Disposable Surgical Face Mask

K223823 · Efofex, Inc. · FXX · Mar 27, 2023 · General, Plastic Surgery

Device Facts

Record IDK223823
Device NameDisposable Surgical Face Mask
ApplicantEfofex, Inc.
Product CodeFXX · General, Plastic Surgery
Decision DateMar 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

This single-use surgical mask EFXPLY3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile.

Device Story

Disposable, non-sterile, flat-pleated surgical mask; three-layer construction (spun-bond polypropylene outer/inner layers, melt-blown polypropylene filter middle layer); secured via elastic polypropylene/spandex ear loops; includes galvanized wire/polyethylene nose bridge for fit. Used in clinical settings or general infection control to provide barrier protection against blood, body fluids, and particulates. Device is single-use; worn over mouth and nose. Provides physical barrier; performance verified against ASTM F2100 Level 2 and 3 standards.

Clinical Evidence

No clinical data. Bench testing only. Performance verified per ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101-19 (bacterial filtration), ASTM F2100-19 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Three-layer mask: spun-bond polypropylene (outer/inner), melt-blown polypropylene (middle). Nose bridge: galvanized wire with polyethylene coating. Ear loops: polypropylene and spandex. Dimensions: 175mm x 95mm. Standards: ASTM F2100, ASTM F1862, ASTM F2101, ASTM F2299, 16 CFR 1610, ISO 10993-5, ISO 10993-10. Non-sterile, single-use.

