K203697 · Bodycad Laboratories, Inc. · HSX · Mar 12, 2021 · Orthopedic
Device Facts
Record ID
K203697
Device Name
BC Reflex Uni Knee System
Applicant
Bodycad Laboratories, Inc.
Product Code
HSX · Orthopedic
Decision Date
Mar 12, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The patient-specific BC Reflex Uni™ is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with: · joint impairment due to osteoarthritis or traumatic arthritis of the knee, · varus deformitv of the knee, and · as an alternative to tibial osteotomy in patients with unicompartmental OA. The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
Device Story
Patient-specific unicompartmental knee system; replaces medial tibiofemoral compartment. Inputs: patient MRI or CT images; surgeon input. Transformation: Bodycad software generates CAD/CAM files for custom femoral/tibial implants and single-use instruments. Output: patient-specific implants (CoCrMo femoral, Ti6Al4V ELI tibial baseplate, UHMWPE insert) and instruments. Used in clinical setting by orthopedic surgeons. Surgeon uses patient-specific guides to facilitate implantation; cemented fixation. Benefits: personalized fit based on patient anatomy; improved surgical workflow.
Clinical Evidence
Bench testing only. Verification and validation activities included software V&V, risk analysis, design control review, and usability testing of updated devices. No clinical data presented.
Indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced medial compartment knee osteoarthritis (OA), traumatic arthritis, or varus deformity, requiring adequate healthy bone for component support; also indicated as an alternative to tibial osteotomy.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date March 12, 2021. The text is in a clear, serif font and is horizontally aligned. The date is written out in full, with the month, day, and year clearly indicated.
Bodycad Laboratories, Inc. Robert Poggie President BioVera, Inc.
65 Promenade Saint Louis Notre Dame de L'lle Perrot, QC J7V 7P2
Re: K203697
Canada
Trade/Device Name: BC Reflex Uni™ Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: December 16, 2020 Received: December 18, 2020
Dear Robert Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorywebsite the DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ting Song, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K203697
Device Name
BC Reflex Uni™ Knee System
Indications for Use (Describe)
The patient-specific BC Reflex Uni™ is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
· joint impairment due to osteoarthritis or traumatic arthritis of the knee,
· varus deformitv of the knee, and
· as an alternative to tibial osteotomy in patients with unicompartmental OA.
The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### The BC Reflex Uni™ Knee System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of the BC Reflex Uni™ Knee System.
#### A. SUBMITTERS INFORMATION
| Submitter Name: | BioVera, Inc. |
|---------------------|------------------------------------------------------------------------------|
| Submitter Address: | 65 Promenade Saint-Louis, Notre-Dame-de-L'Ile-Perrot, QC,<br>J7V 7P2, CANADA |
| Contact Person: | Robert A Poggie, PhD |
| Phone & Fax Number: | (514) 901-0796 |
| Date of Submission: | March 4, 2021 |
#### B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | Laboratories Bodycad, Inc. |
|-----------------------|-----------------------------------------------------------------|
| Manufacturer Address: | 2035 rue du Haut-Bord, Quebec, QC, G1N 4R7, Canada |
| Registration Number: | 3012086398 |
| Contact Name: | Nadine Adia |
| Title: | Specialist, Regulatory Affairs |
| Device Trade Name: | BC Reflex Uni™ Knee System |
| Device Common Name: | Unicondylar knee device |
| Classification Name: | knee joint femorotibial metal/polymer non-constrained, cemented |
| Classification Code: | HSX - Class II |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 CFR section 888.3520 |
## C1. PRIMARY PREDICATE DEVICE
K191996 BC Reflex Uni™ Knee System (primary predicate device)
#### C2. PREDICATE AND REFERENCE DEVICES
| K191150 | BC Reflex Uni™ Knee System (predicate device) |
|------------------|---------------------------------------------------------|
| K181302, K163700 | Bodycad Unicompartmental Knee System (reference device) |
| K193614 | Fine Osteotomy Around the Knee (reference device) |
{4}------------------------------------------------
#### D. DEVICE DESCRIPTION
The BC Reflex Uni™ Knee System is a patient-specific unicompartmental knee system that consists of femoral and tibial implants for replacement of the medial tibiofemoral compartment of the knee. The patient-specific femoral and tibial implants and single-use instruments are manufactured from CAD and CAM files generated from Bodycad software, which are based on MRI or CT images of the patient's knee and input from the surgeon. The BC Reflex Uni™ is for cemented use only and is sterilized by gamma radiation.
