Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D and Stellaris 3D Ceph
Applicant
Fona S.R.L
Product Code
OAS · Radiology
Decision Date
Mar 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Stellaris 2D: Extraoral source dental X-ray system intended to perform panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry. Stellaris 2D Ceph: Extraoral source dental X-ray system intended to perform panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations. Stellaris 3D: Extraoral source dental X-ray system intended to perform 3D and panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry. Stellaris 3D Ceph: Extraoral source dental X-ray system intended to perform 3D, panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
Device Story
Extraoral dental X-ray system; performs panoramic, cephalometric, and 3D (CBCT) imaging. Rotating arm with motor-driven rototranslatory movements follows patient morphology; X-ray generator produces beam based on user-selected technique factors; CMOS or CsI flat panel sensors capture X-rays, converting them to electrical signals. Operated by dental professionals in clinical settings; patient head stabilized via bite blocks, chin rests, and temple supports; exposure controlled by manual hand-switch with dead-man functionality. Ethernet connection transmits data to external PC for image acquisition, processing, and storage using third-party software (e.g., OrisWin DG Suite). Provides diagnostic images for dento-maxillo-facial region; aids surgeons, dentists, and radiologists in diagnosis and treatment planning.
Clinical Evidence
No formal clinical trial; clinical evidence consists of an evaluation of sample clinical images by the manufacturer to confirm diagnostic adequacy per FDA guidance for solid-state digital X-ray panels.
Technological Characteristics
Extraoral X-ray system; CMOS/CsI flat panel sensors; high-frequency multi-pulse generator (ripple ≤4%); 0.5 focal spot (IEC 60336); 60-86 kV tube voltage; 2.5-10 mA tube current. Class I laser aiming lights (650 nm). Ethernet connectivity. Biocompatible patient-contact materials (ISO 10993-1). Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 60601-1-3 (Radiation), EN 60601-2-63 (Performance), EN 62304 (Software).
Indications for Use
Indicated for general and pediatric dentistry patients requiring extraoral dental X-ray imaging, including panoramic, cephalometric, and 3D dento-maxillo-facial exams, and carpal images for clinical investigations.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
March 2, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".
FONA S.r.l % Krupa Srivastava RA Consultant Via G. Galilei 11 Assago, Milan 20057 ITALY
Re: K213579
Trade/Device Name: Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D and Stellaris 3D Ceph Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: January 25, 2022 Received: January 14, 2022
#### Dear Krupa Srivastava:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213579
Device Name
Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D and Stellaris 3D Ceph
#### Indications for Use (Describe)
Stellaris 2D: Extraoral source dental X-ray system intended to perform panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry.
Stellaris 2D Ceph: Extraoral source dental X-ray system intended to perform panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
Stellaris 3D: Extraoral source dental X-ray system intended to perform 3D and panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry.
Stellaris 3D Ceph: Extraoral source dental X-ray system intended to perform 3D, panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for "FONA". The word "FONA" is written in large, bold, gray letters. Above the word "FONA" is a curved yellow line. Below the word "FONA" is the phrase "CLOSER TO YOU" written in smaller, gray letters.
#### 510(k) SUMMARY K213975
| Date Prepared: | January 11, 2022 |
|---------------------------|------------------------------------------------------------------------|
| Submitters Information | |
| Name | FONA s.r.l. |
| Address | Via G.Galilei 11<br>20057 Assago (MI), Italy |
| Contact Person | Luigi Germanò |
| Contact Telephone | +39 0245712171 |
| Device Information | |
| Device/ Trade Name | Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D<br>and Stellaris 3D Ceph |
| Common Name | X-Ray, Tomography, Computed, Dental |
| Classification Name | Computed Tomography X-ray system |
| Classification Regulation | 21 CFR 892.1750 - Computed Tomography<br>X-ray system |
| Product code | OAS |
#### Predicate Devices:
The following legally marketed predicate devices have been chosen for the determination of substantial equivalence:
| Primary Predicate Device<br>Trade Name | ORTHOPHOS SL |
|----------------------------------------|----------------------------------|
| 510(k) Number | K150217 |
| Classification Name | Computed Tomography X-ray system |
| Classification Regulation | 21 CFR 892.1750 |
| Product code | OAS |
Traditional 510(k)- Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph FONA S.r.l
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for FONA. The word "FONA" is written in large, bold, gray letters. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray letters. Above the word "FONA" is a curved yellow line.
