K162060 · New World Medical, Inc. · KYF · Oct 24, 2016 · Ophthalmic
Device Facts
Record ID
K162060
Device Name
Ahmed Glaucoma Valve
Applicant
New World Medical, Inc.
Product Code
KYF · Ophthalmic
Decision Date
Oct 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3920
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162060 · Oct 24, 2016
Ahmed Glaucoma Valve
New World Medical, Inc.
Retrospective medical record/chart review
The sponsor utilized a retrospective, consecutive case series to compare the safety and efficacy of the Ahmed Glaucoma Valve Model FP7 (silicone) against the predicate Model S2 (polypropylene) in a clinical setting.
Comparison of silicone and polypropylene Ahmed glaucoma valves: two-year follow-up (Mackenzie et al., 2007); Single surgeon, retrospective, consecutive case series; Follow-up/Duration: 2 years
Patients undergoing Ahmed glaucoma drainage device implantation; Number of Sites: 1
Ahmed Glaucoma Valve Model S2 (polypropylene)
Safety and efficacy
Indications for Use
The Ahmed™ Glaucoma Valve is intended for use in the surgical management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medications, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.
Device Story
Ahmed Glaucoma Valve Model FP7 is a valved aqueous drainage implant for refractory glaucoma; regulates intraocular pressure (IOP). Device consists of silicone drainage tube connected to valve mechanism (silicone sheet between polypropylene plates) housed in flexible silicone endplate. Implanted surgically by ophthalmologist; endplate placed on eye surface to distribute aqueous humor from anterior chamber. Valve acts as variable resistor: opens at high IOP to allow flow; closes at low IOP to prevent hypotony. Maintains IOP within physiological range. Modification from predicate (Model S2) includes flexible silicone endplate with stiffening ribs to improve biocompatibility and reduce inflammation/scarring compared to rigid polypropylene. Benefits patient by providing long-term IOP control in complex glaucoma cases.
Clinical Evidence
Evidence includes multiple prospective, multicenter, randomized controlled trials and retrospective case series comparing AGV-FP7 to AGV-S2 and Baerveldt implants. Studies (e.g., Ahmed Baerveldt Comparison Study) evaluated long-term IOP outcomes and complication rates over 1-5 years. Results indicate AGV-FP7 provides similar or superior IOP control within physiological range and comparable safety profiles to predicate devices.
Indicated for management of refractory glaucomas in patients where previous surgical treatment failed or is known to be ineffective. Includes neovascular glaucoma, primary open angle glaucoma unresponsive to medications, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.
Regulatory Classification
Identification
An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) “Aqueous Shunts—510(k) Submissions.”
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2016
New World Medical, Inc. Mr. Suhail Abdullah Chief Scientific Officer 10763 Edison Court Rancho Cucamonga, CA 91730
Re: K162060
Trade/Device Name: Ahmed Glaucoma Valve Regulation Number: 21 CFR 886.3920 Regulation Name: Aqueous Shunt Regulatory Class: Class II Product Code: KYF Dated: July 25, 2016 Received: July 26, 2016
Dear Mr. Abdullah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Ahmed Glaucoma Valve Model FP7
#### Indications for Use (Describe)
The Ahmed Glaucoma Valve is indicated for the management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medications, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size: 20px;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size: 20px;"> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
This 510(k) Summary has been prepared in accordance with 21CFR807.92
## Date Prepared and Submitted:
July 25th, 2016
# Address and Registration
Submitter: New World Medical, Inc.
