TIGER 500 SERIES HIGH SPEED HANDPIECES AND ATTACHMENTS
Applicant
Thunder Tiger Corp.
Product Code
EFB · Dental
Decision Date
Apr 27, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
Tiger 500 Series High Speed Handpieces and Attachments are intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.
Device Story
Air-powered dental handpieces; reusable; ergonomically shaped; available with or without fiber optic light system. Used by dentists/dental professionals in clinical settings for tooth preparation and restoration procedures. Connects to various manufacturer couplings (TTBIO, KaVo, NSK Mach, Sirona, Star). Supplied non-sterile; requires sterilization before use. Provides mechanical power to rotary dental burs for cutting/polishing tooth structure; enables precise removal of carious material and finishing of restorations. Benefits patient by facilitating restorative dental procedures.
Clinical Evidence
Bench testing only. Performance and safety verified through conformity with ISO 7785-1 (High-speed air turbine handpieces). No clinical data provided.
Technological Characteristics
Air-powered turbine handpiece; reusable; fiber optic light option. Complies with ISO 7785-1:1997 for high-speed air turbine handpieces. Manufactured under ISO 13485:2003 quality management system. Compatible with multiple industry-standard couplings. Non-sterile, autoclavable.
Indications for Use
Indicated for dental patients requiring removal of carious material, reduction of hard tooth structure, cavity preparation, finishing of preparations and restorations, and tooth polishing. Restricted to use by licensed professionals.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Thunder Tiger. The logo consists of a circle with an airplane inside of it on the left. To the right of the circle is the text "Thunder Tiger" written in a stylized font, with the word "Thunder" on top of the word "Tiger".
THUNDER TIGER CORP. APR 2 7 2011 NO.7, 6th ROAD INDUSTRY PARK, TAICHUNG. TAIWAN. R.O.C. 40755 TEL : 886-4-23591616 FAX : 886-4-23591902 E-mail: tt@thundertiger.com http://www.thundertiger.com
### K10 2517 510(K) SUMMARY
# Tiger 500 Series High Speed Handpieces and Attachments
1. Date Summary Prepared: December 20,2010
2. Submitter Information
510(k) Owner: THUNDER TIGER CORP.
No.7, 6th Road, Industry Park, Taichung, 40755
Taiwan, ROC
Contact Person: Jo S.C. Lee/QA Engineer
jo@thundertiger.com
3. Device Name
-Trade-Name :- Tiger-500 Series-High-Speed-Handpieces-and-Attachments-
Common Name: Dental Handpiece
Classification Name: Handpiece, Air-Powered, Dental
(21 CFR 872.4200, Product Code EFB)
4. Predicate Device: Dental Air-Powered Handpiece, models TIGER 300T, TIGER 300K,
TIGER 300W, TIGER 300B, TIGER 300N (K062812)
SUPERtorque High-Speed Handpieces (K073478)
Tradition high speed Handpiece (K863677)
Rapidd high speed Handpiece (K003518)
5. Device Description:
The Tiger 500 series High Speed Handpieces and Attachments are similar to other high-speed dental handpieces currently on the US dental market in design, function, and intended use. The devices are air-powered handpieces that are reusable and ergonomically shaped, and are provided both with and without a fiber optic light system. The devices are supplied non-sterile, but must be sterilized before use, as with the predicate device. It will be packaged as a single-use bag. Several models could be connected to couplings of manufacturers including TTBIO, KaVo, NSK Mach, Sirona, and Star. According to IS07785-1:1997, Dental Handpiece-- Part 1: High-speed air turbine handpieces; and IS013485:2003, Medical Device Quality Management System to complete the device design steps.
{1}------------------------------------------------
# K102517
Image /page/1/Picture/1 description: The image shows the logo for Thunder Tiger. The logo consists of a circle with an airplane inside of it on the left side. To the right of the circle, the words "Thunder Tiger" are written in a bold, sans-serif font, with the word "Thunder" on top of the word "Tiger".
THUNDER TIGER CORP. NO.7. 6th ROAD INDUSTRY PARK, TAICHUNG, TAIWAN, R.O.C. 40755 TEL : 886-4-23591616 FAX : 886-4-23591902 E-mail: tt@thundertiger.com http://www.thundertiger.com
- Intended Use: 6.
