Neocis Guidance System (NGS) with Intraoral Fiducial Array
Applicant
Neocis, Inc.
Product Code
PLV · Dental
Decision Date
Jul 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments.
Device Story
Neocis Guidance System (NGS) is a dental stereotaxic instrument for dental implantation. Input: CBCT scan of patient anatomy. Transformation: Planning software (Yomi Plan) creates 3D model for virtual implant placement; system uses robotic guide arm to provide haptic feedback, constraining drill motion to the pre-operative plan. Output: Physical, visual, and audible guidance to surgeon. Used in dental clinics by surgeons. Patient tracker assembly monitors movement; system adjusts cutting angle/position/depth in real-time to maintain accuracy. Benefits: Precise osteotomy, avoidance of critical structures (nerves/sinuses), and haptic constraint to prevent plan deviation. Subject device adds Intraoral Fiducial Array (IOFA) accessory to allow use with smaller field-of-view CT scanners.
Clinical Evidence
No clinical or animal testing was conducted. Evidence consists of bench testing, including registration testing, autoclave cycle testing, kinematic mount repeatability, scan artifact testing, fit clearance simulation, total system accuracy, and IOFA deflection testing. End-user validation was performed via nonclinical surrogate simulation.
Indicated for partially edentulous and fully edentulous adult patients who qualify for dental implantation surgery, providing preoperative planning and intraoperative navigational guidance for surgical instruments.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Neocis Inc. William Tapia Vice President Regulatory Affairs/Quality Assurance 2800 Biscayne Blvd Suite 600 Miami, Florida 33137
Re: K211129
Trade/Device Name: Neocis Guidance System (NGS) with Intraoral Fiducial Array Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: PLV Dated: July 1, 2021 Received: July 2, 2021
Dear William Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211129
Device Name
Neocis Guidance System (NGS) with Intraoral Fiducial Array
Indications for Use (Describe)
The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;vertical-align:middle"><svg height="16" width="16"><path d="M2 2 L14 14 M2 14 L14 2" fill="none" stroke="black"></path></svg></div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block;vertical-align:middle"><svg height="16" width="16"></svg></div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary - K211129
#### l. Submitter
Neocis Inc. 2800 Biscayne Blvd. Suite 600 Miami, FL 33137 Tel: 1-855-9NEOCIS
Contact Person: William Tapia, Vice President Regulatory Affairs/Quality Assurance Date Prepared: July 27, 2021
#### II. Device
Trade Name: Neocis Guidance System (NGS) with Intraoral Fiducial Array Common Name: Dental Stereotaxic Instrument Classification Name: Bone cutting instrument and accessories (21 CFR 872.4120) Classification: Class II Product Code: PLV
#### III. Predicate Devices
- . Primary Predicate -
- Neocis Guidance System (NGS) (K161399) .
- Reference Devices: ●
- . Neocis Guidance System (NGS) with Fiducial Array Splint (FAS) (K202348)
- . Neocis Guidance System (NGS) with Clamped Chairside Patient Splint (C-CPS) (K202100)
#### IV. Indications for Use
The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments.
#### V. Device Description
In terms of FDA regulations, the Neocis Guidance System (NGS) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120) (Figure 1).
{4}------------------------------------------------
In terms of previously FDA-cleared indications for use (K200805), the Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
Image /page/4/Picture/4 description: The image shows a Yomi robotic surgery system. The system includes a monitor, planning station PC, lift column, base cart, robotic guide arm, and patient tracker. The monitor and planning station PC display the Yomi logo. The robotic guide arm is positioned above the lift column and base cart.
Figure 1. Overview of the Neocis Guidance System (NGS). Major components list below.
{5}------------------------------------------------
- 1. Monitor: Provides visibility of Yomi Plan for the user and operative teamduring a procedure. Displays the same information as the Planning station laptop PC
- 2. Planning Station Laptop PC: Used to execute Yomi Plan for planning and surgery. This PC is used during the procedure as well to execute commands for the Guide Arm. There is not separate keyboard. (Not intended for removal and use by itself for planning. Remote preplanning shall be executed on a different third party PC that is not connected to the NGS or provided by Neocis.)
