Powder free nitrile examination glove- black, non sterile
K212801 · Hong Seng Gloves Sdn Bhd · LZA · Jan 9, 2022 · General Hospital
Device Facts
Record ID
K212801
Device Name
Powder free nitrile examination glove- black, non sterile
Applicant
Hong Seng Gloves Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jan 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
A patient examination glove is a disposable device made of Nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable nitrile rubber examination glove; ambidextrous; non-sterile; black color. Used in clinical settings by healthcare providers to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Manufactured via nitrile rubber molding; surface treatment applied to inner surface to facilitate donning without powder. Device is single-use. Healthcare providers wear the glove during patient examinations; provides protection against contamination. Benefits include reduced risk of pathogen transmission during routine medical procedures.
Clinical Evidence
No clinical data. Bench testing only. Testing included physical properties (ASTM D412), dimensions (ASTM D3767), watertightness (ASTM D5151), residual powder (ASTM D6124), and biocompatibility (ISO 10993-10, ISO 10993-5, ISO 10993-11).
Technological Characteristics
Material: Nitrile rubber. Standards: ASTM D6319 (physical properties/dimensions), ASTM D6124 (residual powder), ASTM D5151 (watertightness). Design: Ambidextrous, textured finger, powder-free. Dimensions: 230mm minimum length, various palm widths. Biocompatibility: ISO 10993-10 (irritation/sensitization), ISO 10993-5 (cytotoxicity), ISO 10993-11 (acute systemic toxicity).
Indications for Use
Indicated for use as a disposable, non-sterile, ambidextrous nitrile rubber glove worn by medical examiners on hands or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Powder Free Nitrile Examination Glove, Black Color (K190942)
Related Devices
K182528 — Nitrile Glove Powder Free Black · Haining Medical Products Co., Ltd. · Nov 15, 2018
K161099 — Powder Free Nitrile Examination Gloves, Black · Pt. Maja Agung Latexindo · Feb 23, 2017
K213934 — Pro Guard Nitrile Powder Free Examination Glove · Medical Glove Co., Ltd. · Aug 6, 2022
K222720 — Blue Nitrile Powder Free Patient Examination Glove, Non Sterile · Safeskin Retailing (Hk) Limited · Dec 3, 2022
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 9, 2022
Hong Seng Gloves Sdn Bhd % Michael Van Der Woude U.S. Agent Emergo Global Representative LLC 2500 Bee Cave Road, Building 1 Suite 300 Austin, Texas 78746
Re: K212801
Trade/Device Name: Powder free nitrile examination glove- black, non sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: December 2, 2021 Received: December 7, 2021
Dear Michael Van Der Woude:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212801
Device Name POWDER FREE NITRILE EXAMINATION GLOVE - BLACK, NON STERILE
Indications for Use (Describe)
A patient examination glove is a disposable device made of Nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1.0 Sponsor:
2.0
| Company Name: | Hong Seng Gloves Sdn Bhd |
|------------------|--------------------------------------------------------------------------------------------|
| Company Address: | Lot 97, Jalan 10, Kawasan Perusahaan Bakar Arang, 08000<br>Sungai Petani, Kedah, Malaysia. |
| Submitter: | |
| Contact Person: | Mr. Ho Chia Yao |
| Company Name: | Hong Seng Gloves Sdn Bhd |
| Company Address: | Lot 97, Jalan 10, Kawasan Perusahaan Bakar Arang, 08000<br>Sungai Petani, Kedah, Malaysia. |
| Phone No.: | +604-4211555 |
| Fax No.: | +604-4211555 |
Date of Summary Prepared: 20th July 2021 (revised on 8th January 2022)
#### 3.0 Identification of the subject device:
| Trade Name | : Powder Free Nitrile Examination Glove - Black, Non- Sterile |
|-----------------------|---------------------------------------------------------------|
| Common Name | : Patient Examination Gloves |
| Classification Name | : Patient Examination Gloves |
| Device Classification | : 1 |
| Regulation Number | : 21 CFR 880.6250 |
| Product Code | : LZA. |
#### 3.0 Predicate Device:
### K190942
Disposable Powder Free Nitrile Examination Glove, Black Color Company: Ever Growth (Vietnam) Co. Ltd.
