K212307 · Bodycad Laboratories, Inc. · HSX · Oct 12, 2021 · Orthopedic
Device Facts
Record ID
K212307
Device Name
BC Reflex Uni Knee System
Applicant
Bodycad Laboratories, Inc.
Product Code
HSX · Orthopedic
Decision Date
Oct 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The patient-specific BC Reflex Uni™ is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with: · joint impairment due to osteoarthritis or traumatic arthritis of the knee, · varus deformity of the knee, and · as an alternative to tibial osteotomy in patients with unicompartmental OA. The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
Device Story
Patient-specific unicompartmental knee system; replaces medial tibiofemoral compartment. Inputs: MRI or CT images of patient knee; surgeon input. Transformation: Bodycad software generates CAD/CAM files for custom femoral and tibial implants and single-use instruments. Output: Patient-specific implants and instruments. Used in clinical setting by orthopedic surgeons. Implants cemented into place. Benefits: Improved fit and sizing flexibility via patient-specific geometry; enhanced wear resistance with Vit-E HXLPE material option.
Clinical Evidence
Bench testing only. Wear testing per ISO 14243-2/3 compared Vit-E HXLPE vs. standard UHMWPE. FEA verified contact stress equivalence. Material properties verified per ASTM standards (F648, F2759, D695, F2183, F2625, F2214, E647, F2003, F2102, F2381). Locking mechanism integrity verified for 5mm insert. Software V&V performed.
Technological Characteristics
Patient-specific knee implants. Materials: Wrought CoCrMo (ASTM F1537-11), Ti6Al4V ELI (ASTM F136-13), UHMWPE or Vit-E HXLPE. Single-use instruments: additively manufactured Nylon-12. Cemented fixation. Sterilization: SAL 10^-6. Software-driven CAD/CAM design process.
Indications for Use
Indicated for patients with advanced medial compartment knee osteoarthritis (OA) requiring unicompartmental knee arthroplasty. Candidates include those with joint impairment due to OA or traumatic arthritis, varus deformity, or as an alternative to tibial osteotomy, provided there is adequate healthy bone for component support.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
October 12, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Bodycad Laboratories, Inc. % Robert Poggie President BioVera, Inc. 65 Promenade Saint Louis Notre Dame de L'Ile Perrot Ouebe J7V 7P2 CANADA
Re: K212307
Trade/Device Name: BC Reflex Uni™ Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: August 12, 2021 Received: August 13, 2021
Dear Robert Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K212307
Device Name
BC Reflex Uni™ Knee System
Indications for Use (Describe)
The patient-specific BC Reflex Unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
· joint impairment due to osteoarthritis or traumatic arthritis of the knee,
· varus deformitv of the knee, and
· as an alternative to tibial osteotomy in patients with unicompartmental OA.
The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# Traditional 510(k) Notification for the BC Reflex Uni™ Knee System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following is a summary of safety and effectiveness of the BC Reflex Uni™ Knee System.
#### A. SUBMITTERS INFORMATION
| Submitter Name: | BioVera, Inc. |
|---------------------|-------------------------------------------------------------------------------|
| Submitter Address: | 65 Promenade Saint-Louis, Notre-Dame-de-L'Ile-Perrot, QC,<br>J7V 7P2, CANADA |
| Contact Person: | Robert A Poggie, PhD |
| Phone & Fax Number: | (514) 901-0796 |
| Date of Submission: | August 12, 2021 |
### B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | Laboratories Bodycad, Inc. |
|------------------------------|-----------------------------------------------------------------|
| Manufacturer Address: | 2035 rue du Haut-Bord, Quebec, QC, G1N 4R7, CANADA |
| Registration Number: | 3012086398 |
| Contact Name: | Nadine Adia |
| Title: | Specialist, Regulatory Affairs |
| Device Trade Name: | BC Reflex Uni™ Knee System |
| Device Common Name: | Unicondylar knee device |
| Classification Name: | knee joint femorotibial metal/polymer non-constrained, cemented |
| Classification Code: | HSX – Class II |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 CFR section 888.3520 |
| C1. PRIMARY PREDICATE DEVICE | |
| K203697 | Bodycad Inc. BC Reflex Uni™ Knee System |
| C2. PREDICATE DEVICE | |
| K191150 | Bodycad Inc. BC Reflex Uni™ Knee System (Tibia) |
| C3. REFERENCE DEVICES | |
| K033363 | Zimmer Unicompartmental Knee System (ZUK) |
| K163700 | Bodycad Unicompartmental Knee System |
{4}------------------------------------------------
#### D. DEVICE DESCRIPTION
The BC Reflex Uni™ Knee System is a patient-specific unicompartmental knee system that consists of femoral and tibial implants for replacement of the medial tibiofemoral compartment of the knee. The patient-specific femoral and tibial implants and single-use instruments are manufactured from CAD and CAM files generated from Bodycad software, which are based on MRI or CT images of the patient's knee and input from the surgeon. The BC Reflex Uni™ is for cemented use only and is sterilized to a SAL of 10-6.
