K212847 · Pentavest Holdings Sdn Bhd · LZA · Jan 6, 2022 · General Hospital
Device Facts
Record ID
K212847
Device Name
Penta Glove
Applicant
Pentavest Holdings Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jan 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
Blue Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Blue Nitrile Examination Gloves Powder Free are disposable, non-powdered patient examination gloves made from acrylonitrile-butadiene copolymer dispersion. Used in clinical or day-care settings by healthcare examiners to provide a barrier against contamination between the examiner and the patient. The device is worn on the hands; it is a single-use, passive barrier device. No electronic, mechanical, or software components are involved in its function.
Clinical Evidence
Bench testing only. No clinical data was required or provided. Performance was validated against ASTM D6319 (dimensions, physical properties, thickness), ASTM D5151 (water tightness), ASTM D6124 (residual powder), and ISO 10993 (biocompatibility: irritation, sensitization, cytotoxicity, systemic toxicity).
Technological Characteristics
Material: Acrylonitrile-butadiene copolymer (nitrile). Form factor: Disposable, powder-free, blue, finger-textured gloves. Standards: ASTM D6319-2019 (dimensions, physical properties, thickness), ASTM D5151-2019 (water tightness), ASTM D6124-06 (residual powder), ISO 10993 (biocompatibility). Energy source: None. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
JR MEDIC Blue Nitrile Examination Gloves Powder-free (K192333)
Related Devices
K142588 — Nitrile Powder Free Examination Glove (Blue) · Hartalega Sdn Bhd · Dec 18, 2014
K162491 — Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim (Blue) · Hartalega Sdn Bhd · Feb 8, 2017
K040791 — POLYMER COAT, POWDER FREE NITRILE EXAMINATION GLOVE (BLUE) · Pt. Smartglove Indonesia · Jun 3, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 6, 2022
Pentavest Holdings Sdn Bhd % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504
Re: K212847
Trade/Device Name: Penta Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 1, 2021 Received: September 7, 2021
Dear Manoj Zacharias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212847
Device Name
Penta Glove
Indications for Use (Describe)
Blue Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | <table><tr><td>☐ Residential Use (Part 1 CFR 201, Subpart D)</td><td>☑ Day-Care Use (Part 1 CFR 201, Subpart G)</td></tr></table> | ☐ Residential Use (Part 1 CFR 201, Subpart D) | ☑ Day-Care Use (Part 1 CFR 201, Subpart G) |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------|
| ☐ Residential Use (Part 1 CFR 201, Subpart D) | ☑ Day-Care Use (Part 1 CFR 201, Subpart G) | | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY As required by: 21CFR§807.92(c)
| 510(K) Owner's Name | Pentavest Holdings Sdn Bhd |
|---------------------|------------------------------------------------------------------------------------------------|
| Address | No. 9574-9578, Jalan PTB 2 ,<br>Kawasan Perindustrian Tangga,<br>Batu , 76400 Melaka, Malaysia |
| Phone | +601 22332689 |
| Fax | --- |
| E-mail | bjteng@pentavest.com.my |
| Contact Person | Teng Boon Joo |
| Designation | Managing Director |
| Contact Number | +601 22332689 |
| Contact Email | bjteng@pentavest.com.my |
| Date Submitted | 01 September 2021 |
### A. APPLICANT INFORMATION
### B. DEVICE IDENTIFICATION
| Name of the device | Blue Nitrile Examination Gloves Powder Free |
|-----------------------------------|---------------------------------------------|
| Product proprietary or trade name | PENTA GLOVE |
| Common or usual name | Exam Gloves |
| Classification name | Patient Examination Gloves |
| Device Classification | Class-1 |
| Product Code | LZA |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
#### C. PREDICATE DEVICE
| Predicate Device | JR Engineering & Medical Technologies (M) SDN.BHD |
|------------------|---------------------------------------------------|
| 510( K) Number | K192333 |
| Regulatory Class | 1 |
| Product code | LZA |
### D. DESCRIPTION OF THEDEVICE:
Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.
### E. INTENDED USE OF THE DEVICE:
Blue Nitrile Examination Gloves Powder Free is a disposable device intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.
