Neocis Guidance System (NGS) with Yomi Plan v2.0

K202264 · Neocis, Inc. · PLV · Dec 4, 2020 · Dental

Device Facts

Record IDK202264
Device NameNeocis Guidance System (NGS) with Yomi Plan v2.0
ApplicantNeocis, Inc.
Product CodePLV · Dental
Decision DateDec 4, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4120
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous adult patients who qualify for dental implants. When Yomi Plan software is used for preplanning on third party PCs, it is intended to perform the planning (preoperative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures. The output of Yomi Plan is to be used with the Neocis Guidance System (NGS).

Device Story

NGS is a dental stereotaxic instrument for dental implantation surgery; assists in pre-operative planning and intra-operative guidance. Input: CBCT scan of patient anatomy. Operation: Surgeon uses Yomi Plan software to create 3D virtual model; places virtual implants; identifies critical structures (e.g., inferior alveolar nerve). Intra-operatively, NGS robotic arm holds/guides third-party powered bone cutting instrument; provides haptic feedback by constraining motion to the plan; adjusts for patient movement via electromechanical patient tracker (PT) linked to patient splint (CPS/EPS). Output: Physical, visual, and audible feedback to surgeon. Used in dental clinics by surgeons. Benefits: Precise/accurate osteotomy; real-time adjustment to patient movement; haptic constraint prevents deviation from plan. Update v2.0 adds wireless network capabilities for CBCT file transfer and remote access via TeamViewer; wireless disabled during osteotomy phase.

Clinical Evidence

Bench testing only. Software V&V performed per IEC 62304 and FDA guidance. Wireless coexistence validated per AAMI TIR69 and IEEE ANSI C63.27. End-user validation performed in a simulated surgical environment to ensure system design, labeling, and processes meet user needs.

Technological Characteristics

Dental stereotaxic instrument; robotic arm guidance; haptic feedback. Connectivity: Ethernet, Wi-Fi (disabled during osteotomy). Software: Windows 10-based, Moderate Level of Concern. Standards: ANSI AAMI ISO 14971:2019, ANSI AAMI IEC 62304:2006/A1:2016, AAMI TIR69:2017, IEEE ANSI C63.27-2017.