- 3. Lift Column: Supports the Robotic Guide Arm and Patient Tracker assembly and is used to adjust the height of the assembly.
- 4. Base Cart: Supports the NGS, provides mobility with four swivel casters and locks in position with the foot brake lever. Contains control PC and UPS.
- 5. Robotic Guide Arm: Used to assist the surgeon in performing the surgery. Holds dental drill and provides haptic feedback on position with respect to the plan.
- 6. Patient Tracker: Attached to Patient Splint via the End Effector to monitor and relay any changes in patient position.
The system allows the user to plan the surgery virtually in Yomi Plan (K191363-cleared for use alone on third party PCs for preplanning). The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The NGS robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument (K191605).
The patient tracking portion of the NGS is comprised of linkages from the patient to the NGS, which include the Chairside Patient Splint (CPS) (K173402) or Edentulous Patient Splint (EPS) (K200805), the End Effector (EE) and the Patient Tracker (PT). The Patient Splint is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. The EPS is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required).
A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the CPS prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the CPS on the patient, which relays information to the NGS in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy.
For partially edentulous patients, surgeons now have the option to use our new Clamped
{6}------------------------------------------------
Chairside Patient Splint (CCPS) (K202100). The implant process occurs in two phases: (1) The dental surgeon plans the surgical procedure with the planning software, on the day of surgery or sometime prior if a pre-operative CT scan was taken at an earlier visit. A virtual dental implant, selected from the dental implant library or using a generic model, both contained within our planning software (Yomi Plan), is placed at the desired location in the virtual patient model within our planning software. The software highlights critical anatomical structures to avoid, such as the inferior alveolar nerve or maxillary sinuses. (2) When the dental implant plan is optimized, the NGS provides precise and accurate guidance of the dental surgical instruments according to the pre-operative plan. The NGS robotic arm, which holds the powered bone cutting tool, provides haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan. The bone cutting tool's foot pedal control is not connect to the NGS guide arm. Once the guide arm places the bone cutting tool into position according to the operative plan, the surgeon must activate the tool using the provided foot pedal associated with the bone cutting tool control box. The surgeon may modify the plan intraoperatively, if needed, has direct visualization of the patient anatomy, and is always in control of the surgical instrument.
Key safety features include:
- Emergency stop
- Automatic software-controlled safety pause triggered by joint or force limits
- Audio and visual queues
- Drill torque limits provided by the drill control console
- Full surgeon control and direct visualization of the surgical field
The subject of this submission is adding Intraoral Fiducial Array (IOFA) as a new device accessory based on a previously cleared design.
Image /page/6/Picture/11 description: The image shows two views of a black plastic tool. The tool has a flat, wide blade at the bottom and a handle with a hole at the top. There are several small, raised bumps on the handle. Text is visible on the blade of the tool.
Figure 2. LLUR and LRUL Intraoral Fiducial Arrays.
{7}------------------------------------------------
The IOFA is intended for use in clinical sites with reduced scan volumes. The IOFA is designed to reside with all the fiducial beads within the patient's mouth. The IOFA can only be used with the C-CPS.
The Intraoral Fiducial Array corresponds directly to the type of C-CPS being used. There are four basic types of C-CPS.
- . Lower Left/ Upper Right (LLUR) Posterior- sits on teeth 11-15 for the maxilla or teeth 27-31 for the mandible.
- . Lower Right/ Upper Left (LRUL) Posterior- sits on teeth 2-6 for the maxilla or teeth 18-22 for the mandible.
- . Lower Left/ Upper Right (LLUR) Anterior- sits on teeth 6-11 for the maxilla or teeth 22-27 for the mandible.
- . Lower Right/Upper Left (LRUL) Anterior- sits on teeth 6-11 for the maxilla or teeth 22-27 for the mandible.