Tiger 500 Series High Speed Handpieces are intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.
CAUTION: U.S. Federal law restricts the use of this device to licensed professionals.
7. Technological Characteristics:
The functions of Tiger 500 series High Speed Handpieces and Attachments were verified according to ISO7785-1. The substances of Tiger 500 series High Speed Handpieces are same as the predicate device in terms of its intended use, operating principles and functions.
- 8. Non-Clinical Performance/Safety Data:
The performance and safety of the Tiger 500 series High Speed Handpieces and Attachments are based upon conformity with applicable aspects of ISO 7785-1. Bench testing results demonstrate substantially equivalence. Therefore, we conclude that the both safe and effective for its intended use.
| Substantial Equivalence Table | |
|--------------------------------------------------|-------------------------|
| TECHNOLOGICAL CHARACTERISTIC | COMPARISON TO PREDICATE |
| Intended use | Identical |
| Indications for use | Identical |
| Target population | Identical |
| Anatomical sites | Identical |
| Energy used and/or delivered | Similar |
| Human factors | Similar |
| Design | Similar |
| Performance | Similar |
| Standards met | Identical |
| Materials | Similar |
| Biocompatibility | Identical |
| Compatibility with environment and other devices | Identical |
| Sterility | Identical |
| Mechanical safety | Identical |
| Chemical Safety | Identical |
| Thermal safety | Identical |
| Radiation safety | Identical |
9. Substantial Equivalence
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Thunder Tiger. The logo consists of a circle with an airplane in the middle, and the words "Thunder Tiger" written in a stylized font to the right of the circle. The words "Thunder" and "Tiger" are stacked on top of each other, with a line above and below each word.
THUNDER TIGER CORP. NO.7, 6th ROAD INDUSTRY PARK, TAICHUNG, TAIWAN, R.O.C. 40755 FAX : 886-4-23591902 TEL : 886-4-23591616 E-mail: tt@thundertiger.com http://www.thundertiger.com
### 10. Conclusion:
The above descriptions coincide with the substantial equivalence made by Dental Air-Powered Handpiece, models TIGER 300T, TIGER 300K, TIGER 300W, TIGER 300B, TIGER 300N (K062812), SUPERtorque High-Speed Handpieces (K073478), Tradition high speed Handpiece (K863677) and Rapidd high speed Handpiece (K003518). They are same as the predicate device in terms of its intended use, operating principles and functions. Therefore, it can be seen that Tiger 500 series High Speed Handpieces are both safe and effective for their intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Jo S.C. Lee Ouality Assurance Engineer Thunder Tiger Corporation No. 7, 6th Road Industry Park Taichung China (Taiwan) R.O.C. 40755
APR 2 7 2011
Re: K102517
Trade/Device Name: Tiger 500 Series High Speed Handpieces and Attachments Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece Accessories Regulatory Class: I Product Code: EFB Dated: April 19, 2011 Received: April 21, 2011
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Lee
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting . (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm l 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Thunder Tiger. The logo consists of a circle with an airplane inside of it on the left, and the words "Thunder Tiger" on the right. The word "Thunder" is on top of the word "Tiger", and there is a line above the word "Tiger".
、
THUNDER TIGER CORP. NO.7, 6th ROAD INDUSTRY PARK, TAICHUNG, TAIWAN, R.O.C. 40755 TEL : 886-4-23591616 FAX : 886-4-23591902 E-mail: tt@thundertiger.com http://www.thundertiger.com
## Indications for Use
510(K) Number (If Known): ____________________________________________________________________________________________________________________________________________________
Device Name: Tiger 500 Series High Speed Handpieces and Attachments Indications for Use:
Tiger 500 Series High Speed Handpieces and Attachments are intended for removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.
Over-The-Counter Use_ AND/OR Prescription Use_ X (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH Office of Device Evaluation (ODE) | |
|-----------------------------------------------------------------------------------|---------|
| BSBetz DDS for Dr Susan Renner | |
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | |
| 510(k) Number: | K102517 |
age 4-1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.