enable pupil tracking.
- The white and blue LEDs allow the capture of color photos. The retina is uniformly illuminated by a line in a horizontal direction. Along the optical path there is an oscillating mirror which scans the line in order to illuminate the retina with a field of view of 60°.
- The green LED is used as fixation target.
- b) An imaging system collects back-reflected light from the retina and creates a high resolution image. A focusing lens is included in the imaging path to achieve optimal retinal focusing on a CMOS camera having a resolution of 14 megapixels.
- An anterior segment alignment system is included, using two cameras and the two IR LEDs: the c) LEDs illuminate the anterior segment by diffusion, whereas allow a stereoscopic reconstruction of the pupil's position to be obtained with respect to the instrument front lens.
The EIDON device interacts with the patient by directing infrared, white, blue and green wavelength illumination into the patient's eye. The chin-rest and head-rest are the only parts of the device that contact the patient. The chin-rest includes a patient proximity sensor and is motorized for height adjustment.
### Bench tests
The CenterVue EIDON device has been tested and found to be in conformance with IEC 60601-1:2005 and IEC 60601-1-2:2007. In addition, the device meets the requirements of ISO 10940:2009, ISO 15004-1:2006, and ISO 15004-2:2007.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a stylized logo. The logo features a sphere with a dark center, surrounded by a series of gray cubes that appear to be orbiting or partially encasing the sphere. The cubes are arranged in a circular pattern, giving the impression of a three-dimensional structure. The overall design is modern and abstract.
#### Performance data:
| Performance item | Requirements | Test results |
|----------------------|-----------------------------|----------------------------|
| Sensor resolution | 14 megapixels (4608 x 3288) | 4608 x 3288 |
| Resolution on retina | 60 pixel / deg | 59.7 pixel / deg |
| Optical resolution | 15 microns at the center | 15.1 microns at the center |
| Field of view | 60° (H) × 55° (V) | 60.0° x 55.1° |
| Pixel pitch | 4.9 microns max | 4.79 microns |
| Range of focus | -12 D to +15D | -12 D to +15D |
| Minimum pupil size | 3 mm | 3 mm |
#### Comparison with predicate devices:
The predicate devices selected for comparison with the CenterVue EIDON Ophthalmoscope are:
| Predicate Device 1: | EasyScan |
|----------------------|-----------------------------------|
| 510(k) Sponsor: | i-Optics B.V. |
| 510(k) Number: | K111988 |
| Clearance Date: | 28 October 2011 |
| FDA Product Code: | MYC |
| Classification Name: | Scanning Laser Ophthalmoscope |
| Regulation No: | 886.1570 |
| Predicate Device 2: | Digital Retinography system (DRS) |
| 510(k) Sponsor: | CenterVue S.p.A. |
| 510(k) Number: | K101935 |
| Clearance Date: | 27 October 2010 |
| FDA Product Code: | HKI |
| Classification Name: | Ophthalmic Camera |
| Regulation No: | 886.1120 |
The Subject Device and the Predicate Devices share many identical or similar properties and features. Differences include:
- . Field of view
- . Imaging principle
- . Fixation target
- . Working distance
- Sensor size
- . Minimum pupil size
None of these differences have any significant effect on safety or effectiveness of the Subject Device.
#### Conclusion:
Based on the information contained within this submission, it is concluded that the CenterVue EIDON Ophthalmoscope is substantially equivalent to the identified predicate devices already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.