The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-46EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
Device Story
TRAUS ENDO is an AC-powered dental handpiece system comprising a power unit, charging station, and contra-angle handpieces (ACL(B)-41EP, ACL(B)-42EP, ACL(B)-46EP). Used by dentists in clinical settings for endodontic surgery, including grinding, cutting, and polishing. Operates at 600-1,000 rpm. System supports forward, auto-reverse, and reciprocation drive rotation modes. Provides mechanical power to rotary files and drills for root canal preparation. Benefits include controlled mechanical preparation of root canals during endodontic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via software verification/validation (EN 62304), temperature testing (ISO 14457, IEC 80601-2-60), and eccentricity testing (ISO 14457). Biocompatibility and reprocessing validation were leveraged from previously cleared predicates (K143411, K181129).
Indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
June 24, 2022
Saeshin Precision Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612
Re: K212043
Trade/Device Name: TRAUS ENDO Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EKX Dated: May 16, 2022 Received: May 26, 2022
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212043
Device Name TRAUS ENDO
Indications for Use (Describe)
The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-46EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K212043)
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## 1. Date: 6/23/2022
### 2. Applicant / Submitter
Saeshin Precision Co., Ltd. # 52, Secheon-ro 1-gil, Dasa-eup Dalseonggun, Daegu, 711-814, Korea Tel: +82 53 587 2345 / Fax:+82 53 580 0918
### 3. U.S. Designated Agent
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com
### 4. Trade/Proprietary Name:
TRAUS ENDO
## 5. Common Name:
Dental Handpiece
### 6. Classification:
Handpiece, Direct Drive, Ac-Powered (21CFR 872.4200, Product code EKX, Class I, Dental)
## 7. Device Description:
The TRAUS ENDO is an AC-powered device that includes a power unit, charging station and contra-angle handpieces, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-46EP, which are used for grinding, cutting, and polishing work in dental oral use. It is used for endodontic surgery with 600-1,000 rpm.
{4}------------------------------------------------
## 8. Indication for use:
The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-46EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
## 9. Primary Predicate Device:
- . TRAUS ENDO (K143411) by Saeshin Precision Co., Ltd.
## 10. Additional Predicate/Reference Device:
- . STRONG Dental Handpieces (K181129) by Saeshin Precision Co., Ltd.
- ENDO a class (K123582) by Saeyang Microtech Co., Ltd. ●
{5}------------------------------------------------
# 11. Substantial Equivalence:
| | Subject Device | Primary Predicate<br>Device | Predicate Device | Reference Device |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Saeshin Precision Co.,<br>Ltd. | Saeshin Precision Co.,<br>Ltd. | Saeshin Precision Co.,<br>Ltd. | Saeyang Microtech Co.,<br>Ltd. |
| Device Name | TRAUS ENDO | TRAUS ENDO | STRONG Dental<br>Handpieces | ENDO a class |
| 510(k) Number | K212043 | K143411 | K181129 | K123582 |
| Device<br>Classification Name | Handpiece, direct drive,<br>AC-powered | Handpiece, direct drive,<br>AC-powered | Handpiece, direct drive,<br>AC-powered | Handpiece, direct drive,<br>AC-powered |
| Classification<br>Product Code | EKX | EKX | EGS | EKX |
| Device Class | I | I | I | I |
| Regulation Number | 21 CFR 872.4200 | 21 CFR 872.4200 | 21 CFR 872.4200 | 21 CFR 872.4200 |