Indications for Use

Indicated for partially edentulous and fully edentulous adult patients who qualify for dental implants, requiring assistance in pre-operative planning and intra-operative navigational guidance for dental implantation surgery.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The text reads "U.S. FOOD & DRUG ADMINISTRATION". December 4, 2020 Neocis Inc. Thomas E. Claiborne, Ph.D. Regulatory Affairs Manager 2800 Biscayne Blvd Suite 600 Miami, Florida 33137 Re: K202264 Trade/Device Name: Neocis Guidance System (NGS) with Yomi Plan v2.0 Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: PLV Dated: September 4, 2020 Received: September 10, 2020 Dear Thomas Claiborne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202264 Device Name Neocis Guidance System (NGS) with Yomi Plan v2.0 #### Indications for Use (Describe) The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous adult patients who qualify for dental implants. When Yomi Plan software is used for preplanning on third party PCs, it is intended to perform the planning (preoperative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures. The output of Yomi Plan is to be used with the Neocis Guidance System (NGS). X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a gray horizontal line that extends over the "neo" portion of the word. To the right of the word is a gray crosshair symbol, with a small circle at the intersection of the lines. ## 510(k) Summarv K202264 #### -Submitter Neocis Inc. 2800 Biscayne Blvd. Suite 600 Miami, FL 33137 Tel: 1-855-9NEOCIS Contact Person: Thomas Claiborne, Ph.D., Regulatory Affairs Manager Date Prepared: December 3, 2020 #### II. Device | Trade Name: | Neocis Guidance System (NGS) with Yomi Plan v2.0 | |----------------------|-----------------------------------------------------------| | Common Name: | Dental Stereotaxic Instrument | | Classification Name: | Bone cutting instrument and accessories (21 CFR 872.4120) | | Classification: | Class II | | Product Code: | PLV | #### III. Predicate Devices - Primary Predicate: ● - o Neocis Guidance System (NGS) (K161399) - Reference Devices ● - Neocis Planning Software Application for 3rd Party PCs (K191363) O - Neocis Guidance System (NGS) with Edentulous Patient Splint (EPS) O (K200805) #### IV. Indications for Use The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When Yomi Plan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides preoperative planning for dental implantation procedures. The output of Yomi Plan is to be used with the Neocis Guidance System (NGS). #### V. Device Description In terms of FDA regulations, the Neocis Guidance System (NGS) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). In terms of previously FDA-cleared indications for use (K200805), the Neocis Guidance System {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a gray horizontal line that extends to the left and right of the word. A gray vertical line intersects the horizontal line at the center of the word, forming a crosshair-like symbol. (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational quidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. The system allows the user to plan the surgery virtually in Yomi Plan (K191363-cleared for use alone on third party PCs for preplanning). The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The NGS robotic arm holds and quides a standard FDA-cleared third party powered bone cutting instrument (K191605). The patient tracking portion of the NGS is comprised of linkages from the patient to the NGS, which include the Chairside Patient Splint (CPS) (K173402) or Edentulous Patient Splint (EPS) (K200805), the End Effector (EE) and the Patient Tracker (PT). The Patient Splint is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. The EPS is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required). A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the CPS prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the CPS on the patient, which relays information to the NGS in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy. The implant process occurs in two phases: (1) The dental surgeon plans the surgical procedure with the planning software, on the day of surgery or sometime prior if a pre-operative CT scan was taken at an earlier visit. A virtual dental implant, selected from the dental implant library or using a generic model, both contained within our planning software, is placed at the desired location in the patient model. The software highlights critical anatomical structures to avoid, such as the inferior alveolar nerve. (2) When the dental implant plan is optimized, the NGS provides precise and accurate quidance of the dental surgical instruments according to the preoperative plan. The NGS robotic arm, which holds the surgical instrument, provides haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan. The surgeon may modify the plan intraoperatively, if needed, has direct visualization of the patient anatomy, and is always in control of the surgical instrument. Key safety features include: - Emergency stop - Safety pause - Audio and visual queues - Drill torque limits - . Full surgeon control and direct visualization of the surgical field The Neocis Guidance System (NGS) with Yomi Plan v2.0 is a "catch-up" focused on the planning software and presenting changes made from v1.2 (K161399) to the current release {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a thin gray horizontal line that extends beyond the letters on both sides. There is a small gray circle above the "c" with a vertical line extending above and below the circle. v2.0 (wireless network capabilities, interface updates, etc.). The Neocis Guidance System (NGS) contains two software packages: (1) planning and (2) control. Each resides on a separate PC on the device: (1) planning station laptop PC and (2) control PC in the cart