The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-45EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
Device Story
TRAUS ENDO is an AC-powered dental handpiece system comprising a power unit, charging station, and contra-angle handpieces (models ACL(B)-41EP, ACL(B)-42EP, ACL(B)-45EP). Used by dentists in clinical settings for grinding, cutting, and polishing during endodontic procedures. Device utilizes rotary files and Gates-Glidden drills. Operation involves gear-driven rotation at specified ratios (16:1, 20:1, 32:1). Output is mechanical rotation for root canal preparation. Benefits include facilitating standard endodontic instrumentation. Contra-angle handpieces are autoclavable.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, ISO 14457, ISO 17665-1, ISO 1797-1, and ISO 10993-1 demonstrated equivalence.
Technological Characteristics
Materials: Stainless steel (SUS420F, Biodur Trimtite), Brass (C3604BD-F), Aluminum (AL6061). Gear ratios: 16:1, 20:1, 32:1. Power: AC 100-240V, 50/60Hz; DC 4.5V. Chuck: Latch type. Sterilization: Autoclave (132°C for 4 minutes). Standards: IEC 60601-1, IEC 60601-1-2, ISO 14457, ISO 17665-1, ISO 1797-1, ISO 10993-1.
Indications for Use
Indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the left side of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2015
Saeshin Precision Co., Ltd. Mr. Sae Kwan Choi QA Manager 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun Daegu, 711-814 REPUBLIC OF KOREA
Re: K143411 Trade/Device Name: TRAUS ENDO Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: May 26, 2015 Received: May 28, 2015
Dear Mr. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143411
Device Name
TRAUS ENDO
Indications for Use (Describe)
The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-45EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo and contact information for SAESHIN. The logo is a blue geometric design. The contact information includes the address: # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea, and the telephone and fax numbers: Tel: +82 53 587 2341/ Fax: +82 53 580 0999.
# K143411 510(k) Summary
[As required by 21 CFR 807.92] This 510(k) summary is prepared in accordance with 21 CFR807.92
# 1. Date Prepared [21 CFR 807.92(a)(1)]
6/26/2015
# 2. Submitter's Information [21 CFR807.92(a)(1)]
- Name of Sponsor: Saeshin Precision Co., Ltd. .
- Address: # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, -Daegu, 711-814, Republic of Korea
- Contact Name: Sae Kwan, Choi (Mr.) / Quality Manager .
- Telephone No. : +82 53 587 2341 -
- Fax No. : +82 53 580 0999 -
- Email Address : ksqc@saeshin.com -
- . Registration Number: 3007958831
- . Name of Manufacturer: Same as Sponsor
- Address: Same as Sponsor -
# 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- . Trade Name: TRAUS ENDO
- Common Name: Dental Handpieces
- . Classification Name: Dental Handpiece and Accessories
- . Classification Panel: Dental
- . Classification Regulation: 21 CFR 872.4200
- EKX . Product Code:
- Device Class: |
### General Information
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for SAESHIN, along with their address and contact information. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea. The telephone number is +82 53 587 2341, and the fax number is +82 53 580 0999.
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follows:
- . 510(k) Number: K123582
●
Applicant:
- Saeyang Microtech Co., Ltd.
- Common Name: Dental Handpieces ●
- . Device Name: Endo a class
- . 510(k) Number: K111616
- Applicant: Saeshin Precision Co., Ltd
- Dental Handpieces Common Name: ●
- Device Name: E-CUBE
K123582 is the primary predicate and K111616 is the reference predicate. There are no significant differences between the TRAUS ENDO dental handpieces and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in device design, composition of materials and technical specifications.
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The TRAUS ENDO is an AC-powered device that includes a power unit, charging station and contra-angle handpieces, ACL(B)-41EP, ACL(B)-42EP and ACL(B)-45EP, are grinding, cutting, and polishing work in dental oral use.
#### Intended Use [21 CFR 807.92(a)(5)] 6.
The TRAUS ENDO, ACL(B)-41EP, ACL(B)-42EP, and ACL(B)-45EP, are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills (Gates-Glidden).
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
- Model Nos. ACL(B)-41EP, ACL(B)-42EP and ACL(B)-45EP ●
- Gear Ratio: 20:1, 16:1, 32:1
- Dimensions: Ø 16.7 X 63 mm, Ø 16.7 X 63 mm and Ø 16.7 X 63 mm .
- Weight: 34.8 g
- Charger: AC 100 - 240 V, 50 /60 Hz, DC 4.5V
One difference noted between the TRAUS ENDO and its predicates is its slower rotational speed, as a result of the 32:1 gear reduction ratio. However, equivalence to predicates has been demonstrated through bench testing and this difference raises no new concerns.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo and contact information for Saeshin. The logo is a stylized blue cube. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea. The telephone number is +82 53 587 2341 and the fax number is +82 53 580 0999.
### Clinical Tests
A non-clinical evaluation, based on literature research, has been done. The evaluation of the applicable market data showed that TRAUS ENDO Dental handpieces, ACL(B)-41EP, ACL(B)-42EP and ACL(B)-45EP, do not pose known or new clinical risks than similar medical devices currently on the market. Based on those results clinical test have not been executed.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo and contact information for SAESHIN. The logo is a blue geometric shape. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu 711-814, Korea. The telephone number is +82 53 587 2341 and the fax number is +82 53 580 0999.
#### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate devices (K123582 and K111616), the subject device TRAUS ENDO has nearly identical:
- Intended Use .
- Device Design .
- Composition of materials .
