The EMBOLD™ Fibered Detachable Coil system is intended to obstruct or reduce rate of blood flow in the peripheral vasculature. This device is not intended for coronary or neurovascular use.
Device Story
EMBOLD Fibered Detachable Coil System consists of delivery wire, introducer sheath, and permanent implantable fibered embolic coil; compatible with 0.021-0.027 inch microcatheters. Device used in hospital setting by physicians for peripheral vascular embolization. User advances coil through microcatheter to target site; coil can be fully advanced, retracted, and repositioned prior to deployment. Mechanical detachment mechanism uses inner pull wire connected to coupler arms; user activates detachment by retracting pull wire once coil is outside microcatheter. Permanent coil remains in vessel to obstruct blood flow; synthetic fibers promote thrombosis. Benefits include controlled placement and effective vessel occlusion. Device provided sterile via ethylene oxide.
Clinical Evidence
Bench testing and acute GLP animal study performed. Bench testing included biocompatibility and mechanical performance verification. Animal study confirmed safety and effectiveness in vivo. No clinical data provided.
Technological Characteristics
Materials: Platinum/tungsten alloy coil, metal alloy coupler, synthetic fibers. Mechanical detachment via pull wire. Compatible with 0.021-0.027 inch microcatheters. Sterilization: 100% ethylene oxide. MR Conditional. Single-use.
Indications for Use
Indicated for obstruction or reduction of blood flow in peripheral vasculature. Not for coronary or neurovascular use.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Interlock™ Fibered IDC™ Occlusion System (K132578)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2022
Boston Scientific Corporation Heidi Shearer Senior Regulatory Affairs Specialist One Scimed Place Maple Grove, Minnesota 55311
Re: K213398
Trade/Device Name: EMBOLD Fibered Detachable Coil System Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: March 11, 2022 Received: March 14, 2022
Dear Heidi Shearer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Finn Donaldson Assistant Director (Acting) DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213398
Device Name EMBOLD™ Fibered Detachable Coil System
Indications for Use (Describe)
The EMBOLD™ Fibered Detachable Coil system is intended to obstruct or reduce rate of blood flow in the peripheral vasculature.
This device is not intended for coronary or neurovascular use.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:100%;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="1" width="15" x="0" y="0"></rect> <line stroke="black" stroke-width="2" x1="2" x2="13" y1="2" y2="13"></line> <line stroke="black" stroke-width="2" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:100%;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="1" width="15" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
510(K) Summary Complying with 21 CFR 807.92
#### I. SUBMITTER INFORMATION
Submitter Name: Boston Scientific Corporation
### Submitter Address:
One Scimed Place Maple Grove, MN 55311-1566 USA
Telephone: 763-255-0056 Fax: 763-494-2222 e-mail: Heidi.Shearer@bsci.com
Contact person name: Heidi Shearer Date Prepared: April 12, 2022
### II. DEVICE INFORMATION
Proprietary Name: EMBOLD™ Fibered Detachable Coil System Tables 1 and 2 below summarize the relevant subject device information.