{4}------------------------------------------------
## F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| Characteristics | Standards | Device Performance | | Remarks |
|-----------------------|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | | Predicate | Subject | |
| 510(K) Number | | K192333 | | ---- |
| Name of device | | JR MEDIC Blue Nitrile<br>Examination Gloves<br>Powder-free | Blue Nitrile<br>Examination Gloves<br>Powder Free | ---- |
| Dimensions | ASTMD 6319-2019 | Length Min 230 m Width<br>Min 95+/-10<br>mm(for medium size) | Length Min 230 mm<br>Width Min 95+/-10<br>mm(for medium size) | Same |
| Physical Properties | ASTMD 6319-2019 | Before Aging<br>Tensile Strength<br>min 14 Mpa<br>Ultimate Elongation<br>Min 500%<br>After Aging<br>Tensile Strength<br>min 14 Mpa<br>Ultimate Elongation<br>Min 400% | Before Aging<br>Tensile Strength min<br>14 Mpa<br>Ultimate Elongation<br>Min 500%<br>After Aging<br>Tensile Strength<br>min 14 Mpa<br>Ultimate Elongation<br>Min 400% | Same |
| Thickness | ASTMD 6319-2019 | Palm min 0.05 mm<br>Finger min 0.05 mm | Palm min 0.05 mm<br>Finger min 0.05 mm | Same |
| Powder Residue | ASTMD 6319-2019 | ≤2 mg/glove | ≤2 mg/glove | Similar |
| Biocompatibility | Primary Skin<br>Irritation-<br>ISO 10993-<br>10:2010(E) | Under the condition of<br>study not an irritant | Under the condition<br>of study not an<br>irritant | Same |
| | Dermal<br>Sensitization- ISO<br>10993-10:2010(E) | Under the conditions of<br>the study not a sensitizer | Under the conditions<br>of the study not a<br>sensitizer | Same |
| | In vitro cytotoxicity<br>ISO10993-5<br>:2009(E) | Under the conditions<br>of the study, cytotoxic | Under the conditions<br>of the study,<br>cytotoxic | Same |
| | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E) | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity concern | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity<br>concern | Same |
| | Bacterial Endotoxin<br>test USP 42<85> | No data available | <20EU/pair of gloves | ---- |
| Characteristics | Standards | Device Performance | | Remarks |
| | | Predicate | Current | |
| Water Tight (1000 ml) | ASTM D5151-2019 | Passes AQL-1.5 | Passes AQL-1.5 | Similar |
| Intended use | | JR MEDIC Blue Nitrile<br>Examination Gloves<br>Powder-free is<br>disposable devices<br>intended for medical<br>purpose that are won on<br>the examiner's hand to<br>prevent contamination<br>between patient and<br>examiner. | Blue Nitrile<br>Examination Gloves<br>Powder free is a<br>disposable device<br>intended for medical<br>purpose that are won<br>on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner. | Similar |
| Material | ASTMD 6319-2019 | Nitrile | Nitrile | Same |
| Color | - | Blue | Blue | Same |
| Texture | - | Finger Texture | Finger Texture | Same |
| Size | ASTMD<br>6319-2019 | Extra Small, Small,<br>Medium, Large, Extra<br>Large | Extra Small, Small,<br>Medium, Large, Extra<br>Large | Same |
| Single Use | Medical Glove<br>Guidance Manual<br>- Labeling | Single Use | Single Use | Same |
| Manufacturer(s) | | JR Engineering &<br>Medical<br>Technologies<br>(M) SDN.BHD.<br>Malaysia | Pentavest Holdings Sdn<br>Bhd | ------ |
{5}------------------------------------------------
There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6319.
