Spinal Needle (Quincke Bevel and Pencil Point) is intended for injection of local anesthetics into the subarachnoid space to provide spinal anesthesia for pain management. The device is intended to be used in a professional healthcare environment and not tested for MRI safety.
Device Story
Spinal needles (18G-27G) with Quincke bevel or pencil point tips; feature removable stylet to occlude lumen during insertion; luer connector hub. Used by clinicians in professional healthcare environments to access subarachnoid space for local anesthetic delivery. Stylet withdrawn after placement to allow injection. Provides spinal anesthesia for pain management.
Clinical Evidence
No clinical data included; bench testing only.
Technological Characteristics
Stainless steel needle tubing; 18G-27G gauge; 70-120mm lengths; Quincke bevel or pencil point tips; 6% luer taper hub. EO sterilized. Complies with ISO 9626, ISO 7864, ISO 594-1/2, ISO 80369-7, and ISO 6009. Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for patients requiring spinal anesthesia via injection of local anesthetics into the subarachnoid space for pain management. Intended for use in professional healthcare environments.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is placed inside a blue square. The text is stacked in three lines, with "ADMINISTRATION" on the bottom line.
July 29, 2022
Global Medikit Limited % Atonu Dutta CEO Alceon 201-203 Tilak Complex, Jetalpur Road Vadodara, Gujarat 390007 India
Re: K212982
Trade/Device Name: Medispine - Spinal Needle, Glospine - Spinal Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: June 30, 2022 Received: July 5, 2022
Dear Atonu Dutta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212982
Device Name
Medispine - Spinal Needle and Glospine - Spinal Needle
### Indications for Use (Describe)
Spinal Needle (Quincke Bevel and Pencil Point) is intended for injection of local anesthetics into the subarachnoid space to provide spinal anesthesia for pain management. The device is intended to be used in a professional healthcare environment and not tested for MRI safety.
**Type of Use (Select one or both, as applicable)**
| <span style="font-size: 10pt;"> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <div style="text-align: left;">☑</div> | <div style="text-align: left;">☐</div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1. Submission Sponsor
Global Medikit Limited, Khasra no. 323 (MI), Camp Road, Selaqui, Dehradun, Uttarakhand- 248197, India. Contact Person: Mr. Karun Narang Phone: +91 11 27667889 Email: regulatory@globalmedikit.in
#### 2. Submission Correspondent
Alceon 201-203 Tilak Complex, Jetalpur Road, Vadodara, Gujarat-390 007, India. Contact Person: Mr. Atonu Dutta Phone: (+91) 9925023428 Email: regulatory@alceon.in
### 3. Date Summary Prepared:
9th September 2021
#### Device name & classification: 4.
Trade/Proprietary Name: Medispine – Spinal Needle and Glospine – Spinal Needle Common/Usual Name: Anesthesia conduction needle Device classification Regulation: 21CFR Part 868.5150 Product Code: BSP Device Class: Class 2 Classification Panel: Anesthesiology
#### 5. Predicate Device
Trade/Proprietary Name: Spinal Anesthesia Needles Manufacturer: Zhejiang kindly medical devices Co., Ltd Common/Usual Name: Anesthesia conduction needle 510(K): K171518 Device classification Regulation: 21CFR Part 868.5150 Product Code: BSP Device Class: Class 2
### Global Medikit Limited
{4}------------------------------------------------
Classification Panel: Anesthesiology
#### 6. Device Description
Spinal Needles are long, flexible needles of small gauge (typically 18G to 27G) which are available with different types of tips: Pencil point and Quincke bevel. The Spinal Needles are available in a series of combination of needle size and length.
The smaller size needles of Pencil point tip are provided with introducer needles to provide support for accessing the tough tissues.
Spinal needles have a removable stylet that completely occludes the lumen to avoid block. The stylet is withdrawn after the spinal anesthesia needle has penetrated into the subarachnoid space for injecting anesthetic.
The needle hub is luer connector (6% luer taper).
#### 7. Indications for Use
Spinal Needle (Quincke Bevel and Pencil Point) is intended for injection of local anesthetics into the subarachnoid space to provide spinal anesthesia for pain management. The device is intended to be used in a professional healthcare environment and not tested for MRI safety.