Reference Device
| Trade Name | FONA XPan DG, XPan DG Plus, ART Plus & ART Plus C |
|---------------------------|---------------------------------------------------|
| 510(k) Number | K161131 |
| Classification Name | Extraoral Source X-Ray System |
| Classification Regulation | 21 CFR 872.1800 |
| Product code | MUH |
## Indications for use
Stellaris 2D: Extraoral source dental X-ray system intended to perform panoramic exams with production of diagnostic images in the dento-maxillofacial region and in subregions, for general and pediatric dentistry.
Stellaris 2D Ceph: Extraoral source dental X-ray system intended to perform panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial region and in subreqions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
Stellaris 3D: Extraoral source dental X-ray system intended to perform 3D and panoramic exams with production of diagnostic images in the dentomaxillo-facial region and in subregions, for general and pediatric dentistry.
Stellaris 3D Ceph: Extraoral source dental X-ray system intended to perform 3D, panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
## Device Description:
Stellaris 3D is a dental X-ray system for Panoramic and Cone Beam Computed Tomography (CBCT) which allows to perform all the radiographic projections, both 2D and 3D, of most interest for the dentist, the surgeon and the maxilla-facial radiologist.
In addition to the functions of Panoramic and 3D radiography, Stellaris 3D Ceph also allows to perform One-Shot Cephalometric radiographs.
Stellaris 2D is a dental X-ray system for Panoramic which allows to perform all the radiographic projections of most interest for the dentist, the surgeon and the maxilla-facial radiologist.
In addition to the functions of Panoramic, Stellaris 2D Ceph also allows to perform One-Shot Cephalometric radiographs.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for FONA. The word "FONA" is written in large, gray, sans-serif letters. Above the word "FONA" is a yellow swoosh. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray, sans-serif letters.
Except for the panoramic acquisition sensor, the remaining mechanical, electrical and software characteristics of Stellaris 2D are exactly identical to those of Stellaris 3D.
The control panel interface on the unit provides a complete control of the operation and for the setting of the desired technique factors.
Class I LASER aiming lights support positioning of patient's head, which is stabilized through the use of bite blocks, chin rest, and, if required, temple supports.
The user controls the exposure using a manual hand-switch, implementing the dead man functionality.
An Ethernet connection cable allows the FONA Stellaris devices to interface with a computer for image acquisition, processing and storage.
| | Stellaris 3D | Stellaris 3D<br>Ceph | Stellaris 2D | Stellaris 2D<br>Ceph |
|---------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------------------------------|
| X-ray<br>Programs | Pan: 9<br>programs<br>3D: 23<br>programs<br>Ceph: N/A | Pan: 9<br>programs<br>3D: 23<br>programs<br>Ceph: 4<br>programs | Pan: 9<br>programs<br>3D: N/A<br>Ceph: N/A | Pan: 9 programs<br>3D: N/A<br>Ceph:4<br>programs |
| Exposure<br>Times | Pan: max<br>14.2 s<br>3D: max 16.9<br>s<br>Ceph: N/A | Pan: max 14.2<br>s<br>3D: max 16.9 s<br>Ceph: max 4 s | Pan: max<br>14.2 s<br>3D: N/A<br>Ceph: N/A | Pan: max 14.2 s<br>3D: N/A<br>Ceph: max 4 s |
| Sensor<br>Technology | Pan/3D:<br>CMOS sensor<br>15x15 cm<br>Ceph: N/A | Pan/3D:<br>CMOS sensor<br>15x15 cm<br>Ceph: Csl flat<br>panel<br>24.4x30.7 cm | Pan: CMOS<br>sensor<br>15x0.6 cm<br>Ceph: N/A | Pan: CMOS<br>sensor 15x0.6<br>cm<br>Ceph: Csl flat<br>panel 24.4x30.7<br>cm |
| Sensor<br>Characteristics | Pan: pixel<br>size 100 μm<br>3D: voxel size<br>down to 79<br>μm<br>Ceph: N/A | Pan: pixel size<br>100 μm<br>3D: voxel size<br>down to 79 μm<br>Ceph: pixel<br>size 120<br>micron | Pan: pixel<br>size 100 μm<br>3D: N/A<br>Ceph: N/A | Pan: pixel size<br>100 μm<br>3D: N/A<br>Ceph: pixel size<br>120 micron |
| IEC<br>Classification | Class I, Type B | | | |
| Nominal line | 230 V ± 10%, 115 V ± 10%, | | | |
The following table describes the functionality of each model.
