K211886 · Shenzhen Wisonic Medical Technology Co. , Ltd. · IYN · Dec 2, 2021 · Radiology
Device Facts
Record ID
K211886
Device Name
Clivia series Diagnostic Ultrasound System
Applicant
Shenzhen Wisonic Medical Technology Co. , Ltd.
Product Code
IYN · Radiology
Decision Date
Dec 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Clivia series Diagnostic Ultrasound System is applicable for adults, pediatric patients and neonates. It intended for use in Fetal/Obstetrics, Abdominal/GYN, Pediatrics, Small Organ(breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric), peripheral vessel and Urology exams. The operator for Clivia series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics. Modes of operation include: B-Mode, Color Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, 3D/4D, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging(TDI), Anatomic M(AMM) and combined mode.
Device Story
General-purpose diagnostic ultrasound system; acquires, processes, and displays ultrasound data for clinical diagnosis. Inputs: ultrasound transducer signals; optional barcode reader; ECG trace synchronization. Operation: controlled via touch screen or control panel; supports B-Mode, M-Mode, Color/Power Doppler, PW/CW Doppler, 3D/4D, Holo PW, Tissue Harmonic Imaging, TDI, and AMM. Output: real-time ultrasound images and anatomical measurements displayed on screen; supports DICOM storage/archiving and printing. Used in hospitals/clinics by professional clinical staff. Battery-powered for portable use. Healthcare providers use images and analysis packages to inform clinical decision-making. Benefits: non-invasive diagnostic imaging for various patient populations and anatomical structures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation), electrical safety, EMC, and software verification/validation.
Technological Characteristics
General-purpose ultrasound system; Track 3. Materials evaluated per ISO 10993. Modes: B, M, Color, Power, PW/CW Doppler, 3D/4D, Holo PW, TDI, AMM. Connectivity: wireless, DICOM, barcode reader. Power: battery-operated. Safety standards: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2. Acoustic output per NEMA UD 2-2004 (R2009).
Indications for Use
Indicated for adults, pregnant women, pediatric patients, and neonates for diagnostic ultrasound imaging in Fetal/Obstetrics, Abdominal/GYN, Pediatrics, Small Organ (breast, thyroid, testes), Neonatal/Adult Cephalic, Trans-rectal, Trans-vaginal, Musculoskeletal, Thoracic/Pleural, Cardiac, peripheral vessel, and Urology exams.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic Ultrasound System (K173471)
Navi e/Navi s/Navi X Diagnostic Ultrasound System (K180461)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen Wisonic Medical Technology Co., Ltd. % Jiang Xiaosan Management Representative 1st, 2nd, 5th & 6th Floor, NO.6 Building, Pingshan Technology Park, Taoyuan Street, Nanshan District Shenzhen, Guangdong 518055 CHINA
December 2, 2021
### Re: K211886
Trade/Device Name: Clivia series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: October 25, 2021 Received: November 1, 2021
### Dear Jiang Xiaosan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211886
Device Name Clivia series Diagnostic Ultrasound System
#### Indications for Use (Describe)
The Clivia series Diagnostic Ultrasound System is applicable for adults, pediatric patients and neonates. It intended for use in Fetal/Obstetrics, Abdominal/GYN, Pediatrics, Small Organ(breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Musculoskeletal (Conventional), Musculoskeletal (Superticial), Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric), peripheral vessel and Urology exams.
The operator for Clivia series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics.
Modes of operation include: B-Mode, Color Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, 3D/4D, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging(TDI), Anatomic M(AMM) and combined mode.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K211886
# 510(k) Summary
- 1. Submitter
| Manufacturer: | Shenzhen Wisonic Medical Technology Co., Ltd. | | |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--|
| Address: | 1st, 2nd, 5th & 6th Floor, NO.6 Building, Pingshan Technology Park,<br>Taoyuan Street, Nanshan District, Shenzhen. Guangdong, 518055, P.R.<br>CHINA | | |
| Contact person: | Name: | Jiang Xiaosan | |
| | Phone: | +86-755-86007788 | |
| | Fax: | +86-755-86007799 | |
| Date prepared: | June 08, 2021 | | |
| 2. Device | | | |
| Name of Device: | Clivia series Diagnostic Ultrasound System | | |
| Models: | Clivia 90Elite, Clivia 90Exp, Clivia 90Nova, Clivia 90, Clivia 90T,<br>Clivia 90Pro, Clivia 90Plus, Clivia 90Go | | |
| Common/Usual Name: | Diagnostic Ultrasound System | | |
| Regulatory Class | II | | |
| Product Code: | IYN, IYO, ITX | | |
### 3. Device Description
The proposed Clivia series Diagnostic Ultrasound System is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement and analysis of the human body and fluid that provides digital acquisition, processing and display capabilities.
