MagnetOs Easypack Putty is intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. In the posterolateral spine, MagnetOs Easypack Putty must be used with autograft as a bone graft extender. The osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. MagnetOs Easypack Putty resorbs and is replaced with bone during the healing process.
Device Story
MagnetOs Easypack Putty is a synthetic, resorbable, osteoconductive bone void filler; composed of 65–75% tri-calcium phosphate and 25-35% hydroxyapatite granules mixed with a synthetic polymeric binder. Supplied in open-ended syringes for clinical use; manipulated by surgeons to conform to bony defect contours. Operates via osteoconduction; provides scaffold for new bone deposition; resorbs and is replaced by host bone during healing. Used as autograft extender in posterolateral spine. Benefits include filling voids to facilitate natural bone remodeling. Gamma-sterilized for single use.
Clinical Evidence
No clinical studies conducted. Substantial equivalence supported by bench testing (physicochemical, crystallographic, and material characterization) and a rabbit posterolateral spine fusion animal model. Animal study compared subject device, primary predicate, and autograft control over 12 weeks using manual palpation, flexibility testing, radiography, micro-CT, and histology/histomorphometry.
Technological Characteristics
Composition: 65-75% tri-calcium phosphate, 25-35% hydroxyapatite granules, synthetic polymeric binder. Form factor: putty in open-ended syringes. Sterilization: gamma radiation. Biocompatibility: ISO 10993 compliant. Testing standards: FDA Class II Special Controls Guidance for Resorbable Calcium Salt Bone Void Filler.
Indications for Use
Indicated for filling bony voids or gaps of the skeletal system, specifically the posterolateral spine, in patients requiring bone graft extension. Must be used with autograft in the posterolateral spine. Defects may be surgically created or traumatic in origin; not for use in defects intrinsic to bony structure stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
September 2, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The logos are displayed side-by-side.
Kuros Biosciences B.V. % Angela Paterson Senior Consultant Compliance Solutions Ltd Suite 10, Dunswood House, 1 Dunswood Road, Cumbernauld Glasgow G67 3EN United Kingdom
Re: K211201
Trade/Device Name: MagnetOs Easypack Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: July 28, 2021 Received: August 3, 2021
Dear Angela Paterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211201
Device Name MagnetOs Easypack Putty
Indications for Use (Describe)
MagnetOs Easypack Putty is intended to fill bony voids or gaps of the skeletal system, i.e. posterolateral spine. In the posterolateral spine, MagnetOs Easypack Putty must be used with autograft as a bone graft extender. The osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the bony structure.
MagnetOs Easypack Putty resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:inline-block"><span style="font-size:20px">☒</span></div> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block"><span style="font-size:20px">☐</span></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MagnetOs Easypack Putty
## 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
## Submitter's Name:
Kuros Biosciences B.V.
## Submitter's Address:
Kuros Biosciences B.V. Prof. Bronkhorstlaan 10, building 48 3723 MB Bilthoven The Netherlands
## Establishment Registration Number:
3008147766
## Contact Person:
Angela Paterson
Telephone + 44 7884 274 220
## Date Prepared:
2nd September 2021
{4}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MaqnetOs Easypack Putty
The Device Classification Information regarding MagnetOs Easypack Putty is summarized below.
## Primary Product Code:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|---------------------------------------------|-----------------|-----------------|-------------------------|
| 888.3045 | Resorbable calcium salt<br>bone void filler | 2 | MQV | Orthopedic |
#### Device Trade Name:
MagnetOs Easypack Putty
## Device Common Name:
MagnetOs Easypack Putty
#### Intended Use:
MagnetOs Easypack Putty is intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. In the posterolateral spine, MagnetOs Easypack Putty must be used with autograft as a bone graft extender. The osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure.
MagnetOs Easypack Putty resorbs and is replaced with bone during the healing process.
