The Incisive CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Incisive CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Device Story
Philips Incisive CT is a whole-body computed tomography system; utilizes continuously rotating x-ray tube, detector array, and gantry to acquire transmission data. System reconstructs data into cross-sectional images; includes signal analysis, display equipment, and patient support couch. Operated by clinicians in diagnostic imaging environments. Features include iterative reconstruction (iDose4), cardiac gating, and various clinical applications (e.g., Brain Perfusion, Lung Nodule Analysis, Vessel Analysis). Output consists of diagnostic images and quantitative data for clinical decision-making. Benefits include improved image quality, artifact reduction (O-MAR), and workflow efficiency. System supports advanced features like Precise Image (deep learning noise reduction), Precise Cardiac (motion compensation), and Precise Position (automated patient positioning).
Clinical Evidence
No clinical study required. Substantial equivalence established through non-clinical performance data, design verification, and design validation testing. System verification conducted against system requirement specifications (SRS) using manual/automated tests and analysis. Validation included workflow assessment and compliance with international consensus standards (IEC 60601 series, ISO 14971).
Technological Characteristics
Whole-body CT system with 72cm bore, 50cm scan FOV, 64/128 slice capability. Components: rotating x-ray tube, generator, detector array, gantry, patient couch, and console. Connectivity: DICOM compliant. Software: iterative reconstruction (iDose4), deep learning-based reconstruction (Precise Image), and automated positioning (Precise Position). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, IEC 62304, ISO 14971, NEMA XR 25/28/29.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages. Includes low-dose CT lung cancer screening for early detection of lung nodules in accordance with established professional or governmental protocols.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
April 27, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services seal. To the right of the symbol is a blue square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Philips Healthcare (Suzhou) Co., Ltd. % Shiguang An Regulatory Affairs Engineer No. 258, Zhongyuan Road, Suzhou Industrial Park Suzhou. Jiangsu 215024 CHINA
## Re: K212441
Trade/Device Name: Philips Incisive CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: March 16, 2022 Received: March 16, 2022
### Dear Shiguang An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K212441
Device Name Philips Incisive CT
Indications for Use (Describe)
The Incisive CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Incisive CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
| Type of Use (Select one or both, as applicable) | <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text 'K212441' in a simple, sans-serif font. The text is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, making the text easily readable.
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
# 510(k) Summary of Safety and Effectiveness
[As required by 21 CFR 807.92(c)]
| Date Prepared: | July 30, 2021 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, Zhongyuan Road, Suzhou Industrial Park,<br>Suzhou Jiangsu, CHINA, 215024<br>Establishment Registration Number: 3009529630 |
| Primary Contact Person: | Shiguang An<br>Regulatory Affairs Engineer<br>Phone: +86-24-28206367<br>E-mail: shiguang.an@philips.com |
| Secondary Contact Person | Erhong Wang<br>Senior Manager Regulatory Affairs<br>Phone: +86-512-67336804<br>E-mail: ErHong.WANG@philips.com |
| Device Name: | Philips Incisive CT |
| Classification: | Classification name: Computed tomography x-ray system<br>Classification Regulation: 21CFR 892.1750<br>Classification Panel: Radiology<br>Device Class: Class II<br>Primary Product Code: JAK |
| Primary Predicate Device: | Trade name: Philips Ingenuity CT<br>Manufacturer: Philips Medical Systems (Cleveland), Inc.<br>510(k) Clearance: K160743<br>Classification Regulation: 21CFR 892.1750<br>Classification name: Computed tomography x-ray system<br>Classification Panel: Radiology<br>Device class: Class II<br>Product Code: JAK |
| Secondary Predicate Device: | Trade name: Philips Incisive CT<br>Manufacturer: Philips Healthcare (Suzhou) Co., Ltd.<br>510(k) Clearance: K180015<br>Classification Regulation: 21CFR 892.1750<br>Classification name: Computed tomography x-ray system<br>Classification Panel: Radiology<br>Device class: Class II<br>Product Code: JAK |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
