Automatic segmentation of structures for radiation therapy
Convolutional neural networks (CNNs)
Dice metric (large organs) and HD95 (small organs) compared against predicate/reference device thresholds
All organs passed acceptance criteria and demonstrated noninferiority
—
>1 (dosimetrist, medical physicist and/or radiation oncologist)
Test data from multiple sources covering head and neck, thorax, abdomen, and male pelvis regions, independent of training data.
>1 (dosimetrist, medical physicist and/or radiation oncologist)
Indications for Use
INTContour provides a machine learning-based approach for the automatic segmentation of structures including treatment targets and organs at risk to support the radiation therapy treatment planning process. INTContour is intended as an initial method to segment and contour study series; therefore, this software must be used in conjunction with an appropriate software to edit the segmentation results if necessary. It is not intended to replace a thorough review by qualified medical professionals. INTContour is developed for use by dosimetrists, medical physicists, and radiation oncologists. The currently supported anatomical regions for automatic segmentation are head and neck, thorax, abdomen, and male pelvis.
Device Story
INTContour is a software-only device for automatic segmentation of medical images (CT/MRI) in radiation therapy planning. It uses convolutional neural networks (CNNs) trained on expert-contoured libraries to identify treatment targets and organs at risk. Deployed as a service within a hospital's private network on a GPU-equipped workstation, it is accessed via a web-based interface. Dosimetrists, medical physicists, and radiation oncologists use the software to generate initial contours, which are then exported in DICOM RTSTRUCT format. The device acts as a pre-processing tool; clinicians must review and edit the output using external software. By automating the initial contouring process, the device aims to support and streamline the radiation therapy treatment planning workflow.
Clinical Evidence
Bench testing only. Performance evaluated using multi-source datasets covering head and neck, thorax, abdomen, and male pelvis (patient ages 18–76). Ground truth established by at least two trained professionals (dosimetrist, physicist, or oncologist). Metrics included Dice coefficient (for large organs) and 95% Hausdorff Distance (HD95, for small organs). Statistical analysis demonstrated non-inferiority compared to predicate/reference devices, with all organs meeting pre-defined acceptance criteria.
Indicated for automatic segmentation of treatment targets and organs at risk in radiation therapy planning. Intended for use by dosimetrists, medical physicists, and radiation oncologists. Supported anatomical regions: head and neck, thorax, abdomen, and male pelvis. Must be used with secondary software for manual editing/review.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Smart Segmentation – Knowledge Based Contouring (K141248)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Carina Medical LLC % Xue Feng Chief Executive Officer 1233 Litchfield Ln Lexington, KY 40513
Re: K212274
Trade/Device Name: INT Contour Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ, QKB Dated: July 20, 2021 Received: July 20, 2021
Dear Xue Feng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Sullivan. Ph.D. Assistant Director Nuclear Medicine and Radiation Therapy Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center For Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212274
Device Name INT Contour
#### Indications for Use (Describe)
INTContour provides a machine learning-based approach for the automatic structures including treatment targets and organs at risk to support the radiation therapy treatment planning process. INTContour is intended as an initial method to segment and contour study series; therefore, this software must be used in conjunction with an appropriate software to edit the segmentation results if necessary. It is not intended to replace a thorough review by qualified medical professionals. INTContour is developed for use by dosimetrists, and radiation oncologists. The currently supported anatomical regions for automatic segmentation are head and neck, thorax, and male pelvis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "CARINA" in a bold, sans-serif font. The letters "A" and "I" are colored red, while the remaining letters are black. The word is presented in a clean and modern style.