#### Table 1: EMBOLD™ Fibered Detachable Coil System UPNs
| Universal Product Numbers<br>(UPNs) | Coil Secondary<br>Wind OD (mm) | Coil Length<br>(cm) |
|-------------------------------------|--------------------------------|---------------------|
| M001394240020040 | 2 | 4 |
| M001394240020060 | 2 | 6 |
| M001394240030060 | 3 | 6 |
| M001394240030120 | 3 | 12 |
| M001394240040080 | 4 | 8 |
| M001394240040150 | 4 | 15 |
| M001394240050080 | 5 | 8 |
| M001394240050150 | 5 | 15 |
| M001394240060100 | 6 | 10 |
| M001394240060200 | 6 | 20 |
| M001394240080200 | 8 | 20 |
| M001394240100200 | 10 | 20 |
| M001394240100300 | 10 | 30 |
| M001394240100500 | 10 | 50 |
| M001394240120200 | 12 | 20 |
| M001394240120300 | 12 | 30 |
| M001394240120500 | 12 | 50 |
| M001394240140200 | 14 | 20 |
| M001394240140300 | 14 | 30 |
| M001394240140500 | 14 | 50 |
| M001394240180500 | 18 | 50 |
| M001394240200500 | 20 | 50 |
| M001394240220600 | 22 | 60 |
| M001394240320600 | 32 | 60 |
{4}------------------------------------------------
### Table 2: Additional Device Information
| Common or<br>Usual name | Regulatory<br>Number | Regulatory<br>Name | Product Code | Product Code<br>Name | Regulatory<br>Class |
|------------------------------------|-------------------------|------------------------------------|--------------|-------------------------------------------------------|---------------------|
| Vascular<br>embolic coil<br>system | 21 CFR Part<br>870.3300 | Vascular<br>Embolization<br>Device | KRD | Device,<br>Vascular, For<br>Promoting<br>Embolization | II |
# III. PREDICATE DEVICE IDENTIFICATION
### Name of Predicate Device
Interlock™ Fibered IDC™ Occlusion System (K132578, Cleared September 13, 2013) Predicate device referenced above has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
The EMBOLD™ Fibered Detachable Coil System consists of a delivery wire, removable introducer sheath, and permanent implantable fibered embolic coil. The device is compatible with 0.021 inch to 0.027 inch microcatheters. In addition, the device features a modified mechanical detachment mechanism consisting of an inner pull wire connecting the coil and delivery wire coupler arms, activated by the user through a proximal laser etched outer wire perforation. The delivery wire design allows the coil to be fully advanced, retracted, and deployed prior to final placement.
The EMBOLD™ Fibered Detachable Coil System is provided sterile, using 100% ethylene oxide (EO) gas sterilization, and is intended for hospital and single use only.
Table 3 below provides information on the differences between the EMBOLD™ Fibered Detachable Coil System models, in relation to specific features such as coil secondary OD, length, and shape. The delivery wire and introducer sheath utilize the same design throughout the product matrix.
| Coil Secondary Wind OD (mm) | Coil Length (cm) | Coil Shape |
|-----------------------------|------------------|-----------------|
| 2 | 4 | Helical |
| 2 | 6 | |
| 3 | 6 | 2-Diameter (2D) |
| 3 | 12 | |
| 4 | 8 | |
| 4 | 15 | |
| 5 | 8 | |
| 5 | 15 | |
| 6 | 10 | |
| 6 | 20 | |
| 8 | 20 | |
| 10 | 20 | |
| 10 | 30 | |
| 10 | 50 | |
| 12 | 20 | |
| 12 | 30 | |
| 12 | 50 | |
| 14 | 20 | |
| 14 | 30 | |
| 14 | 50 | |
| 18 | 50 | |
| 20 | 50 | |
### Table 3: EMBOLD™ Fibered Detachable Coil System Device Features
{5}------------------------------------------------
| Coil Secondary Wind OD (mm) | Coil Length (cm) | Coil Shape |
|-----------------------------|------------------|------------|
| 22 | 60 | |
| 32 | 60 | |
# V. INTENDED USE AND INDICATION FOR USE
Predicate and subject device intended use/indication for use is the same.
# EMBOLD™ Fibered Detachable Coil System Intended Use / Indication for Use
The EMBOLD™ Fibered Detachable Coil system is intended to obstruct or reduce rate of blood flow in the peripheral vasculature.
