The electrical-driven SURGmatic S15 L Pro handpiece is a dental handpiece in accordance with 21 CFR § 872.4200 (dental handpieces and accessories) designed for use by a trained professional in the field of general dentistry only.
Device Story
Electrically driven, reusable dental handpiece; fiber optic light system; connects to treatment unit via ISO 3964 coupling; powered by electrical motor. Converts electrical energy to mechanical rotary movement; force transmitted via shafts/gears to head housing; accepts ISO 1797-1 burs via push-button collet. Used by dentists/oral surgeons in clinical settings for surgical and restorative procedures. Provides illumination and mechanical power for cutting/processing tooth structure. Sterilizable via steam autoclave. Benefits patient through precise, illuminated, and efficient surgical/restorative dental intervention.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and adherence to international standards, including ISO 14457 (handpieces), ISO 10993 (biocompatibility), ISO 17665 (sterilization), and ISO 3964 (couplings).
Technological Characteristics
Stainless steel construction; includes pad printing ink and glass rod fiber. Fiber optic illumination (~25,000 Lux). ISO 3964 coupling; ISO 1797-1 bur compatibility. Push-button chuck. Steam sterilizable. Electric motor-driven rotary mechanism. Gear ratio 1:5. No software/algorithm.
Indications for Use
Indicated for oral, jaw and facial surgery, tooth extraction, root resection, osteotomy, removal of carious material, tooth/cavity/crown preparations, and processing of fillings. For use by trained professionals in general dentistry and/or oral surgery.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
December 21, 2021
KaVo Dental GmbH Kerstin Henseler Regulatory Affairs Manager Bismarckring 39 Biberach, 88400 GERMANY
Re: K213139
Trade/Device Name: SURGmatic S15 L Pro Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EGS Dated: September 24, 2021 Received: September 27, 2021
Dear Kerstin Henseler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K213139
Device Name SURGmatic S15 L Pro
Indications for Use (Describe)
The SURGmatic S15 L Pro handpiece is intended for:
- Oral, jaw and facial surgery
- Tooth extraction procedures
- Root resection
- Osteotomy
- Removal of carious material
- Tooth, cavity and crown preparations
- Processing of fillings
The SURGmatic S15 L Pro is for use by a trained professional in the field of general dentistry and/or oral surgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Kavo Dental Excellence. The logo consists of the word "KAVO" in a bold, sans-serif font, with the "A" stylized to resemble a dental tool. Below the word "KAVO" is the phrase "Dental Excellence" in a smaller, lighter font. The logo is simple and modern, and it conveys a sense of professionalism and quality.
# 510(k) Summary K213139
### SURGmatic S15 L Pro handpiece
- 1. Submitter Information:
KaVo Dental GmbH Bismarckring 39 88400 Biberach, Germany
Contact Person: E-Mail: Telephone Number: Kerstin Henseler kerstin.henseler@kavo.com (+49) 7351-56-2431
Date Prepared: September 24, 2021
- 2. Device Name:
| • Proprietary Name: | SURGmatic S15 L Pro |
|---------------------|---------------------|
|---------------------|---------------------|
- . KaVo Dental GmbH Manufacturer:
- Dental handpiece, Contra-Angle Attachment Common Name: ●
- o Classification Name: Dental Handpiece and Accessories
- CFR Number: 21 CFR §872.4200 ●
- . Device Class: 1
- Product Code: EGS ●
#### 3. Predicate Device:
Predicate Device (Primary):
- Proprietary Name: SURGmatic (K140308) o
- Manufacturer: KaVo Dental GmbH O
- 0 Common Name: Dental Handpiece, Contra-Angle Attachment
- Classification Name: Dental Handpiece and Accessories O
- 872.4200 CFR Number:
- Device Class: I O
- Product Code: EBW, EGS 0
# Reference Device:
O
- O Proprietary Name: SUPERtorque, GENTLEpower Handpieces (K073478)
- Manufacturer: KaVo Dental GmbH O
- Common Name: Dental Handpiece, Air-Powered O
- Classification Name: Dental Handpiece and Accessories O
- CFR Number: 872.4200 O
- O Device Class: I
- EFB Product Code: O
{5}------------------------------------------------