base. There are no changes to the control software or the NGS hardware in this submission. The use of TeamViewer has been implemented to access NGS systems that connected to external networks to examine system performance for postmarket. ### VI. Comparison of Technological Characteristics This submission is focused on an update to our planning software, Yomi Plan v2.0, and adds wireless network capabilities to the system primarily for CBCT file transfer during the planning phase of the procedure. During the osteotomy (guidance) phase of the system disables the wireless network connection. Once the osteotomy portion of the procedure is completed and exited in the software application, the wireless network connection is reenabled. | Feature | Subject Device:<br>NGS with Yomi Plan<br>v2.0 | Predicate 1: NGS<br>K161399 | Predicate 2: Neocis<br>Planning and<br>Guidance Software<br>for 3rd Party PCs<br>K191363 | Predicate 3: NGS<br>with Edentulous<br>Patient Splint<br>K200805 | SE<br>Analysis | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| | Indications for<br>use | The Neocis Guidance<br>System (NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both the<br>planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides navigational<br>guidance of the<br>surgical instruments.<br>The NGS is intended<br>for use in partially<br>edentulous and fully<br>edentulous adult<br>patients who qualify<br>for dental implants.<br>When Yomi Plan<br>software is used for<br>preplanning on third<br>party PCs, it is<br>intended to perform<br>the planning (pre-<br>operative) phase of<br>dental implantation<br>surgery. Yomi Plan<br>provides pre-<br>operative planning for<br>dental implantation<br>procedures. The<br>output of Yomi Plan is<br>to be used with the<br>Neocis Guidance<br>System (NGS). | The Neocis<br>Guidance System<br>(NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both<br>the planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides navigational<br>guidance of the<br>surgical instruments. | The Neocis Planning<br>Software Application<br>(NPSA) for 3rd Party<br>PCs is intended to<br>perform the planning<br>(pre-operative)<br>phase of dental<br>implantation<br>surgery. The NPSA<br>provides pre-<br>operative planning<br>for dental<br>implantation<br>procedures. The<br>output of the NPSA<br>is to be used with<br>the Neocis<br>Guidance System. | The Neocis Guidance<br>System (NGS) is a<br>computerized<br>navigational system<br>intended to provide<br>assistance in both the<br>planning (pre-<br>operative) and the<br>surgical (intra-<br>operative) phases of<br>dental implantation<br>surgery. The system<br>provides software to<br>preoperatively plan<br>dental implantation<br>procedures and<br>provides navigational<br>guidance of the<br>surgical instruments.<br>The NGS is intended<br>for use in partially<br>edentulous and fully<br>edentulous adult<br>patients who qualify<br>for dental implants. | Subject<br>device<br>combines<br>indications<br>from the<br>predicates.<br>Rebrand as<br>"Yomi Plan." | | Feature | Subject Device:<br>NGS with Yomi Plan<br>v2.0 | Predicate 1: NGS<br>K161399 | Predicate 2: Neocis<br>Planning and<br>Guidance Software<br>for 3rd Party PCs<br>K191363 | Predicate 3: NGS<br>with Edentulous<br>Patient Splint<br>K200805 | SE<br>Analysis | | Yomi Plan<br>Software<br>Version | v2.0 | v1.2 | v1.8.1 | v1.10 (not a subject<br>of the submission) | Updated<br>GIU,<br>network<br>connectivity,<br>remote<br>access<br>added | | OS | Windows 10 | Windows 7 | Windows 7 | Windows 10 | Remains<br>Windows<br>based | | PC<br>Requirements | PC with 64-bit<br>Windows 10 OS or<br>newer with a<br>minimum of 4 GB of<br>RAM and a 2 GHz<br>dual core processor.<br>Local memory (hard<br>drive) should be a<br>minimum of 100 GB<br>with 7200 RPM or<br>SSD. Connectivity<br>requirements include<br>ethernet, Wi-Fi, USB,<br>or CD drive. | PC with 64-bit<br>Windows 7 OS or<br>newer with a<br>minimum of 4 GB of<br>RAM and a 2 GHz<br>dual core processor.<br>Local memory (hard<br>drive) should be a<br>minimum of 100 GB<br>with 7200 RPM or<br>SSD. USB or CD<br>drive. | PC with 64-bit<br>Windows 7 OS or<br>newer with a<br>minimum of 4 GB of<br>RAM and a 2 GHz<br>dual core processor.<br>Local memory (hard<br>drive) should be a<br>minimum of 100 GB<br>with 7200 RPM or<br>SSD. USB or CD<br>drive. | PC with 64-bit<br>Windows 7 OS or<br>newer with a<br>minimum of 4 GB of<br>RAM and a 2 GHz<br>dual core processor.<br>Local memory (hard<br>drive) should be a<br>minimum of 100 GB<br>with 7200 RPM or<br>SSD. USB or CD<br>drive. | Added<br>network<br>connectivity<br>via ethernet<br>or Wi-Fi | | Yomi Plan<br>Functions | Load CT<br>● Scanned Image<br>● Optimize Image<br>● Plan Procedure<br>(place implant)<br>● Save Surgical<br>Plan<br>Connect to<br>● Control<br>software<br>Provide<br>● Feedback to<br>Surgeon<br>regarding<br>physical<br>location of Drill<br>and Drill<br>components<br>Select Surgical<br>● Phase<br>Set areas for<br>● mechanical<br>restriction<br>during surgical<br>operation<br>Visualize CT<br>● Scanned Image<br>with 2D Slices<br>Generate<br>● Panoramic<br>reconstruction<br>along arch<br>Visualize<br>● Panoramic<br>reconstruction<br>with cross<br>sections along<br>panoramic arch<br>Map Splint<br>● coordinate | Load CT<br>● Scanned Image<br>● Optimize Image<br>● Plan Procedure<br>● (place implant)<br>Save Surgical<br>● Plan<br>Connect to<br>● Control<br>software<br>Provide<br>● Feedback to<br>Surgeon<br>regarding<br>physical<br>location of Drill<br>and Drill<br>components<br>Select Surgical<br>Phase<br>Set areas for<br>● mechanical<br>restriction<br>during surgical<br>operation<br>Visualize CT<br>● Scanned Image<br>with 2D Slices<br>Generate<br>● Panoramic<br>reconstruction<br>along arch<br>Visualize<br>Panoramic<br>reconstruction<br>with cross<br>sections along<br>panoramic arch<br>Map Splint<br>● coordinate | Load CT<br>● Scanned<br>Image<br>● Optimize<br>Image<br>● Plan<br>Procedure<br>(place