#### 4.0 Description of The Device:
Powder Free Nitrile Examination Glove - Black, Non-Sterile meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250.
The powder free nitrile examination glove is manufactured from Nitrile rubber. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the qloves without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand and it is a single use device.
| Design | : Ambidextrous (i.e., fit either hand) |
|-------------|--------------------------------------------------|
| Finish | : Texture Finger |
| Performance | : See Section 7 Summary of Non-Clinical Testing. |
### Storage:
The product is kept away from direct sun and fluorescent lighting and stored in an environment with temperature not exceeding 40-degree C.
{4}------------------------------------------------
### Information For "Powder Free" Claim:
The finished powder free gloves meet ASTM D6319 requirements and are tested in according to ASTM D6124 method for powder measurement (less than 2mg per glove).
### Glove Size and dimension:
Measurement is done as per ASTM D6319. Length is measured from the tip of the middle finger to the outside edge of the cuff. Width is measured at a level between the base of the index finger and the base of the thumb.
| Size | Palm Width (mm) | Length (mm) |
|---------|-----------------|-------------|
| X-Small | 70±10 | Minimum 230 |
| Small | 80±10 | Minimum 230 |
| Medium | 95±10 | Minimum 230 |
| Large | 110±10 | Minimum 230 |
| X-Large | 120±10 | Minimum 230 |
#### 5.0 Indication for use:
A patient examination glove is a disposable device made of Nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### 6.0 Technological Characteristics Comparison of the Device:
Provided below is the technological comparison of the subject device vs the predicate device as shown in Table 1.
{5}------------------------------------------------
### Table 1
| | | DEVICE PERFORMANCE | | |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| CHARACTERIST<br>ICS | STANDARDS | PREDICATE | SUBJECT DEVICE | COMPARISON<br>ANALYSIS |
| | | BLACK | BLACK | |
| 510(k) Number | - | K190942 | K212801 | |
| Manufacturer(s) | - | Ever Growth Enterprise<br>Corporation | Hong Seng Gloves Sdn Bhd | Same |
| Material | ASTM D6319 | Nitrile | Nitrile | Same |
| Color | - | Black | Black | Same |
| Physical<br>Properties | ASTM D6319 | | | |
| Before Aging<br>Tensile Strength:<br>Ultimate<br>Elongation: | | 14Mpa, min<br>500% min | 29.7Mpa<br>585% | Different but<br>within the ASTM<br>standard |
| After Aging<br>Tensile Strength:<br>Ultimate<br>Elongation: | | 14Mpa, min<br>400% min | 33.1Mpa<br>554% | Different but<br>within the ASTM<br>standard |
| Physical<br>Dimensions: | | | | |
| Glove Length: | | Complies with ASTM D6319 -<br>230mm minimum | Complies with ASTM D 6319 -<br>230mm minimum. | Similar |
| Glove<br>Thickness:<br>- Finger<br>- Palm | ASTM D6319 | 0.05mm min<br>0.05mm min | 0.10mm<br>0.07mm | Different but<br>within the ASTM<br>standard |
| Glove Palm<br>Width:<br>X-Small<br>Small<br>Medium<br>Large<br>X-Large | | Complies with ASTM D6319 | Complies with ASTM D6319 | Similar |
| Powder Free | ASTM D6124 | < 2mg per glove | 0.20 mg/glove | Different but<br>within the ASTM<br>standard |
| | Primary Skin<br>Irritation - ISO<br>10993-10:2010 (E)<br>& Consumer Product<br>Safety Commission<br>Title 16. Chapter II,<br>Part 1500 | Passes | The test material did not cause an<br>irritant response. The Primary<br>Irritant Response Category is<br>deemed 'Negligible' | Similar |
| Biocompatibility | Dermal<br>Sensitization- ISO<br>10993-10: 2010 (E)<br>& Consumer Product<br>Safety Commission,<br>Title 16, Chapter II,<br>Part 1500.3 (c) (4) | Passes | The test material did not produce<br>a skin sensitization effect in the<br>guinea pigs. | Similar |