Traditional 510(k)- Stellaris 2D, Stellaris 2D Ceph, Stellari 3D & Stellaris 3D Ceph FONA S.r.
FONA S.r.l. – Socio Unico - Sede legale e operativa: Via Galilei, 11 - 20057 Assago (MI) Italia - Capitale Sociale 93.000 Euro - Codice Fiscale, Partita I.V.A.: 12983700159 - Iscrizione Registro Imprese di Milano REA:1606385 - Telefono: +39 02 45712171 - Fax: +39 02 45703385 - E-mail: info@fonaitaly.com - Web: www.fonadental.it
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for FONA, a company with the tagline "Closer to You". The word "FONA" is displayed in large, bold, gray letters. Above the word "FONA" is a curved yellow line. Below the word "FONA" is the tagline "CLOSER TO YOU" in smaller, gray letters.
| | Stellaris 3D | Stellaris 3D Ceph | Stellaris 2D | Stellaris 2D Ceph |
|----------------------------|------------------------------------------------------------------------------|-------------------|--------------|-------------------|
| voltage | | | | |
| Nominal line frequency | 50/60 Hz | | | |
| Nominal current absorption | MAX 10 A @ 230 V, MAX16 A @ 115 V | | | |
| Line fuse | T8 A 250V for 230 V version; T15 A 250V for 115 V version | | | |
| Mains Resistance | ≤ 0.8 Ohm at 230 V, ≤ 0.4 Ohm at 115 V | | | |
| Rating | 1450W | | | |
| Curve form of high voltage | High frequency multi-pulse, ripple ≤ 4% | | | |
| X-ray Tube | OPX/105 | | | |
| Tube voltage | 60 - 86 kV ± 5%, constant potential | | | |
| Tube current | 2.5 - 10 mA ± 10%, direct current (DC) | | | |
| Focal spot | 0.5 IEC 60336 | | | |
| Total filtration | > 2.5 mm Al /70 kV IEC 60522 | | | |
| Maximum duty cycle | 1/8 | | | |
| Anatomical selection | 4 Patient size levels: Small, Medium, Large, Extra Large | | | |
| kV setting | 14 positions in 2 kV steps: from 60 to 86kV | | | |
| mA setting | 7 positions according to R'10 scale: 2.5, 3.2, 4, 5, 6.3, 8, 10 mA | | | |
| Aiming lights | 3 laser planes: Median Sagittal Vertical, Canine and Frankfurt<br>Horizontal | | | |
| Laser beam Class | I | | | |
| Laser wavelength | 650 nm | | | |
| Laser Output Power | < 0.15 mW at 100 mm | | | |
FONA provides technical and maintenance support for this device to ensure proper operation and to answer any questions regarding the functioning of the device. Contact details are provided to all end users and in the user manual.
## Substantial Equivalence:
The FONA Stellaris 3D has a rotating arm which performs motor-driven rototranslatory movements that drive the tube housing assembly and the image detector around the patient head according to different orbits that follow the morphologic profile of the patient.
