The Electric breast pump is intended to be used by lactating women to express and collect milk from their breast. It is intended for a single user.
Device Story
Electric breast pump (models RH-228, RH-338) for single-user, home-use milk expression. Device utilizes diaphragm-type vacuum pump driven by microprocessor to control vacuum pressure and cycle speed. RH-228 features stimulate/expression modes; RH-338 adds massage mode. User interface includes keypad with LED (RH-228) or LCD (RH-338) display for mode/pressure adjustment. RH-228 powered by 5Vdc USB; RH-338 powered by 3.7Vdc lithium battery or 5Vdc USB. Components include breast shield, cylinder, valve, tubing, feeding bottle, nipple kit, and massage cushion (RH-338). User-contacting materials are polypropylene and silicone. Device assists lactating women in milk collection; provider/user adjusts settings via interface to optimize comfort and expression efficiency.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2), battery safety (IEC 62133), biocompatibility (ISO 10993-5, ISO 10993-10), software verification, and performance testing (vacuum pressure, cycle rate, backflow prevention, and battery life).
Technological Characteristics
Diaphragm-type vacuum pump; microprocessor-controlled. Materials: Polypropylene (breast shield), silicone (massage cushion). Power: 5Vdc USB (RH-228); 3.7Vdc Li-ion battery/5Vdc USB (RH-338). Connectivity: None. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 62133, ISO 10993. Software: Microprocessor-based control of vacuum/cycle parameters.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Youha electric breast pump (Model: YH-8006IV) (K163136)
Related Devices
K172989 — Electric Breast Pump · Joytech Healthcare Co. , Ltd. · Dec 20, 2017
K210124 — Synergy Gold (SG) Portable (Models MM011450 and MM011460) · Uzinmedicare Co. · Oct 19, 2021
K252342 — Electric Breast Pump (YM-8807, YM-8805, YM-8806, YM-8810) · Guangdong Youmeng Electrical Technology Co., Ltd. · Dec 19, 2025
K212955 — Fit Pro Breast Pump · Hygeia II Medical Group, Inc. · Mar 15, 2022
K251512 — Hi bebe super (BT-150B) · Bistos Co., Ltd. · Sep 25, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font, with "FDA" in a blue box.
April 9, 2021
Cixi Ruihong Electric Appliance Co., Ltd. Kelong Zhang General Manager No. 18, Xiamaojia Road, Shuangqingpu Village, Xinpu Town, Cixi Ningbo. Zhejiang 315322 China
Re: K201903
Trade/Device Name: Electric breast pump (Models: RH-228 and RH-338) Regulation Number: 21 CFR& 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: February 9, 2021 Received: March 8, 2021
Dear Kelong Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201903
Device Name
Electric breast pump (Models: RH-228 and RH-338)
Indications for Use (Describe)
The Electric breast pump is intended to be used by lactating women to express and collect milk from their breast. It is intended for a single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### K201903
### 1. Submitter Information:
| Name: | Cixi Ruihong Electric Appliance Co., Ltd. |
|----------|---------------------------------------------------------------------------------------------------|
| Address: | No.18, Xiamaojia Road, Shuangqingpu Village,<br>Xinpu Town, Cixi, Ningbo, 315322, Zhejiang, China |
| Contact Person: | Mr. Kelong Zhang |
|-----------------|-------------------|
| Position: | General Manager |
| Phone: | +86-574-63581761 |
| Fax: | +86-574-63581761 |
| Email: | 1020666000@qq.com |
# Date 510(k) Summary Prepared: April 7, 2021
# 2. Device Information
Trade/Proprietary Name: Electric breast pump (Models: RH-228 and RH-338)
| Regulation Name: | Powered breast pump |
|--------------------|-----------------------------|
| Regulation Number: | 21 CFR 884.5160 |
| Common Name: | Powered breast pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
| Review Panel: | Obstetrics/Gynecology |
# 3. Predicative Device
| 510(k) Number: | K163136 |
|----------------|--------------------------------------------------------|
| Manufacturer: | Ningbo Youhe Electrical Appliance Technology Co., Ltd. |
| Device Name: | Youha electric breast pump (Model: YH-8006IV) |
The predicate device has not been subject to a design-related recall.
# 4. Device Description:
The Electric breast pump is a single-patient reusable device, intended to be used by lactating women to express and collect milk from their breasts. The device uses a diaphragm-type vacuum pump driven by a microprocessor. The microprocessor provides control over vacuum pressure and cycle speed.
{4}------------------------------------------------
The Electric breast pump has two models: RH-228 and RH-338. Model RH-228 is capable of single pumping and has two modes - stimulate mode and expression mode. The RH-228 user interface consists of a panel keypad and LED display in which the user switches between modes and controls the vacuum pressure. Model RH-338 is capable of single pumping and has three modes - stimulate mode, massage mode and expression mode. The RH-338 user interface consists of a panel keypad and LCD display in which the user switches between modes and controls the vacuum pressure.
