Pain Therapy Device(Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I)

K212948 · Guangzhou Xinbo Electronic Co., Ltd. · NGX · Nov 1, 2021 · Physical Medicine

Device Facts

Record IDK212948
Device NamePain Therapy Device(Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I)
ApplicantGuangzhou Xinbo Electronic Co., Ltd.
Product CodeNGX · Physical Medicine
Decision DateNov 1, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Pain Therapy Device (Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I): TENS: - For relief of pain and management of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise, work or normal household work activities by applying transcutaneous electrical current to stimulate the nerves. - For relief of pain associated with arthritis. - For symptomatic relief and management of chronic intractable pain EMS: - To stimulate healthy muscles in order to improve and facilitate muscle performance. - To temporarily increase local blood circulation in healthy muscles. NMES: - For relaxation of muscle spasm. - Increase of blood flow circulation. - Prevention or retardation of disuse muscle atrophy. - Muscle re-education. - Maintaining or increasing range of motion. - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. - For Users with conditions or diseases that are associated with impaired (poor) blood flow in the legs/ankle/feet. NMES through the foot electrodes is intended to use as an adjunctive treatment (as an additional treatment to your existing treatment) to temporarily reduce lower extremity pain, swelling and cramping.

Device Story

Pain Therapy Device (Models DH-P.T.S-I, II, IIA, IIB, CP-I) is an OTC electrical stimulator. Device delivers transcutaneous electrical current to nerves and muscles via electrodes. User operates device to select stimulation parameters for pain relief, muscle conditioning, or circulation improvement. Output consists of electrical pulses applied to skin. Device aids in managing chronic pain, muscle atrophy, and post-surgical recovery. Benefits include non-pharmacological pain management, improved muscle function, and reduced swelling/cramping in lower extremities.

Clinical Evidence

Bench testing only.

Technological Characteristics

Powered muscle stimulator; electrical stimulation via electrodes; battery-operated; multi-model form factor; non-implantable; software-controlled pulse generation.

Indications for Use

Indicated for adults seeking relief of pain associated with sore/aching muscles (back, neck, extremities) from strain, arthritis, or chronic intractable pain. Also indicated for healthy individuals to improve muscle performance, increase local blood circulation, relax muscle spasms, prevent disuse atrophy, facilitate muscle re-education, maintain range of motion, and provide post-surgical calf stimulation to prevent venous thrombosis. Adjunctive treatment for users with impaired blood flow in legs/ankles/feet to reduce pain, swelling, and cramping.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. November 1, 2021 Guangzhou Xinbo Electronic Co., Ltd. % Cassie Lee Official Correspondent Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District Guangzhou, Guangdong 511400 China Re: K212948 Trade/Device Name: Pain Therapy Device(Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, NUH, IPF, NYN, GZJ Dated: September 14, 2021 Received: September 16, 2021 Dear Cassie Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K212948 #### Device Name The Pain Therapy Device (Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I) Indications for Use (Describe) The Pain Therapy Device (Model: DH-P.T.S-I, DH-P.T.S-II, DH-P.T.S-IIA, DH-P.T.S-IIB, DH-CP-I): TENS: - For relief of pain and management of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise, work or normal household work activities by applying transcutaneous electrical current to stimulate the nerves. - For relief of pain associated with arthritis. - For symptomatic relief and management of chronic intractable pain ## EMS: - To stimulate healthy muscles in order to improve and facilitate muscle performance. - To temporarily increase local blood circulation in healthy muscles. ## NMES: - For relaxation of muscle spasm. - Increase of blood flow circulation. - Prevention or retardation of disuse muscle atrophy. - Muscle re-education. - Maintaining or increasing range of motion. - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. - For Users with conditions or diseases that are associated with impaired (poor) blood flow in the legs/ankle/feet. NMES through the foot electrodes is intended to use as an adjunctive treatment (as an additional treatment to your existing treatment) to temporarily reduce lower extremity pain, swelling and cramping. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...