K211183 · Mizuho America, Inc. · HCH · Mar 2, 2022 · Neurology
Device Facts
Record ID
K211183
Device Name
Sugita AVM Microclips
Applicant
Mizuho America, Inc.
Product Code
HCH · Neurology
Decision Date
Mar 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures. Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.
Device Story
Sugita AVM Microclips are permanent surgical implants used to occlude fragile capillary vessels during arteriovenous malformation (AVM) surgery. Manufactured from Elgiloy (cobalt-chromium-molybdenum alloy), the clips are available in 2mm, 3mm, 4mm, and 5mm sizes. They are delivered to the surgical site using a dedicated Sugita Microclip Applier. The clips exert a predetermined holding force of 50 to 70g to provide mechanical occlusion. The device is intended for use by neurosurgeons in an operating room setting. The current modification updates the device labeling to reflect MR Conditional status for both 1.5-Tesla and 3-Tesla environments and provides validated steam sterilization parameters for end-user reprocessing.
Clinical Evidence
No clinical data. Bench testing only. MRI compatibility testing performed per ASTM standards (F2052-15, F2213-17, F2182-19, F2119-13) demonstrated safety in 1.5T and 3T environments, with a maximum temperature rise of 1.5°C and imaging artifacts extending approximately 10mm. Sterilization validation performed per ISO 17665-1 achieved a sterility assurance level (SAL) of 10^-6.
Technological Characteristics
Materials: Elgiloy (cobalt-chromium-molybdenum alloy, ASTM F1058). Form factor: 2mm, 3mm, 4mm, 5mm clips. Mechanism: Mechanical occlusion with 50-70g holding force. Connectivity: None. Sterilization: Steam (pre-vacuum, 132°C/4 min or 134°C/3 min). MRI Safety: MR Conditional (1.5T and 3T).
Indications for Use
Indicated for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures in patients requiring surgical intervention.
Regulatory Classification
Identification
An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.
{0}------------------------------------------------
March 2, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Mizuho America, Inc. Richard Chadwick Senior Manager, Operations and QA 30057 Ahern Avenue Union City, California 94587
#### Re: K211183
Trade/Device Name: Sugita AVM Microclips Regulation Number: 21 CFR 882.5200 Regulation Name: Aneurysm Clip Regulatory Class: Class II Product Code: HCH Dated: January 25, 2022 Received: January 31, 2022
#### Dear Richard Chadwick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradvan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K211183
Device Name Sugita AVM Microclips
Indications for Use (Describe)
The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.
Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
[X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Mizuho America, Inc. The logo features a blue circular graphic on the left, followed by the word "MIZUHO" in large, bold, blue letters. Below "MIZUHO" is the text "Medical Innovation" in a smaller font. Underneath that is the text "Mizuho America, Inc." in a similar font size.
# K211183
# 510(k) SUMMARY
## Applicant Information:
| Owner Name: | Mizuho America, Inc. |
|------------------------------------|--------------------------------------------|
| Address: | 30057 Ahern Avenue<br>Union City, CA 94587 |
| Phone number: | +1(510) 324-4500 |
| Establishment Registration Number: | 1223656 |
| Contact Person: | Richard Chadwick |
| Date Prepared: | February 28, 2022 |
| Device Information: | |
| Device Classification: | Class II |
| Trade Name: | Sugita AVM Microclips |
| Common name: | Aneurysm Clip |
| Classification name: | Aneurysm Clip |
| Regulation number: | 882.5200 |
| Product Code: | HCH |
## Predicate Device:
Sugita AVM Microclips (K960037).
## Device Modification:
The following changes have been made to the Sugita AVM Microclips:
- . The device has been demonstrated to be safe for imaging in both 1.5- Tesla and 3-Tesla MR systems.
- New parameters for steam sterilization are provided for end users.
