The NightOwl is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adult patients suspected of sleep apnea. The device is intended for the clinical and home setting use under the direction of a Healthcare Professional (HCP).
Device Story
NightOwl is a wearable home sleep apnea test (HSAT) device. It consists of a fingertip sensor and cloud-based analysis software. The sensor uses red/infrared LEDs for photoplethysmography (PPG) and an accelerometer for movement detection. Data is recorded overnight, stored on-board or streamed via Bluetooth to a smartphone app, then uploaded to a cloud-based signal processing suite. The software analyzes the PPG and actigraphy signals to calculate sleep-related breathing parameters, including pAHI (pediatric/peripheral apnea-hypopnea index) using either AASM 1A or 1B scoring rules. Results are presented in a dashboard for review by a prescribing healthcare professional to assist in diagnosing sleep-related breathing disorders. The device is intended for use in home or clinical settings under HCP direction.
Clinical Evidence
Clinical validation compared the modified device (1B rule) against gold-standard polysomnography (PSG). Performance metrics at AHI cutoffs of 5, 15, and 30 were reported. At AHI cutoff 5: sensitivity 1.000, specificity 0.823. At AHI cutoff 15: sensitivity 0.973, specificity 0.886. At AHI cutoff 30: sensitivity 0.840, specificity 0.979. The pAHI correlation was 0.946.
Technological Characteristics
Wearable fingertip sensor with red/IR LEDs and accelerometer. Biocompatible enclosure and adhesive. Battery-powered (coin cell). Connectivity via Bluetooth to smartphone or offline storage. Cloud-based signal processing. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (environmental), and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for adult patients (22+ years) suspected of sleep apnea to aid in the evaluation of sleep-related breathing disorders.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 16, 2021
Ectosense nv Bart Van Pee Product Manager Bosbessenlaan 19 A Rotselaar, Vlaams Brabant 3110 Belgium
Re: K213463
Trade/Device Name: NightOwl Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: October 26, 2021 Received: October 27, 2021
Dear Bart Van Pee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213463
Device Name NightOwl
Indications for Use (Describe)
The NightOwl is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adult patients suspected of sleep apnea.
The device is intended for the clinical and home setting use under the direction of a Healthcare Professional (HCP).
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) SUMMARY
#### NightOwl
#### 1. SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON AND DATE PREPARED
Ectosense nv Bosbessenlaan 19A Rotselaar Vlaams-Brabant 3110 Belgium Phone: +32 496 74 46 12
Contact Person: Bart Van Pee
Date Prepared: 15th of December 2021
#### 2. DEVICE
#### Name of Device
NightOwl
#### Trade Name
NightOwl
#### Common or Usual Name
Ventilatory Effort Recorder
#### Classification Name/Product Code/CFR Reference
Class II, Ventilatory Effort Recorder, Product Code: MNR; 21 CFR 068.2375
{5}------------------------------------------------
#### 3. PREDICATE DEVICE
Predicate: Ectosense's NightOwl (K191031)
#### 4. DEVICE DESCRIPTION
The NightOwl is prescribed by a Health Care Professional for the patient to use in the home as a 'home sleep apnea test' (HSAT).
The NightOwl comprises a sensor that is worn on the fingertip (the "NightOwl Sensor") over-night whilst the subject is sleeping and cloud-based analysis software (the "NightOwl Software").
The NightOwl Sensor has a small biocompatible enclosure. The sensor has 2 LEDs, one in the red spectrum and the other in the infrared spectrum, and an accelerometer. The sensor is attached to the fingertip by single-use biocompatible adhesive tape, with the sensor window applied against the fingerprint area of the fingertip. The sensor measures the reflected red/infrared signals to record the photoplethysmograph (PPG) signal. The accelerometer is used to detect movement.
The data recorded by the NightOwl Sensor can either be stored in on-board memory ("Offline" mode) or streamed via a Bluetooth link to an Ectosense app on a smartphone ("Streaming" mode)
- if the data is stored on the device, the data is retrieved when the NightOwl sensor . is returned to the prescribing HCP and passed up to a cloud-based signal processing suite, the NightOwl Software.
- If the device is used in Streaming mode, the data is stored by the Ectosense app . on the smartphone during the recording. At the end of the recording, it is then passed directly up to the cloud-based signal processing suite.
- The NightOwl Software signal processing algorithms produce a number of sleep . and sleep-disordered breathing related traces and parameters. The trace and parameter information are passed to a company-managed database for storage and access by the prescribing Health Care Professional in the Ectosense Dashboard.
{6}------------------------------------------------
#### 5. INTENDED USE / INDICATIONS FOR USE
The NightOwl is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adult patients suspected of sleep apnea. The device is intended for the clinical and home setting use under the direction of a Healthcare Professional (HCP).