#### 8. Comparison to predicate device
The following table compares the subject device to the predicate device with respect to their indications for use, technology and performance testing, thus demonstrating the basis for determination of substantial equivalence between these devices.
| Description | Subject Device | Predicate Device | Comparison |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|------------|
| Trade/Device Name | Spinal Needle | Spinal Anesthesia<br>Needles | - |
| Manufacturer | Global Medikit Limited | Zhejiang kindly medical<br>devices Co., Ltd | - |
| 510 (K) | Applied | K171518 | - |
| Class | Class 2 | Class 2 | - |
| Regulatory number | 21 CFR Part 868.5150 | 21 CFR Part 868.5150 | - |
| Product code | BSP | BSP | - |
| Clinical | | | |
| Intended Use | Spinal Needle (Quincke<br>Bevel and Pencil Point) | The Spinal Anesthesia<br>Needles are | Same |
| Description | Subject Device | Predicate Device | Comparison |
| | is intended for injection<br>of local anesthetics into<br>the subarachnoid space<br>to provide spinal<br>anesthesia for pain<br>management. | intended to be used for<br>injection of<br>local anesthetic agent<br>into the<br>subarachnoid cavity for<br>pain<br>management. | |
| Technical specification | | | |
| | Stylet Hub<br>Stylet<br>Needle Tube<br>Needle Hub<br>Needle Hub Insert<br>Protective Cap of<br>Needle | Stylet Hub<br>Stylet<br>Needle Tube<br>Needle Hub<br>Needle Hub Insert<br>Protective Cap of<br>Needle | |
| Product<br>Configuration | | | Same |
| Needle gauge | 18G-27G | 18G-27G | Same |
| Needle length | 70 mm, 90 mm, 120<br>mm | 40mm, 50mm, 65mm,<br>70mm, 80mm,<br>90mm, 120mm | Same |
| Needle tip<br>configuration | Quincke bevel and<br>Pencil point | Quincke and Pencile | Same |
| Sterility | EO sterilized | EO Sterilized | Same |
| Single use | Single use | Single Use | Same |
| Shelf life | 5 years | 3 years | Similar# |
| Label/Labeling | Conform with 21 CFR<br>801 | Conform with 21 CFR<br>801 | Same |
| Product<br>Performance | Complies with:<br>ISO 9626:2016,<br>ISO 7864:2016,<br>ISO 594-1:1986,<br>ISO 594-2:1998<br>ISO 80369-7:2016<br>ISO 6009:2016 | Complies with:<br>ISO 9626:1991 AMD<br>1:2001,<br>ISO 7864:1993,<br>ISO 594-1:1986,<br>ISO 594-2:1998 | Same |
| Biological | | | |
| Needle Material | Stainless steel | Stainless steel | Same |
| Description | Subject Device | Predicate Device | Comparison |
| Biocompatibility | Conform with ISO<br>10993 | Conform with ISO<br>10993 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
The shelf life of subject device is verified through the stability studies conducted for the device and hence, there are no additional risk generated due to the difference observed in the shelf life of the proposed device compared to the predicate device.
Hence, the device is considered as substantially equivalent to the predicate device.
### 9. Summary of non-clinical testing
Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- ISO 7864:2016 4th edition: Sterile hypodermic needles for single use - Requirements and test methods
- o ISO 9626:2016 2nd edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 594-1:1986 1st edition: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
- ISO 594-2:1998 2nd edition: Conical fittings with a 6% (Luer) taper for syringes, needles ● and certain other medical equipment - Part 2: Lock fittings
- ISO 80369-7:2016 1st edition: Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- ISO 6009:2016 4th edition: Hypodermic needles for single use - Colour coding for identification.
The biocompatibility tests were performed in accordance with the following standards:
- . ISO 10993-1:2018 5th edition: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- . ISO 10993-4:2017 3rd edition: Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood
- ISO 10993-5:2009 3rd edition: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-10:2021 4th edition: Biological evaluation of medical devices - Part 10: Tests for skin sensitization
{7}------------------------------------------------
- . ISO 10993-23:2021:15 edition: Biological evaluation of medical devices – Part 23: Tests for irritation
- . ISO 10993-11:2017 3rd edition: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
The device is sterilized by Ethylene oxide sterilization method, the sterilization process was validated as per ISO 11135:2014/AMD 1:2018 2nd edition: Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices.
The EO residual were tested which meets the requirements of ISO 10993-7:2008 2nd edition: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
The Bacterial endotoxin testing of subject devices was performed by "Gel-Clot Method" and meets the requirement of USP.
Packaging validation and transport study tests were conducted in accordance with the following standard:
- . ISO 11607-1:2019 2nd edition: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ISO 11607-2: 2019 2nd edition: Packaging for terminally sterilized medical devices -Part 2: Validation requirements for forming, sealing and assembly processes
- . ASTM D4169-16: Standard Practice for Performance Testing of Shipping Containers and Systems
### 10. Summary of Clinical Testing
No clinical data was included in this premarket application submission.
### 11. Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.