#### 4. Description of Device:
The electrical-driven SURGmatic S15 L Pro handpieces in accordance with 21 CFR § 872.4200 (dental handpieces and accessories) designed for use by a trained professional in the field of general dentistry only. The device is an electrically driven handpiece that is reusable and provided with a fiber optic light system. The electrical motor includes the light source and the SURGmatic S15 L Pro handbiece is provided with a fiber optic light transmitter to transport the light to the bur side. DIN EN ISO 14457 requlates the minimum illuminance. The dental handpieces can be sterilized in a steam sterilizer (autoclave). SURGmatic handpieces equipped with a handpiece connector in accordance with ISO 3964 are connected to a treatment unit by means of a hose and the electrical motor and receive the energy for the gear, cooling water and air for conservative dental treatment as well as the light for illumination of the operation area through corresponding output openings. Burs in accordance with ISO 1797-1 must be used with the SURGmatic handpieces. Based on the INTRAmatic connection that meets the ISO 3964 standard, the SURGmatic handpieces fit with any electrical dental motor which is produced in accordance to this standard. The SURGmatic handpieces interact with the patient through a rotating bur with the patient teeth according to the intended use.
The electrical motor rotates with an idle speed of 40.000 1/min and the gear ratio of the SURGmatic S15 L Pro instrument amounts to 1:5 (exactly 1:4,71).
The bur is held in the rotating head drive with the help of a collet and can be released by a push-button mechanism. DIN EN ISO 14457 requlates the minimum extraction force of the bur.
### Principle of Operation / Mechanism of Action:
Rotatory movement. Surgical instruments for the mouth, jaw and facial surgery are straight and contra-angle handpieces which are powered by electrical surgical motors. Electrical energy is converted into mechanical energy through the motor. This mechanical energy is transferred to the drive trains in the contra-angle handpiece in the form of a rotary movement via a coupling interface in accordance with DIN EN ISO 3964. The force is transmitted through shafts and gear components to another coupling interface located in the head housing, which is designed for safe insertion and removal of tools in accordance with DIN EN ISO 1797. The mechanical energy is passed on to the cutting part of these tools and thus to the treatment site by this interface.
#### 5. Indications for Use:
The SURGmatic S15 L Pro handpiece is intended for:
- Oral, jaw and facial surgery
- Tooth extraction procedures
- Root resection
- Osteotomy
- Removal of carious material
- Tooth, cavity and crown preparations
- Processing of fillings
The SURGmatic S15 L Pro is for use by a trained professional in the field of general dentistry and/or oral surgery.
{6}------------------------------------------------
#### 6. Description of Substantial Equivalence:
# Details of the similarities between subject and predicate device:
The similarities between the subject and predicate device is mainly the indications for use and the intended use with the only difference being choice of similar wording. It can be said that the handpieces are for use in the field of general dentistry and/or oral surgery. Furthermore, equally similar is the drive by a motor, which drives trains in the contra-angle handpiece in the form of a rotary movement via a coupling interface in accordance with DIN EN ISO 3964. The force is transmitted through shafts and qear components to another coupling interface located in the head housing. Technical Specifications like light illumination, type of chuck, power source, dental bur, speed range, light intensity are the same. To maintain the handpiece the same automated handpiece lubricant system and lubricant are used. The SURGmatic S15 L Pro handpiece has the same type of chuck as the both predicate devices which is push button.
Furthermore, the proposed device and the predicate devices complies with performance standards ISO 14457; Dentistry - Handpieces and motors as well as Biocompatible according to ISO 10993-1 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management system).