#### Summary 510(k)
#### 1. Applicant:
- Carina Medical LLC
- 1233 Lichfield Ln
- Lexington, KY, 40513
- USA
- Contact Name: Xue Feng Chief Executive Officer
- Phone: 434-284-1073
- Fax: 855-615-2856
- E-mail: xfeng@carinaai.com
#### 2. Device:
Trade Name: INTContour
Common Name: Intelligent Organ Contouring System
Model Number: v1.0
Product Code: MUJ
Regulation Description: Medical charged-particle radiation therapy system
Regulation Number: 21 CFR 892.5050
Device Class: II
#### 3. Predicate Devices:
Trade Name: Smart Segmentation – Knowledge Based Contouring
info@carinaai.com
+1 434-284-1073
{4}------------------------------------------------
# CARINA
Manufacturer: Varian Medical Systems, Inc.
Address: 3100 Hansen Way, Palo Alto, CA, 94303
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical charged-particle radiation therapy system
Device Class: II
Product Code: MUJ
- 510(k) Number: K141248
- 510(k) Clearance Date: 09/05/2014
#### 4. Reference Device:
Trade Name: AccuContour
Manufacturer: Xiamen Manteia Technology LTD.
Address: 1903, B Tower, Zijin Plaza, No.1811 Huandao East Road, Xiamen, China
Regulation Number: 21 CFR 892.2050
Regulation Name: Picture archiving and communications system
Device Class: II
Product Code: QKB
510(k) Number: K191928
510(k) Clearance Date: 02/28/2020
## 5. Device Description
INTContour is a software-only product that uses a machine learning-based approach to perform automatic segmentation of structures in medical images, coupled with tools for visualizing the seqmentation results. A library of previously contoured expert cases serves as inputs to train the machine learning algorithms, specifically, convolutional networks (CNNs), to perform
info@carinaai.com
{5}------------------------------------------------
# CARINA
automatic segmentation. The results of the automatic segmentation will be stored in the DICOM Radiotherapy Structure Set (RTSTRUCT) format, which can be sent to desired destinations via the DICOM protocol.
INTContour is intended to be used by dosimetrists, medical physicists, and radiation oncologists, and serves as an initial method to segment and contour study series. It must be used in conjunction with appropriate software to edit the segmentation results if necessary. The currently supported anatomical regions for automatic segmentation are head and neck, thorax, abdomen, and male pelvis. The supported structures for each region are shown below in section 10.
INTContour software is intended to be deployed within a hospital's private network on a workstation with an advanced graphics processing unit (GPU) and runs as a service. A webbased interface is used to access the service and manage the transfer of data, automatic segmentation, and visualization.
#### 6. Intended Use Statement
INTContour provides a machine learning-based approach for automatic segmentation of structures including treatment targets and organs at risk to support the radiation therapy treatment planning process.
#### 7. Indications for Use Statement
INTContour provides a machine learning-based approach for the automatic segmentation of structures including treatment targets and organs at risk to support the radiation therapy treatment planning process. INTContour is intended as an initial method to segment and contour study series; therefore, this software must be used in conjunction with an appropriate software to edit the segmentation results if necessary. It is not intended to replace a thorough review by qualified medical professionals. INTContour is developed for use by dosimetrists, medical physicists, and radiation oncologists. The currently supported anatomical regions for
{6}------------------------------------------------
automatic segmentation are head and neck, thorax, abdomen, and male pelvis. The supported structures for each region are shown below in Section 10.
## 8. Summary of Technological Characteristics Comparison
Table 1 shows the similarities and differences between the technological characteristics of the two products. The key difference is the detailed implementation of the automated segmentation algorithms. Testing demonstrates that the differences do not raise new questions of safety or effectiveness.
