K960037 · Mizuho America, Inc. · HCH · Feb 5, 1997 · Neurology
Device Facts
Record ID
K960037
Device Name
SUGITA AVM MICROCLIP
Applicant
Mizuho America, Inc.
Product Code
HCH · Neurology
Decision Date
Feb 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5200
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Occlusion of small vessels as a compliment to bipolar coagulation.
Device Story
Sugita AVM Microclip; miniaturized version of Sugita Aneurysm Clip; spring-action mechanism; used for occlusion of small vessels; adjunct to bipolar coagulation; neurosurgical application; physician-operated; provides mechanical vessel occlusion to control bleeding during AVM procedures.
Clinical Evidence
Bench testing only. Sterilization cycle testing (50 cycles at 250°F) showed <10% change in closing force. Multiple opening testing (100 cycles) showed <10% change in holding force. Twist testing confirmed structural return to initial configuration. MRI compatibility testing at 1.5 Tesla showed deflection angle <3.5°.
Technological Characteristics
Spring-action aneurysm clip; miniaturized form factor; material and production methods identical to predicate Sugita Aneurysm Clip; mechanical operation; MRI compatible (1.5T).
Indications for Use
Indicated for the occlusion of small vessels as a compliment to bipolar coagulation in patients requiring control of bleeding on arteriovenous malformations or similar venous structures.
Regulatory Classification
Identification
An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.
{0}
FEB - 5 1997
Mizuho
K960037
Mizuho America Inc.
123 Brimbal Avenue,
Beverly, MA 01915
800-699-CLIP(2547)
TEL: 508-921-1718
FAX: 508-921-4003
# 510(k) SUMMARY
## Sugita AVM Microclip
Applicant:
Mizuho America Inc.
123 Brimbal Avenue
Beverly, MA 01915
Contact: Brian Connelly
Telephone: 508-921-1718
FAX: 508-921-4003
Submitted by:
Curtin & Associates International
1200 East Street
Dedham, MA 02026
Contact: Donald A. Lincoln
Telephone: 617-329-1955
FAX: 617-329-2529
Date this Summary was prepared: INSERT DATE
Trade Name: Sugita AVM Microclip
Common Name: AVM Clip
Classification Name: Could not be determined
Legally Marketed Devices to Which Substantial Equivalence is claimed:
Sundt AVM Micro Clip - K902544
SUGITA The leading name in neurosurgical instrumentation
{1}
# 510(k) SUMMARY-continued Sugita AVM Microclip
Description of Device: These devices are miniaturized versions of the currently marketed Sugita Aneurysm Clip. They are used to occlude small vessels when controlling bleeding on arteriovenous malformations or other similar venous structures.
Intended Use of the Device: Occlusion of small vessels as a compliment to bipolar coagulation.
Summary of Technological Characteristics: The devices are based in design upon a currently marketed spring action aneurysm clip. The material is the same, the production methods are the same, but the clip size is reduced to allow for the intended use.
Summary of test data:
CHANGES DURING STERILIZATION: The sample devices were subjected to sterilization at 250°F using a 20 minute cycle. This was repeated 50 times with the closing force being measured before the first cycle and at the conclusion of the test. The change in closing force for all clips tested was less than 10%.
MULTIPLE OPENING TESTING: Sample clips were subjected to 100 openings and closings. The results showed that holding force changed less than 10%.
TWIST TESTING: Samples were subjected to a twisting motion to determine the clips ability to return to its initial configuration. All clips passed.
MRI COMPATIBILITY: Sample clips were subject to testing in a 1.5 Tessla field. The angle of deflection for the clips tested was less than 3.5°.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.