The RH-228 is externally powered by 5Vdc USB. The RH-338 is internally powered by 3.7Vdc lithium battery or externally powered by 5Vdc USB. The external adapter also charges the battery.
Both models of the Electric breast pump are provided with a breast shield, cylinder cap, cylinder, valve, vacuum tubing, feeding bottle, and a nipple kit. Model RH-338 also includes a massage cushion and a dust cap. The user-contacting materials are limited to polypropylene (breast shield) and silicone (massage cushion).
# 5. Indications for Use
The Electric breast pump is intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
# 6. Comparison of Intended Use and Technological Characteristics
The following table compares the subject device to the predicate device with respect to the indications for use and technological characteristics:
| Characteristics | Subject Device<br>Electric breast pump<br>K201903 | Predicate Device<br>Youha Electric Breast Pump<br>ModelYH-8006IV<br>K163136 | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Indications for Use | The Electric breast pump is<br>intended to be used by<br>lactating women to express<br>and collect milk from their<br>breasts. It is intended for a<br>single user. | The Youha electric breast pump<br>is intended to be used by<br>lactating women to express and<br>collect milk from their breasts. It<br>is intended for a single user. | |
| Single-user | Yes | Yes | |
| Provided | Yes | Yes | |
| Non-sterile | | | |
| Re-usable | | Yes | Yes |
| Direct user contact | | Yes | Yes |
| Patient population | | Lactating women | Lactating women |
| Anatomical Sites | | Breast | Breast |
| Suction strength<br>(mmHg) | RH-228 | Expression mode:<br>120-275<br>Stimulate mode:<br>60-150<br>Massage mode:<br>135-255 | High expression mode:125-280<br>Massage mode: 50-190 |
| | RH-338 | Stimulate mode:<br>45-165<br>Expression mode:<br>110-300 | |
| Suction Levels | RH-228 | Expression mode: 4<br>levels<br>Stimulate mode: 5<br>levels | 10 levels |
| | RH-338: 9 levels | | |
| Cycle Speed<br>(cycles/min) | RH-228 | Expression mode:<br>28-52<br>Stimulate mode:<br>58-90<br>Massage mode:<br>56-72 | High expression mode:32-57<br>Massage mode: 71-100 |
| | RH-338 | Stimulate mode:<br>115-155<br>Expression mode:<br>25-52 | |
| Backflow<br>protection | | Yes | Yes |
| Single or Double<br>pumping | | Single pumping | Single Pumping |
| Visual Indicator | | RH-228: LED<br>RH-338: LCD | LCD |
| Pump Type | | Diaphragm | Diaphragm |
{5}------------------------------------------------
{6}------------------------------------------------
The subject and predicate device have identical indications for use statements. Therefore, they have the same intended use, i.e., for collection of breast milk from the breasts of a single user.
The subject and predicate device have different technological features, including the user interface, expression levels, suction strength, cycle speed, and power source. These technological differences do not raise different questions of safety or effectiveness.
# 7. Summary of Non-clinical Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications and that the subject devices are substantially equivalent to the predicate device.
- a. Electrical Safety and Electromagnetic Compatibility:
- IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+A1:2012,Medical . electrical equipment - Part 1: General requirements for basic safety, and essential performance with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012.
- . IEC 60601-1-11:2015. Medical electrical equipment - Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility
- IEC 62133:2012 Secondary cells and batteries containing alkaline or other ● non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- b. Biocompatibility:
Biocompatibility testing in accordance with the FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020. Testing included the following assessments:
- Cytotoxicity per ISO 10993-5:2009 Biological evaluation of medical devices . -- Part 5: Tests for in vitro cytotoxicity
- . Sensitization per ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
{7}------------------------------------------------
- Irritation per ISO 10993-10:2010 Biological evaluation of medical devices --. Part 10: Tests for irritation and skin sensitization
The user-contacting materials were shown to be biocompatible.
- c. Software Verification:
- . Software verification and validation in accordance with the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005.
- d. Performance testing:
- . Vacuum pressure and cycle rate testing was conducted at all settings for each device model and demonstrated that the devices met their specifications.
- . Backflow testing was conducted to demonstrate that liquid does not backflow into the tubing/pump.
- Battery performance testing was conducted to demonstrate that the battery . remains functional during its stated battery use-life.
- Battery status indicator testing was conducted to demonstrate that the battery . status indicator remains functional during its stated battery life.
# 8. Conclusion:
The comparison and analysis above demonstrate that the Electric breast pump is as safe and effective as the predicate device and supports a determination of substantial equivalence to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.