| Device Description: | The proposed Philips Incisive CT is a whole-body<br>computed tomography (CT) X-Ray System featuring a<br>continuously rotating x-ray tube, detectors, and gantry with<br>multi-slice capability. The acquired x-ray transmission data<br>is reconstructed by computer into cross-sectional images of<br>the body taken at different angles and planes. This system<br>also includes signal analysis and display equipment, patient<br>and equipment supports, components, and accessories.<br>The Philips Incisive CT has a 72cm bore and includes a<br>detector array that provides 50cm scan field of view (FOV). | Traditional 510k |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | The main components (detection system, the reconstruction<br>algorithm, and the x-ray system) that are used in the<br>proposed Philips Incisive CT have the same fundamental<br>design characteristics and are based on comparable<br>technologies as the current market predicate Philips<br>Ingenuity CT (K160743, 08/08/2016). | |
| | The main system modules and functionalities are: | |
| | 1. Gantry. The Gantry consists of 4 main internal units: | |
| | a. Stator - a fixed mechanical frame that carries HW<br>and SW | |
| | b. Rotor - A rotating circular stiff frame that is<br>mounted in and supported by the stator. | |
| | c. X-Ray Tube (XRT) and Generator, - fixed to the<br>Rotor frame | |
| | d. Data Measurement System (DMS) – a detectors<br>array, fixed to the Rotor frame | |
| | 2. Patient Support (Couch) - carries the patient in and out<br>through the Gantry bore synchronized with the scan | |
| | 3. Console - Containing a Host computer and display that is<br>the primary user interface. | |
| | In addition to the above components and the software<br>operating them, each system includes hardware and<br>software for data acquisition, display, manipulation, storage<br>and filming as well as post-processing into views other than<br>the original axial images. Patient supports (positioning aids)<br>are used to position the patient. | |
| | Traditional 510(k) | |
| Indications for Use: | The Incisive CT is a Computed Tomography X-Ray System<br>intended to produce images of the head and body by<br>computer reconstruction of x-ray transmission data taken at<br>different angles and planes. These devices may include<br>signal analysis and display equipment, patient and<br>equipment supports, components and accessories. The<br>Incisive CT is indicated for head, whole body, cardiac and<br>vascular X-ray Computed Tomography applications in<br>patients of all ages. | |
| | These scanners are intended to be used for diagnostic<br>imaging and for low dose CT lung cancer screening for the<br>early detection of lung nodules that may represent cancer*.<br>The screening must be performed within the established<br>inclusion criteria of programs / protocols that have been<br>approved and published by either a governmental body or<br>professional medical society. | |
| | *Please refer to clinical literature, including the results of the<br>National Lung Screening Trial (N Engl J Med 2011;<br>365:395-409) and subsequent literature, for further<br>information. | |
| Fundamental<br>Scientific<br>Technology: | The proposed Philips Incisive CT is advanced continuous-<br>rotation computed tomography systems suitable for a wide<br>range of computed tomographic (CT) applications. | |
| | The proposed Philips Incisive CT is used clinically as a<br>diagnostic patient imaging device that produces images that<br>correspond to tissue density. The quality of the images<br>depends on the level and amount of X-ray energy delivered<br>to the tissue. CT imaging displays both high-density tissue,<br>such as bone, and soft tissue. | |
| | The principal technological components (rotating x-ray tube,<br>detector and gantry) of the proposed Philips Incisive CT<br>substantially equivalent to the currently marketed predicate<br>device Philips Ingenuity CT (K160743, 08/08/2016) | |
| | Based on the information provided above, the proposed<br>Philips Incisive CT does not raise different questions of<br>safety and effectiveness compare to the currently marketed<br>predicate device Philips Ingenuity CT (K160743,<br>08/08/2016). | |
| Summary of Non-Clinical Performance Data: | The proposed Philips Incisive CT comply with the following international and FDA-recognized consensus standards: | |
| | • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R)2012 and A2:2010/(R) 2012 (Consolidated<br>Text) Medical Electrical Equipment - Part 1: General<br>Requirements For Basic Safety And Essential Performance<br>(IEC 60601-1:2005, MOD).<br>FDA/CDRH recognition number 19-4 | |
| | • IEC 60601-1-2:2014, Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests<br>FDA/CDRH recognition number 19-8 | |
| | • IEC 60601-1-3:2013 Medical electrical equipment -- Part<br>1-3: General requirements for basic safety - Collateral<br>standard: Radiation protection in diagnostic X-ray<br>equipment<br>FDA/CDRH recognition number 12-269 | |
| | • IEC 60601-1-6:2013 Medical electrical equipment -- Part<br>1-6: General requirements for basic safety and essential<br>performance – Collateral standard: Usability<br>FDA/CDRH recognition number 5-89 | |
| | • IEC 60601-2-44:2016 Medical electrical equipment - Part<br>2-44: Particular requirements for the safety and essential<br>performance of X-ray equipment for computed tomography<br>FDA/CDRH recognition number 12-302 | |
| | • IEC 62304:2015 Medical device software -- Software life<br>cycle processes<br>FDA/CDRH recognition number 13-79 | |
| | • IEC 62366-1:2015, Medical devices -- Part 1: Application<br>of usability engineering to medical devices<br>FDA/CDRH recognition number 5-114 | |
| | • ISO14971 Medical devices - Application of risk<br>management to medical devices (Ed. 3.0, 2019)<br>FDA/CDRH recognition number 5-125 | |
| | • NEMA XR 25-2010 Computed Tomography Dose Check<br>FDA/CDRH recognition number 12-225 | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The font is sans-serif and appears to be a standard typeface.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill the majority of the frame. The background is plain white, which makes the blue text stand out. The font appears to be sans-serif.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and it is centered.
Traditional 510(k) • NEMA XR 28-2013 Supplemental Requirements for User Information and System Function Related to Dose in CT FDA/CDRH recognition number 12-287 • NEMA XR 29-2013 Standard Attributes on CT Equipment Related to Dose Optimization and Management · Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005, document number 337). • Guidance for Industry and FDA Staff — Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (issued October 2, 2014) • Guidance for Industry and FDA Staff – Use of International Standard ISO 10993-1. "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016) · Guidance for Industry and FDA Staff - Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices (issued July 11, 2016) The systems comply with industry guidance and performance standards for Computed Tomography (CT) Equipment and Laser products (21 CFR 1020.33 and 21 CFR 1040.10, respectively). The systems performed a comparison to the predicate using these technological characteristics and image quality metrics to establish that the subject device is substantially equivalent to the predicate for its intended use. The systems pass the design verification, design validation and consensus standards test as nonclinical tests. The system verification is conducted against the system requirement specifications (SRS). System verification activities demonstrate the system meet the established system design input requirements. System requirements may be verified by manual test, automated test, inspection/analysis, or any combination of the three. Non-Clinical design validation testing covered the intended use and commercial claims. Validation testing included workflow validation.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The word is centered and takes up most of the frame.
| | The test results demonstrate that the proposed <b>Philips Incisive CT</b> meets the acceptance criteria and is adequate for its intended use. Additionally, the risk management activities show that all risks are sufficiently mitigated, that no new risks are introduced, and that the overall residual risks are acceptable. |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Clinical Data: | The proposed <b>Philips Incisive CT</b> did not require clinical study since substantial equivalence to the legally marketed predicate device was proven with the verification/validation testing. |
## Substantial Equivalence
| Scan characteristics Comparison | | | |
|---------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proposed Philips<br>Incisive CT | Predicate Device<br>Philips Ingenuity<br>CT(K160743) | Conclusion |