The address and registration number of the manufacturer and sterilization site of all Ahmed™ Glaucoma Valve Models are:
| Manufacturer | Sterilization Site |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| New World Medical, Inc.<br>10763 Edison Court<br>Rancho Cucamonga, CA 91730<br>Phone: 909-466-4304<br>Fax: 909-466-4305<br>Contact Person: Suhail Abdullah | Sterigenics U.S., LLC<br>344 Bonnie Circle<br>Corona, CA 92880 |
| FDA REGISTRATION #: 1000125279 | FDA REGISTRATION #: 2029275 |
## Device Name
The device trade name and common/classification name are:
| Device Trade Name | Common Name | Classification Name |
|--------------------------------------------|--------------------------|------------------------------------------------------|
| Ahmed™ Glaucoma Valve<br>Implant Model FP7 | Glaucoma Drainage Device | Aqueous Shunt (21 CFR 886.3920,<br>Product Code KYF) |
### Predicate Device
The two predicate devices are the Ahmed™ Glaucoma Valve Model S2 which was first marketed in 1993 and the Baerveldt Glaucoma Implant, first marketed in 1991. The Ahmed™ Glaucoma Valve Model FP7 is a modification to the Ahmed™ Glaucoma Valve Model S2 (510(k) Number: K925636) in which the plate design and material have been modified.
Predicate devices information 510(k) Number: K925636 Device Name: Ahmed™ Glaucoma Valve Model S2 Decision Date: 11/12/1993
510(k) Number: K905129 Device Name: Baerveldt Glaucoma Implant Decision Date: 02/11/1991
{5}------------------------------------------------
# Device Class
Ahmed™M Glaucoma Valve Implants have been classified as Class II in the Ophthalmic panel (21CFR886.3920) with product code KYF.
# Intended Use
The Ahmed™ Glaucoma Valve is intended for use in the surgical management of refractory glaucomas, where previous surgical treatment has failed, or by experience is known not to provide satisfactory results. Such refractory glaucomas can include neovascular glaucoma, primary open angle glaucoma unresponsive to medications, congenital or infantile glaucoma, and refractory glaucomas resulting from aphakia or uveitis.
#### This is the same intended use as the previously cleared Ahmed™ Glaucoma Valve S2, 510(k) Number K925636
# Device Description and Technological Characteristics
The Ahmed™ Glaucoma Valve Model FP7 (AGV-FP7) is a valved aqueous drainage implant designed to regulate intraocular pressure in eyes suffering from intractable glaucoma. The Ahmed™ device is comprised of a silicone drainage tube that is connected to a valve mechanism. This valve mechanism is the same in the AGV-FP7 and the predicate AGV-S2. The valve mechanism consists of a silicone sheet folded and pressed between two complimentary polypropylene plates. The valve mechanism is securely positioned in a pocket inside of a silicone endplate that serves to distribute the aqueous humor from the anterior chamber of the eye over the surface of the endplate. The valve in the AGV-FP7 behaves like a variable resistor, decreasing resistance to allow more flow when intraocular pressure is high. When pressure is low, the resistance to fluid outflow is high and the valve closes, thereby preventing hypotony. By means of the valve mechanism, the AGV-FP7 maintains intraocular pressure within the appropriate physiological range.
In both the AGV-S2 and AGV-FP7, the silicone sheet is folded and pressed between two polypropylene plates. In the AGV-S2, the bottom polypropylene plate is comprised of the polypropylene endplate body. In the AGV-FP7, the polypropylene bottom plate is separate from the silicone endplate material. The valve mechanism is inserted into a pocket in the silicone endplate to fixate the valve components to the endplate. Additional differences include stiffening ribs in the posterior half of the AGV-FP7 to add stiffness to the flexible endplate, The other predicate device, the Baerveldt Glaucoma Implant (BGI) also consists of a flexible silicone endplate which shares some features with the AGV-FP7, though the BGI endplate is larger in area. The AGV-FP7 endplate has the same curvature as the average human eye at its equator and also protects the valve from blockage by fibrous tissue. The endplate is made of flexible silicone. Inflammation and scarring around flexible silicone implants in animal ocular tissue was less pronounced than that found around rigid polypropylene.
### Testing
Extensive non-clinical testing of the AGV-FP7 has been performed per the FDA Guidance on Aqueous Shunts and includes the following:
{6}------------------------------------------------
- . Physical Stability testing to ensure that the device maintains its performance characteristic and structural integrity after exposure to an aqueous environment at body temperature.