# 6. DEVICE MODIFICATIONS AND COMPARISON OF TECHNOLOGICAL
#### CHARACTERSITICS WITH THE PREDICATE DEVICE
We have now incorporated an alternative tuning of the NightOwl algorithm to include hypopnea as scored using the 1B rule. The HCP can now choose to display either the 1A-tuned or the 1B-tuned pAHI (or both) on the NightOwl Examination report. As per the American Academy of Sleep medicine (AASM):
1A rule: a hypopnea is scored when:
- 1. There is an airflow reduction of 30% for at least 10 seconds
- 2. There is an oxygen desaturation of at least 3 percent or an arousal.
1B rule: a hypopnea is scored when:
- 1. There is an airflow reduction of 30% for at least 10 seconds
- 2. There is an oxygen desaturation of at least 4 percent
| Characteristic | NightOwl (K191031) | NightOwl (K213463) | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|------------|
| Intended Use/<br>Indications for<br>Use | The NightOwl is a wearable device intended for use in<br>the recording, analysis, displaying, exporting, and<br>storage of biophysical parameters to aid in the<br>evaluation of sleep-related breathing disorders of adult<br>patients suspected of sleep apnea. The device is<br>intended for the clinical and home setting use under the<br>direction of a Healthcare Professional (HCP). | | Equivalent |
| Intended<br>Environment | Recording in the home environment with the report<br>interpretation performed in the clinical setting. | | Equivalent |
| Prescription | Prescription only | | Equivalent |
{7}------------------------------------------------
| Target<br>Population | 22 years old and older | | Equivalent |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Channels | 1. PAT<br>2. Pulse rate<br>3. Oximetry<br>4. Actigraphy | | Equivalent |
| Sensors | Optical plethysmography sensor, accelerometer | | Equivalent |
| Wearable sensor<br>location | The photoplethysmography (PPG) sensor and accelerometer<br>components are worn on the fingertip. | | Equivalent |
| Sensor Software | Firmware is limited to control the recording and<br>communications processes. No presentation of test results<br>to the patient. Data analyzed and presented in a separate<br>software suite. | | Equivalent |
| Analysis Software<br>- location | Analysis performed off the recording device, exclusively<br>cloud-based by the NightOwl software. | | Equivalent |
| Analysis Software<br>- algorithm | pAHI calculation tuned to<br>the AASM's '1A Rule' for<br>the scoring of hypopnea | pAHI calculation tuned to<br>the AASM's '1A Rule' for<br>the scoring of hypopnea<br>AND<br>pAHI calculation tuned to<br>the AASM's '1B Rule' for<br>the scoring of hypopnea | Substantially equivalent<br>Addition of pAHI<br>calculation tuned to the<br>AASM's '1B Rule' for the<br>scoring of hypopnea<br>The addition of the pAHI<br>calculation tuned to the<br>AASM's '1B Rule' for the<br>scoring of hypopnea<br>does not alter the<br>device's intended use<br>and does not introduce<br>any change to the safety<br>and effectiveness of the<br>originally cleared device<br>(predicate device). As<br>such, the algorithm of<br>the altered device<br>remains substantially<br>equivalent to that of the<br>originally cleared device<br>(predicate device). |
| Data transfer | Data transfer through a smartphone by wireless connection. | | Equivalent |
| Power Source<br>recorder | Battery powered by coin cell | | Equivalent |
| Patient<br>isolation | Device has no galvanic connections to mains as it is a<br>battery-operated device. | | Equivalent |
| Sterilization | Non-sterile | | Equivalent |
| Bio-<br>compatibility | Assessed to ISO1099-1:2009 requirements for<br>sensitization, irritation, and cytotoxicity | | Equivalent |
| EMC | IEC 60601-1-2:2014 | | Equivalent |
| Electrical<br>Safety | IEC 60601-1:2005 +AMD1:2012 | Equivalent | |
| Environmental<br>Testing | IEC 60601-1-11:2010 | Equivalent | |
{8}------------------------------------------------
# 7. PERFORMANCE DATA
Software validation: In support of the substantial equivalence determination of the modification made to the device, software validation was performed.
Clinical Study: The clinical validation of the NightOwl performance using 1B rule was compared to the gold standard polysomnography (PSG) in the US population. The following table compares the performance of the subject and predicate devices.
| Endpoint Parameters | 1A rule performance of<br>originally cleared device<br>(Predicate K191031) | 1B rule performance of<br>modified device<br>(Subject) |
|------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------|
| Sensitivity at AHI cutoff 5 | 0.936 | 1.000 |
| Specificity at AHI cutoff 5 | 0.727 | 0.823 |
| Sensitivity at AHI cutoff 15 | 0.978 | 0.973 |
| Specificity at AHI cutoff 15 | 0.704 | 0.886 |
| Sensitivity at AHI cutoff 30 | 0.964 | 0.840 |
| Specificity at AHI cutoff 30 | 0.844 | 0.979 |
| pAHI correlation | 0.909 | 0.946 |
## 8. CONCLUSIONS
Based on (i) the results of the clinical validation testing, (ii) the results of the software validation testing, and (iii) the unchanged product risk analysis, the NightOwl has been demonstrated to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.