| No. of Slices | 64/128 | 64/128 | Identical.<br>Therefore, substantially<br>equivalent. |
| Scan Modes | Surview<br>Axial Scan<br>Helical Scan | Surview<br>Axial Scan<br>Helical Scan | Identical.<br>Therefore, substantially<br>equivalent. |
| Minimum Scan<br>Time | 0.35 sec for 360°<br>rotation | 0.42 sec for 360° rotation | The proposed Philips<br>Incisive CT rotation<br>speed faster than<br>Ingenuity CT to meet<br>the wider heart rate<br>application.<br>Safety and<br>effectiveness are not<br>affected.<br>Therefore,<br>demonstrating<br>substantial<br>equivalence. |
| Image (Spatial)<br>Resolution | High resolution mode:<br>16 lp/cm<br>Standard resolution<br>mode: 13 lp/cm | High resolution mode:16 lp/cm<br>Standard resolution mode:<br>13 lp/cm | Identical.<br>Therefore, substantially<br>equivalent. |
| Image Noise | 0.27% at 120 kV, 230 mAs, 10 mm slice thickness | 0.27% at 120 kV, 250 mAs, 10 mm slice thickness | Identical.<br>Therefore, substantially<br>equivalent. |
| Slice<br>Thicknesses | Helical: 0.67mm – 5mm<br>Axial: 0.625mm – 10.0mm | Helical: 0.67mm – 5mm<br>Axial: 0.625mm – 12.5mm | Essentially the same slice thickness, does not affect safety and effectiveness.<br>Therefore, demonstrating substantial equivalence. |
| Scan Field of View | Up to 500 mm | Up to 500 mm | Identical.<br>Therefore, substantially<br>equivalent. |
| Image Matrix | Up to 1024 * 1024 | Up to 1024 * 1024 | Identical.<br>Therefore, substantially<br>equivalent. |
| Display | 1920 * 1080 | 1024 * 1280 | The proposed Philips Incisive CT provide higher resolution than Ingenuity CT.<br>Safety and effectiveness are not affected.<br>Therefore, demonstrating substantial equivalence. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is white, which makes the blue letters stand out.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The font is sans-serif and appears to be a standard typeface.
| Host<br>Infrastructure | Windows 10 | Windows 7 | Same supplier, same<br>technology and similar<br>function.<br><br>Therefore, substantially<br>equivalent. |
|-------------------------------------|----------------------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Communication | Compliance with<br>DICOM | Compliance with DICOM | Identical.<br><br>Therefore, substantially<br>equivalent. |
| Dose Reporting<br>and<br>Management | Compliance with NEMA<br>XR25 , XR28 and XR29 | Compliance with NEMA<br>XR25 and XR29 | Compliance with more<br>NEMA standard.<br><br>Safety and<br>effectiveness are not<br>affected.<br><br>Therefore,<br>demonstrating<br>substantial<br>equivalence. |
| Imaging features Comparison | | | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------|
| Incisive CT<br>Features<br>Name | Feature description | Secondary<br>Predicate<br>Device Philips<br>Incisive<br>CT(K180015) | Conclusion<br>(Function/ User<br>interface/<br>Workflow) |
| 2D Viewer | In 2D Viewer mode operator can review<br>original axial images as acquired by the<br>scanner. | Yes | Identical. |
| MPR | Use the MPR mode to view three-plane<br>orthogonal images. In this mode, the three<br>shown planes can be easily correlated.<br>Three orthogonal cut<br>planes are shown:<br>• Axial Orientation<br>• Coronal Orientation<br>• Sagittal Orientation | Yes | Identical. |
| 3D(volume<br>mode) | The volume mode is used to display CT<br>scanner data in a full volume image. It | Yes | Identical. |
| | provides basic tools for image editing and<br>generation of cine movies. | | |
| Virtual<br>Endoscopy<br>(Endo) | The CT Endo viewer is a review function<br>that allows you to perform a general<br>flythrough of any suitable anatomical<br>structure that is filled with air or with<br>contrast material, including general vessels,<br>cardiac vessels, the bronchus, and the<br>colon. | Yes | Identical. |
| Filming | The Filming application is used for viewing,<br>rearranging, windowing and zooming<br>images prior to sending them to be printed. | Yes | Identical. |
| Image matrix | 1024 * 1024 | Yes | Identical. |
| O-MAR | O-MAR stands for orthopedic metal artifact<br>reduction. This post processing capability<br>reduces metal induced artifacts and is<br>directed for large orthopedics metals that<br>cause photon starvation of the rays that<br>pass through the metal object. | Yes | Identical. |
| Dose<br>Modulation | Dose-Modulation is a scanner function<br>which modulates the tube current in two<br>ways (angular and longitudinal modulation)<br>simultaneously. | Yes | Identical. |
| Scan<br>Preparation | iPlanning / Manual | Yes | Identical. |