| | Cytotoxicity - MEM<br>Elution, ISO 10993-<br>5: 2009 (E) | Passes | The test material demonstrated a<br>cytotoxic effect under the<br>condition of this study. Additional<br>test i.e. Acute Systemic Toxicity<br>was tested. | Different - but<br>additional test of<br>Acute Systemic<br>Toxicity is<br>conducted. |
| | DEVICE PERFORMANCE | | | |
| CHARACTERISTICS | STANDARDS | PREDICATE | SUBJECT DEVICE | COMPARISON<br>ANALYSIS |
| | | BLACK | BLACK | |
| Biocompatibility | Acute Systemic<br>Toxicity, ISO 10993-<br>11:2017 (E) | Not Applicable | The test item did not induce<br>any systemic toxicity in Swiss<br>albino mice. | Different. |
| Watertight (1000ml) | ASTM D5151:2019 | In accordance with ASTM<br>D6319-10 and ASTM D5151-06<br>(reapproved 2011), G-1, AQL<br>2.5 | Gloves passed AQL 1.5 | Different, but<br>within the ASTM<br>standard. |
| Indication for use | - | The Nitrile Powder Free patient<br>examination glove is a non-<br>sterile disposable device<br>intended for medical purpose<br>that is worn on the examiner's<br>hands or finger to prevent<br>contamination between patient<br>and examiner. | A patient examination glove is a<br>disposable device made of<br>nitrile rubber intended for<br>medical purposes that is worn<br>on the examiner's hand or<br>finger to prevent contamination<br>between patient and examiner. | Same |
| Size | Medical Glove<br>Guidance Manual -<br>Labeling | X Small<br>Small<br>Medium<br>Large<br>X Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Same |
| Single use | Medical Glove<br>Guidance Manual -<br>Labeling | Yes | Single Use | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
There are no significant differences between the two products and they are the same or similar in terms of intended use, materials design, physical properties, thickness and biocompatibility test.
#### 7.0 Summary of Non-Clinical Testing
The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.
{9}------------------------------------------------
| Test<br>Method | Standard | Purpose of Testing | Acceptance Criteria | | Results | | Status |
|---------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|----------------------------------|--------------|--------------|-------------|--------|
| | | | Before aging | After aging | Before aging | After aging | |
| Physical Properties | ASTM D412<br>(Standard Test<br>Method for<br>Vulcanized Rubber<br>and Thermoplastic<br>Elastomers-Tension) | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile strength<br>Min 14.0 MPa | Min 14.0 MPa | 29.7Mpa | 33.1Mpa | Pass |
| | | | Ultimate elongation<br>Min 500% | Min 400% | 585% | 554% | Pass |
{10}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Glove<br>Size | Acceptance Criteria | Results | Status | |
|----------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|----------------------------------|----------------------------------|----------------------|------|
| | | To measure the<br>length, width and<br>thickness<br>of<br>glove | X-Small | Length<br>Min 240 mm | Length<br>250 mm | Pass | |
| | | | | Width<br>$70 \pm 10$ mm | Width<br>78.0 mm | Pass | |
| | | | | Thickness<br>Finger - min 0.05mm | Thickness<br>0.10 mm | Pass | |
| | | | | Palm - min 0.05mm | 0.07 mm | Pass | |
| | ASTM D3767<br>Standard<br>Practice for<br>Rubber—<br>Measurement<br>of Dimensions | To measure the<br>length, width and<br>thickness<br>of<br>glove | Small | Length<br>Min 240 mm | Length<br>250 mm | Pass | |
| | | | | | Width<br>$80 \pm 10$ mm | Width<br>88.0 mm | |
| Dimension | | | | | Thickness<br>Finger - min 0.05mm | Thickness<br>0.10 mm | |
| | | | | | | Palm - min 0.05mm | |