A LLUR C-CPS requires an LLUR Intraoral Fiducial Array, and an LRUL C-CPS requires an LRUL Intraoral Fiducial Array. See Figure 2 for LLUR and LRUL Fiducial Arrays.
The IOFA uses kinematic features to attach to the C-CPS and therefore has only one correct orientation.
{8}------------------------------------------------
#### VI. Substantial Equivalence (SE) Discussion
The indications for use (IFU) and contraindications of the subject device have remained the same as the predicate and reference devices. Acrylates allergy contraindication was removed since it is no longer applicable. Contraindications remain the same as the previously cleared reference device K202100. The subject device represents a design change to the predicate device. They are functionally equivalent.
### a. Comparison of the Indications for Use and Contraindications
The indications for use (IFU) and contraindications of the subject device have remained the same as the predicate and reference devices. Acrylates allergy contraindication was removed since it is no longer applicable. Contraindications remain the same as the previously cleared reference device K202100. The subject device represents a design change to the predicate device. They are functionally equivalent.
{9}------------------------------------------------
| Technological<br>Characteristics | K211129<br>IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | Substantial<br>Equivalence<br>Analysis |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use (IFU) | The Neocis Guidance<br>System (NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both the<br>planning (pre-operative)<br>and the surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and provides<br>navigational guidance of<br>the surgical instruments. | The Neocis Guidance<br>System (NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both<br>the planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides<br>navigational<br>guidance of the<br>surgical instruments. | The Neocis<br>Guidance System<br>(NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both<br>the planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides accurate<br>navigational<br>guidance of surgical<br>instruments, with<br>regard to pre-<br>operative planning | The Neocis Guidance<br>System (NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both<br>the planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides<br>navigational | No difference. |
| Technological<br>Characteristics | K21xxxx<br>IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | Substantial<br>Equivalence<br>Analysis |
| Contraindicati<br>ons | The Neocis Guidance<br>System is not intended<br>for use with patients<br>that have insufficient<br>bone or teeth to retain a<br>Neocis Clamped<br>Chairside Patient Splint<br>(C-CPS) rigidly<br>throughout a surgical<br>procedure.<br>The Neocis Chairside<br>Patient Splint (C-CPS) | The Neocis Guidance<br>System is not intended<br>for use with patients<br>that have insufficient<br>bone or teeth to<br>retain a Splint rigidly<br>throughout a surgical<br>procedure.<br>Patients with allergies<br>to methyl<br>methacrylates. | The Neocis<br>Guidance System is<br>not intended for use<br>with patients that<br>have insufficient<br>bone or teeth to<br>retain a FAS rigidly<br>throughout a<br>surgical procedure. | The Neocis Guidance<br>System is not<br>intended for use<br>with patients that<br>have insufficient<br>bone or teeth to<br>retain a Neocis<br>Clamped Chairside<br>Patient Splint (C-<br>CPS) rigidly<br>throughout a<br>surgical procedure. | Acrylates allergy<br>contraindication<br>not needed.<br>Contraindications<br>remain the same as<br>the previously<br>cleared reference<br>device K202100. |
| Technological<br>Characteristics | K21xxxx<br>IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | Substantial<br>Equivalence<br>Analysis |
| | should not be affixed to<br>patients that exhibit:<br>• Periodontal disease<br>to include loose<br>teeth and inflamed<br>tissue<br>• Fixed orthodontic<br>appliances, bridges,<br>or dental implants<br>• Patients with a<br>history of jaw or TMJ<br>pain | | The FAS should not<br>be affixed to<br>patients that exhibit:<br>• Periodontal<br>disease to include<br>loose teeth and<br>inflamed tissue<br> | The Neocis Clamped<br>Chairside Patient<br>Splint (C-CPS) should<br>not be affixed to<br>patients that exhibit:<br>• Periodontal<br>disease to<br>include loose<br>teeth and<br>inflamed tissue<br>• Fixed<br>orthodontic<br>appliances,<br>bridges, or<br>dental implants<br>• Patients with a<br>history of jaw or<br>TMJ pain | |
# Table 1. Comparison of the Indications for Use and Contraindications
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## b. Comparison of Technology to Predicate Devices
There are no changes to the non-splint NGS hardware or software in this submission.