The primary reasons for this 510(k) notification are to notify the FDA: (1) that the tibial insert will be manufactured from vitamin E stabilized and 100 kGy crosslinked UHMWPE (Vit-E HXLPE), (2) that a 5 mm thick tibial insert was added as a thickness option, and (3) of minor design updates to the implants and single use instruments.
Materials: Wrought Cobalt-28Chromium-6Molybdenum Alloy (ASTM F1537-11) for the femoral component, wrought Titanium-6Aluminum-4Vanadium ELI Alloy (Ti6AI4V ELI; ASTM F136-13) for the tibial baseplate and locking pin, Ultra-High-Molecular-Weight Polyethylene (UHMWPE) or Vit-E HXLPE for the tibial insert.
#### E. INTENDED USE
The patient-specific BC Reflex Uni™ is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
- · joint impairment due to osteoarthritis or traumatic arthritis of the knee,
- · varus deformity of the knee, and
- · as an alternative to tibial osteotomy in patients with unicompartmental OA.
The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
## F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The BC Reflex Uni™ Knee System described in this 510(k) is essentially the same device as the primary predicate device cleared in K203697. The technological characteristics that remain the same for the BC Reflex Uni™ Knee System are:
- Patient specific knee implants and instruments that are designed and manufactured from MRI or CT images of the patient's knee. The implant components include CoCr femoral and Ti alloy tibial implants, polyethylene inserts, a Ti alloy locking pin, and single use additively manufactured Nylon-12 bone models and cutting guides.
- · The clinical indications for use and the intended use are identical.
- · The CoCrMo, titanium alloy, and Nylon-12 materials are identical.
- · The manufacturing processes are the same.
- · The size envelope for the tibial and femoral components are the same.
- · Single use, patient specific kits are provided sterile with SAL of 10-6.
{5}------------------------------------------------
The similarities in technological characteristics for the subject BC Reflex Uni Knee System and the ZUK reference device are:
- The minimum thickness of the tibial insert (5 mm) is the same for the reference ZUK and subject BC Reflex Uni Knee System, and both are unconstrained articulation.
The primary differences between the subject and primary predicate BC Reflex Uni devices are:
- Vitamin E stabilized and 100 kGy crosslinked UHMWPE (Vit-E HXLPE) material was added as an option for manufacture of the tibial inserts.
- A new thickness tibial insert (5 mm) was added to the BC Reflex Uni Knee System.
- · Minor design updates were made to the femoral and tibial implant components to increase flexibility in sizing and fit to the patient, in keeping with original design envelope for the primary predicate and reference ZUK devices.
- Updates to the single use instruments were made to facilitate surgery with the new thickness tibial insert and improve usability.
## G. PERFORMANCE DATA
The following performance testing and verification activities were performed:
- Wear testing per ISO 14243-3 and ISO 14243-2 of worst-case subject BC Reflex Uni and ● reference ZUK tibial inserts made of standard UHMWPE (non-vitamin E, 25-40 kGy gamma radiation) and Vit-E HXLPE. The results showed less wear of the Vit-E HXLPE inserts relative to the standard UHMWPE inserts, and no difference in wear rate between the 5 and 6 mm tibial inserts made from standard UHMWPE.
- Finite element analysis verified substantial equivalence of contact stress of the 5 and 6 mm tibial inserts for the subject, primary predicate, and reference devices.
- Verification of substantial equivalence and/or superiority of performance characteristics ● and materials properties of the Vit-E HXLPE relative to standard UHMWPE for: density per ASTM F648 and D792, mechanical properties per ASTM F648, F2759, D695, F2183, melting point, crystallinity, and enthalpy of fusion per ASTM F2625, swell ratio and crosslink density per ASTM F2214, fatigue crack propagation and coefficient per ASTM E647, oxidation challenge per ASTM F2003 and analysis per ASTM F2102, ESR testing for residual free radical content, and Transvinylene index (TVI) per ASTM F2381.
- Testing and verification of integrity of the locking mechanism of a 5mm thick tibial insert.
- Software V&V accounting for all changes per Bodycad standard operating procedures. .
- o Risk analysis and design control review confirming no new or changed risks relative to the indications for use and efficacy of the subject device.
## H. CONCLUSION
The results of performance testing, V&V analyses, and engineering review demonstrated substantial equivalence of the subject BC Reflex Uni™ Knee System to the primary predicate, predicate, and reference devices cited in this 510(k) notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.