Disposable Medical Surgical Mask

K212726 · Hubei Huaqiang High-Tech Co., Ltd. · FXX · Mar 23, 2022 · General, Plastic Surgery

Device Facts

Record IDK212726
Device NameDisposable Medical Surgical Mask
ApplicantHubei Huaqiang High-Tech Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateMar 23, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

Device Story

Disposable, flat-pleated surgical mask; 3-layer construction (polypropylene spunbond outer/inner layers, polypropylene melt-blown middle layer); secured via elastic ear loops (nylon/spandex); includes internal pliable nose piece (polypropylene/metal wire). Used in clinical settings for infection control; worn by healthcare personnel and patients to provide barrier protection against blood, body fluids, and particulates. Device is single-use, sterile, and disposable. Provides ASTM F2100 Level 2 barrier performance.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100 Level 2 requirements: Fluid resistance (120 mmHg), particulate filtration efficiency (≥98% at 0.1 µm), bacterial filtration efficiency (≥99%), differential pressure (<5.8 mmH2O/cm²), and flammability (Class 1). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-7 (EtO/ECH residuals). Shelf-life validated via ASTM F1980 (2 years).

Technological Characteristics

3-layer non-woven polypropylene construction; flat-pleated design; elastic ear loops; internal nose piece. ASTM F2100 Level 2 compliant. Sterilized via Ethylene Oxide (EO) per ISO 11135. Biocompatibility per ISO 10993 series. No electronic or software components.