Traditional 510(k)- Stellaris 2D, Stellaris 2D Ceph, Stellari 3D & Stellaris 3D Ceph FONA S.r.l
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for FONA. The word "FONA" is in large, gray, sans-serif font. Above the word is a yellow swoosh. Below the word "FONA" is the text "CLOSER TO YOU" in a smaller, gray, sans-serif font.
The tube housing assembly includes an X-ray generator that generates a beam of X-ray as per the inputs selected by the user. Depending on the imaging selection i.e. 2D/panoramic, cephalometric or 3D volume reconstructions, the sensors transform the X-rays into electrical signals.
The Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices contain the same type of X-rav tube as the FONA XPan DG. FONA XPan DG Plus, FONA ART Plus and FONA ART Plus C devices. The X-ray generator in the ORTHOPHOS SL device is similar in technology as compared to the FONA Stellaris devices.
The construction of the panoramic image of the dental arch is made with a thin X-ray vertical beam from the left to the right of the patient. 3D volume reconstruction is obtained from the acquisition of two-dimensional radiographic images (15x15 cm) via the cone-beam technology, then elaborated with a reconstruction algorithm.
Stellaris 2D is equipped with a CMOS sensor that allows the acquisition of panoramic images. The reconstruction is obtained from the acquisition of twodimensional radiographic images (15x 0.6 cm) then elaborated with a reconstruction algorithm.
Stellaris 2D Ceph and Stellaris 3D Ceph, in addition, is equipped with a cephalometric arm mountable on Stellaris 2D/ Stellaris 3D respectively, which includes a one shot X-Ray sensor and a cephalostat to hold the patient in position during the examination.
Accessories like bite block, chin rest, nasal support, temporal resting bars are provided to permit correct patient positioning.
The cephalometric models come with nasion reference and side bars with ear plugs to stabilize the head during exposure.
These patient positioning items are likely to enter in contact with the patient during the normal use of the device. The use of standard hygienic protective sleeves (not included in the scope of this submission) to avoid crosscontamination among patients, operators and other persons, is mandatory.
The Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices with its associated accessories fulfil the biocompatibility requirements of ISO 10993-1:2018.
The panoramic and cephalometric exposures are obtained by standard technology that have been used in dental medicine for a very long time. The principle of operation of the Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices remains the same as the predicates. The difference is the type of sensors used which work on similar principles as the sensors found in the predicate devices.
Traditional 510(k)- Stellaris 2D, Stellaris 2D Ceph, Stellari 3D & Stellaris 3D Ceph FONA S.r.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for FONA, a company whose slogan is "Closer to You". The logo features the word "FONA" in large, bold, gray letters. Above the word "FONA" is a curved, yellow line that resembles a smile or an arc. Below the word "FONA" is the slogan "CLOSER TO YOU" in smaller, gray letters.
The ORTHOPHOS SL predicate device utilises the SIDEXIS software tool for image acquisition, administration, analysis & diagnosis and the images can be stored in the SIDEXIS database.
The Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices can be used with the OrisWin DG Suite software for image acquisition and database (not part of this submission), same as the FONA XPan DG, FONA XPan DG Plus, FONA ART Plus and FONA ART Plus C devices.
The use of this software is not mandatory. Any software that allows the viewing and management of radiographic images in DICOM format can also be used.
Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices have more volume sizes, positions and collimations than the predicate devices. This results in more possibilities to get an exposure of relevant dento maxillofacial areas.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for FONA. The word "FONA" is written in large, gray, sans-serif letters. Above the word "FONA" is a yellow swoosh. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray, sans-serif letters.