### c. Head-to-Head Comparison
| VII. | Table 2. Head-to-Head Comparison | |
|------|----------------------------------|--|
| | | |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
|-------------------------------------------------------------------------------|------------------------------------------|-------------------------------|------------------------------------|--------------------------------------|-------------|
| Patient Contacting<br>Materials | Ixef®-HC-1022<br>Avaspire AV-651<br>CF30 | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| NGS Power Supply | 120VAC/60 Hz | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Type of<br>Protection<br>against Electric<br>Shock | Class I Equipment | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Equipment<br>Suitable for use<br>in the presence<br>of Flammable<br>Mixtures? | No | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
{13}------------------------------------------------
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Electrical Safety | ANSI AAMI<br>ES60601-<br>1:2005/(R)2012<br>and A1:2012,<br>C1:2009/(R)2012<br>and<br>A2:2010/(R)2012<br>(Consolidated<br>Text) Medical<br>electrical<br>equipment - Part<br>1: General<br>requirements for<br>basic safety and<br>essential<br>performance (IEC<br>60601-1:2005,<br>MOD) | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
| Electromagnetic<br>Disturbances | IEC 60601-1-2<br>Edition 4.0 2014-<br>02<br>Medical electrical<br>equipment - Part<br>1-2: General<br>requirements for<br>basic safety and<br>essential<br>performance -<br>Collateral<br>Standard:<br>Electromagnetic<br>disturbances -<br>Requirements<br>and tests | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Ingress<br>Protection | IPXO | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Mode of<br>Operation | Continuous<br>Operation | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| System Lateral<br>Accuracy | RMS < 1 mm | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| System Depth<br>Accuracy | RMS < 1 mm | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
| System Angular<br>Accuracy | RMS < 6.0° | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| CT Scan Quality<br>Requirements | 0.3 mm Voxel,<br>0.3 mm Slice<br>Thickness,<br>Matrix 512 x<br>512, Full 13<br>cm 21 sec,<br>Multi 2 DICOM<br>format. | Same as the<br>subject device | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| F/T Sensor Force<br>Measurement<br>Range | +/- 30 N | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| F/T Sensor<br>Torque<br>Measurement<br>Ra nge | +/- 2 Nm | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| F/T Sensor Single<br>Axis Force<br>Overload Limit | 200 N | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| F/T Sensor Single<br>Axis Torque<br>Overload Limit | 20 Nm | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
| Upper limit<br>specification for<br>Guidance Arm<br>Translation Speed | 1.25 m/s | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| Storage<br>Requirements | Store powered<br>at Room<br>Temperature<br>(68°F to 76°F or<br>20°C to 24.4°C)<br>and standard<br>ambient<br>humidity (5% to<br>95%) in a dust<br>free, clean<br>environment. | Same as the<br>subject device | Same as the<br>subject device | Same as the subject<br>device | Same |
| Patient Tracking | Patient Tracking<br>Arm | Same as the<br>subject device with<br>CPS | Same as the<br>subject device | Same as the subject<br>device | Same |
| Affixation of<br>tracking<br>technology to<br>patient | Clamped<br>Patient Splint | Same as the<br>subject device<br>with CPS | Same as the<br>subject device<br>with FAS | Same as the<br>subject device | Same |
| Patient<br>attachment<br>removal | Standard<br>dental<br>techniques | Same as the<br>subject device. | Same as the<br>subject device. | Same as the subject<br>device. | Same |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
| Fiducials | The Fiducial<br>Array attaches<br>to the splint<br>during the CT<br>scan to provide<br>a reference in<br>the image. | The Fiducial Array<br>attaches to the<br>splint during the CT<br>scan to provide a<br>reference in the<br>image. | Not required | The Fiducial Array<br>attaches to the splint<br>during the CT scan to<br>provide a reference<br>in the image. | Same |