## Device Description:
The Sugita AVM Microclips were cleared by the FDA through 510(k) K960037. The Sugita AVM Microclips are manufactured from Elgiloy, a cobalt-chromiummolybdenum alloy. They are designed for the occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures. The Sugita AVM Microclips are available in the following sizes: 2mm, 3mm, 4mm and 5mm. Other than the modifications identified above, there have been no other changes to the Sugita AVM Microclips.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Mizuho Medical Innovation. The logo features a blue circular graphic with white lines emanating from the center, resembling a stylized globe or sun. To the right of the graphic is the word "MIZUHO" in a bold, blue sans-serif font. Below "MIZUHO" are the words "Medical Innovation" in a smaller, lighter font, followed by "Mizuho America, Inc." in a similar font.
### Indications for Use:
The Sugita AVM Microclips have the same Indications for Use as the predicate device.
The Sugita AVM Microclips are designed for occlusion of fragile capillary vessels to stop bleeding from arteriovenous malformations and other analogous venous structures.
Sugita AVM Microclips are not intended as a replacement for bipolar electrocoagulation of smaller vessels nor do they negate the need for immediate post-operative angiography following arteriovenous malformation surgery.
### Comparison of Technological Characteristics to the Predicate Device:
Both the Sugita AVM Microclips and the predicate device have the same intended use. Both are intended for occlusion of small vessels in Arteriovenous Malformations (AVM's) and other similar structures.
The materials, design, technological characteristics and operating principles of the current device are unchanged from the predicate cleared through K960037. See the table below.
Substantial equivalence in materials, technological characteristics, and performance of the Sugita AVM Microclips to the predicate device is outlined in the table below:
| Product | Sugita AVM Microclips | Sugita AVM Microclips | Conclusion |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------|
| 510(k) number | Subject Device – K211183 | K960037 | |
| Manufacturer | Mizuho America, Inc. | Mizuho America, Inc. | |
| Materials | Elgiloy (cobalt-chromium-<br>molybdenum alloy – ASTM<br>F1058) | Elgiloy (cobalt-chromium-<br>molybdenum alloy – ASTM<br>F1058) | SAME |
| Technological<br>Characteristics | | | |
| Designed with a<br>predetermined<br>holding force | Holding force of 50 to 70g | Holding force of 50 to 70g | SAME |
| Various Sizes for<br>specific<br>requirements | Four Sizes Available –<br>2mm, 3mm, 4mm and 5mm | Four Sizes Available –<br>2mm, 3mm, 4mm and 5mm | SAME |
| Delivery to<br>Surgical Site | Sugita Microclip Applier<br>designed specifically for<br>Sugita AVM Microclips to<br>aid in delivery to surgical<br>site | Microclip Applier designed<br>specifically for Sugita AVM<br>Microclips to aid in delivery<br>to surgical site | SAME |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Mizuho Medical Innovation. The logo consists of a circular graphic on the left and the word "MIZUHO" in blue, bold letters on the right. Below the word "MIZUHO" is the phrase "Medical Innovation" in smaller, lighter-colored letters. The circular graphic appears to be a stylized representation of a medical device or technology.
| Product | Sugita AVM Microclips | Sugita AVM Microclips | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------|
| 510(k) number | Subject Device – K211183 | K960037 | |
| Manufacturer | Mizuho America, Inc. | Mizuho America, Inc. | Conclusion |
| Steam Sterilized<br>by end user | Cycle Type: Pre-vacuum<br>• Temperature: 132°C<br>• Exposure Time: 4 min.<br>• Drying Time: 20 min.<br><br>• Temperature: 134°C<br>• Exposure Time: 3 min.<br>• Drying Time: 20 min. | Cycle Type: Pre-vacuum<br>• Temperature: 252°F<br>• Exposure Time: 20 minutes | Substantially<br>Equivalent |
| Performance<br>Testing | | | |
| MRI Safety | MR Conditional for Use in<br>Both 1.5T and 3T MR<br>Systems Based on<br>Testing:<br>-Magnetic Field<br>Interactions<br>-MRI Related Heating<br>-MRI Artifact Test | MR Conditional for Use in<br>1.5T MR Systems | Substantially<br>Equivalent |
#### Non-clinical Performance Testing
Performance testing was conducted to demonstrate the safety of the Sugita AVM Microclips for imaging in both 1.5-Tesla and 3-Tesla MR Systems and to provide labeling for MRI safety of the device conforming to ASTM F2503-13 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment and Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, Guidance for Industry and Food and Drug Administration Staff, document issued on May 20, 2021.