This device is not intended for coronary or neurovascular use.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The EMBOLD™ Fibered Detachable Coil System has similar technological characteristics to the predicate device as shown in the table below.
| Characteristic | Subject Device<br>EMBOLD™ Fibered<br>Detachable Coil System | Predicate Device<br>Interlock™ Fibered IDC ™<br>Occlusion System<br>(K132578) | Comparison |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Coil Primary<br>Wind OD<br>(inches) | 0.015 | 0.012 | SIMILAR<br>Minimal dimensional<br>differences support<br>additional microcatheter |
| Microcatheter<br>Compatibility<br>(inches) | 0.021 (0.53mm) to 0.027<br>(0.6mm) | 0.0021 (0.53mm) | compatibility. |
| Coil Lengths<br>(cm) | 4-60 | 4-60 | SAME |
| Secondary<br>Coil OD (mm) | 2-32 | 2-22 | SIMILAR<br>Additional OD treats<br>same anatomy as<br>predicate. |
| Coil Shapes | 2mm: Helical<br>3-32mm: 2D | 2-22mm: 2D | SIMILAR<br>Helical loops same as<br>2D, with no smaller<br>loops. |
| Delivery Wire<br>Length | 177cm effective length | 175cm effective length | SIMILAR<br>Subject device delivered<br>through same standard<br>microcatheter lengths as<br>predicate. |
| Coil<br>Detachment<br>Mechanism | Mechanical, user activated,<br>interlocking coupler arms<br>detach when outside of<br>microcatheter and when user<br>manually retracts inner pull<br>wire from coupler arms | Mechanical, user activated,<br>interlocking coupler arms<br>detach when pushed out of<br>microcatheter | SIMILAR<br>Same mechanical<br>detachment with coupler<br>arm separation, with<br>additional feature. |
| Introducer<br>Sheath | 1 length<br>Reinforced tapered tip<br>Separate lock component | 2 lengths<br>Tapered tip<br>Integrated lock | SIMILAR<br>Same function as<br>predicate with<br>modifications to align<br>with EMBOLD Fibered |
{6}------------------------------------------------
| Characteristic | Subject Device<br>EMBOLDTM Fibered<br>Detachable Coil System | Predicate Device<br>InterlockTM Fibered IDCTM<br>Occlusion System<br>(K132578) | Comparison |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MRI<br>Compatibility | MR Conditional | MR Conditional | SAME |
| Sterilization<br>Method | EO | EO | SAME |
| SAL | 10-6 | 10-6 | SAME |
| Coil Materials | Platinum /Tungsten alloy coil<br>Metal alloy coil coupler arm<br>Synthetic fibers | Platinum /Tungsten alloy coil<br>Metal alloy coil coupler arm<br>Synthetic fibers | SAME |
| Delivery Wire<br>Materials | Metal and polymer materials<br>Metal alloy delivery wire<br>coupler arm | Metal and polymer materials<br>Metal alloy delivery wire<br>coupler arm | SIMILAR<br>Basic delivery function is<br>the same as the<br>predicate with material<br>differences to facilitate<br>detachment and<br>microcatheter<br>compatibility. |
| Packaging<br>Materials and<br>Configuration | Package holds one EMBOLD<br>Fibered device. No RHV is<br>included. Sterile package<br>contained in shelf carton.<br><br>An eIFU exists for the<br>EMBOLD Fibered product.<br>No physical IFU included<br>with package. | Package holds one Interlock<br>Fibered IDC device. RHV is<br>included inside sterile<br>barrier. Sterile package and<br>paper IFU contained in shelf<br>carton. | SIMILAR<br>The RHV is no longer<br>provided with the device<br>as available through<br>ancillary devices or<br>hospital. The IFU is<br>available outside the<br>package configuration. |
## VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
A full suite of biocompatibility testing, bench testing, and an acute GLP animal study were performed to support a determination of substantial equivalence between the EMBOLD™ Fibered Detachable Coil System and the predicate Interlock™ Fibered IDC™ Occlusion System (K132578). The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or effectiveness issues were raised during the device testing.
### VIII. CONCLUSION
Based on the intended use, technological characteristics, and the safety and performance data provided, the EMBOLD™ Fibered Detachable Coil System is considered appropriate for the intended use and substantially equivalent to the predicate device technological and material differences do not raise new questions of safety or effectiveness, and the subject devices are substantially equivalent to the predicate Interlock™ Fibered IDC™ Occlusion System (K132578).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.