# Details of the differences between subject and predicate device:
There are no major differences however there are a few minor differences such as the course of air and water ports, direct and indirect patient-contacting materials and dimensions. The SURGmatic S15 L Pro handpiece is available with External spray which is transferred internally into the spray insert. The other handpieces have an external spray instead. For all handpieces the material stainless steel is used for nearly all direct and indirect patient-contacting parts with the only difference by identifying now also the pad printing ink & the qlass rod fiber as a direct / indirect patient-contacting part. The dimensions differ in a few millimeters.
# Conclusion:
Based on a comparison of intended use, indications for use, technological characteristics, principle of operation, features, and performance data, the SURGmatic S15 L Pro handbiece is deemed to be substantially equivalent to the predicate devices as it satisfies all criteria of substantial equivalence and does not raise new concerns regarding substantial equivalence: (1) Indications for Use, (2) Technological Characteristics, and (3) Performance Data. The new device does not introduce a fundamentally new scientific technology and the nonclinical tests demonstrate that the device is substantial equivalent.
# Device Comparison Table:
### Table 5.1
| Descriptive<br>Information | Subject Device<br>SURGmatic S15 L Pro | Primary Predicate<br>Device<br>SURGmatic (K140308) | Reference Device<br>SUPERtorque,<br>GENTLEpower<br>Handpieces and<br>Attachement<br>(K073478) | Comparison |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | KaVo Dental GmbH | KaVo Dental GmbH | KaVo Dental GmbH | Same |
| Pictorial<br>Representation | | | | N/A |
| Regulatory Classification | | | | |
| Regulation # | 21 CFR §872.4200 | 21 CFR §872.4200 | 21 CFR §872.4200 | Same |
| Regulation Title | Dental Handpiece and<br>Accessories | Dental Handpiece and<br>Accessories | Dental Handpiece and<br>Accessories | Same |
| Regulation Class | Class I | Class I | Class I | Same |
| Product Code | EGS | EBW / EGS | EFB | Similar to<br>Primary<br>Predicate |
| Indications for<br>Use | Indications for Use / Intended Use | | | |
| | The SURGmatic S15 L Pro<br>handpiece is intended for:<br>- Oral, jaw and facial surgery<br>- Tooth extraction procedures<br>- Root resection<br>- Osteotomy<br>- Removal of carious material<br> | The medical device is<br>intended for the following<br>uses: Surgery such as<br>setting an implant, bone<br>augmentation, sinus lift,<br>tooth extraction<br>procedures, implantology,<br>osteotomia, root<br>resection and oral, jaw<br>and facial surgery. | The GENTLEpower<br>Handpieces and<br>Attachments are<br>intended for the removal<br>of carious material,<br>excess filling material,<br>reducing hard tooth<br>structure, cavity and<br>crown preparation, root<br>canal preparation,<br>finishing tooth<br>preparations,<br>restorations and for<br>polishing teeth. They are<br>designed for use by a<br>trained professional in<br>the field of general<br>dentistry. | Same as<br>Primary<br>Predicate and<br>Reference<br>Device with<br>the only<br>difference<br>being choice<br>of similar<br>wording. |
| Intended Use | The electrical-driven<br>SURGmatic S15 L Pro<br>handpiece is a dental<br>handpiece in accordance with<br>21 CFR § 872.4200 (dental<br>handpieces and accessories)<br>designed for use by a trained<br>professional in the field of<br>general dentistry only. | The electrical-driven<br>SURGmatic handpieces<br>are dental handpieces in<br>accordance with 21 CFR §<br>872.4200 (dental<br>handpieces and<br>accessories) and is only<br>intended for dental<br>treatment. | The GENTLEpower<br>Handpieces and<br>Attachments are dental<br>instruments for use by a<br>trained professional in<br>general dentistry. The<br>handpieces and<br>attachments are<br>powered by either an<br>air-motor or electric<br>motor. | Same |