## Summary of Substantial Equivalence:
MagnetOs Easypack Putty is substantially equivalent in indications and design principles to the following cleared and legally marketed predicate devices:
- . MagnetOs Putty (K181958, K171563), the primary predicate, is highly similar to the subiect device used for the comparison animal spine study. It consists of the identical unmodified MagnetOs granules component. which is responsible for the mode of action of the subject device, together with a synthetic polymeric binder, which is highly similar to the binder used in the subject device.
- . Osteo3 ZP Putty (K193075) is packed in identical primary packaging and is indicated for the same intended use as MagnetOs Easypack Putty.
- . Actifuse ABX (K082575) is packed in comparable primary packaging and has similar indications for use, as a bone void filler in spine.
- MagnetOs Granules (Xpand Biotechnology B.V., K161859) is cleared for the same . intended use as MagnetOs Easypack Putty. Identical MagnetOs granules are present in MagnetOs Easypack Putty and MagnetOs Putty.
Note: Xpand Biotechnology B.V. is the previous company name of Kuros Biosciences B.V.
{5}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MagnetOs Easypack Putty
The subject device, MagnetOs Easypack Putty, and predicate devices MagnetOs Putty, Osteo3 ZP Putty, Actifuse ABX, and MaqnetOs Granules have the same intended use, the same product classification and product code (MQV) and have similar Indications for Use.
| Property | 510(k) Subject<br>Device | Primary<br>Predicate Device | Equivalent<br>Device | Equivalent<br>Device | Equivalent<br>Device |
|--------------------------|------------------------------|------------------------------|----------------------|----------------------|------------------------------|
| Device Name | MagnetOs<br>Easypack Putty | MagnetOs Putty | Osteo³ ZP<br>Putty | Actifuse<br>ABX | MagnetOs<br>Granules |
| Device<br>Manufacturer | Kuros<br>Biosciences<br>B.V. | Kuros<br>Biosciences<br>B.V. | SIRAKOSS<br>Ltd. | Baxter | Kuros<br>Biosciences<br>B.V. |
| 510(k) | No | K181958,<br>K171563 | K193075 | K082575 | K161859 |
| Device<br>Classification | II | II | II | II | II |
| Product Code | MQV | MQV | MQV | MQV | MQV |
| Use<br>Environment | Medical/<br>Clinical | Medical/<br>Clinical | Medical/<br>Clinical | Medical/<br>Clinical | Medical/<br>Clinical |
MagnetOs Easypack Putty is substantially equivalent to the predicate devices MagnetOs Putty, Osteo3 ZP Putty, Actifuse ABX, and MagnetOs Granules with respect to design, structure, materials, and mechanism of action, and has similar Indications for Use in posterolateral spine. MagnetOs Putty is the primary predicate device used for animal performance testing studies.
For the three MagnetOs devices, the MagnetOs granules component is identical.
Between the MagnetOs Easypack Putty subject device and the MagnetOs Putty, Osteo3 ZP Putty, Actifuse ABX, and MagnetOs Granules predicate devices, there are limited differences. These aspects do not affect the safety and biocompatibility of MagnetOs Easypack Putty because of the identical nature of the building blocks of these materials and the site of application (bone).
The subject device and the Osteo3 ZP Putty and Actifuse ABX predicate devices are provided in equivalent primary packaging (open-ended syringes).
The granules used in MagnetOs Easypack Putty subject device are identical to the granules used in the MagnetOs Putty primary predicate device. The polymeric binders in these two devices share the same identity. The only difference between the two binders is the slightly different molecular weight to adjust softness. The nature and essential characteristics of the granules remain unaltered when mixed with the polymeric binder.
As a result, MagnetOs Easypack Putty, and all predicate devices, including MagnetOs Granules (which consist of granules alone, without a binder/carrier), are equivalent in terms of mechanism of action (same fundamental technology): they all provide calcium salt bone void fillers which resorb and are replaced by bone during the natural process of bone healing. All devices have equivalent Indications for Use.
{6}------------------------------------------------
# Kuros Biosciences Traditional 510(k) For MaqnetOs Easypack Putty
## Device Description:
MagnetOs Easypack Putty is a synthetic, resorbable, osteoconductive bone void filler for the repair of bony defects.
