MagnetOs is an implant intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. MagnetOs must be used with autograft as a bone graft extender in the posterolateral spine. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. MagnetOs resorbs and is replaced with bone during the healing process.
Device Story
Synthetic, osteoconductive, resorbable bone void filler; composed of Tri-Calcium Phosphate (TCP) and Hydroxyapatite (HA). Implanted into bony voids or gaps in the posterolateral spine; used as bone graft extender with autograft. Resorbs over time; replaced by natural bone during healing. Used by surgeons in clinical settings for orthopedic repair. Provides scaffold for bone growth; facilitates skeletal defect healing.
Clinical Evidence
No human clinical data. Substantial equivalence supported by non-clinical bench testing (chemical composition, physical properties, biocompatibility per ISO 10993 series) and a comparative posterolateral spine fusion animal model. Bench testing demonstrated equivalent performance to predicate devices.
Technological Characteristics
Synthetic bone void filler; 65-75% Tri-Calcium Phosphate (Ca3(PO4)2) and 25-35% Hydroxyapatite (Ca5(PO4)3OH). Granular or chip form. Gamma sterilized. Single-use. Evaluated via XRD, FTIR, ICP/MS, SEM, and mercury intrusion porosimetry.
Indications for Use
Indicated for patients requiring bone void filling in the posterolateral spine. Used as a bone graft extender with autograft for surgically created or traumatic osseous defects not intrinsic to bony structure stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2017
Xpand Biotechnology B.V. Ms. Yvonne P. Bovell QA/RA Manager Professor Bronkhorstlaan 10D, Building 48 3723 MB Bilthoven The Netherlands
Re: K161859
Trade/Device Name: MagnetOs Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: January 19, 2017 Received: January 23, 2017
Dear Ms. Bovell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Division Director Division of Orthopaedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES. Food and Drug Administration
### Indications for Use
510(k) Number (if known) K161859
Device Name MagnetOs
### Indications for Use (Describe)
MagnetOs is an implant intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. MagnetOs must be used with autograft as a bone graft extender in the posterolateral spine. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. MagnetOs resorbs and is replaced with bone during the healing process.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{4}------------------------------------------------
# 510(k) Summary MagnetOs
Date: 19 January 2017
# ADMINISTRATIVE INFORMATION
Submitter details:
| Name | Xpand Biotechnology B.V. |
|-----------|----------------------------------------------------------------------------------|
| Address | Professor Bronkhorstlaan 10, building 48<br>3723 MB Bilthoven<br>The Netherlands |
| Telephone | +31-30-2297280 |
| Fax | +31-30-2297299 |
Contact person:
| Name | Yvonne P. Bovell |
|-----------|----------------------------|
| Position | QA/RA Manager |
| Telephone | +31-30-2297212 |
| Email | yvonne.bovell@signifix.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | MagnetOs |
|----------------------------|-------------------------------------------------|
| Classification Name | Resorbable calcium salt bone void filler device |
| Classification Regulations | 21 CFR 888.3045, Class II |
| Product Code | MQV |
| Classification Panel | Orthopaedic and Rehabilitation Devices Panel |
| Reviewing Branch | Restorative Devices Branch |
{5}------------------------------------------------
## INTENDED USE
MagnetOs is an implant intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. MaqnetOs must be used with autograft as a bone graft extender in the posterolateral spine. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure.
MagnetOs resorbs and is replaced with bone during the healing process.
# DEVICE DESCRIPTION
MagnetOs is a synthetic, osteoconductive and resorbable bone void filler for the repair of bony defects, containing 65-75% Tri-Calcium Phosphate (TCP, Ca3(PQ4)2) and 25-35% Hydroxyapatite (HA, Cas(PO4)3OH).
MagnetOs is gamma sterilized, comes in various sizes in granular or chip form and is sterile packaged for single use only.
## EQUIVALENCE TO MARKETED DEVICE
MagnetOs is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
- Progentix Orthobiology B.V., CuriOs™, cleared under K090641;
- Biomatlante SA, MBCP, cleared under K032268, and ●
- . IsoTis N.V.. OsSatura™ BCP. cleared under K030131.
The subject device and predicate devices K090641, K032268 and K030131 have the same intended use, the same product classification and product code (MQV), and have similar Indications for Use.
MagnetOs granules are substantially equivalent to the predicate devices CuriOs™ (K090641), MBCP (K0302268) and OsSatura™ BCP (K030131) with respect to design, structure, materials, mechanism of action, and similar Indications for Use in the posterolateral spine.
Between MagnetOs and the CuriOs™, MBCP and OsSatura™ BCP predicate devices, the limited difference are the total porosity, ratio of beta-tricalcium phosphate and hydroxyapatite.
These aspects do not affect the safety and biocompatibility of MagnetOs because of the identical nature of the building blocks of these materials and the site of application (bone).
{6}------------------------------------------------
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included chemical composition, physical properties, biocompatibility, and performance characteristics. Biocompatibility of the device was assessed using the methodology described in ISO 10993-1, ISO10993-3, ISO 10993-5, ISO 10993-6. ISO 10993-10 and ISO10993-11.
Material characterization performed included the following:
- . chemical composition and crystallinity was analyzed by x-ray diffraction (XRD), and Fourier transform infrared spectroscopy (FTIR),
- . trace elemental analysis was performed by inductively coupled plasma/mass spectroscopy (ICP/MS).
- surface microstructure was evaluated by scanning electron microscopy (SEM), and ●
- physical properties including porosity by mercury intrusion porosimetry, and dissolution.
The analytical characterization demonstrated equivalent chemical composition, physical properties and performance characteristics for the subject MagnetOs and the predicate devices.
The performance of the subject MagnetOs was compared to that of the predicate devices in a posterolateral spine fusion animal model. The results of the study demonstrated that the performance of the subject device was equivalent to that of the predicate.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Test results included in Sections 15, 18 and 19 of this submission demonstrate that MagnetOs generated equivalent results in the comparative bench, biocompatibility and pre-clinical animal testing performed compared to the predicate devices.
Overall. MagnetOs has the following similarities to the predicate devices:
- has the same intended use,
- has the same product classification and product code (MQV), ●
- has similar Indications for Use,
- uses the same operating principle,
- incorporates the same basic design,
- . incorporates the same or very similar materials, and is manufactured at the same facility using very similar processes as predicate, CuriOs™ Granules (K090641).
### CONCLUSION
The above testing demonstrates that MagnetOs is as safe, as effective and performs as well as or better than the legally marketed predicate devices CuriOs™ (Progentix Orthobiology B.V.), OsSatura™ BCP (IsoTis N.V.) or MBCP (Biomatlante SA).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.