implant)<br>● Save Surgical<br>Plan<br>Visualize CT<br>Scanned<br>Image with 2D<br>Slices<br>Generate<br>● Panoramic<br>reconstruction<br>along arch<br>Visualize<br>Panoramic<br>reconstruction<br>with cross<br>sections along<br>panoramic<br>arch<br>Define<br>● anatomical<br>planes<br>Clip CT<br>● Scanned<br>Images<br>Define Arch for<br>● generating<br>panoramic<br>reconstruction<br>Provide the<br>● user with a<br>means to<br>define a nerve | Load CT<br>● Scanned Image<br>● Optimize Image<br>● Plan Procedure<br>● (place implant)<br>Save Surgical<br>● Plan<br>Connect to<br>● Control<br>software<br>Provide<br>● Feedback to<br>Surgeon<br>regarding<br>physical<br>location of Drill<br>and Drill<br>components<br>Select Surgical<br>● Phase<br>Set areas for<br>mechanical<br>restriction<br>during surgical<br>operation<br>Visualize CT<br>● Scanned Image<br>with 2D Slices<br>Generate<br>● Panoramic<br>reconstruction<br>along arch<br>Visualize<br>● Panoramic<br>reconstruction<br>with cross<br>sections along<br>panoramic arch<br>Map Splint<br>● coordinate | No Changes | | Feature | Subject Device:<br>NGS with Yomi Plan<br>v2.0 | Predicate 1: NGS<br>K161399 | Predicate 2: Neocis<br>Planning and<br>Guidance Software<br>for 3rd Party PCs<br>K191363 | Predicate 3: NGS<br>with Edentulous<br>Patient Splint<br>K200805 | SE<br>Analysis | | | system to<br>structures in CT<br>Scan<br>• Define<br>anatomical<br>planes<br>• Clip CT<br>Scanned<br>Images<br>• Define Arch for<br>generating<br>panoramic<br>reconstruction<br>• Provide the<br>user with a<br>means to define<br>a nerve<br>• Allow the user<br>to plan multiple<br>implants<br>• Measure<br>distances and<br>angles in the<br>plan | system to<br>structures in<br>CT Scan<br>• Define<br>anatomical<br>planes<br>• Clip CT<br>Scanned<br>Images<br>• Define Arch for<br>generating<br>panoramic<br>reconstruction<br>• Provide the<br>user with a<br>means to<br>define a nerve<br>• Allow the user<br>to plan multiple<br>implants<br>• Measure<br>distances and<br>angles in the<br>plan | • Allow the user<br>to plan<br>multiple<br>implants<br>• Measure<br>distances and<br>angles in the<br>plan | system to<br>structures in CT<br>Scan<br>• Define<br>anatomical<br>planes<br>• Clip CT<br>Scanned<br>Images<br>• Define Arch for<br>generating<br>panoramic<br>reconstruction<br>• Provide the<br>user with a<br>means to define<br>a nerve<br>• Allow the user<br>to plan multiple<br>implants<br>• Measure<br>distances and<br>angles in the<br>plan | | | Level of<br>Concern | Moderate | Moderate | Moderate | Moderate | Identical | | Installation | Windows Installer<br>.msi file | Executable file .exe | Windows Installer<br>.msi file | Windows Installer<br>.msi file | Upgrade to<br>use<br>Windows<br>Installer | | OTS Software | TeamViewer | N/A | N/A | N/A | Add<br>TeamViewer<br>Identical | | NGS<br>Hardware | No changes<br>(activate/use pre-<br>existing network<br>features) | See predicate 510(k)<br>Summary | See predicate<br>510(k) Summary | See predicate 510(k)<br>Summary | Identical | | Wireless data<br>transmission<br>over LAN | Yes, via integrated<br>hardware, tested<br>according to:<br>• AAMI TIR69:<br>2017 Technical<br>Information<br>Report Risk<br>management of<br>radio-frequency<br>wireless<br>coexistence for<br>medical devices<br>and systems.<br>• IEEE ANSI<br>C63.27-2017<br>American<br>National Standard<br>for Evaluation of<br>Wireless<br>Coexistence | N/A | N/A | N/A | Add wireless<br>transmission<br>of files over<br>LAN | | Interface | Windows based GUI | Windows based GUI | Windows based GUI | Windows based GUI | Updated<br>look and<br>organization | #### Table 1: Comparison of technological characteristics to the predicates {6}------------------------------------------------ # K202264 Image /page/6/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a horizontal gray line, and to the right of the word is a gray line with a circle in the middle. The logo is simple and modern, and the colors are eye-catching. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a horizontal gray line that extends over the "neo" portion of the word. To the right of the "cis" portion of the word is a vertical gray line with a circle at the top, intersecting the horizontal line. # VII. Performance Testing {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font in yellow color. A thin gray line is above the word, and a gray circle with a plus sign through it is above the "i" in "neocis". The logo is simple and modern. This submission only includes the planning software and wireless transmission coexistence. Software V&V has been fully executed according to the following: • ANSI AAMI ISO 14971: 2019 Medical devices - Applications of risk management to medical devices · ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes [Including Amendment 1 (2016)] · Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005 · Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2. 2014 Wireless Coexistence was testing according to the following: · AAMI TIR69: 2017 Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. · IEEE ANSI C63.27-2017 American National Standard for Evaluation of Wireless Coexistence · End User Validation testing was performed in a simulated use environment that is representative of the Surgical Environment and represents use with a patient to ensure that the system in its entirety, inclusive of design, manufacture, labeling, and processes developed for use meet the needs of the user, a validation activity will be performed. #### VIII. Conclusion This submission includes an update to our planning software and adds networking capabilities to the system. There are no technological changes to the hardware (NGS) in this submission. There are no changes to the intended use in this submission. There are no fundamental changes to the technology. Our performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...