Indications for Use

Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text. March 27, 2023 Efofex, Inc. % Timothy Kania Official Correspondent mdi Consultants. Inc. 55 Northern Blvd, Suite 200 Great Neck. New York 11021 Re: K223823 Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 20, 2022 Received: December 21, 2022 Dear Timothy Kania: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) #### K223823 Device Name Disposable Surgical Mask Indications for Use (Describe) This single-use surgical mask EFXPLY3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary # Contact Details K223823 ## A. Applicant Details | Applicants Name | Efofex, Inc. | |--------------------|-----------------------------------------| | Applicants Address | 1300 N Dixieland Road, Rogers, AR 72758 | | Contact Person | Siddhant Doshi | | Phone Number | 917-213-2740 | | E-mail | sid@efofexinc.com | | Date Prepared | 12/14/22 | ## Official Correspondent | Contact Person | Tim Kania | |----------------|-----------------------------------------------------| | Company Name | mdi Consultants, Inc. | | Address | 55 Northern Blvd. Suite 200<br>Great Neck, NY 11021 | | Phone | 516-482-9001 | | Email | Tim@mdiconsultants.com | ## B. Device Information | Trade Name | Disposable Surgical Mask | |---------------------|--------------------------| | Common Name | Disposable Surgical Mask | | Model | EFX3PLYSMSK | | Classification | II | | Classification Name | Mask, Surgical | | Product Code | FXX | | Regulation No. | 21 CFR 878.4040 | ## C. Legally Marketed Primary Predicate Device | Trade Name | Surgical Face Mask, Model GFYY95 | |--------------|-------------------------------------------------------| | 510 (K) | K202761 | | Product Code | FXX | | Manufacturer | Guangdong Good Feeling Hygiene Material Tec Co., LTD. | {4}------------------------------------------------ #### D. Indications for use of the device This single-use surgical mask EFXPL Y3SMSK is intended to be worn to protect both the patient and health care personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of blood and body fluids. This is a single use, disposable device, non-sterile. #### E. Device description The device is a flat pleated type mask the outer layer being blue and the inner layer being white, utilizing ear loops to secure the mask in place, and a nose bridge is also incorporated for a proper fit around the nose. The device number is EFX3PLYSMSK. The device is manufactured with three layers, the inner and outer layers are made of spun bond polypropylene, and the middle layer is made of melt-blown polypropylene filter material. The model EFX3PLYSMSK device is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of polypropylene and spandex. The nose bridge contained within the device is placed between the layers of the surgical mask to allow the user to fit the surgical mask properly around their nose. The nose bridge is made up of galvanized with polyethylene. {5}------------------------------------------------ | DEVICE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | COMPARISON | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Manufacturer | Efofex, Inc<br>1300 N Dixieland Road<br>Rogers AR 72758 | Guangdong Good Feeling<br>Hygiene Material Tec Co. LTD.<br>10 Phoenix Av. Eco-Industrial<br>Pk. Guangdong Province, China | NA | | 510K | K223823 | K202761 | NA | | Classification | Class II Device, FXX<br>(21CFR878.4040) | Class II Device, FXX<br>(21CFR878.4040) | Similar | | Indications for<br>use | This single use surgical mask<br>EFX3PLY3SMSK is intended<br>to be worn to protect both<br>the patient and health care<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material.<br>These face masks are<br>intended for use in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This is<br>a single use, disposal device,<br>provided non-sterile. | This single use surgical mask<br>GFYY95 is intended to be worn<br>to protect both the patient and<br>health care personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood<br>and body fluids. This is a single<br>use, disposal device, provided<br>non-sterile. | Similar | | Description | Ear loops, flat pleated, three<br>layers | Ear loops, flat pleated, three<br>layers | Similar | | Outer layer | Spun-bond polypropylene | Spun-bond polypropylene | Similar | | Middle layer | Melt-blown polypropylene<br>filter | Melt-blown polypropylene<br>filter | Similar | | Inner layer | Spun- bound polypropylene | Spun- bound polypropylene | Similar | | Nose bridge | Galvanized wire with<br>polyethylene coating | Galvanized wire with<br>polyethylene coating | Similar | | Ear loops | Polypropylene & Spandex | Nylon & Spandex | Different¹ | | Color | Blue | Blue | Similar | | Length | 175 mm | 180mm | Different² | | Width | 95 mm | 95mm | Similar | | OTC use | Yes | Yes | Similar | | Sterility | No | No | Similar | | Use | Single use, Disposable | Single use, Disposable | Similar | | ASTM F2100<br>level | Level 2 & 3 | Level 3 | Similar3 | | Biocompatibility | ISO10993 | ISO10993 | Similar | - F. Technological Characteristics Comparison with primary predicate device {6}------------------------------------------------ #### Discussion of Similarities and Differences 1: Differences in Ear Loop material have no effect on the safety or efficacy of the device. The subject device was tested for Biocompatibility and passed. Please see table 3 below for a summary of the Biocompatibility testing. 2: Differences in Dimensions have no effect on the safety or efficacy of the device. Please see table 2 below for a summary of the performance testing performed on this device. 3: The Level 2 and Level 3 masks are comprised of the same material which has passed performance testing for Level 3 protection. This has no effect on the safety or performance of the device. #### G. Nonclinical test conclusion Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device meets the acceptance criteria for the following standards and the requirements stated in the guidance for industry and FDA staff; surgical mask-premarket notification ( 510 K ). ISO 10993-5: 2009 Biological evaluation of medical devices - part 5: Test for In vitro cytotoxicity ISO 10993-10: 2010 Biological Evaluation of Medical Devices – part 10: Test for irritation and skin sensitization ASTM F2100, Standard Specification for performance of Materials Used in Face Masks. ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood ( Horizontal Projection of Fixed Volume at A Known Velocity); EN 14683, Medical Face Masks -- Requirements and Test Methods; ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials Using a Biological Aerosol of Staphylococcus Aureus; ASTM F2299, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres; 16 CFR 1610, Standard for the Flammability of clothing textiles; {7}------------------------------------------------ | Table 2 – Performance Testing Level 3 | | | | |-----------------------------------------------------|---------------------------------------------------------------|------------------------------|--------| | Test Methodology | Proposed Device<br>EFX3PLYSMSK | Accepted Criteria<br>Level 3 | Result | | Fluid Resistance<br>Performance ASTM<br>F1862 | 95 of 96 samples<br>passed | 160mm Hg. | Pass | | Particulate Filtration<br>Efficiency<br>ASTM F2299 | 99.79% Filtration<br>average efficiency<br>across 96 samples | ≥98% | Pass | | Bacterial Filtration<br>Efficiency<br>ASTM F2101-19 | 99.8% Filtration<br>average efficiency<br>across 96 samples | ≥98% | Pass | | Differential Pressure<br>(Delta P) AMST<br>F2100-19 | 3.56 mm H2O/cm²<br>average pressure drop<br>across 96 samples | <6.0mm H2O /cm² | Pass | | Flammability<br>16 CFR 1610 | 96 of 96 samples did<br>not ignite | Class 1 | Pass | | Table 3 – Performance Testing Level 2 | | | | |-----------------------------------------------------|---------------------------------------------------------------|------------------------------|--------| | Test Methodology | Proposed Device<br>EFX3PLYSMSK | Accepted Criteria<br>Level 2 | Result | | Fluid Resistance<br>Performance ASTM<br>F1862 | 96 of 96 samples<br>passed | 120mm Hg. | Pass | | Particulate Filtration<br>Efficiency<br>ASTM F2299 | 99.79% Filtration<br>average efficiency<br>across 96 samples | ≥98% | Pass | | Bacterial Filtration<br>Efficiency<br>ASTM F2101-19 | 99.8% Filtration<br>average efficiency<br>across 96 samples | ≥98% | Pass | | Differential Pressure<br>(Delta P) AMST<br>F2100-19 | 3.56 mm H2O/cm²<br>average pressure drop<br>across 96 samples | <6.0mm H2O /cm² | Pass | | Flammability<br>16 CFR 1610 | 96 of 96 samples did<br>not ignite | Class 1 | Pass | {8}------------------------------------------------ | Table 4 – Biocompatibility Testing | | | | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------------|--------| | Item | Proposed Device | Accepted Criteria | Result | | Cytotoxicity | Under the conditions<br>of the study, the<br>subject device extract<br>did not show<br>potential toxicity to L<br>929 cells | No potential<br>cytotoxicity | Pass | | Irritation | Under the condition<br>of the study, the<br>device is non-irritating | Non-irritating | Pass | | Sensitization | Under the conditions<br>of the study, the<br>device is non<br>sensitizing | Non-sensitizing | Pass | - H. Clinical Test Conclusions No clinical tests is included in this submission - l. Conclusion Based on the nonclinical tests performed, the subject device is as safe, is effective, and performs as well or better than the legally marketed primary predicate device number K202761, manufactured by Guangdong Good Feeling Hygiene Material Tec Co. LTD, single use surgical mask.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...