| On line MPR | Use the on line MPR mode to view three-<br>plane orthogonal images. In this mode, the<br>three shown planes can be easily<br>correlated. | Yes | Identical. |
| Control<br>Panel | iStation (Touch Panel) | Yes | Identical. |
| iBatch | iBatch application enables the system to<br>assist the user to identify the lumbar disk<br>space automatically and creating a batch<br>based on the protocol selected. | Yes | Identical. |
| Bolus<br>Tracking | The Bolus tracking function maximizes the<br>efficiency of CT scans that are enhanced<br>through the use of a contrast agent. This is<br>done by preceding the Clinical scan with<br>Locator and Tracker scans. | Yes | Identical. |
| SAS(Spiral<br>Auto Start) | This feature enable the usage of the injector<br>scan trigger. | Yes | Identical. |
| Worklist | The Worklist displays patient information<br>provided by the HIS/RIS. | Yes | Identical. |
| MPPS | If the patient is from the Worklist and the<br>MPPS function is enabled, feedback<br>regarding the study status of the patient can<br>be sent to the hospital HIS/RIS. | Yes | Identical. |
| Reporting | | | Traditional 510(k) |
| Reporting | The Reporting package allows you to create customized reports using pre-formatted templates. A template is a specially designed formatting document that places the analytical information and images that you send from an application into an organized report which can be printed and saved. | Yes | Identical. |
| CCT(Continuous CT) | Continuous CT (CCT) is a scanning mode that allows the physician to perform extended, low-dose scans while performing a biopsy. The resulting images display on a remote monitor in the scan room, providing visual feedback during the biopsy. | Yes | Identical. |
| Brain Perfusion | Brain Perfusion is a blood flow imaging application that analyzes the uptake of injected contrast in order to determine perfusion-related information about one or more regions of interest. | Yes | Identical. |
| Dental (Dental planning) | Dental application uses to create true-size (life size) film images of the mandible and maxilla for assisting oral surgeons in planning implantation of prostheses. Using a special dental planning procedure, and the images will be created from this scan which can be input into the Dental planning application. | Yes | Identical. |
| iDose4 | iDose4 is an iterative reconstruction technique that improves image quality through artifact prevention and increased spatial resolution at low dose. | Yes | Identical. |
| Helical Retrospective Tagging | Helical retrospective cardiac scanning enables the system to acquire a volume of data while the patient's ECG is recorded. The acquired data is tagged and reconstructed retrospectively at any desired phase of the cardiac cycle. | Yes | Identical. |
| Axial Prospective Gating calcium scoring | Axial prospective gating uses an external ECG gating system to synchronize individual axial scans with the patient's heartbeat. The ECG-triggered scans significantly minimize heart-motion artifacts. | Yes | Identical. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
Traditional 510(k) Step & Shoot Cardiac provides high quality Step & Yes Identical. CT images of the coronary arteries and Shoot heart anatomy at very low radiation dose levels. During Step & Shoot Cardiac, X-rays are generated only during the cardiac phase of interest. The Cardiac Calcium Scoring application is CCS(Cardia Yes ldentical. used to quantify the buildup of calcium c calcium plaque on the walls of the patient's coronary scoring) arteries and other relevant locations. The potential calcifications are highlighted by the application during launch. Precise image reconstruction is a recon Precise No Different: mode where the system uses a image safety and trained deep learning neural network to effectiveness generate noise reduction images refer to and improve low contrast detectability K210760 for with reduced dose compared with Precise Image standard FBP recon mode. Precise Cardiac is a reconstruction Different: Precise No technique with the potential to cardiac safety and provide compensation for cardiac effectiveness refer to motion. K203020 for Precise Cardiac Precise Position is a camera based Precise No Different: workflow designed to assist with position Safety and positioning the patient automatically effectiveness from console or OnPlan, it can: refer to • automatically select patient orientation. K203514 for • automatically set vertical centering & Precise Position positioning of the patient to the Surview start and end positions. • support editing Surview start & end range and scan direction. In Precise Intervention viewer there are Precise No Different: Precise several tools, they will help intervention intervention is a you to navigate the needle safely during function that the intervention. support user to view the distance, angle between the injector and the object, system validation report proved that