Indications for Use

Indicated for use by healthcare personnel and patients to protect against the transfer of microorganisms, body fluids, and particulate material in infection control settings.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 23, 2022 Hubei Huaqiang High-Tech Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District, Beijing 102401 China Re: K212726 Trade/Device Name: Disposable Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 8, 2022 Received: March 11, 2022 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K212726 Device Name Disposable Medical Surgical Mask #### Indications for Use (Describe) The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile. Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary K212726 This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. - 1. Date of Preparation: 2022/03/23 - 2. Sponsor Identification Hubei Huaqiang High-Tech CO., LTD. No.499 DongLin Road, Yichang, Hubei Province, 443000, China Contact Person: Jing Yang Position: Vice Director of Tech Center Tel: +86-13545751006 Fax: +86-717-6347336 Email: 409281039@gq.com - 3. Designated Submission Correspondent Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China Tel: +86-18910677558 Fax: +86-10-56335780 Email: Ray.Wang@believe-med.com - 4. Identification of Proposed Device Trade Name: Disposable Medical Surgical Mask Common Name: Mask, Surgical Regulatory Information Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 878.4040 Review Panel: General Hospital {4}------------------------------------------------ Indication for use Statement: The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile. #### Device Description The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile. The proposed device(s) are Blue color, and Flat Pleated type mask(s), utilizing Ear Loops to secure to the users head and a pliable Nose Piece for fitting the facemask around the nose. The proposed device(s) are manufactured with three layers, the inner and outer layers are made of polypropylene spunbond fabric, and the middle layer is made of polypropylene melt-blown fabric. The Medical Surgical Masks-Sterile is held in place over the users' mouth and nose by two elastic ear loops affixed to the facemask. The elastic ear loops are made with nylon or spandex. The nose piece contained in the proposed device(s) is contained within the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polypropylene and metal wire. The proposed device(s) are sold sterile and are intended to be single-use, disposable devices. - 5. Identification of Predicate Device(s) 510(k) number: K202904 Device Name: Surgical Face Mask Manufacturer: Jiangxi Feilikang Medical Technology Co., Ltd. #### Technological Characteristics Comparison 6. #### Table 1 General Comparison | ITEM | Proposed Device K212726 | Predicate Device K202904 | Remark | | |---------------------------------------------|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------| | | ASTM F2100 Level 2 | ASTM F2100 Level 2 | | | | Intended Use /<br>Indication for Use | | The Disposable Medical Surgical Mask<br>is intended to be worn to protect both<br>the patient and healthcare personnel<br>from the transfer of microorganisms,<br>body fluids, and particulate material.<br>The Disposable Medical Surgical Mask<br>is intended for use in infection control<br>practices to reduce the potential<br>exposure to blood and body fluids. This<br>is a single-use, disposable device(s),<br>provided sterile. | Surgical Face Mask is intended for<br>single use by operating room personnel<br>and other general healthcare workers to<br>protect both patients and healthcare<br>workers against transfer of<br>microorganisms, blood and body fluids,<br>and particulate materials. | Similar | | Basic Design | | Ear Loops, Flat Pleated, 3 layers | Ear Loop, Flat-pleated, 3 layers | SAME | | Materials | Outer Facing<br>Layer | polypropylene spunbond fabric | Polypropylene Non-Woven Fabric | Differences<br>addressed by<br>performance | | | Middle Layer | polypropylene melt-blown fabric | Polypropylene Melt-Blown Fabric | | | | Inner Facing Layer | polypropylene spunbond fabric | Polypropylene Non-Woven Fabric | and | | | Nose Piece | polypropylene and metal wire | Polypropylene and Iron Wire | biocompatibility | | | Ear Loops | nylon or spandex | Nylon and Spandex | testing | | | | | | | | | | | | | | Color | | Blue | Blue | SAME | | Dimension (Length) | | $17.5 \pm 5\%$ cm | 17.5 cm | SAME | | Dimension (Width) | | $9.5 \pm 5\%$ cm | 9.5 cm | SAME | | OTC use | | Yes | Yes | SAME | | Single Use | | Yes | Yes | SAME | | Sterile | | Sterile | Sterile | SAME | | Sterilization method | | EO | EO | SAME | | Sterilization residues | | EtO and ECH residues meet the<br>requirements of ISO 10993-7 | EtO and ECH residues meet the<br>requirements of ISO 10993-7 | SAME | | Shelf Life | | 2 years | Unknown | Difference<br>addressed<br>by<br>Shelf-Life<br>validation | | Packaging | | 10 pcs/bag | Unknown | SAME | | Packaging<br>Configuration and<br>materials | | Medical dialysis paper (Tyvek) and PP<br>(polyethylene and polypropylene) film. | | | {5}------------------------------------------------ ## Table 2 Performance Characteristic Comparison | ITEM | Proposed Device<br>K212726 | Predicate Device<br>K202904 | ASTM F2100<br>Requirements | Remark | |-----------------------------------------------|----------------------------|-----------------------------|----------------------------|--------------------------------------| | ASTM F2100 Level | Level 2 | Level 2 | Level 2 | SAME | | Fluid Resistance<br>Performance ASTM<br>F1862 | 120 mmHg | 120 mmHg | 120 mmHg | Both meet<br>requirements.<br>No new | | Particulate Filtration | ≥98% | Average 98.74% at | ≥ 98% | issues of | {6}------------------------------------------------ | Efficiency ASTM<br>F2299 | | 0.1 µm | | safety or<br>effectiveness | |-------------------------------------------------------------------------|----------------|----------------------|-----------------|----------------------------| | Bacterial Filtration<br>Efficiency ASTM<br>F2101 | ≥99% | Average 99.65% | ≥ 98% | | | Differential Pressure<br>(Delta P)<br>EN 14683-2019<br>+AC:2019 Annex C | < 5.8mmH2O/cm² | Average 4.6mmH2O/cm² | < 6.0 mmH2O/cm2 | | | Flammability<br>16 CFR 1610 | Class 1 | Class 1 | Class 1 | SAME | Table 3 Biocompatibility Comparison | ITEM | Proposed Device K212726 | Predicate Device K202904 | Remark | |---------------|-------------------------|--------------------------|--------| | Cytotoxicity | No Cytotoxicity | No Cytotoxicity | SAME | | Irritation | No Irritation | No Irritation | SAME | | Sensitization | No Sensitization | No Sensitization | SAME | #### 7. Non-Clinical Test Conclusion The test results demonstrated that the proposed device complies with the following standards: - > ISO 11135-2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices - > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices – Part 5: Tests For In Vitro Cytotoxicity - > ISO 10993-7: 2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals - ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests For Irritation > And Skin Sensitization. - > ASTM F2100-19, Standard Specification For Performance Of Materials Used In Medical Face Masks. - > ASTM F1862-17, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity) - > EN 14683-2019+AC:2019 Annex C, Medical face masks – Requirements and test methods; - ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency A (Bfe) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus; - > ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres; {7}------------------------------------------------ | Test Method | Purpose | Acceptance Criteria | Results | |---------------------|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | ASTM F1862 | Resistance to penetration by synthetic blood | 120 mm Hg | 120 mm Hg | | ASTM F2299 | Sub-micron particulate filtration efficiency at 0.1 micron | ≥ 98% | ≥ 98% | | ASTM F2101 | Bacterial Filtration Efficiency | ≥ 98% | ≥ 99% | | EN 14683<br>Annex C | Differential Pressure | < 6.0 mm H2O/cm² | < 5.7 mm H2O/cm² | | 16 CFR 1610 | Flammability | Class 1 | Class 1 | | ISO 10993-10 | Irritation<br><br>Sensitization | No irritation effect<br><br>No sensitization effect | Under the conditions of the study, no irritation effect<br><br>Under the conditions of the study, no sensitization effect | | ISO 10993-5 | Cytotoxicity | No cytotoxicity effect | Under the conditions of the study, no cytotoxicity effect | | ISO 10993-7 | EtO and ECH Residual | The average daily dose of EO to patient shall not exceed 4 mg;<br>The average daily dose of ECH to patient shall not exceed 9 mg; | EO average daily dose (mg/d) < 0.006;<br>ECH average daily dose (mg/d) < 0.03 | | ASTM F1980 | Shelf-Life Validation | Meet the requirements of Level 2 barrier performance at the end of shelf-life claimed | Meet the requirements of Level 2 barrier performance after 2 years accelerated aging. The aging temperature is 60°C and the aging time is 56 days. | - > 16 CFR 1610, Standard for the Flammability of clothing textiles; - 8. Clinical Test Conclusion No clinical study is included in this submission. - 9. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device, K212726, Disposable Medical Surgical Mask, is as safe, as effective, and performs as well as or better than the legally marketed device, K202904, Surgical Face Mask by Jiangxi Felikang Medical Technology Log., Ltd.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%