| | Subject Device | Predicate Device | Reference Device |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Stellaris 2D, Stellaris 2D Ceph,<br>Stellaris 3D and Stellaris 3D<br>Ceph | ORTHOPHOS SL 2D, ORTHOPHOS<br>SL 2D Ceph, ORTHOPHOS SL 3D and<br>ORTHOPHOS SL 3D Ceph | FONA XPan DG, FONA XPan DG Plus,<br>FONA ART Plus & FONA ART Plus C |
| Manufacturer<br>Name | FONA S.r.l | Sirona Dental Systems GmbH | FONA S.r.l |
| Device<br>Classification<br>Regulation | 21 CFR 892.1750 | 21 CFR 892.1750 | 21 CFR 872.1800 |
| Classification<br>Name | Computed Tomography X-ray<br>system | Computed tomography x-ray system | Extraoral Source X-Ray System |
| Product Code(s) | OAS | OAS | MUH |
| Regulatory Class | 2 | 2 | 2 |
| 510(k) | K213579 | K150217 | K161131 |
| Indications for Use | Stellaris 2D: Extraoral source<br>dental X-ray system intended to<br>perform panoramic exams with<br>production of diagnostic images in<br>the dento-maxillo-facial region<br>and in subregions, for general and<br>pediatric dentistry.<br><br>Stellaris 2D Ceph: Extraoral<br>source dental X-ray system<br>intended to perform panoramic<br>and cephalometric exams with<br>production of diagnostic images in<br>the dento-maxillo-facial region<br>and in subregions, for general and<br>pediatric dentistry, as well as<br>carpal images for dental clinical<br>investigation. | The X-ray system creates data for digital<br>exposures in the maxillofacial area and<br>in subareas for dentistry, including<br>pediatric dentistry, for hard-tissue<br>diagnostics within ENT medicine, and<br>for carpus exposures. | The FONA Pan/Ceph device, comprising<br>models FONA XPan DG, FONA XPan<br>DG Plus, FONA ART Plus, and FONA<br>ART Plus C, is an extraoral source<br>dental X-ray system intended to perform<br>panoramic or cephalometric exams with<br>production of diagnostic images in the<br>dento-maxillo-facial region and in<br>subregions, for general and pediatric<br>dentistry, as well as carpal images. |
| | Subject Device | Predicate Device | Reference Device |
| Device Name | Stellaris 2D, Stellaris 2D Ceph,<br>Stellaris 3D and Stellaris 3D<br>Ceph | ORTHOPHOS SL 2D, ORTHOPHOS<br>SL 2D Ceph, ORTHOPHOS SL 3D and<br>ORTHOPHOS SL 3D Ceph | FONA XPan DG, FONA XPan DG Plus,<br>FONA ART Plus & FONA ART Plus C |
| | Stellaris 3D: Extraoral source<br>dental X-ray system intended to<br>perform 3D and panoramic exams<br>with production of diagnostic<br>images in the dento-maxillo-facial<br>region and in subregions, for<br>general and pediatric dentistry. | | |
| | Stellaris 3D Ceph: Extraoral<br>source dental X-ray system<br>intended to perform 3D,<br>panoramic and cephalometric<br>exams with production of<br>diagnostic images in the dento-<br>maxillo-facial region and in<br>subregions, for general and<br>pediatric dentistry, as well as<br>carpal images for dental clinical<br>investigations. | | |
| Panoramic<br>Programs | P1: Adult Full<br>P2: Child Full<br>P3: Adult Partial Front<br>P4: Adult Partial Left<br>P5: Adult Partial Right<br>P6: Adult Dentition<br>P7: Sinuses<br>P8: Bitewing<br>P9: TMJ Open/Closed Mouth | P1: Panoramic exposure<br>P1A: Panoramic exposure, artifact-<br>reduced<br>P1C: Panoramic exposure, constant<br>1.25x magnification<br>P2: Panoramic exposure, without<br>ascending rami<br>P2A: Artifact-reduced panoramic<br>exposure without ascending rami | P1 Standard Panoramic Adult<br>P2 Panoramic Child<br>P3 Left Dentition<br>P4 Right Dentition<br>P5 Anterior Dentition<br>P6 TMJ lateral view open/closed mouth<br>P7 Frontal View of Maxillary Sinuses<br>P8 Bitewing (only FONA Art Plus /<br>FONA ART Plus C) |