| Indication for Use | The TRAUS ENDO,<br>ACL(B)-41EP, ACL(B)-<br>42EP, ACL(B)-46EP, are<br>indicated for use by<br>dentists in standard<br>endodontic procedures<br>using rotary files and<br>rotary endodontic drills<br>(Gates-Glidden). | The TRAUS ENDO,<br>ACL(B)-41EP, ACL(B)-<br>42EP, and ACL(B)-45EP,<br>are indicated for use by<br>dentists in standard<br>endodontic procedures<br>using rotary files and<br>rotary endodontic drills<br>(Gates- Glidden). | The Strong Dental<br>Handpiece, ACL(B)-<br>46EP is intended for wide<br>range of dental<br>procedures, including:<br>A. Implant placement,<br>including<br>1. Preparation of the<br>osteotomy site<br>2. Bone contouring,<br>osteoplasty<br>B. Periodontal surgeries<br>1. Bone contouring &<br>alveoplasty around living<br>teeth<br>2. Removal of exostosis<br>C. Bone grafting<br>1.Preparation of the donor<br>site (for e.g. symphysis<br>and ascending rames etc.)<br>2. Harvesting autogen<br>living bone<br>Sinus elevation &<br>grafting of alveolar<br>sockets<br>3. D. Removal and<br>sectioning of teeth and<br>teeth bone for e.g.<br>impacted third molars and<br>complicated extractions. | This application area<br>extends to endodontic<br>procedures using a root<br>canal instrument which is<br>intended by the<br>manufacturer for use in<br>the mechanical and<br>rotary preparation of root<br>canals. |
| Components | Contra-Angle Handpiece<br>Motor Handpiece Battery<br>Charger<br>Power Cord | Contra-Angle Handpiece<br>Motor Handpiece Battery<br>Charger<br>Power Cord | Contra-Angle Handpiece | Contra-Angle Handpiece<br>Motor Handpiece Battery<br>Charger<br>Power Cord |
| Gear | 10:1<br>16:1<br>20:1 | 16:1<br>20:1<br>32:1 | 10:1 | 4:1<br>10:1<br>16:1<br>20:1 |
{6}------------------------------------------------
| Chuck Type<br>Shank Type | Latch Type<br>Type 1 | Latch Type<br>Type 1 | Latch Type<br>Type 1 | Latch Type<br>Type 1 |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------|
| Composition of<br>Material | Gear: Stainless steel Shank: Stainless steel (Trimrite) & Brass Head: Stainless steel Chuck: Aluminum | Gear: Stainless steel Shank: Stainless steel (Trimrite) & Brass Head: Stainless steel Chuck: Aluminum | Gear: Stainless steel Shank: Stainless steel (Trimrite) & Brass Head: Stainless steel Chuck: Aluminum | Stainless Steel |
| Supply voltage | 100-240V AC 50/60Hz | 100-240V AC 50/60Hz | - | 100-240V AC 50/60Hz |
| Operational voltage | DC 4.5V | DV 4.5V | - | - |
| Torque moto speed | 1-4 Ncm | 1-4 Ncm | 2 Ncm | 1-4 Ncm |
| Sterilization | Contra-angle at 132°C for 4 minutes in Autoclave<br>conforms IEC 60601-1<br>IEC 60601-1-2<br>ISO 14457 | Contra-angle at 132°C for 4 minutes in Autoclave<br>conforms IEC 60601-1<br>IEC 60601-1-2<br>ISO 14457 | Contra-angle at 132°C for 4 minutes in Autoclave<br>conforms ISO 14457 | conforms IEC 60601-1<br>IEC 60601-1-2 |
| Electrical safety | | | | |
| Allows selection of<br>forward or Auto<br>reverse drive rotation | Yes | Yes | - | Yes |
| Allows reciprocation drive | Yes | No | - | Yes |
The indications for use and the technological characteristics of the subject device is the same as the predicate device, the unmodified devices. The ACL(B)-41EP and the ACL(B)-42EP are cleared under K143411 and the ACL(B)-46EP is cleared under K181129. The modification is adding reciprocation mode and we identified a reference device (K123582) which offers the same mode.
We performed risk analysis and verification/validation tests per modifications, and the test results support that the modification raises no new safety and effectiveness questions.
## 12. Performance Data:
The following tests were conducted, and the device passed all of the tests based on predetermined Pass/Fail criteria.
- · Software Verification/Validation (EN 62304).
- · Temperature Test (ISO 14457, IEC 80601-2-60)
- · Eccentricity Test (ISO 14457)
We also referenced the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" while preparing the software documentation.
Biocompatibility Testing per ISO 10993 was assessed under K143411, the previous cleared device, which has the same patient contacting material and manufacturing processes as the subject device.
Reprocessing Validation (i.e., Cleaning, Sterilization per ISO 17665-1) was assessed under K143411 and K181129, the previously cleared devices.
{7}------------------------------------------------
## 13. Conclusion:
The subject device is substantially equivalent in the areas of technical characteristics, general function, application, and indications for use. The subject device does not introduce a fundamentally new scientific technology, and the device has been validated through system level test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.