| Kinematic<br>mount | KM integrated<br>into splint. | The Kinematic<br>Mount attaches to<br>the splint to<br>provide a mounting<br>point for the<br>Fiducial Array and<br>Patient Tracker. | KM integrated into<br>the FAS | KM integrated into<br>splint. | Same |
| Biocompatibility | Yes<br>(ISO 10993-1, -5,<br>10, -12) | Yes<br>(ISO 10993-1, -5, -<br>10, -12) | Same as the<br>subject device | Same as the subject<br>device | Same |
| Sterilization | Steam (ISO<br>17665-1) | Steam (ISO 17665-<br>1) | Same as the<br>subject device | Same as the subject<br>device | Same |
| Planning Software | Neocis Planning<br>Software<br>Application v2.0<br>(K202264) | Same as the<br>Subject Device | Same as the<br>Subject Device | Same as the Subject<br>Device | Yomi Plan v2.0<br>cleared under<br>K202264.<br>SE under K161399 |
| Technological<br>Characteristics | NGS with IOFA<br>Subject Device | K161399<br>Predicate | K200348<br>FAS<br>Reference Device | K202100<br>C-CPS<br>Reference Device | SE Analysis |
| Software Level of<br>Concern | Moderate | Same as the<br>Subject Device | Same as the<br>Subject Device | Same as the Subject<br>Device | Same |
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
The design changes in this submission involve adding a new Intraoral Fiducial Array (IOFA), which is a modified version of the Fiducial Array. The IOFA is a separate part that mounts to the Clamped Chairside Patient Splint (C-CPS) and contains fiducial markers for registration in CBCT scans prior to surgery. The IOFA is intended to sit intra-orally, which allows for use smaller field of view CT scanners. We have implemented the same affixation methods and technology as the predicate device. We have adopted the same indications for use as the predicate device.
We think that the subject device:
- Has a legally marketed predicate, .
- . Has the same intended use as the predicate,
- Has somewhat different technological characteristics that do not raise different questions of safety and effectiveness, ●
- For which we have provided adequate performance testing to show that ●
- · Those different technological characteristics are substantially equivalent to the predicate device.
Therefore, the subject device is substantially equivalent.
{19}------------------------------------------------
#### VIII. Performance Testing
Use of FDA-Recognized Consensus Standards
A risk analysis, sterilization, and biocompatibility testing were conducted on the final finished device per the following standards:
- ANSI AAMI ISO 14971:2007/(R)2010 (Corrected 4 October 2007) Medical devices -Applications of risk management to medical devices
- . ANSI AAMI ISO 17665-1:2006/(R)2013 Sterilization of health care products -- Moistheat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices
- ANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ANSI AAMI ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity
- ANSI AAMI ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- ANSI AAMI ISO 10993-12:2012 Biological evaluation of medical devices - Part 12:Sample preparation and reference materials
- Center for Devices and Radiological Health (CDRH)(2015). Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. Guidance for Industry and Food and Drug Administration Staff (FDA Amended June 9, 2017)
## Verification
- . Registration Testing
- Autoclave Cycle Testing
- Kinematic Mount Repeatability
- Scan Artifact Testing
- Fit Clearance Simulation
- Total System Accuracy
- IOFA Deflection Test
### Validation
- . IOFA End User Validation
- To validate the user requirements of the Intraoral Fiducial Array, as performed by a surgeon (end user). This validation activity is a nonclinical surrogate that simulates the process of applying, qualitatively evaluating rigidity, and removing a Intraoral Fiducial Array directly to a patient.
Animal or clinical testing was not conducted for the subject device.
### Conclusion:
The IOFA is substantially equivalent to the predicate. There are no changes to the intended use or to the fundamental technology, and there are no new risks to patients or users.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.