| Topic | Predicate Device | Subject Device |
|------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Physical<br>Characteristics | Software package that operates on<br>off-the-shelf hardware | Software package that operates on a<br>virtual machine within the off-the-shelf<br>hardware |
| Computer | PC Compatible | Same |
| Operating<br>System | Windows | Ubuntu (18.04 & 20.04) |
| DICOM Standard<br>Compliance | The software processes DICOM<br>compliant image data, including<br>RTSTRUCT | Same |
| Modalities | CT & MRI | Same |
| User Interface | The software is designed for use on a<br>radiotherapy workstation with a native<br>user interface. | The software is designed for use on a<br>radiotherapy workstation with a web-<br>based user interface |
| Segmentation<br>Structures | Organs at risk and target volumes | Same |
## Table 1. Summary of Technological Characteristic Comparison
info@carinaai.com
{7}------------------------------------------------
| Overall<br>Segmentation<br>Method | Model-based approach using a library<br>of expert contours | Same |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Detailed<br>Implementation<br>of Segmentation<br>Algorithm | Smart Segmentation – knowledge-based contouring used an atlas-based method to deform expert contours to the target images to generate contours | INTContour used a machine learning-based method to train CNNs from expert contours to perform segmentation on the target images to generate contours |
| Support of<br>Manual Editing | Yes | No |
| Provide Expert<br>Case Library to<br>Users | Yes | No |
#### 9. Performance Data
The safety and performance of INTContour has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Non-clinical verification and validation test results, including model performance and software usability, established that the device meets its design requirements and intended use, that it is as safe and as effective as the predicate devices, and that no new issues of safety and effectiveness were raised.
Testing data was acquired from multiple sources than the training data that covers head and neck, thorax, abdomen, and male pelvis regions. The cases were selected from patients who went through radiation treatment with age 18 – 76, both male and various types of cancers. Ground truth was performed by at least two trained personnel including dosimetrist, medical physicist and/or radiation oncologist to minimize human bias in segmentation.
{8}------------------------------------------------
# CARINA
Further, during the development, potential hazards were controlled by a risk management plan including risk analysis, risk mitigation, verification and validation.
## 10. Results
The Dice metric and 95% Hausdorff Distance (HD95) were calculated for each organ. Statistical analysis on the test data was performed by comparing the calculated metrics of INTContour against the predicate/reference device. Dice metric was used for large organs and HD95 was used for small organs to define the acceptance criteria due to the large variations on Dice metric on small organs. By comparing the lower bound (Dice) or upper bound (HD95) of the performance differences between INTContour and the predicate/reference device with the threshold values, all organs have passed the acceptance criteria and demonstrated the noninferiority against the predicate/reference device.
| Head and Neck | Thorax | Abdomen | Male Pelvis | |
|----------------|-----------------|---------------|--------------|----------------|
| Brainstem | Pituitary | SpinalCanal | Spleen | Bladder |
| OpticChiasm | InnerEar_L | Lung_R | Kidney_R | Femur_Head_L |
| Bone_Mandible | InnerEar_R | Lung_L | Kidney_L | Femur_Head_R |
| OpticNrv_L | Lens_L | Heart | Gallbladder | PenileBulb |
| OpticNrv_R | Lens_R | Esophagus | Esophagus | Prostate |
| Parotid_L | Lobe_Temporal_L | Trachea | Liver | SeminalVesicle |
| Parotid_R | Lobe_Temporal_R | BronchialTree | Stomach | Rectum |
| Glnd_Submand_L | Larynx | SpinalCord | A_Aorta | |
| Glnd_Submand_R | Cavity_Oral | | V_Venacava_I | |
| Eye_L | Pharynx | | PortalVein | |
The table below shows the list of all supported organs:
{9}------------------------------------------------
| Eye_R | Brain | Pancreas |
|----------------|-----------------|----------------|
| MidEar_L | Cochlea_L | AdrenalGland_R |
| MidEar_R | Cochlea_R | AdrenalGland_L |
| Joint_TM_L | Glnd_Lacrimal_L | SplenicVein |
| Joint_TM_R | Glnd_Lacrimal_R | |
| BrachialPlex_L | BrachialPlex_R | |
## 11. Substantial Equivalence Conclusion
INTContour is an automatic segmentation software to support the radiation treatment planning process which has similar intended use and indications for use statement as the predicate device. The two devices have similar technological characteristics: both algorithms rely on a library of expert contours and use model-based approach to generate contours on new images. This 510(k) submission includes information on the INTContour technological characteristics, as well as performance data and verification and validation activities demonstrating that INTContour is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.