- · Technical Specifications
The TRAUS ENDO is similar to other commercially available products based on the intended use, the claims, the material composition employed. The following tables outlines and differences between the TRAUS ENDO and the predicate devices.
| Parameter | Proposed<br>TRAUS ENDO | Predicate<br>ENDO a class | Predicate<br>E-CUBE |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K143411 | K123582 | K111616 |
| Manufacturer | Saeshin Precision Co., Ltd | Saeyang Microtech Co., Ltd. | Saeshin Precision Co., Ltd |
| Intended Use | TRAUS ENDO, ACL(B)-41EP,<br>ACL(B)-42EP, and ACL(B)- 45EP are indicated for use by dentists in standard endodontic procedures using rotary files and rotary endodontic drills(Gates- Glidden). | ENDO a class is indicated for application area extends to endodontic procedures using a root canal instrument which is intended by the manufacturer for use in mechanical and rotary preparation of root canals. | E-CUBE is indicated for use by dentists in standard endodontic procedures using rotary endodontic files and rotary endodontic drills(Gates- Glidden). |
| Device Design<br>Components | Contra-Angle Handpiece<br>Motor Handpiece Charging Station<br>Power Cord | Contra-Angle Handpiece<br>Motor Handpiece<br>Charging Stand<br>Motor Handpiece Cap<br>Power Cord<br>Oil Guide | Contra Box<br>Micro Motor<br>Contra-Angle Handpiece<br>Foot-Control Pedal<br>Adapter |
| Parameter | Proposed<br>TRAUS ENDO | Predicate<br>ENDO a class | Predicate<br>E-CUBE |
| Operational Mode | AC power and Gear driven | AC power and Gear driven | AC power and Gear driven |
| Reduction Ratio | 16:1, 20:1, 32:1 | 4:1, 10:1, 16:1 and 20:1 | 1:1, 4:1, 10:1, 16:1, 20:1 |
| Coupling Dimension | N/A | ISO 3964 connection | ISO 3964 connection |
| Chuck Type | Latch Type | Latch Type | Latch Type |
| Shank Type | Type 1 | Type 1 | Type 1 |
| Composition of Materials | | | |
| Gear | Stainless steel (SUS420F) | Not Known | Stainless steel (SUS420F) |
| Shank | Stainless steel (Biodur<br>Trimtite) | | Stainless steel (Biodur<br>Trimtite) |
| Head | Brass (C3604BD-F) | | Brass (C3604BD-F) |
| Chuck | Stainless steel (SUS420F) | | Stainless steel (SUS420F) |
| Handle | Aluminum (AL6061) | | Aluminum (AL6061) |
| Patient-Contacting | Chuck, Head | | Chuck, Head |
| Operator-Contacting | Handle, Head | | Handle, Head |
| Technical Specification | | | |
| Supply Voltage | 100-240 V AC 50/60 Hz | 100-240 V | 100-240 V AC 50/60 Hz |
| Operational Voltage | DC 4.5 V | DC 2.4 V | DC 4.5 V, 0.4 VA |
| Gear ratio | 16:1, 20:1 and 32:1 | 4:1, 10:1, 16:1 and 20:1 | 1:1, 4:1, 10:1, 16:1 and 20:1 |
| Torque | 1 -4 Ncm | 0.1 - 4.0 Ncm | 0.6 – 5.2 Ncm |
| Motor Speed | 100 – 12,000 rpm | 140 - 500 rpm | 1,000 - 13,000 rpm |
| Lubricant | | | |
| Chemical Composition | N/A | N/A | K073353 (DO-ALL Dental<br>handpiece Lubricant/ProDrive<br>Systems Inc.) |
| 510k# | | | |
| Biocompatible<br>Delivery system | | | |
| Sterilization | Contra-angle at 132 °C for 4<br>minutes in Autoclave. | Contra-angle at 121 °C for 20<br>minutes or 132 °C for 15<br>minutes in Autoclave. | Contra-angle at 134 °C for 4<br>minutes in Autoclave. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo and contact information for SAESHIN. The logo is a blue geometric design. The contact information includes the address '# 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 711-814, Korea' and the telephone number '+82 53 587 2341' and fax number '+82 53 580 0999'.
{8}------------------------------------------------
| Parameter | Proposed<br>TRAUS ENDO | Predicate<br>ENDO a class | Predicate<br>E-CUBE | |
|-----------------------|------------------------|---------------------------|---------------------|---------|
| Performance Standards | IEC60601-1 | Conform | Conform | Conform |
| | IEC60601-1-2 | Conform | Conform | Conform |
| | ISO14457 | Conform | Conform | Conform |
## 9. Bench Testing
The following bench tests were performed on the TRAUS ENDO Dental Handpieces, ACL(B)-41EP, ACL(B)-42EP, to verify conformity to standards and demonstrate substantial equivalence to the predicates:
- IEC 60601-1: Medical Electrical Equipment Part 1 (2005/(R)2012 AND a1:2012) .
- IEC 60601-1-2: Medical Electrical Equipment Part 1-2 (2007-03) .
- ISO 14457: Dentistry-Handpieces and Motors (2012-09-15) .
- ISO 17665-1: Sterilization of health care products-Moist heat Part1 (2006-08-15) .
- . ISO 1797-1: Shanks for rotary instrument - Part1 (1st edition Amendment)
- ISO 10993-1: Biological Evaluation of Medical Devices Part 1 (4th Edition) .
TRAUS ENDO Dental Handpieces, ACL(B)-41EP, ACL(B)-45EP samples were compliant with standards and demonstrated substantial equivalence to the predicates.
# 10. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic, 21 CFR Part 807, and based on the information provided in this premarket notification Saeshin Presicion Co., Ltd., we conclude that the TRAUS ENDO has been determined to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.