| Device Design | | | | |
| Principles of<br>Operation | Electrical energy is converted<br>into mechanical energy<br>through the motor. This<br>mechanical energy is<br>transferred to the drive trains<br>in the contra-angle handpiece<br>in the form of a rotary<br>movement via a coupling<br>interface in accordance with<br>DIN EN ISO 3964. The force is<br>transmitted through shafts and<br>gear components to another | Through the micro motor<br>connected to the dental<br>treatment unit the<br>contra-angle handpieces<br>equipped with a<br>handpiece connection<br>according to ISO 3964<br>(Dentistry - Coupling<br>dimensions for handpiece<br>connectors - ISO<br>3964:1982) receives the<br>energy, the cooling water<br>and air for treatment and | Through the micro<br>motor connected to the<br>dental treatment unit<br>the contra-angle<br>handpieces equipped<br>with a handpiece<br>connection according to<br>ISO 3964 (Dentistry -<br>Coupling dimensions for<br>handpiece connectors -<br>ISO 3964:1982)<br>receives the energy, the<br>cooling water and air for | Similar to<br>Primary<br>Predicate and<br>Reference<br>Device |
| | the head housing, which is<br>designed for safe insertion and<br>removal of tools in accordance<br>with DIN EN ISO 1797. The<br>mechanical energy is passed<br>on to the cutting part of these<br>tools and thus to the treatment<br>site by this interface. | the light for illumination<br>the operating area. | treatment and the light<br>for illumination the<br>operating area. | |
| Air / water ports | External spray which is<br>transferred internally into the<br>spray insert | External Spray | Internal Spray | Similar to<br>Primary<br>Predicate and<br>Reference<br>Device |
| Composition of Materials | | | | |
| Direct / Indirect<br>patient-<br>contacting<br>materials | Stainless Steel<br>Pad printing ink<br>Glass Rod Fiber | Stainless Steel | Stainless Steel | Similar to<br>Primary<br>Predicate and<br>Reference<br>Device |
| Technical Specifications | | | | |
| Light<br>illumination | Fibreoptic | Fibreoptic | Fibreoptic | Same |
| Dimension | Headsize-High: 14.7mm<br>Headsize-Diameter: 10.2mm<br>Handpiece length: 97.5mm | Headsize-High: 13.8mm<br>Headsize-Diameter: 9.8mm<br>Handpiece length: 96.5mm | Headsize-Height: 24mm<br>Headsize-Diameter:<br>10.5mm<br>Handpiece length: 80-100mm | Similar to<br>Primary<br>Predicate and<br>Reference<br>Device |
| Type of chuck | Push Button | Push Button, lever chuck | Push Button | Same as<br>Reference<br>Device |
| Power Source | Electric Motor | Electric Motor | Electric Motor | Same |
| Dental Bur | Burs in accordance with ISO<br>1797-1 can be used | Burs in accordance with<br>ISO 1797-1 can be used | Burs in accordance with<br>ISO 1797-1 can be used | Same |
| Speed Range | Up to 40,000 rpm | Up to 40,000 rpm | High-speed: Approx.<br>200,000 rpm<br>Low-speed: 100 -<br>40,000 rpm | Same as<br>Primary<br>Predicate |
| Light Intensity | Approx. 25,000 Lux | Approx. 25,000 Lux | Approx. 25,000 Lux | Same |
| Lubricant | | | | Same |
| Automated<br>handpiece<br>Lubricant system | KaVo QUATTROcare (K071288) | KaVo QUATTROcare<br>(K071288) | KaVo QUATTROcare<br>(K071288) | Same |
| | KaVo Spray America | KaVo Spray America | KaVo Spray America | Same |
| Lubricant | QUATTROcare Plus Spray<br>America | QUATTROcare Plus Spray<br>America | QUATTROcare Plus<br>Spray America | |
| | (K071288) | (K071288) | (K071288) | |
| Performance Testing | | | | |
| Complies to<br>Standards | Complies with:<br>ISO 14457 (Dentistry -<br>Handpieces and motors) | Complies with:<br>ISO 14457 (Dentistry -<br>Handpieces and motors) | Complies with:<br>ISO 7785-2 (Dental<br>handpieces - Part 2: | Same |
| | | | | |