| | Subject Device | Predicate Device | Reference Device |
| Device Name | Stellaris 2D, Stellaris 2D Ceph,<br>Stellaris 3D and Stellaris 3D<br>Ceph | ORTHOPHOS SL 2D, ORTHOPHOS<br>SL 2D Ceph, ORTHOPHOS SL 3D and<br>ORTHOPHOS SL 3D Ceph | FONA XPan DG, FONA XPan DG Plus,<br>FONA ART Plus & FONA ART Plus C |
| | | P2C: Panoramic exposure without<br>ascending rami at a constant<br>magnification of 1.25x | |
| | | P10: Panoramic exposure for children | |
| | | P10A: Panoramic exposure for children,<br>without ascending rami, artifact-reduced | |
| | | P10C: Panoramic exposure for children,<br>without ascending rami, constant 1.25x<br>magnification | |
| | | P12: Thick slice, anterior tooth region | |
| | | BW1: Bitewing exposures in the<br>posterior tooth region | |
| | | BW2: Bitewing exposures in the anterior<br>tooth region | |
| | | TM1.1/TM1.2: Temporomandibular joints<br>from a lateral aspect with the mouth<br>open and closed, two-part exposure | |
| | | TM3: Temporomandibular joints lateral,<br>ascending rami | |
| | | S1: Paranasal sinuses | |
| | | S3: Paranasal sinuses, linear slice<br>orientation | |
| Cephalometric<br>Programs | C1: LL (Latero-Lateral) large view<br>C2: LL (Latero-Lateral) small view<br>C3: AP (Anterior-Posterior) / PA<br>(Postero-Anterior) view<br>C4: CEPH Carpus | C1: Posterior-anterior exposure,<br>symmetrical<br>C2: Anterior-posterior exposure,<br>symmetrical<br>C3: Lateral exposure<br>C3F: Full-format exposure, lateral | FONA XPan DG Plus:<br>• P8 Antero-posterior<br>• P9 Letro-lateral<br>• P10 Carpus<br>FONA ART Plus C<br>• P9 Antero-posterior |
| | Subject Device | Predicate Device | Reference Device |
| Device Name | Stellaris 2D, Stellaris 2D Ceph,<br>Stellaris 3D and Stellaris 3D<br>Ceph | ORTHOPHOS SL 2D, ORTHOPHOS<br>SL 2D Ceph, ORTHOPHOS SL 3D and<br>ORTHOPHOS SL 3D Ceph | FONA XPan DG, FONA XPan DG Plus<br>FONA ART Plus & FONA ART Plus C |
| | | C4: Carpus view, symmetrical | P10 Latero-lateral P11 Carpus |
| 3D<br>Reconstructions<br>(for 3D model only) | V1: Adult Full View | VOL1 (HS, SD, Low): Volume exposure<br>with a diameter of approx. 8 cm and a<br>height of approx. 8 cm or 5.5 cm<br>collimated. | Not Available |
| | V2: Adult Upper Jaw – Surgical/<br>Implant | VOL2 (HS, SD, Low): Volume exposure<br>with a diameter of about 5 cm and a<br>height of about 5.5 cm for upper or lower<br>mandible. | |
| | V3: Adult Lower Jaw - Surgical/<br>Implant | VOL3 (HS, SD, Low) (optional): Volume<br>exposure with a diameter of about 11<br>cm and a height of about 10 cm or<br>selection of upper quadrant collimated to<br>7.5 cm and selection lower quadrant<br>collimated to8.0 cm | |
| | V4: Adult Sinuses | | |
| | V5: Child Full | | |
| | V6: Child Upper Jaw - Surgical/<br>Implant | | |
| | V7: Child Lower Jaw - Surgical/<br>Implant | | |
| | V8: Child Sinuses | | |
| | V9: Upper+Lower Jaws -<br>Surgical/Implant Front | | |
| | V10: Upper+Lower Jaws -<br>Surgical/Implant Left | | |
| | V11: Upper+Lower Jaws -<br>Sur…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.