### NON-CLINICAL TESTING SUMMARY
## PERFORMANCE DATA
| Test Method | Purpose | Acceptance Criteria | Result |
|--------------------------------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>length of the gloves | Min 230 mm for all sizes | X-Small:- 246 mm<br>Small:- 246 mm<br>Medium:- 248 mm<br>Large:- 248 mm<br>X-Large:- 250 mm |
| ASTM D6319-2019 Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To determine the<br>width of the gloves | X-small:- $70\pm 10$ mm<br>Small:- $80\pm 10$ mm<br>Medium:- $95\pm 10$ mm<br>Large:- $110\pm 10$ mm<br>X-Large:- $120\pm 10$ mm | X-small- 68 mm<br>Small:- 80 mm<br>Medium:- 92 mm<br>Large:- 105 mm<br>X-Large:- 115 mm |
{6}------------------------------------------------
| Test Method | Purpose | Acceptance Criteria | Result |
|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D6319-2019 Standard Specification for Nitrile Examination Gloves for Medical Application | To determine the thickness of the gloves | Palm 0.05 mm min<br>Finger 0.05 mm min for all sizes | Size<br>X-Small 0.12mm 0.20mm<br>Small 0.12mm 0.20mm<br>Medium 0.12mm 0.20mm<br>Large 0.12mm 0.20mm<br>X-Large 0.12mm 0.20mm |
| ASTM D6319-2019 Standard Specification for Nitrile Examination Gloves for Medical Application | To Determine the physical properties-<br>Tensile strength | Before Ageing<br>Tensile Strength 14Mpa Min for all sizes<br>After Ageing<br>Tensile Strength 14Mpa Min for all sizes | Size<br>Before ageing<br>X-Small 18.45Mpa 17.54Mpa<br>Small 18.54Mpa 17.56Mpa<br>Medium 18.62Mpa 17.68Mpa<br>Large 18.67Mpa 17.75Mpa<br>X-Large 18.72Mpa 17.80Mpa |
| ASTM D6319-2019 Standard Specification for Nitrile Examination Gloves for Medical Application | To Determine the physical properties-<br>Ultimate Elongation | Before Ageing<br>Ultimate Elongation 500% Min for all sizes<br>After Ageing<br>Ultimate Elongation 400% Min for all sizes | Size<br>Before ageing<br>X-Small 686% 659%<br>Small 690% 665%<br>Medium 694% 668%<br>Large 702% 670%<br>X-Large 705% 674% |
| Test Method | Purpose | Acceptance<br>Criteria | Result |
|-----------------------------------------------------------------------------------------------------|------------------------------------------------------|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D5151-2019 Standard<br>Test Method for Detection of<br>Holes in Medical Gloves | To determine the<br>holes in the gloves | AQL 1.5 | Gloves Passes AQL 1.5 |
| ASTM D6124-06 (Reapproved<br>2017) Standard Test Method<br>for Residual Powder on<br>Medical Gloves | To determine the<br>residual powder in<br>the gloves | 2 Mg/Glove Max | Size<br>X-small<br>Small<br>Medium<br>Large<br>X-Large<br><br>Residual Powder<br>Content<br>0.21mg/glove<br>0.21mg/glove<br>0.22 mg/glove<br>0.22 mg/glove<br>0.22 mg/glove |
{7}------------------------------------------------
#### BIO-COMPATIBILITY DATA
| Test Method | Purpose | Acceptance Criteria | Result |
|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| ISO 10993-10 Biological<br>Evaluation of Medical Devices<br>Test for Irritation and Skin<br>Sensitization. Test done for<br>irritation. | To determine the potential of<br>the material under test to<br>produce dermal irritation in<br>Rabbits | Under the condition<br>of study not an<br>irritant | Under the condition of<br>study not an irritant |
| ISO 10993-10 Biological<br>Evaluation of Medical Devices<br>Test for Irritation and Skin<br>Sensitization. Test done Skin<br>sensitization. | To determine the skin<br>sensitization potential of the<br>material both in terms of<br>induction and elicitation in<br>Guinea Pig. | Under the<br>conditions of the<br>study not a<br>sensitizer | Under the conditions of<br>the study not a<br>sensitizer |
| ISO 10993-5:2009 biological<br>evaluation of medical devices -<br>part 5, tests for in vitro<br>cytotoxicity. | To evaluate the in vitro<br>cytotoxic potential of the test<br>item (both inner and outer<br>surface) Extracts in L-929<br>mouse fibroblasts cells using<br>elution method. | Under the<br>conditions of study<br>non cytotoxic | Under the conditions of<br>the study cytotoxic. |
| ISO 10993-11:2017 biological<br>evaluation of medical devices -<br>part 11, tests for systemic<br>toxicity. | To determine the acute<br>systemic toxicity potential of<br>the test item extracts (both<br>inside and outer surfaces) in<br>Swiss Albino mice. | Under the<br>conditions of study<br>the device extracts<br>do not pose a<br>systemic toxicity<br>concern | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity<br>concern |
| Bacterial Endotoxin test<br>USP 42<85> | To determine the Bacterial<br>Endotoxin limit in the glove | NMT 20 EU/pair of<br>gloves | <20 EU/pair of gloves |
## G. Clinical Testing Summary
Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(K) process.
## H. CONCLUSION
The conclusions drawn from the non clinical test demonstrate that the subject device in 510(K) submission, Blue Nitrile Examination Gloves Powder Free is as effective, and performs as well as than the legally marketed predicate device K192333.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.