SUPERTORQUE HIGH-SPEED HANDPIECES AND GENTLEPOWER HANDPIECES AND ATTACHEMENTS
Applicant
Sybron Dental Specialties, Inc.
Product Code
EFB · Dental
Decision Date
Sep 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The SUPERtorque High-Speed Handpieces are intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry. The GENTLEpower Handpieces and Attachments are intended for the removal of carious material, excess filling material, reducing hard tooth structure, cavity and crown preparation, root canal preparation, finishing tooth preparations, restorations and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
Device Story
SUPERtorque and GENTLEpower are reusable, air-powered or electric-motor-driven dental handpieces; used by trained dental professionals in clinical settings for tooth preparation, restoration, and polishing; devices feature ergonomic designs and optional fiber optic lighting; sterilized via steam autoclave; output is mechanical rotation for dental burs/instruments; aids clinicians in precise removal of tooth structure and restorative materials; benefits patients through efficient, controlled dental procedures.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Air-powered or electric-motor-driven dental handpieces; reusable; ergonomic form factor; optional fiber optic light system; steam autoclave sterilization.
Indications for Use
Indicated for trained dental professionals to perform removal of carious/filling material, tooth structure reduction, cavity/crown/root canal preparation, finishing, and polishing in general dentistry patients.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
Siemens AG (now Sirona), T1 Control High Speed Dental Handpieces (K965238)
DEN-TAL-EZ, Inc., STAR DENTAL 430 Series High Speed Handpieces (K960719)
A-dec/W&H, Synea Air-Driven Highspeed Handpieces and Attachments (K070663)
Submission Summary (Full Text)
{0}------------------------------------------------
K073478
Image /page/0/Picture/1 description: The image shows the logo for Sybron Dental Specialties, which includes the letters 'sds' in a stylized font with a swooping line above them. Below the logo, the text 'SYBRON DENTAL SPECIALTIES' is printed in a smaller, sans-serif font. The text 'Section III - 510(k) Summary of Safety and Effectiveness' is located below the company name.
SEP - 3 2008
Submitter:
Sybron Dental Specialties, Inc. 1717 West Collins Drive Orange, CA 92656 Colleen Boswell - Contact Person Date Summary Prepared: November 2007
#### Device Name:
- Trade Namc SUPERtorque High-Speed Handpieces �
- Common Name Dental Handpiece .
- Classification Name Handpiece, Air-Powered, Dental, per 21 CFR § 872.4200 .
## Devices for Which Substantial Equivalence is Claimed:
- Siemens AG (now Sirona), T1 Control High Speed Dental Handpieces (K965238) .
- DEN-TAL-EZ, Inc., STAR DENTAL 430 Series High Speed Handpieces (K960719). .
## Device Description:
The SUPERtorque High-Speed Handpieces are dental handpieces for use by a trained professional in general dentistry. The devices are air-powered handpieces that are reusable and ergonomically shaped, and are provided both with and without a fiber optic light system. The devices can be sterilized by the steam autoclave method.
## Intended Use of the Device:
The SUPERtorque High-Speed Handpieces are intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
## Substantial Equivalence:
The SUPERtorque High-Speed Handpieces is substantially equivalent to other legally marketed devices in the United States. The intended use of the devices is identical to that of the predicate. The SUPERtorque High-Speed Handpieces are substantially equivvalent in design, application and function to the TI Control High Speed Dental Handpiees marketed by Siemens AG (now owned by Sirona) and STAR DENTAL 430 Series High Speed Handpieces marketed by DEN-TAI -EZ, Inc.
{1}------------------------------------------------
# K073478
Image /page/1/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a simple, sans-serif font. The logo is black and white.
Section III - 510(k) Summary of Safety and Effectiveness
## Submitter:
Sybron Dental Specialties, Inc. 1717 West Collins Drive Orange, CA 92656 Colleen Boswell - Contact Person Date Summary Prepared: November 2007
#### Device Namc:
- . Trade Name - GENTLEpower Handpieces and Attachments
- . Common Name - Dental Handpiece and Accessories
- Classification Name Handpiece, Air-Powered, Dental, per 21 CFR § 872.4200 .
## Devices for Which Substantial Equivalence is Claimed:
- A-dec/W&H, Synea Air-Driven Highspeed Handpieces and Attachments (K070663) .
#### Device Description:
The GENTLEpower Handpieces and Attachments are dental instruments for use by a trained professional in general dentistry. The handpieces and attachments are powered by either an air-motor or electric motor. The devices are re-usable and ergonomically shaped, and can be sterilized by the steam autoclave method.
#### Intended Use of the Device:
The GENTLEpower Handpieces and Attachments are intended for the removal of carious material, excess filling material, reducing hard tooth structure, cavity and crown preparation, root canal preparation, finishing tooth preparations, restorations and for polishing tooth.
#### Substantial Equivalence:
The GENTLEpower Handpieces and Attachments are substantially equivalent to other legally marketed devices in the United States. The intended use of the devices is identical to that of the predicate. The GENTLEpower Handpieces and Attachments are substantially equivalent in design, application and function to the Synea Air-Driven Highspeed Handpieces and Attachments marketed by A-Dec/W&H.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## SEP - 3 2008
Ms. Colleen Boswell Vice President, Regulatory Affairs Sybron Dental Specialtics, Incorporated 1717 West Collins Avenue Orange, California 92867
Re: K073478
Trade/Device Namc: SUPERtorque High-Speed Handpieces and GENTI.Epower Handpieces and Attachments Regulation Number: 872.4200 Regulation Name: Dental Handpiecc and Accessories Regulatory Class: II Product Code: EFB Dated: August 29, 2008 Received: September 2, 2008
Dear Ms. Boswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Boswell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labcling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
MKAnnelle Kendon for //
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known):
Device Name: SUPERtorque High-Speed Handpieces
#### Indications for Use:
The SUPERtorque High-Speed Handpieces are intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
3
Conettreance of CDRH, Office of Device Evaluation (ODE)
Surenkane
(Division Sign-Off)
Division of Anesthes:ology, General Hospital Infection Control, Dental Devices
510(k) Number: KC73078
Page 1 of 1
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known):
Device Name: GENTLEpower Handpieces and Attachments
#### Indications for Use:
The GENTLEpower Handpieces and Attachments are intended for the removal of carious material, excess filling material, reducing hard tooth structure, cavity and crown preparation, root canal preparation, finishing tooth preparations, restorations and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concusrence of CDRH, Office of Device Evaluation (ODE)
Susan Fung
Division Sign-Off) ivision of Anesthesiology, General Hospital nrection Control, Dental Devices
10(k) Number:
4
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.