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
| Direct<br>results | Direct Result-With Direct Result the user<br>is able to choose a desired<br>result during scan planning phase and get<br>the result for diagnosis without<br>further intervention. | No | Different: Direct<br>result is<br>workflow<br>enhancement to<br>make user direct<br>get derived<br>image after<br>acquisition no<br>impact on<br>system safety<br>and effectiveness | there is no<br>impact on<br>system safety<br>and effectiveness |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Parallel<br>workflow | Parallel workflow<br>The system support Parallel workflow using<br>Dual monitor as below:<br>- main monitor: Patients, scan, service, "show<br>all" for scan planning, Help.<br>- extend monitor: Completed, viewers, Analysis,<br>recon, filming, report | No | The Parallel<br>workflow of<br>Incisive CT is the<br>workflow<br>update, no<br>impact safety<br>and<br>effectiveness. | |
| Supplementary Imaging features Comparison | | | |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------|
| Incisive CT<br>Features<br>Name | Feature description | Secondary<br>Predicate<br>Device Philips<br>Incisive<br>CT(K180015) | Conclusion<br>(Function/ User<br>interface/<br>Workflow) |
| CTC(CT<br>Colonoscopy) | CT Colonoscopy (CTC) application enables<br>fast and easy visualization of colon scans,<br>using acquired CT images. | Yes | Identical. |
| VA(Vessel<br>Analysis) | Vessel Analysis (VA) offers a set of tools for<br>general vascular analysis. With VA the user<br>can easily remove bone, and extract<br>vessels. User also can perform<br>measurements such as intraluminal<br>diameter, cross-sectional lumen area,<br>length. | Yes | Identical. |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and centered in the image.
| LNA(Lung<br>Nodule<br>Analysis) | The Lung Nodule Analysis (LNA)<br>application assists the radiologist with the<br>detection and quantification of pulmonary<br>nodules and lesions. | Yes | Identical. |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|------------|
| CAA(Cardiac<br>Artery<br>Analysis) | The Coronary Artery Analysis provides<br>viewing and measuring tools that allow you<br>to perform dimensional and quantitative<br>measurements of the coronary arteries to<br>help you identify and examine the patient<br>study for stenosis. | Yes | Identical. |
| CFA(Cardiac<br>Function<br>Analysis) | Cardiac Function Analysis (CFA) application<br>is used to assess the state of the left<br>ventricle (LV) and to analyze functional<br>heart data. | Yes | Identical. |
| DE(Dual<br>Energy) | Spin / Spin scan mode<br><br>Dual energy Viewer is an application for<br>review and analysis of CT dual-energy<br>scans. User need to load CT dual-energy<br>scan data which is two series with similar<br>KV. It provides registration function and can<br>generate different weighted KV images.<br>User can use the tools to separate<br>materials. | Yes | Identical. |
| | Substantial<br>Equivalence<br>Conclusion: | The design, intended use, technology and principal technological<br>components (Tube, Generator, Detector) of the proposed Philips<br>Incisive CT substantially equivalent to the currently marketed |
|--|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | primary predicate device Philips Ingenuity CT (K160743,<br>08/08/2016).Based on the information provided above, the |
| | | proposed Philips Incisive CT does not raise different questions<br>of safety and effectiveness compare to the currently marketed<br>primary predicate device Philips Ingenuity CT (K160743,<br>08/08/2016). The proposed Philips Incisive CT is identical to the<br>primary predicate device Philips Ingenuity CT (K160743,<br>08/08/2016), and therefore is considered substantially equivalent. |
| | | Additionally, substantial equivalence was demonstrated with non-<br>clinical performance, V&V and consensus standards tests, which<br>complied with the requirements specified in the international and<br>FDA-recognized consensus standards. |
| | | The results of these tests demonstrate that the proposed Philips<br>Incisive CT meet the acceptance criteria and is adequate for its<br>intended use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.