The Clivia series Diagnostic Ultrasound System can be controlled both by touch screen or the control panel. Its basic function is to acquire and display ultrasound data in B-Mode, M-Mode, Color-Mode, Power(Dirpower)-Mode, PW-Mode, 3D/4D, Holo PW, Anatomic M(AMM), Tissue Doppler Imaging(TDI) and the combined modes. The system can also measure anatomical structures and offer software analysis packages performance to provide information based on which the competent health care professionals can make the diagnosis.
Ten different models of probes are available for the Clivia series, that is C6-1B-H, C6-2-H, C9-3-H, L15-4NB-H, L22-10-H, LH15-6-H, SP5-1-H, P8-3-H, EV11-3-H, D7-2-H.
The Clivia series Diagnostic Ultrasound System has the capability for displaying the patient's ECG and trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. ECG is not intended for monitoring or diagnosis.
{4}------------------------------------------------
The Clivia series Diagnostic Ultrasound System has four batteries in total that allows for scanning or other operation without the need to plug in to an electrical outlet. The system is capable of wireless communication and a barcode reader is available to be used as an input device. System meets DICOM requirements to support users image storage and archiving needs and allows for output to printing devices.
- 4. Indications for Use
The Clivia series Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It intended for use in Fetal/Obstetrics, Abdominal/GYN, Pediatrics, Small Organ(breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric), peripheral vessel and Urology exams.
The operator for Clivia series Diagnostic Ultrasound System is intended for professional clinical staff or the qualified and trained personnel with experience in the use of ultrasound diagnostic equipment. The device is intended to be used in hospital or clinics.
Modes of operation include: B-Mode, Color Mode, Power (Dirpower)-Mode, PW Doppler Mode, CW Doppler Mode, 3D/4D, Holo PW, Tissue Harmonic Imaging, Tissue Doppler Imaging(TDI), Anatomic M(AMM) and combined modes.
| Predicate<br>Device | Manufacturer | Device Name | 510(k) Number |
|---------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------------|---------------|
| Predicate<br>device | GE<br>Medical Systems<br>Ultrasound and Primary Care<br>Diagnostics, LLC | Venue | K180599 |
| Reference<br>device | Shenzhen<br>Mindray<br>Bio-medical Electronics Co.,<br>LTD. | DC-80/DC-80<br>PRO/DC-80<br>EXP/DC-80S/DC-85 Diagnostic<br>Ultrasound System | K173471 |
| Reference<br>device | Shenzhen Wisonic Medical<br>Technology Co., Ltd. | Navi e/Navi s/Navi X Diagnostic<br>Ultrasound System | K180461 |
5. Comparison of Technological Characteristics between Proposed Subject Device and Predicate Device
Clivia series Diagnostic Ultrasound System is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness:
- The Proposed device and predicate device Venue (K180599) have similar intended use, the intend ● use of the Proposed device are covered in the intended use of the predicate device.
- The proposed device has similar operation modes to the predicate device Venue (K180599), except for Holo PW. The Holo PW has been cleared in reference device (K180461);
- The proposed device has the same function with the predicate device Venue (K180599), except for wiGuide. For wiGuide, it has identical operating principals and specifications as the NGS function of the reference device (K180461), the difference between wiGuide and NGS is only the name.
{5}------------------------------------------------
# wisonic
- The proposed device has similar ports to the predicate device Venue (K180599)
- The proposed device has similar peripheral devices to the predicate device Venue (K180599), except for Magnetized Cup and Foot switch. For Magnetized Cup, it has been cleared in reference device (K180461). For Foot switch, it can be considered same as reference device (K173471).
- The acoustic power levels of the proposed device are below the limits of FDA, which is the same as the predicate device (K180599).
- The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The proposed device and predicate device Venue (K180599) have been designed in compliance with approved electrical and physical safety standards.
The difference in technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device.
### 6. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 6.1 Biocompatibility testing
The biocompatibility evaluation for the Clivia series Diagnostic Ultrasound System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,' and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
6.2 Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Clivia series Diagnostic Ultrasound System. The system complies with the IEC 60601-1 and IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC.
## 6.3 Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "Moderate" level of concern.
{6}------------------------------------------------
# wisonic
Acoustic testing is conducted on the Clivia series Diagnostic Ultrasound System in accordance with the NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic ultrasound equipment revision 3.
# 6.5 Clinical study
The subject of this premarket submission, Clivia series Diagnostic Ultrasound System, did not require clinical studies to support substantial equivalence.
# 7. Conclusion
The differences between the Clivia series Diagnostic Ultrasound System and its predicate device do not raise different questions of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the Clivia series Diagnostic Ultrasound System should perform as intended in the specified use conditions.
From the results of performance data described, Wisonic concludes that the Clivia series Diagnostic Ultrasound System is substantially equivalent to the predicate device in terms of safety and effectiveness.