| | ISO 1797 (Dentistry - Shanks<br>for rotary and oscillating<br>instruments)<br>ISO 3964 (Dentistry - Coupling<br>dimensions for handpiece<br>connectors) | ISO 1797-1 (Dentistry -<br>Shanks for rotary<br>instruments - Part 1:<br>Shanks made of metals)<br>ISO 3964 (Dentistry -<br>Coupling dimensions for<br>handpiece connectors) | Straight and geared<br>angle handpieces)<br>ISO 1797 (Dentistry -<br>Shanks for rotary<br>instruments)<br>ISO 3964 (Dentistry -<br>Coupling dimensions for<br>handpiece connectors) | |
| Sterilization | Sterilizable according to ISO<br>17665-1 (Sterilization of health<br>care products – Moist heat –<br>Part 1: Requirements for the<br>development, validation and<br>routine control of a sterilization<br>process for medical devices on<br>the final, finished device | Sterilizable according to<br>ISO 17665-1<br>(Sterilization of health<br>care products – Moist<br>heat - Part 1:<br>Requirements for the<br>development, validation<br>and routine control of a<br>sterilization process for<br>medical devices on the<br>final, finished device | Sterilizable according to<br>ISO 17665-1<br>(Sterilization of health<br>care products - Moist<br>heat - Part 1:<br>Requirements for the<br>development, validation<br>and routine control of a<br>sterilization process for<br>medical devices on the<br>final, finished device | Same |
| Biocompatibility | Biocompatible according to ISO<br>10993-1<br>(Biological evaluation of<br>medical devices - Part 1:<br>Evaluation and testing within a<br>risk management system). | Biocompatible according<br>to ISO 10993-1<br>(Biological evaluation of<br>medical devices - Part 1:<br>Evaluation and testing<br>within a risk management<br>system). | Biocompatible according<br>to ISO 10993-1<br>(Biological evaluation of<br>medical devices - Part 1:<br>Evaluation and testing<br>within a risk<br>management system). | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Non-Clinical Test Data:
Performance bench testing according to international standards for Dental Handpieces has been conducted to determine conformance in regards to:
- . Biocompatibility has been completed for the applicable components.
- o Comparative performance testing of the functions of the Proposed device compared to the cleared stand-a-lone device.
Furthermore, the performance of the SURGmatic S15 L Pro has been verified utilizing the following standards:
- ISO 10993-1:2018: Biological evaluation of medical devices Part 1: Evaluation and o testing within a risk management process.
- . ISO 10993-5:2009: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10:2010: Biological evaluation of medical devices -Part 10: Tests for irritation . and skin sensitization.
- ISO 7405:2018 Dentistry Evaluation of biocompatibility of medical devices used in ● dentistry.
- ISO 14971:2007: Medical devices Application of risk management to medical devices. ●
- IEC 62366-1:2015-02: Medical devices Part 1: Application of usability engineering to ● medical devices [Including CORRIGENDUM 1 (2016)]
- . ISO 17664:2017: Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices.
- . ISO 14457:2017: Dentistry - Handpieces and motors
{10}------------------------------------------------
- . AAMI / ANSI ST79:2017: Comprehensive quide to steam sterilization and sterility assurance in health care facilities
- ISO 3964:2016: Dentistry - Coupling dimensions for handpiece connectors
- ISO 1797:2017: Dentistry - Shanks for rotary and oscillating instruments
- o ISO 17665-1:2006: Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices on the final, finished device
### Clinical Performance Data:
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product.
### Conclusion as to Substantial Equivalence:
Based on a comparison of intended use, indications, technological characteristics, principle of operation, and performance data, the SURGmatic S15 L Pro is deemed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.