The subject device of this 510(k) is the same as the primary predicate device. The purpose of this Special 510(k) Device Modification is to notify the FDA of minor changes to the design and contents of the patient specific kits and reusable instruments for the BC Reflex Uni™ Knee System.
Materials: Wrought Cobalt-28Chromium-6Molybdenum Alloy (CoCrMo; ASTM F1537-11) for the femoral component, wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (Ti6Al4V ELI; ASTM F136-13) for the tibial baseplate and locking pin, Ultra-High-Molecular-Weight Polyethylene (UHMWPE; F648-14) for the tibial insert.
## E. INTENDED USE
The patient-specific BC Reflex Uni™ is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
- joint impairment due to osteoarthritis or traumatic arthritis of the knee,
- · varus deformity of the knee, and
- as an alternative to tibial osteotomy in patients with unicompartmental OA.
The patient-specific BC Reflex Uni™ components fit within an envelope of dimensions that are specific to each patient. The BC Reflex Uni™ femoral component and tibial baseplate are intended for cemented fixation.
## F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The BC Reflex Uni™ Knee System described in this 510(k) notification is essentially the same device as the primary predicate device cleared by FDA in K191996 and the predicate device described in K191150. Most technological characteristics remain unchanged from the two predicate and subject devices. Commonalities between the subject and predicate devices are:
- Patient specific knee implants and instruments are designed and manufactured from MRI or CT images of the patient's knee.
- · Patient specific components including CoCr femoral and Ti alloy tibial implants, polyethylene inserts, and locking pin for assembly of the tibial insert and baseplate, and single use additively manufactured Nylon-12 bone models and guides to facilitate surgery.
- · The clinical indications for use are identical.
- · The CoCrMo, titanium alloy, UHWMPE, and Nylon-12 materials are identical.
- · The manufacturing processes are identical.
- · Single use, patient specific kit provided sterile with SAL of 10-6.
{5}------------------------------------------------
The minor differences between the subject and predicate devices are:
- · Updates to Bodycad software to improve efficiency in design and manufacture of patient specific components of the BC Reflex Uni™ Knee System.
- · Added option in patient specific design process to use validated Bodycad software, or previously validated off the shelf software.
- · Minor design updates to implant components to improve clarity in design specifications, manufacturability, and/or usability per the intended use.
- Updates to the single use / disposable instruments and the reusable instrument kit to improve workflow in surgery and improve function, in keeping with intended / indications for use for the subject and predicated device (same).
The minor changes to the BC Reflex Uni™ Knee System do not raise new issues of safety or effectiveness. The information presented in this Special 510(k) Device Modification demonstrates substantial equivalence of the subject and predicate device.
# G. PERFORMANCE DATA
The following verification and validation (V&V) activities were performed:
- . Software V&V accounting for all changes per Bodycad procedures, which are the same procedures presented to FDA previously for the predicate and reference devices.
- . Risk analysis and design control review confirming no new or changed risks relative to the indications for use and efficacy of product.
- Verification testing of usability of updated devices.
The results of V&V testing and engineering review, risk analysis, and design control activities demonstrated substantial equivalence of the subject and predicate BC Reflex Uni™ Knee System.
## H. CONCLUSION
The BC Reflex Uni™ subject components of 510(k) notification are substantially equivalent and share the same intended use as the predicate devices cleared in K191150 and K191996.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.