The Electric Breast Pump (Models 918, HF918) is a powered breast pump to be used by lactating women to express and collect milk from their breast. The Electric Breast Pump is intended for a single user.
Device Story
Electric breast pump (Models 918, HF918) facilitates expression and collection of breast milk for lactating women. Device operates via diaphragm pump mechanism powered by 3.7V lithium-ion battery. User selects vacuum levels and cycle speeds via buttons on pump unit; LED visual indicator displays status. Available in two configurations: standard (918) with breast shield and bottle, or hands-free (HF918) with milk collection cup designed for placement inside bra. Device offers five modes: Stimulation, Expression, Two-in-one, Dual-Frequency, and Simulation, each with varying suction strengths and cycle speeds. Intended for single-user, home-healthcare environment. Backflow protection prevents milk from entering pump motor. Output is expressed milk collected in bottle or cup. Benefits include efficient milk expression and stimulation of milk secretion.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including electrical safety, electromagnetic compatibility, biocompatibility (cytotoxicity, sensitization, irritation), and performance verification (suction strength, backflow prevention, cycle speed, and battery performance).
Technological Characteristics
Powered diaphragm breast pump; 3.7V lithium-ion battery; LED interface. Materials evaluated per ISO 10993-5 and ISO 10993-10. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 62133. Features backflow protection. Software verification performed per FDA guidance.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Electric breast pump (Models: RH-228 and RH-338) (K201903)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 11, 2022
Shantou Huihengqi Electronic Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, Shanghai 200120 China
Re: K211024
> Trade/Device Name: Electric Breast Pump (Models 918, HF918) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: January 8, 2022 Received: January 11, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211024
Device Name Electric Breast Pump (Models 918, HF918)
Indications for Use (Describe)
The Electric Breast Pump (Models 918, HF918) is a powered breast pump to be used by lactating women to express and collect milk from their breast. The Electric Breast Pump is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary K211024
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR 807.92.
- Date of 510(k) Summary Preparation: February 10, 2022 1.
- 2. Sponsor Identification
### Shantou Huihengqi Electronic Technology Co., Ltd.
501, 5/F, Block C, 14A Industrial Zone, Longhu District, Shantou, 515000, China
Establishment Registration Number: Not registered
Contact Person: Gensong Yang Position: Quality Supervisor Tel: +86-0754-88893348 Email: 357732164@qq.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Tingting Su (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{5}------------------------------------------------
Page 2
Trade Name: Electric Breast Pump (Models 918, HF918) Common Name: Electric Breast Pump Models: 918, HF918
Regulatory Information Classification Name: Powered Breast Pump Product Code: HGX: Classification: II Regulation Number: 21 CFR 884.5160
- Identification of Predicate device న్.
510(k) Number: K201903 Product Name: Electric breast pump (Models: RH-228 and RH-338)
The predicate device has not been subject to a design-related recall.
#### Device Description: 6.
The Electric Breast Pump (Model 918 and HF918) are powered breast pumps intended to express and collect milk from a lactating woman's breast. Pumping can be performed on one breast (single pumping) or both breasts (double pumping) at the same time. The Electric Breast Pump is provided non-sterile and should be cleaned and disinfected according to the instructions for use prior to first use. The user employs buttons on the pump unit to select the pumping mode and vacuum level.
The Electric Breast Pump has five basic modes: Stimulation mode, Expression mode, Two-in-one mode, Dual-Frequency mode, and Simulation mode has 5 suction levels (vacuum levels: 60-217.5 ± 20 mmHg; cycle speed: 39-123 ± 5 cycles/min), expression mode (vacuum levels: 105-285 ± 20 mmHg; cycle speed: 24-84 ± 5 cycles/min), dual-frequency mode (vacuum levels: 67.5-277.5 ± 20 mmHg; cycle speed: 39-85 ± 5 cycles/min), and two-in-one mode (vacuum levels: 75-285 ± 20 mmHg; cycle speed: 59-123 ± 5 cycles/min) have seven suction levels, and simulation mode (vacuum levels: 187.5 ± 20 mmHg; cycle speed: 14 ± 5 cycles/min) has one suction level. Stimulation mode is for massaging the breast to stimulate breast milk secretion. Expression mode is for pumping breast milk. Dual Frequency mode is for cycling between strong and weak pumping strengths. Two-in-One mode cycles between massaging for 5 cycles and expressing breast milk for 1 cycle. Simulation mode operates at a single suction level to stimulate milk secretion.
The subject devices are available in two models: 918 and HF918. The hands-free (HF) model (HF918) is designed to be placed inside a user's bra and connect to the external pump unit. The milk collection
{6}------------------------------------------------
cup is located on the breast. The HF model (HF918) includes breast shield assemblies, milk collection cups, single and double tubing sets, a bottle set, power adapter, and a pump unit. The non-HF model (918) includes a breast shield connected to a milk bottle. The non-HF model includes breast shield assemblies, bottles, bottle stands, single and double tubing sets, power adapter, and a pump unit.
The Electric Breast Pumps are intended for a single user.
- 7. Indication for Use:
The Electric Breast Pump (Models 918, HF918) is a powered breast pump to be used by lactating women to express and collect milk from their breast. The Electric Breast Pump is intended for a single user.