| | | | To measure the<br>length, width and<br>thickness<br>of<br>glove | Medium | Length<br>Min 240 mm | Length<br>250 mm | |
| | | | | | Width<br>$95 \pm 10$ mm | Width<br>98.0 mm | |
| | | | | | Thickness<br>Finger - min 0.05mm | Thickness<br>0.10 mm | |
| | | | | | Palm - min 0.05mm | 0.07 mm | |
| | | | To measure the<br>length, width and<br>thickness<br>of<br>glove | Large | Length<br>Min 240 mm | Length<br>249 mm | |
| | | | | | Width<br>$110 \pm 10$ mm | Width<br>108 mm | |
| | | | Thickness | Finger – min 0.05mm | Thickness | 0.10 mm | Pass |
| | | | | Palm - min 0.05mm | | 0.07 mm | |
| | | Length | | Min 240 mm | Length | 250 mm | Pass |
| | | X-Large | Width | $120 \pm 10$ mm | Width | 118 mm | Pass |
| | | | Thickness | Finger – min 0.05mm | Thickness | 0.10 mm | Pass |
| | | | | Palm - min 0.05mm | | 0.07 mm | |
{11}------------------------------------------------
{12}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|----------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holes<br>in Medical Gloves) | To detect holes<br>that leak water<br>and thereby<br>compromise the<br>usefulness of the<br>glove. | Sample size: 500 pcs<br>Inspection level: G1<br>AQL: 1.5, Acceptance No. 10 | The batch size for this<br>sampling is 150,001 to<br>500,000. Hence,<br>according to the single<br>sampling plan GI, the<br>sample to be drawn is<br>under code M<br>equivalent to 315<br>pieces with accept 10<br>and reject 11 to be<br>accepted under AQL<br>1.5. During the test, 0<br>piece was found with<br>leaks. Hence it falls<br>within the acceptance<br>criteria. | Pass |
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|--------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------|-------------------------------------------------------------------------|--------|
| Residual<br>Powder | ASTM D6124<br>(Standard Test<br>Method for Residual<br>Powder on Medical<br>Gloves) | To determine the<br>amount of<br>residual powder<br>and non-powder<br>solids found on<br>gloves. | Less than 2 mg per glove | Sample size : 5 pcs<br>Requirement : <2mg/glove<br>Result :0.20mg/glove | Pass |
{13}------------------------------------------------
| Test Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|------------------|---------------------------------------------------|-------------------------------------------------------------------------|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Biocompatibility | Primary Skin Irritation - ISO 10993-10:2010 | To assess the potential of the material producing skin | Non-Irritant | The test material did not cause an irritant response. The Primary Irritant Response Category is deemed 'Negligible' | Pass |
| | Dermal Sensitization- ISO 10993-10: 2010 | irritation and skin sensitization | Non-Irritant | The test material did not produce a skin sensitization effect in the guinea pigs. | Pass |
| | Cytotoxicity - MEM Elution, ISO 10993-5: 2009 (E) | To assess the in vitro cytotoxicity of the material | Non-Toxicity | The test material demonstrated a cytotoxic effect under the condition of this study. Additional test i.e. Acute Systemic Toxicity was tested. | Fail |
| | Acute Systemic Toxicity, ISO 10993-11:2017 (E) | To assess the potential of material to cause adverse systemic reactions | Non-Toxicity | The test item did not induce any systemic toxicity in Swiss albino mice. | Pass |
{14}------------------------------------------------
#### 8.0 Summary of Clinical Testing:
No clinical study is included in this submission.
#### 9.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrate that the subject Powder Free Nitrile Examination Glove - Black, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K190942.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.