MagnetOs Easypack Putty consists of 65–75% tri-calcium phosphate (TCP – Саз(РО4)2) and 25-35% hydroxyapatite (HA - Ca10(PO4)6(OH)2) granules, premixed with a synthetic polymeric binder that provides cohesion between the granules.
New bone will be deposited on the surface of the graft when placed next to viable host bone. The graft resorbs and is replaced by bone during the natural process of bone remodeling.
MagnetOs Easypack Putty is a ready-to-use product. Pressure applied by manipulation allows users to shape MagnetOs Easypack Putty to conform to the contours of bony defects.
MagnetOs Easypack Putty is provided in open-ended syringes in a range of product volumes. MagnetOs Easypack Putty is gamma-sterilized and sterile packaged for single use only.
#### Technological Characteristics:
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included chemical composition, physical properties, biocompatibility, and performance characteristics in accordance with FDA recognized consensus standards and FDA guidance documents as applicable.
## Non-Clinical Tests (Performance/Physical Data):
Non-clinical testing data were submitted according to the quidance documents Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device (issued June 2003) and Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (issued January 2016). The non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included chemical composition, physical properties, sterilization, material-mediated pyrogenicity, sterile barrier shelf life, product shelf life, and biocompatibility.
Bacterial endotoxin testing of MagnetOs Easypack Putty was conducted and met endotoxin limits.
{7}------------------------------------------------
# Kuros Biosciences Traditional 510(k) For MaqnetOs Easypack Putty
Bench testing and animal data demonstrated that the safety and effectiveness of the MagnetOs Easypack Putty is equivalent to the predicate devices. A summary is given below.
Testing was performed to evaluate the physicochemical and crystallographic characteristics of MagnetOs Putty. Characterization tests included:
- Product visual inspection
- Composition, crystallinity, and Ca/P ratio (X-ray diffraction (XRD)) ●
- Microstructure (SEM) ●
- Identity (Fourier Transform Infrared (FTIR) spectroscopy) ●
- Porosity (mercury intrusion porosimetry)
- Polymer inherent viscosity (Ubbelohde Viscometry) ●
- Polymer chemical composition (NMR) ●
- . Organic volatile impurities (OVI)
- Water content (Coulometric Karl Fischer Titration)
- Trace elemental analysis (ICP-MS)
- In vitro dissolution and pH ●
- Polymer irrigation test ●
- . Calcium ion release (ICP-MS)
#### Biocompatibility Testing
Biocompatibility of the subject device was assessed using the methodology described in ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11, ISO 10993-12, ISO 10993-18, USP <87>, USO <88>, USP 43 - NF 38 and European Pharmacopeia, 10th edition, 2020.
#### Animal Studies
MagnetOs Putty is the primary predicate device used for animal performance testing studies. The performance of the MagnetOs Easypack Putty was established in a posterolateral spine fusion animal model justifying that MagnetOs Easypack Putty can be used as an autograft extender (1:1) in posterolateral spine.
Animal testing was performed in a rabbit posterolateral spine fusion model to demonstrate substantial equivalence to the primary predicate device (K181958, K171563). Animals were evaluated after implantation with the subject device, the primary predicate device, and autograft (positive control) up to 12 weeks. Evaluation endpoints included manual palpation. range of motion/flexibility testing, plain and high-resolution radiography, micro-computed tomography (micro-CT) imaging, undecalcified histologic evaluation, and histomorphometric analysis.
Decalcified paraffin histology sections also were graded according to ISO 10993-6 (Annex E).
## Clinical Studies
No clinical studies were conducted to prove substantial equivalence as part of the submission.
{8}------------------------------------------------
# Kuros Biosciences Traditional 510(k) For MagnetOs Easypack Putty
## Safety and Effectiveness/Conclusion:
Based on the information presented in this 510(k) premarket notification, MagnetOs Easypack Putty is considered substantially equivalent to predicate devices. MagnetOs Easypack Putty is as safe and effective as currently marketed predicate devices.
It is our determination that the MagnetOs Easypack Putty is as safe, as effective and performs within its design specifications, and is substantially equivalent to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.