Additionally, sterilization validation testing was conducted to provide recommended steam sterilization parameters for the device.
| Test | Test Summary | Conclusions |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| MRI Compatibility Testing –<br>Magnetic Field Interaction (force) | Testing according to ASTM<br>F2052-15 -translational<br>attraction determined to be 14°. | Force no greater than<br>gravitational field. |
| MRI Compatibility Testing –<br>Magnetic Field Interaction<br>(torque) | Testing according to ASTM<br>F2213-17 – torque showed no<br>movement or alignment. | Magnetically induced torque is<br>not substantial and requires no<br>further evaluation. |
| MRI Compatibility Testing – MRI<br>Related Heat Testing | Testing according to ASTM<br>F2182-19 – maximum<br>temperature rise of 1.5°C after 15<br>min. of continuous scan. | Maximum Whole Body Averaged<br>SAR 2-W/Kg for 60 min.<br>continuous RF exposure. |
#### Non-clinical Performance Testing Summary
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Mizuho Medical Innovation. The logo consists of a blue circular graphic on the left, followed by the word "MIZUHO" in large, bold, blue letters. Below the word "MIZUHO" are the words "Medical Innovation" in smaller, thinner, black letters. The circular graphic on the left appears to be a stylized representation of a globe or sphere with curved lines.
| MIZUNO America, Inc.<br>MRI Compatibility Testing -<br>Imaging Artifact Testing | Testing according to ASTM<br>F2119-2013 – image artifact<br>extends approx. 10mm from<br>implant with 3-Tesla gradient<br>echo pulse sequence. | The presence of this implant<br>produces an imaging artifact.<br>Therefore, should carefully select<br>pulse sequence parameters if the<br>implant is located in the area of<br>interest. |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MRI Compatibility Testing -<br>Spatial Gradient Magnetic Field<br>Testing | 5mm Sugita AVM Microclips were<br>attached to a porcine blood vessel<br>in order to conduct a digital force<br>gauge-based Pull-Test. | Testing supports using a value of<br>2,000 gauss/cm (20-T/m) for the<br>MRI related labeling of the<br>Sugita AVM Microclips. |
| Steam Sterilization Validation<br>Testing | Testing according to ISO 17665-<br>1 was performed to validate<br>steam sterilization parameters for<br>the Sugita AVM Microclips using<br>an FDA cleared nonwoven wrap<br>to a sterility assurance level of<br>(SAL) of 10-6. | Validated steam sterilization<br>parameters are provided in the<br>product labeling. |
### Shelf-life
The Sugita AVM Microclips do not have a shelf life because they are provided non-sterile and are constructed of inert materials. Performance data are not needed to establish maintenance of device performance over the shelf life of the Sugita AVM Microclips because the materials of construction of these products are highly stable metal allovs with a long history of use in surgical implants.
### Biocompatibility
The proposed changes do not impact the contact duration or biocompatibility profile of the Sugita AVM Microclips.
#### Conclusion
The modified Sugita AVM Microclips subject to this submission are substantially equivalent to the predicate. Sugita AVM Microclips, in the following ways: they have the same intended use and Indications for Use, the same technological characteristics and operating principles, and incorporate the same design and materials.
Performance testing has demonstrated that the Sugita AVM Microclips are substantially equivalent to the predicate and are safe for imaging in both 1.5- Tesla and 3-Tesla MR systems. In addition, validated parameters for steam sterilization and cleaning are provided to update the instructions for end users of the device.
Page 4 of 4
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.