- Comparison of Technological Characteristics 8.
| ITEM | Subject Device<br>Electric Breast Pump (918, HF918)<br>K211024 | Predicate Device<br>Electric Breast Pump (Models:<br>RH-228, RH-338)<br>K201903 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | HGX | HGX |
| Class | II | II |
| Regulation No. | 21 CFR 884.5160 | 21 CFR 884.5160 |
| Indication for Use | The Electric Breast Pump (Models 918,<br>HF918) is a powered breast pump to be<br>used by lactating women to express<br>and collect milk from their breast. The<br>Electric Breast Pump is intended for a<br>single user. | The Electric breast pump is intended<br>to be used by lactating women to<br>express and collect milk from their<br>breasts. It is intended for a single<br>user. |
| Patient Population | Breastfeeding women | Breastfeeding women |
| Anatomical Sites | Breast | Breast |
| Single User | Yes | Yes |
| Provided Non-sterile | Yes | Yes |
| Direct user contact | Yes | Yes |
| Re-usable | Yes | Yes |
Table 1 General Comparison
#### Table 2 Technological Comparison
| ITEM | Subject Device | Predicate Device |
|------|--------------------------------------|-------------------------------|
| | Electric Breast Pump (918,<br>HF918) | Electric Breast Pump (Models: |
| | K211024 | RH-228, RH-338) |
| | | K201903 |
{7}------------------------------------------------
| Pump Type | | Diaphragm | Diaphragm |
|---------------------|---------------------|----------------------------|--------------------------------------|
| Pumping Options | | Single and Double Pumping | Single Pumping |
| Visual Indicator | | 918 & HF918: LED | RH-228: LED<br>RH-338: LCD |
| Backflow Protection | | Yes | Yes |
| Suction<br>strength | Stimulation mode | 918 & HF918: 60-217.5 | RH-228:60-150<br>RH-338: 45-165 |
| | Expression mode | 918 & HF918: 105-285 | RH-228:120-275<br>RH-338: 110-300 |
| | Two-in One mode | 918 & HF918: 75-285 | N/A |
| | Dual-frequency mode | 918 & HF918: 67.5-277.5 | N/A |
| | Simulation mode | 918 & HF918:187.5 | N/A |
| Suction level | Massage mode | N/A | RH-338: 135-255 |
| | Stimulation mode | 918 & HF918: 5 levels | RH-228: 5 levels<br>RH-338: 9 levels |
| | Expression mode | 918 & HF918: 7 levels | RH-228: 4 levels<br>RH-338: 9 levels |
| | Two-in One mode | 918 & HF918: 7 levels | N/A |
| | Dual-frequency mode | 918 & HF918: 7 levels | N/A |
| | Simulation mode | 918 & HF918: 1 level | N/A |
| Cycling Speed | Massage mode | N/A | RH-338: 9 levels |
| | Stimulation mode | 918 & HF918: 39-123 | RH-228: 58-90<br>RH-338: 115-155 |
| | Expression mode | 918&HF918: 24-84 | RH-228: 28-52<br>RH-338: 25-52 |
| | Two-in One mode | 918 & HF918: 59-123 | N/A |
| | Dual-frequency mode | 918 & HF918: 39-85 | N/A |
| | Simulation mode | 918 & HF918:14 | N/A |
| | Massage mode | N/A | RH-338: 56-72 |
| Power Source | | 3.7 V, Lithium ion battery | 3.7 V, Lithium ion battery |
| User Control | | On pump body | On pump body |
The subject and predicate device have same indications for use statements, and the same intended use. They are both used to express and collect milk from a lactating woman's breast.
The differences between the subject devices and the predicate devices are pumping options, suction levels, suction strength, and cycle speed. These differences do not raise different questions of safety and effectiveness.
{8}------------------------------------------------
#### Non-Clinical Testing 9.
Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device.
Electrical Safety and Electromagnetic Compatibility
- > IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012, Medical electrical equipment -Part 1: General requirements for basic safety, and essential performance.
- > IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility
- A IEC 60601-1-11:2010, AMD1:2013, Medical electrical equipment -- Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- > IEC 62133:2012, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
Biocompatibility testing in accordance with the FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020. Testing included the following assessments:
- A Cytotoxicity per ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- > Sensitization and Irritation per ISO 10993-10:2010 Biological evaluation of medical devices --Part 10: Tests for irritation and skin sensitization
Software Verification
- ▲ Software verification in accordance with the FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005.
Additionally, the following non-clinical tests were conducted:
Suction strength of subject devices was tested. All the test results complied with the design specifications of the subject device throughout the use life.
Backflow testing was conducted to ensure that even if the bottle is over-filled, no liquid will backflow into the tubing, and therefore no liquid can backflow into the pump motor. The test results showed that there was no backflow during the test.
{9}------------------------------------------------
Cycle speed of subject devices was tested. All the test results complied with the design specifications of the subject devices throughout the use life.
Battery performance testing was conducted to demonstrate that the battery remains functional during its stated use-life.
Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
- 10. Conclusion
The comparison and analysis above demonstrate that the Electric Breast Pump (Models 918, HF918) is as safe and effective as the predicate device and supports a determination of substantial equivalence to the predicate device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.