- Chemical analysis and aqueous aging chemical testing to assess any potential chemical hazards and . stability of device materials
- . Validation of the packaging system
- Sterilization validation
- Distribution simulation and subsequent validation of the stability of the packaging system and device
- Accelerated aging and subsequent validation of the stability of the packaging system and device
- . Biocompatibility testing
A published animal study also indicated that the AGV-FP7 valve mechanism is functional in-vivo, as expected, and that the Foreign Body Reaction to the AGV-FP7 is consistent with bleb formation around predicate implant materials such as silicone and polypropylene.
Furthermore, there are several published clinical studies comparing the AGV-FP7 to the predicate AGV-S2 and the BGI. In these studies, the AGV-FP7 group of patients had similar IOP and complication rates to the predicate devices. Particularly, compared to the AGV-S2, the IOP of the AGV-FP7 groups was lower (within the acceptable physiological range) and fewer complications were reported. This clinical data which includes safety and effectiveness data and information on adverse effects and complications associated with the AGV-FP7 can be found in the referenced citations below.
- 1. The Ahmed Baerveldt Comparison Study. This study compared the long-term outcomes and complication of the Ahmed Glaucoma Valve model FP7 (AGV-FP7) and the Baerveldt glaucoma implant. This was a multicenter, prospective, randomized, controlled clinical trial with several relevant publications which have been cited here.
Barton K, Gedde SJ, Budenz DL, Feuer WJ, Schiffman J. The Ahmed Baerveldt Comparison Study methodology, baseline patient characteristics, and intraoperative complications. Ophthalmology. 2011 Mar;118(3):435-42. PubMed PMID: 20932581; NIHMSID: NIHMS233027; PubMed Central PMCID: PMC3020244.
Budenz DL, Barton K, Feuer WJ, Schiffman J, Costa VP, et al. Treatment outcomes in the Ahmed Baerveldt Comparison Study after 1 year of follow-up. Ophthalmology. 2011 Mar;118(3):443-52. PubMed PMID: 20932583; NIHMSID: NIHMS233028; PubMed Central PMCID: PMC3020266.
Budenz DL. Barton K. Gedde SJ. Feuer WJ. Schiffman J. et al. Five-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology. 2015 Feb;122(2):308-16. PubMed PMID: 25439606; NIHMSID: NIHMS625185; PubMed Central PMCID: PMC4306613.
Budenz DL, Feuer WJ, Barton K, Schiffman J, Costa VP, et al. Postoperative Complications in the Ahmed Baerveldt Comparison Study During Five Years of Followup. Am J Ophthalmol. 2016 Mar;163:75-82.e3. PubMed PMID: 26596400; NIHMSID: NIHMS746915; PubMed Central PMCID: PMC4901387.
- 2. Comparison of Polypropylene and Silicone Ahmed Glaucoma Valves. This study was conducted to evaluate and compare clinical outcomes after implantation of the silicone plate Ahmed
{7}------------------------------------------------
Glaucoma Valve Model FP7 (AGV-FP7) and the polypropylene plate Ahmed Glaucoma Valve Model S2 (AGV-S2). This was a prospective, multicenter, comparative study.
Ishida K. Netland PA. Costa VP. Shiroma L. Khan B. et al. Comparison of polypropylene and silicone Ahmed Glaucoma Valves. Ophthalmology. 2006 Aug;113(8):1320-6. PubMed PMID: 16877071.
- 3. Comparison of silicone and polypropylene Ahmed glaucoma valves: two-year follow-up. The purpose of this study was to compare the safety and efficacy of the silicone Ahmed glaucoma drainage device (AGV-FP7) and the polypropylene Ahmed glaucoma drainage device (AGV-S2). This study was a single surgeon, retrospective, consecutive case series.
Mackenzie PJ, Schertzer RM, Isbister CM. Comparison of silicone and polypropylene Ahmed glaucoma valves: two-year follow-up. Can J Ophthalmol. 2007 Apr;42(2):227-32. PubMed PMID: 17392844.
# Substantial Equivalence
The change from a rigid plate material (AGV-S2) to a softer plate material in the AGV-FP7 does not raise new questions of safety and efficacy. There is no change in intended use and clinical studies have shown similar efficacy and safety profiles compared predicate devices. In summary, the Ahmed™ Glaucoma Valve Model FP7 described in this submission is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.